TELIGEN System Peripheral Motor Nerve Stimulation Indications
K223438 · Medos International SARL · PDQ · Feb 10, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K223438
Device Name
TELIGEN System Peripheral Motor Nerve Stimulation Indications
Applicant
Medos International SARL
Product Code
PDQ · Ear, Nose, Throat
Decision Date
Feb 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine. The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which: · the use of stereotactic surgery may be appropriate, and · reference to a rigid anatomical structure, such as the pelvis or a vertebrac can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays. These procedures include but are not limited to spinal fusion. TELIGEN Clear, when used with UNAS, can be precalibrated with the Brainlab Navigation System. TELIGEN Access Probe, when used with UNAS, can be pre-calibrated with the Brainlab Navigation System, where other navgation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer. The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.
Device Story
TELIGEN System is a sterile, single-use kit for spinal surgery; includes camera, ports, retractor, and navigation-ready instruments. TELIGEN Access Probe functions as a nerve stimulator; connects to SENTIO MMG System. During surgery, probe delivers electrical stimulation to peripheral motor nerves/spinal nerve roots; SENTIO MMG sensors detect mechanomyographic (MMG) response. System alerts surgeon to nerve proximity via MMG detection. Used by surgeons in OR; provides real-time feedback for nerve location/identification. Navigation-ready when used with Universal Navigation Adaptor Set (UNAS); integrates with navigation systems (e.g., Brainlab) to assist in locating anatomical structures relative to pre-acquired images (CT/MR/Fluoroscopy). Benefits include improved surgical precision and nerve avoidance during minimally invasive spinal procedures.
Clinical Evidence
Bench testing only. Evaluations included electrical safety (IEC 60601-1), functional performance (connector compatibility, insulation effectiveness, stimulation density), and sterility validation (ISO 11135). Biocompatibility assessed per ISO 10993-1 (cytotoxicity, systemic toxicity, irritation, sensitization, pyrogenicity).
Technological Characteristics
Stainless steel probe with Parylene C coating; inline blunt tip geometry. Electrical stimulation via proximal connector to SENTIO MMG System. Sterile, single-use, ethylene oxide sterilized. Navigation-ready via UNAS interface. Complies with IEC 60601-1 and ISO 10993-1.
Indications for Use
Indicated for patients undergoing spinal surgery requiring minimally invasive access, visualization, discectomy, or navigation-assisted procedures. The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services - USA emblem. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 10, 2023
Medos International, SARL % Daria Bochenek Senior Regulatory Affairs Specialist Synthes Gmbh Eimattstrasse 3 4436 Oberdorf Switzerland
## Re: K223438
Trade/Device Name: TELIGEN System Peripheral Motor Nerve Stimulation Indications Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ, OLO, HRX Dated: November 14, 2022 Received: November 14, 2022
Dear Daria Bochenek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Patrick Antkowiak -S
for
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223438
## Device Name
TELIGEN System Nerve Stimulation Indications
## Indications for Use (Describe)
The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.
The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:
· the use of stereotactic surgery may be appropriate, and
· reference to a rigid anatomical structure, such as the pelvis or a vertebrac can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.
These procedures include but are not limited to spinal fusion. TELIGEN Clear, when used with UNAS, can be precalibrated with the Brainlab Navigation System. TELIGEN Access Probe, when used with UNAS, can be pre-calibrated with the Brainlab Navigation System, where other navgation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.
The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div></span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div></span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| A. | Submitter Information | | | |
|----|----------------------------|---------------------------------------------------------------------------------------------------|--|--|
| | 510(k) Sponsor: | Medos International, SARL | | |
| | Contact Person: | Daria Bochenek, Senior Regulatory Affairs | | |
| | | Specialist | | |
| | | Eimattstrasse 3 | | |
| | | 4436 Oberdorf | | |
| | | Switzerland | | |
| | Telephone: | +41 61 965 61 54 | | |
| | Fax: | +41 61 965 66 16 | | |
| | Email: | dbochene@its.jnj.com | | |
| B. | Date Prepared | 01 February 2023 | | |
| C. | Device Name | | | |
| | Trade/Proprietary Name: | TELIGEN System Nerve Stimulation Indications | | |
| | Common/Usual Name: | PDQ - Neurosurgical Nerve Locator<br>OLO - Orthopedic Stereotaxic Instrument<br>HRX - Arthroscope | | |
| | Device Classification and | Class II | | |
| | Regulation: | PDQ – 21 CFR §874.1820 | | |
| | | OLO – 21 CFR §882.4560 | | |
| | | HRX – 21 CFR §888.1100 | | |
| | Classification Product and | PDQ - Neurology | | |
| | Panel Code | OLO, HRX – Orthopedic | | |
| D. | Predicate Device Names | | | |
| | Primary Predicate Device: | SENTIO MMG Pedicle Access Needles (K173134) – PDQ | | |
Secondary Predicate Device:
{4}------------------------------------------------
TELIGEN System Navigation Ready Indications (TELIGEN Access Probe, TELIGEN Clear) (K223108) – OLO, HRX
#### Device Description E.
The TELIGEN Procedure Kit Pro is a sterile, single use kit intended for use in surgical spinal procedures allowing for access, visualization, discectomy, graft delivery, navigation and peripheral motor nerve stimulation.
The TELIGEN Procedure Kit Pro includes a camera, ports and port holder, TELIGEN Clear, an Access Probe, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments.
The TELIGEN Access Probe is part of the DePuy Synthes Navigation Ready Instruments Portfolio and is designed for navigated and non-navigated use. Navigation of this instrument is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS). For further details on UNAS, refer to the UNAS labeling.
The TELIGEN Access Probe is intended to stimulate peripheral motor nerves, including spinal nerve roots, for location and identification during surgery. It contains an insulated probe compatible with the SENTIO MMG System. The SENTIO MMG sensors serve to measure a mechanomyographic (MMG) response and the SENTIO MMG System's detection of this response alerts the surgeon of nerve location in the proximity of the TELIGEN Access Probe's distal tip.
#### F. Indications for Use
The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.
The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:
● the use of stereotactic surgery may be appropriate, and
. reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.
These procedures include but are not limited to spinal fusion. TELIGEN Clear, when used with UNAS, can be pre-calibrated with the Brainlab Navigation System. TELIGEN Access Probe, when used with UNAS, can be pre-calibrated with the Brainlab Navigation System,
{5}------------------------------------------------
where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.
The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.
## G. Summary of Similarities and Differences in Technological Characteristics, Performance, and Indications for Use
The technological characteristics, including design, material and performance as well as indications for use of the Nerve Stimulating TELIGEN Access Probe are consistent with those of the predicate devices.
{6}------------------------------------------------
| Attribute | Subject Device<br>TELIGEN System Peripheral<br>Motor Nerve Stimulation<br>Indications | Primary Predicate Device<br>SENTIO MMG Pedicle Access<br>Needles (K173134) | Substantially Equivalent |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(peripheral motor<br>nerve stimulation<br><i>specific text in italic</i> ) | The TELIGEN System is indicated<br>to provide minimally invasive<br>access, visualization, illumination,<br>magnification and discectomy of the<br>surgical area of the spine.<br>The TELIGEN Access Probe and<br>TELIGEN Clear are Navigation<br>Ready Instruments when used with<br>the compatible Universal Navigation<br>Adaptor Set (UNAS) and are<br>intended to assist the surgeon in<br>locating anatomical structures in<br>either open or percutaneous<br>procedures. The navigation feature<br>is used in surgical spinal procedures,<br>in which:<br>• the use of stereotactic<br>surgery may be appropriate, and<br>• reference to a rigid<br>anatomical structure, such as the<br>pelvis or a vertebrae can be<br>identified relative to the acquired<br>image (CT, MR, 2D fluoroscopic<br>image or 3D fluoroscopic image<br>reconstruction) and/or an image data<br>based model of the anatomy using a<br>navigation system and associated<br>tracking arrays. | | Yes, substantially equivalent for<br>peripheral motor nerve stimulation<br>indications. |
| Attribute | Subject Device<br>TELIGEN System Peripheral<br>Motor Nerve Stimulation<br>Indications | Primary Predicate Device<br>SENTIO MMG Pedicle Access<br>Needles (K173134) | Substantially Equivalent |
| These procedures include but are not<br>limited to spinal fusion. TELIGEN<br>Clear, when used with UNAS, can<br>be pre-calibrated with the Brainlab<br>Navigation System. TELIGEN<br>Access Probe, when used with<br>UNAS, can be pre-calibrated with<br>the Brainlab Navigation System,<br>where other navigation systems<br>require manual calibration and<br>tracking arrays supplied by the<br>navigation system manufacturer.<br>The TELIGEN Access Probe is<br>indicated for stimulation of<br>peripheral motor nerves, including<br>spinal nerve roots, for location and<br>identification during surgery. | The SENTIO MMG Pedicle Access<br>Needles are indicated for use in<br>pedicle pilot hole preparation and<br>stimulation of peripheral motor<br>nerves for location and<br>identification during surgery,<br>including spinal nerve roots. | | |
| Classification | Class II, Product Code PDQ,<br>added with this submission | Class II, Product Code PDQ | Yes |
| Instruments | Access Probe<br>Stimulation Wire | Pedicle Access Needle (Trocar and<br>Cannula)<br>Stimulation Wire | Yes |
| Compatible Nerve<br>Stimulation System | SENTIO MMG System | SENTIO MMG System | Yes |
| Stimulation device<br>distal tip geometry | Inline blunt tip | Trocar tip<br>Diamond tip<br>Beveled tip | Yes. Since the Pedicle Access<br>Needles provide the function of pilot<br>hole preparation (in addition to |
| Attribute | Subject Device<br>TELIGEN System Peripheral<br>Motor Nerve Stimulation<br>Indications | Primary Predicate Device<br>SENTIO MMG Pedicle Access<br>Needles (K173134) | Substantially Equivalent |
| | | | access), the tip design is defined as<br>trocar, diamond or bevel compared<br>to the blunt tip of TELIGEN Access<br>Probe. This difference does not raise<br>any new safety or performance<br>questions. |
| Electrical Insulation | The TELIGEN Access Probe has<br>biocompatible electrical insulation<br>applied to select portions. The<br>proximal and distal surface of the<br>instrument is selectively non-<br>insulated stainless steel to connect to<br>the stimulation wire and for tissue<br>stimulation, respectively. | The Pedicle Access Needles have<br>biocompatible electrical insulation<br>applied to select portions. The<br>proximal and distal surface of the<br>instrument is selectively non-<br>insulated stainless steel to connect to<br>the stimulation wire and for tissue<br>stimulation, respectively. | Yes |
| Proximal Stimulator<br>Connector | Yes | Yes | Yes |
| Patient contacting<br>materials | Stainless Steel, Parylene C Coating | Stainless Steel, Parylene C Coating | Yes |
| Delivery | Sterile via Ethylene Oxide<br>Single Use | Sterile via Ethylene Oxide<br>Single Use | Yes |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### H. Materials
- . TELIGEN Access Probe: Stainless Steel, Parvlene C Coating
#### I. Performance Data
The performance data for the subject device consists of the following evaluations:
Electrical Safety Evaluation, Functional Performance Testing and Verification Analysis as well as Sterility. The table below summarizes testing and/or evaluations that were performed on the subject device to show substantial equivalence to the predicate device. Results demonstrated that the subject device is appropriate for peripheral motor nerve stimulation and is substantially equivalent when compared to the legally marketed primary predicate device cleared by FDA.
| Test | Test Method Summary | Results |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Electrical Safety Evaluation | Evaluation was performed in accordance with IEC 60601-1: 2005(R)2012. | All evaluations passed acceptance criteria. |
| Functional Performance Testing and Verification Analysis | Connector Compatibility Insulation Effectiveness SENTIO MMG System Compatibility Electrical Stimulation Density | All samples and evaluations passed acceptance criteria. |
| Sterility | Ethylene Oxide sterilization validation performed per ISO 11135. | All samples passed acceptance criteria. |
#### J. Biocompatibility Data
The TELIGEN Access Probe is manufactured from the same materials as the predicate devices.
A risk analysis was performed taking into account the categorization of body contact and duration of use of the medical device. Existing data, endpoint-specific testing, and endpoint assessments were analyzed to cover the identified test methods in Annex A of ISO 10993-1. Additionally, data was leveraged by other means (e.g., authorized use of Master File, predicate and reference devices, well known and characterized materials) to support the biocompatibility of the subject devices.
{10}------------------------------------------------
Biocompatibility endpoints evaluated per ISO 10993-1:
- Physical and/or chemical information
- Cytotoxicity
- Acute Systemic Toxicity
- Irritation or Intracutaneous Reactivity
- Sensitization
- Material mediated pyrogenicity
#### K. Conclusion
The indications for use of the Nerve Stimulating TELIGEN Access Probe and Stimulation Wire are consistent with those of the predicate devices. The technological characteristics of the Nerve Stimulating TELIGEN Access Probe and Stimulation Wire in terms of design, materials and performance are consistent with those of the predicate devices. The Nerve Stimulating TELIGEN Access Probe and Stimulation Wire are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.