# Contour Knee (K223429)

_Quality Electrodynamics, LLC · MOS · Jan 13, 2023 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223429

## Device Facts

- **Applicant:** Quality Electrodynamics, LLC
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Jan 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Contour Knee is intended for use with Siemens 0.55T MR systems to produce diagnostic images of knee anatomy that can be interpreted by a trained physician.

## Device Story

Contour Knee is a receive-only, 12-channel phased array RF coil for MRI. It features a flexible blanket-like enclosure for anterior elements and a rigid plastic housing for posterior elements. Used in clinical settings with Siemens 0.55T MR systems, it is operated by trained MRI technicians or physicians. The device captures RF signals from knee anatomy, which are processed by the MR system to generate diagnostic images. These images are reviewed by physicians to aid in clinical decision-making and diagnosis of knee conditions. The device benefits patients by providing high-quality diagnostic imaging of the knee.

## Clinical Evidence

Bench testing included SNR and image uniformity analysis per NEMA MS-9 standards, meeting all acceptance criteria. Clinical evidence consisted of volunteer scanning of knee anatomy on a Siemens 0.55T MR system to demonstrate diagnostic image quality. No adverse events were reported.

## Technological Characteristics

Receive-only 12-channel phased array RF coil. Materials: Polycarbonate, aramid felt, polyurethane-coated nylon fabric. Features active PIN diode switching and passive blocking circuitry. Hybrid design: flexible anterior blanket, rigid posterior housing. Compatible with Siemens 0.55T MR systems. Tested per NEMA MS-9 and IEC standards for electrical safety and electromagnetic compatibility.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Contour 24 ([K183111](/device/K183111.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, and the full name of the agency, "U.S. Food & Drug Administration," is in a smaller font size below the acronym.

January 13, 2023

Quality Electrodynamics, LLC Eric Yeh Senior Regulatory Affairs Specialist 6655 Beta Drive Suite 100 Mayfield Village, Ohio 44143

Re: K223429

Trade/Device Name: Contour Knee Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: November 11, 2022 Received: November 14, 2022

Dear Eric Yeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

D. G. K.

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K223429

Device Name

Contour Knee

Indications for Use (Describe)

The Contour Knee is intended for use with Siemens 0.55T MR systems to produce diagnostic images of knee anatomy that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| K223429                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                    | 510(k) Summary | Prepared on: 2023-01-12                      |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|----------------|----------------------------------------------|
| Contact Details                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                    |                | <div align="right">21 CFR 807.92(a)(1)</div> |
| Applicant Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Quality Electrodynamics, LLC                                       |                |                                              |
| Applicant Address                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 6655 Beta Drive, Suite 100 Mayfield Village OH 44143 United States |                |                                              |
| Applicant Contact Telephone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 440-484-2940                                                       |                |                                              |
| Applicant Contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Mr. Eric Yeh                                                       |                |                                              |
| Applicant Contact Email                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | eric.yeh@qualedyn.com                                              |                |                                              |
| Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                    |                | <div align="right">21 CFR 807.92(a)(2)</div> |
| Device Trade Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Contour Knee                                                       |                |                                              |
| Common Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Magnetic resonance diagnostic device                               |                |                                              |
| Classification Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Coil, Magnetic Resonance, Specialty                                |                |                                              |
| Regulation Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 892.1000                                                           |                |                                              |
| Product Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | MOS                                                                |                |                                              |
| Legally Marketed Predicate Devices <div align="right">21 CFR 807.92(a)(3)</div>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                    |                |                                              |
| Predicate #                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Trade Name (Primary Predicate is listed first)           | Product Code   |                                              |
| K183111                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Contour 24                                                         | MOS            |                                              |
| Device Description Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | <div align="right">21 CFR 807.92(a)(4)</div>                       |                |                                              |
| The Contour Knee is a receive-only, 12-channel phased array coil designed for magnetic resonance imaging (MRI) using the Siemens             0.55T MR systems. The Contour Knee is intended to be used for imaging knee anatomy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                    |                |                                              |
| Intended Use/Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | <div align="right">21 CFR 807.92(a)(5)</div>                       |                |                                              |
| The Contour Knee is intended for use with Siemens 0.55T MR systems to produce diagnostic images of knee anatomy that can be             interpreted by a trained physician.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                    |                |                                              |
| Indications for Use Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | <div align="right">21 CFR 807.92(a)(5)</div>                       |                |                                              |
| The Indications for Use statement for the Contour Knee is not identical to that of the predicate device (Contour 24); however, the             differences do not affect the safety or effectiveness of the device relative to the predicate device. Both Indications for Use statements for             the proposed Contour Knee and predicate Contour 24 indicate that the device is intended to be used in conjunction with a MR system to             produce images of human anatomy and that the images can be interpreted by a trained physician. The indications for use statements             differ only in that the proposed Contour Knee is intended to be used with 0.55T Siemens MR systems instead of 1.5T Siemens MR             systems, and intended for use to image knee anatomy instead of general human anatomy. |                                                                    |                |                                              |
| Technological Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | <div align="right">21 CFR 807.92(a)(6)</div>                       |                |                                              |
| At a high level, the proposed and predicate device are based on the following same technological elements: <ul><li>Receive-only phased array RF coil</li><li>Active PIN diode switching blocking circuitry. Passive blocking circuitry.</li></ul>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                    |                |                                              |

{4}------------------------------------------------

· Materials used for flame retardancy and biocompatibility: Polycarbonate and aramid felt with a polyurethane coated nylon fabric cover

The following technological differences exist between the proposed and predicate device:

· Number of channels (12 (proposed device) versus 24 (predicate device))

· Intended for use (knee anatomy (proposed device) versus general human anatomy (predicate device))

· Flexible blanket-like enclosure for anterior coil elements, rigid plastic housing for posterior coil elements (proposed device) versus flexible blanket-like enclosure (predicate device))

· Compatible with Siemens 0.55T MR systems (proposed device) versus compatible with Siemens (predicate device)

## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The signal-to-noise ratio (SNR) and image uniformity of the Contour Knee were measured on a 0.55T Siemens MR System, manufactured by Siemens Healthineers. The SNR and uniformity of the Contour Knee were analyzed per NEMA MS-9 (using alternate method 2.5 from MS-6) and uniformity was analyzed using NEMA MS-9 (primary method from MS-6) and was found to conform to predetermined acceptance criteria.

In accordance with the FDA Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, clinical images from volunteer scanning of knee anatomy were obtained from a Siemens 0.55T MR system. These images were used to demonstrate that the Contour Knee produces diagnostic quality images of the intended anatomy. No adverse events were reported or recorded.

The electrical safety and electromagnetic compatibility data support the safety of the Contour Knee and the bench testing per the IEC standards and diagnostic quality sample clinical images demonstrates the performance and the device under the specified use conditions. This testing demonstrates that the Contour Knee performs as well as or better than the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K223429](https://fda-staging.innolitics.com/device/K223429)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).