Models: uLumb-9530A, uLumb-9531A, uLumb-9532A, uLumb-9533A, LQ-9525B: Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of neck, arm, and leg, due to strain from exercise or normal household and work activities. Note1: The electrode on the stimulator is used in the waist. Note2: The additional gel patch is used in shoulder, waist, back, back of neck, arm, and leg, Heating: To be used for temporary relief of minor aches and pains. Models: uNeck-9512A, uNeck-9515A, uNeck-9517B, uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck-9529A, LQ-9535A: Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the back of neck, due to strain from exercise or normal household and work activities. Heating: To be used for temporary relief of minor aches and pains.
Device Story
Portable, battery-powered TENS device; provides electrical pulse stimulation and heating. Inputs: user-selected stimulation modes/levels and heat settings via controller or wireless remote. Operation: device controller generates symmetric biphasic pulsed electrical waveforms and heat (max 43°C) through built-in metal electrodes or external gel electrodes. Used in home/clinic settings by patients for muscle pain relief. Output: electrical stimulation to muscles/nerves; heat for comfort. Healthcare providers/patients use output to manage pain; device provides temporary relief of sore/aching muscles. Includes rechargeable battery, adapter, and gel electrodes.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility (ISO 10993-5, -10), electrical safety/EMC (IEC 60601-1, -1-2, -1-11, -2-10), software verification/validation, and performance verification (waveform/output testing, electrode shelf-life per ASTM F1980-07).
Technological Characteristics
Materials: ABS/PC plastic enclosure, metal electrodes, conductive hydrogel. Energy: 7.4V/3.7V rechargeable battery. Waveform: Pulsed, symmetric, biphasic (rectangular). Connectivity: Wireless remote control. Sterilization: Not specified. Software: Microprocessor-controlled, moderate level of concern. Standards: IEC 60601 series, ISO 10993.
Indications for Use
Indicated for temporary relief of pain in shoulder, waist, back, neck, arm, and leg muscles due to exercise or household/work strain; also indicated for temporary relief of minor aches and pains via heating function. For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Health Expert Electronic Stimulator/AST-300L (K190783)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 8, 2023
Shenzhen Leqing Medical Instrument Co., Ltd % Rain Yip Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm2401, ZhenYe International Center, No.3101-90 Qianhai Road Nanshan District Shenzhen, Guangdong 518000 China
Re: K223428
Trade/Device Name: Electronic Stimulator (Models: uLumb-9530A, uLumb-9531A, uLumb-9532A, uLumb-9533A. LO-9525B. uNeck-9512A. uNeck-9515A. uNeck-9517B. uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck-9529A, LQ-9535A) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, IRT Dated: January 17, 2023 Received: April 12, 2023
Dear Rain Yip:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Amber T. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223428
#### Device Name
Electronic Stimulator (Models: uLumb-9530A, uLumb-9532A, uLumb-9533A, LQ-9525B, uNeck-9512A, uNeck-9515A, uNeck-9517B, uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck-9529A, LQ-9535A)
#### Indications for Use (Describe)
Models: uLumb-9530A、uLumb-9531A、uLumb-9532A、uLumb-9533A、LQ-9525B:
Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of neck, arm, and leg, due to strain from exercise or normal household and work activities.
Notel : The electrode on the stimulator is used in the waist.
Note2: The additional gel patch is used in shoulder, waist, back, back of neck, arm, and leg,
Heating: To be used for temporary relief of minor aches and pains.
Models: uNeck-9512A, uNeck-9517B, uNeck-9517B, uNeck-9519A, uNeck-9521A, uNeck-9529A, LO-9535 A: Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the back of neck, due to strain from exercise or normal household and work activities. Heating: To be used for temporary relief of minor aches and pains.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# "510(k) Summary" as required by 21 CFR Part 807.92.
# Date: May 8, 2023
#### I. Submitter
Shenzhen Leqing Medical Instrument Co.,LTD 2-3/F, Building D, No.31 Fangkeng Road, Longgang, Shenzhen, Guangdong, China Post code: 518000 Tel .: +86 755 2839 2373
Shaodong Wang Management Representative Tel: +86 135 9035 9865 Email: wangsd@pentasmart.com.cn
## II. Device
Trade Name: Electronic Stimulator (Models: uLumb-9530A, uLumb-9531A, uLumb-9532A, uLumb-9533A, LQ-9525B, uNeck-9512A, uNeck-9517B, uNeck-9517B, uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck- 9529A, LQ-9535A) Common Name: Transcutaneous electrical nerve stimulator Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Regulatory Class: II Product Code: NUH. IRT Regulation Number: 21 CFR 882.5890
## III. Predicate Device
Primary Predicate device: 510(k) number: K211942 Manufacturer: Shenzhen OSTO Technology Company Limited Trade name: Heating Lumbar Therapy Cushion/AST-622 Product code: NGX, NUH, IRT Approval date: December 13, 2021
Secondary predicate device: 510(k) number: K190783 Manufacturer: Shenzhen OSTO Technology Company Limited Trade name: Health Expert Electronic Stimulator/AST-300L Product code: NUH, NGX Approval date: March 14, 2020
{4}------------------------------------------------
## IV. Device Description
The Electronic Stimulator is a portable and battery-powered TENS device with multiple models, offering both electrical pulse stimulation and heating function in one device.
The device consists of a controller that has built-in metal electrodes and accessories of a remote control, an adapter, and/or a pair of gel electrodes, an external connection coupling. The external connection coupling is used to connect the gel electrodes to the device. All models are powered by internal rechargeable battery and charged by the adapter. And all accessories, including the remote control, adapter, gel electrodes, external connection coupling can only be changed or replaced by a qualified person.
The functions of the device are controlled by the device controller and the remote control, where the remote control establishes a connection with the device controller through wireless signals. The device has five stimulation modes and one heating mode, through the metal electrodes to provide users with waist or nape stimulation and heat generation to warm and comfortable waist or nape muscles. And the heating function is controlled by a heat/temperature key and can only be run on metal electrodes, it is cannot be operated alone. The heating temperature does not exceed 43 degrees Celsius at the same time. In addition, the device is also equipped with a pair of gel electrodes, which can be used to stimulate different parts of the body (shoulder, waist, back, back of neck (nape), arm, and leg).
The device has 13 models: uLumb-9530A, uLumb-9531A, uLumb-9532A, uLumb-9533A, LQ-9525B, uNeck-9512A, uNeck-9515A, uNeck-9517B, uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck-9529A, LQ-9535A, the 13 models differ in appearance, size, battery capacity, number of keys on the stimulator, the specification of electrodes equipped, and some output parameters, the details can refer to the "Model list form" of the Device Description file in this application.
## V. Indications for Use
Models: uLumb-9530A, uLumb-9531A, uLumb-9532A, uLumb-9533A, LQ-9525B: Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of neck, arm, and leg, due to strain from exercise or normal household and work activities.
Note1: The electrode on the stimulator is used in the waist.
Note2: The additional gel patch is used in shoulder, waist, back, back of neck, arm, and leg. Heating: To be used for temporary relief of minor aches and pains.
Model): uNeck-9512A, uNeck-9515A, uNeck-9517B, uNeck-9518A, uNeck-9519A, uNeck-9521A, uNeck-9529A, LQ-9535A:
Electrical simulation: To be used for temporary relief of pain associated with sore and aching muscles in the back of neck, due to strain from exercise or normal household and work activities. Heating: To be used for temporary relief of minor aches and pains.
## VI. Comparison of Technological Characteristics With the Predicate Device
{5}------------------------------------------------
Compare with predicate devices, the Electronic Stimulator is very similar in the same technical characteristics, features, specifications, materials, mode of operation, intended use, and the applicable standards. The differences between subject device and the predicate devices do not raise new questions of safety or effectiveness.
{6}------------------------------------------------
| Comparison items | Subject device | Secondary Predicate devices | Primary predicate device | | |
|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 510(k) number | Pending | K190783 | K211942 | | |
| Device name | Electronic Stimulator | Health Expert Electronic Stimulator | Heating Lumbar Therapy Cushion | | |
| Model(s) | uLumb-9530A,<br>uLumb-9531A,<br>uLumb-9532A,<br>uLumb-9533A, LQ-<br>9525B | uNeck-9512A,<br>uNeck-9515A,<br>uNeck-9517B,<br>uNeck-9518A,<br>uNeck-9519A,<br>uNeck-9521A,<br>uNeck-9529A, LQ-<br>9535A | AST-300L | AST-622 | |
| Manufacturer | Shenzhen Leqing Co.,LTD | Medical Instrument | Shenzhen OSTO Technology Company Limited | Shenzhen OSTO Technology Company Limited | |
| Indication for<br>use/Intended use | Electrical simulation:<br>To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist, back,<br>back of neck, arm, and<br>leg, due to strain from<br>exercise or normal<br>household and work<br>activities.<br>Notel: The electrode<br>on the stimulator is<br>used in the waist.<br>Note2: The additional<br>gel patch is used in<br>shoulder, waist, back,<br>back of neck, arm | Electrical simulation:<br>To be used<br>for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the back of<br>neck, due to strain<br>from exercise or<br>normal household and<br>work activities.<br>Heating: To be used<br>for temporary relief of<br>minor aches and<br>pains. | PMS (Mode 1~8)<br>It is intended to stimulate healthy<br>muscles in order to improve and<br>facilitate muscle performance.<br>TENS (Mode 9~25)<br>To be used for temporary relief of pain<br>associated with sore and aching<br>muscles in the shoulder, waist, back,<br>back of the neck, arm, leg, and foot due<br>to strain from exercise or normal<br>household work activities by applying<br>current to stimulate nerve. | PMS(1 ~ 2): It is intended to stimulate<br>healthy muscles in order to improve and<br>facilitate muscle performance.<br>TENS(3 ~ 5): To be used for temporary<br>relief of pain associated with sore and<br>aching muscles in the shoulder, waist,<br>back, arm and leg due to strain from<br>exercise or normal household work<br>activities by applying current to<br>stimulate nerve.<br>Heating: The Heating Lumbar Therapy<br>Cushion is intended for temporary relief<br>of minor aches and pains. | |
| Comparison items | Subject device | Secondary Predicate devices | Primary predicate device | | |
| | and leg.<br>Heating: To be used<br>for temporary relief of<br>minor aches and<br>pains. | | | | |
| Location for use | OTC | OTC | OTC | | |
| Power source(s) | Powered by the<br>internal rechargeable<br>battery, 7.4Vd.c.,<br>2600mAh | Powered by the<br>internal rechargeable<br>battery, 3.7Vd.c.,<br>700mAh (uNeck-<br>9512A, uNeck-<br>9515A, uNeck-<br>9517B, uNeck-<br>9518A, uNeck-<br>9519A)<br><br>Powered by the<br>internal rechargeable<br>battery, 3.7Vd.c.,<br>$280mAh*2$ (uNeck-<br>9521A)<br><br>Powered by the<br>internal rechargeable<br>battery, 3.7Vd.c.,<br>1200mAh (uNeck-<br>9529A, LQ-9535A) | Adaptor Input: 100-240Vac, 50-60Hz,<br>0.1A, Output: 5Vdc, 2A<br>Unit Input: 5Vdc, 2A | Power by adapter, 100-240VAC, 50-60Hz, Output: 5Vdc, 2A<br>Unit Input: 5Vdc, 2A | |
| Charged by the<br>external adapter: | Charged by the<br>external adapter: | | | | |
| Comparison items | Subject device | | Secondary Predicate devices | Primary predicate device | |
| | Input: 100-240Va.c.,<br>50/60Hz, 0.4A Max.<br>Output: 5.0Vd.c.,<br>2.0A | Input: 100-240Va.c.,<br>50/60Hz, 0.4A Max.<br>Output: 5.0Vd.c.,<br>1.0A | | | |
| -Method of Line<br>Current Isolation | Not applicable | | Type BF Applied Part | Type BF Applied Part | |
| -Patient<br>Leakage<br>Current | NC<br>(μA) | <50μA | AC: 54.5μA<br>DC: 0.5μA | AC: 54.5μA<br>DC: 0μA | |
| | SFC<br>(μA) | <100μA | AC: 120μA<br>DC: 0.6μA | AC: 13.4μA<br>DC: 0μA | |
| Average DC current<br>through electrodes<br>when device is on but<br>no pulses are<br>being applied (μA) | 0μA | | <0.01μA | Not publicly available | |
| Function and design | Electrical stimulation and heat | | Electrical stimulation and heat | Electrical stimulation and heat | |
| Function output<br>patterns | Electrical stimulation only<br>Electrical stimulation + Heat simultaneously | | Electrical stimulation only<br>Alternating combinations of Electrical<br>stimulation and Heat | Electrical stimulation only<br>Heat only<br>Electrical stimulation + Heat<br>simultaneously | |
| Heating setting | (+/-2°C)<br>Level LOW: 39°C<br>Level HIGH: 41°C | (+/-2°C)<br>Level LOW: 39°C<br>Level HIGH: 41°C | Adjustable<br>Heating temperature is from 30~40℃ | Level 1: 37°C<br>Level 2: 41°C<br>Level 3: 43°C | |
| Maximum<br>temperature setting | 43°C | | 40°C | 43°C | |
| Number of Output<br>Modes | Electrical simulation: 5 modes<br>Heat: 1 mode | | 25 modes | For stimulation: 5 modes<br>For heating: 1 mode | |
| Number of Output<br>Levels | Electrical simulation:<br>60 levels<br>Heat: 2 levels | Electrical simulation:<br>30 levels<br>Heat: 2 levels | 99 levels | For stimulation: 50 levels<br>For heating: 3 levels | |
| Number of Output | 2 channels | 1 channel | 2 channels | 1 channel | |
| Comparison items | Subject device | Secondary Predicate devices | Primary predicate device | | |
| Channels | | | | | |
| Number<br>of Output<br>Channels | Synchronous or<br>Alternating? | Alternating | Synchronous | Synchronous | |
| | Method of<br>Channel<br>Isolation | Potential transformer,<br>RELAY, Software<br>isolation | Voltage Transform<br>Isolation<br>"Body+" and "Body-" buttons for<br>body channel, "Sole+" and "Sole-"<br>buttons for feet channel | Not applicable | |
| Regulated Current or<br>Regulated Voltage? | | Regulated Voltage | Voltage Control | Regulated Voltage | |
| Software/Firmware/<br>Microprocessor<br>Control? | | Yes | Yes | Yes | |
| Automatic Overload<br>Trip? | | No | No | No | |
| Automatic No-Load<br>Trip? | | No | No | No | |
| Automatic Shut Off? | | Yes | Yes | Yes | |
| User<br>Control? | Override | No | Yes | Yes | |
| | On/Off<br>Status? | Yes | Yes | Yes | |
| Indicator<br>Display | Low<br>Battery? | Yes | No | No…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.