MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium. surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults.
Device Story
System processes 3D CT images to create patient-specific anatomical models, surgical guides, and planning reports for maxillofacial reconstructive surgery; involves harvesting bone from fibula or ilium. Operated by trained MedCAD employees using COTS software; physician provides input/feedback during planning. Outputs produced via additive manufacturing. Used in clinical settings to assist surgeons in pre-operative simulation and surgical execution. Benefits include patient-specific surgical planning and guidance, potentially improving surgical accuracy and outcomes.
Clinical Evidence
Bench testing only. Wear debris testing (ASTM F1877-16) on titanium guides confirmed safety. Fit and form validation performed using worst-case CT inputs and anatomical models; all samples met acceptance criteria.
Technological Characteristics
Materials: Biocompatible polymers and Ti-6Al-4V ELI Titanium Alloy (ASTM F136). Additive manufacturing process. Steam sterilization (SAL 10^-6). COTS software used for design. Connectivity: Standalone system for image processing and additive manufacturing.
Indications for Use
Indicated for adolescents (>12 to 21 years) and adults requiring maxillofacial reconstructive surgery involving bone grafts from the fibula or ilium.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
CenterMed Patient Matched Assisted Surgical Planning (ASP) System (K211614)
{0}------------------------------------------------
September 20, 2023
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
MedCAD % Justin Gracyalny Regulatory Affairs Manager Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock. Georgia 30188
Re: K223421
Trade/Device Name: MedCAD® AccuPlan® Orthopedics System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: PBF, LLZ Dated: August 18, 2023 Received: August 18, 2023
Dear Mr. Gracyalny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223421
Device Name MedCAD® AccuPlan® Orthopedics System
#### Indications for Use (Describe)
MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium. surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K223421 510(k) Summary MedCAD® AccuPlan® Orthopedics System
### September 18, 2023
#### Sponsor
MedCAD 501 S 2nd Ave, Suite A-1000 Dallas, TX 75226 (214) 453-8864 x305
## Contacts
Secure BioMed Evaluations Justin Gracyalny, MSE Linda Braddon, Ph.D. 7828 Hickory Flat Highway, Suite 120 Woodstock, GA 30188 770-837-2681 Regulatory(@SecureBME.com
### Name of Device and Classification Name
Device Name: MedCAD® AccuPlan® Orthopedics System Common Device: Orthopaedic Surgical Planning And Instrument Guide Regulation Name and Number: 21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories, 21 CFR 892.2050 Medical image management and processing system Product Code: PBF, LLZ Classification Panel: Orthopedic
### Predicate Device CenterMed Patient Matched Assisted Surgical Planning (ASP) System (K211614)
Reference Devices MedCAD® AccuPlan® System (K192282 / K223024)
### Indications for Use:
MedCAD® AccuPlan® Orthopedics System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the system and the result is an output data file. This file may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models of the fibula and ilium, surgical guides for harvesting bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults.
{4}------------------------------------------------
# K223421
## Device Description
The MedCAD® AccuPlan® Orthopedics System is a collection of software and associated additive manufacturing equipment intended to provide a variety of outputs to support harvesting of bone to support maxillofacial reconstructive surgeries. The system uses electronic medical images of the patient's anatomy with input from the physician to manipulate original patient images for planning and executing surgery. The patient specific outputs from the system include anatomical models, surgical guides, and patient-specific case reports.
Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical Computed Tomography (CT) images to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during and prior to delivery of the final outputs. While the process and dataflow vary somewhat based on the requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how they interact to produce the defined system outputs. It should be noted that the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.
The MedCAD® AccuPlan® Orthopedics System is made up of two individual pieces of software for the design and various manufacturing equipment integrated to provide a range of anatomical models (physical and digital), surgical guides, and patient-specific planning reports for harvesting of bone from the fibula and ilium for use in maxillofacial reconstructive surgeries.
The MedCAD® AccuPlan® Orthopedics System requires an input 3-D image file from medical imaging systems (i.e., CT). This input is then used, with support from the prescribing physician to provide the following potential outputs to support maxillofacial reconstructive surgery. Each system output is designed with physician inout and reviewed by the physician prior to finalization. All outputs are used only with direct physician involvement to reduce the criticality of the outputs.
System outputs include:
- Anatomical Models ●
- Surgical Guides
- Patient-Specific Case Reports ●
# Sterilization Validation
Sterilization validation was conducted in accordance with international standard ISO 17665 and FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling." to a sterility Assurance Level (SAL) of 1x10 . All test method acceptance criteria were met.
# Biocompatibility Validation
Biocompatibility validation was conducted in accordance with international standard ISO 10993-1 and FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The results of the testing adequately address biocompatibility for the output devices and their intended use.
{5}------------------------------------------------
# K223421
# Performance Testing
Performance testing for the MedCAD® AccuPlan® Orthopedics System is summarized in the table below:
| Test | Test Method Summary | Results |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wear Debris<br>Testing | Cutting / drilling instruments were used on a<br>worst-case titanium surgical guide. The quantity<br>of wear debris generation was characterized.<br>Wear particles were characterized per ASTM<br>F1877-16. | PASS<br>The quantity and<br>morphology of wear<br>debris generated by the<br>subject device under<br>worst case use<br>conditions aligns with<br>values as reported in the<br>literature to be safe. |
| Fit and Form<br>Validation | Subject devices were manufactured from worst-<br>case CT scan input data. The manufactured<br>devices were optically scanned to verify<br>alignment with the 3D model. Evaluation of fit<br>was also validated by fitting the guide over the<br>corresponding defect in a representative<br>anatomical model. | PASS<br>All samples met the<br>predetermined<br>acceptance criteria. |
# Substantial Equivalence
MedCAD® AccuPlan® Orthopedics System is substantially equivalent to the identified predicate based on indications for use, principles of operation, technological characteristics, inputs, and outputs.
| Property | MedCAD® AccuPlan®<br>Orthopedics System | CenterMed Patient Matched<br>Assisted Surgical Planning (ASP)<br>System |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K223421 | K211614 |
| Indications for Use | MedCAD® AccuPlan® Orthopedics<br>System is intended for use as a<br>software system and image<br>segmentation system for the transfer<br>of imaging information from a<br>medical scanner such as a CT based<br>system. The input data file is<br>processed by the system and the<br>result is an output data file. This file<br>may then be provided as digital<br>models or used as input to a rapid<br>prototyping portion of the system that<br>produces physical outputs including<br>anatomical models of the fibula and<br>ilium, surgical guides for harvesting | CenterMed Patient Matched Assisted<br>Surgical Planning (ASP) System is<br>intended for use as a software system<br>and image segmentation system for<br>the transfer of imaging information<br>from a medical scanner such as a CT<br>based system. The input data file is<br>processed by the ASP system and the<br>result is an output data file. This file<br>may then be provided as digital<br>models or used as input to a rapid<br>prototyping portion of the system that<br>produces physical outputs including<br>anatomical models of the Fibula and<br>Ilium, surgical guides for harvesting |
| Property | MedCAD® AccuPlan®<br>Orthopedics System | CenterMed Patient Matched<br>Assisted Surgical Planning (ASP)<br>System |
| | bone grafts from the fibula or ilium, and surgical planning case reports for use in maxillofacial reconstructive surgeries. MedCAD® AccuPlan® Orthopedics System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. The MedCAD® AccuPlan® Orthopedics System is indicated for use in adolescents (greater than 12 to 21 years of age) and adults. | bone grafts from the Fibula or Ilium, and surgical planning case reports for use in Maxillofacial reconstructive surgeries. CenterMed Patient Matched ASP System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. |
| Preoperative software | Yes | Yes |
| Additive<br>manufacturing of<br>surgical guides and<br>models | Yes | Yes |
| Data inputs | Images from medical scanners | Images from medical scanners |
| Data outputs | Output for patient-specific surgical<br>guides and anatomical models | Output for patient-specific surgical<br>guides and anatomical models |
| Physical outputs | Patient-specific surgical guides,<br>anatomical models, and patient-<br>specific case reports | Patient-specific surgical guides,<br>anatomical models, and patient-<br>specific case reports |
| Materials | Biocompatible polymers and<br>Ti-6Al-4V ELI Titanium Alloy per<br>ASTM F136 | Polyamide-12 (PA-12) with 316L<br>stainless steel sleeves when used as<br>cutting guides |
| Recommended<br>Temporary Screw<br>Diameter | 1.5mm – 2.1mm | 1.5mm – 2.1mm |
| Recommended<br>Temporary Screw<br>Length | 5mm – 22mm | 5mm – 22mm |
| Recommended<br>Temporary Screw<br>Style | Non-Locking | Drill-Free, Tapping-Free |
| Sterilization | Provided non-sterile and is steam<br>sterilized by the end-user (SAL 10⁻⁶) | Provided non-sterile and is steam<br>sterilized by the end-user (SAL 10⁻⁶) |
| Manufacturing<br>Method | Additive Manufacturing | Additive Manufacturing |
| Patient Contact | Implant – Tissue / Bone; Limited (<<br>24 hours) | Implant – Tissue / Bone; Limited (<<br>24 hours) |
| Patient Population | Adults and Pediatrics (12+) | Adults |
{6}------------------------------------------------
{7}------------------------------------------------
# K223421
# Comparison of Technological Characteristics with the Predicate Device
MedCAD® AccuPlan® Orthopedics System is substantially equivalent to its predicate device (K211614 CenterMed Patient Matched Assisted Surgical Planning (ASP) System).
## Similarities to Predicate
The MedCAD® AccuPlan® Orthopedics System has the same intended use and similar technological characteristics as the identified predicate device. The system employs similar fundamental technologies as the identified predicate including software image transfer, manipulation, and surgical planning. The principles of operation and technological characteristics are either identical or substantially equivalent to the predicate. The system has similar technological characteristics including:
- System inputs: Images from medical scanner systems (CT)
- System outputs: Additively manufactured physical and/ or digital outputs such as patientspecific anatomical models and surgical guides for use in harvesting bone to support maxillofacial reconstructive surgeries
- Software: Use of commercial off-the-shelf software to support the design phase of the device. ●
- Sterility assurance level of 1x10-6 ●
The intended use of the subject device and the predicate both provide tools and accessories (software for image manipulation, anatomical models, and guides) for use in maxillofacial reconstructive surgery. Additionally, both the subject and the predicate device are intended to be used by trained personnel with active support from the surgeon.
## Differences to Predicate
The only differences between the subject and predicate device relate to the addition of the 12+ pediatric population for the subject device and minor geometrical differences. A pediativ risk analysis was performed to support the change in patient population. Performance testing demonstrates that the minor material and geometrical differences do not raise new questions for safety and effectiveness.
### Conclusion
The MedCAD® AccuPlan® Orthopedics System is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.