← Product Code [IYN](/productcode/IYN) · K223407

# Versana Essential (K223407)

_GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Feb 15, 2023 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223407

## Device Facts

- **Applicant:** GE Medical Systems Ultrasound and Primary Care Diagnostics
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Feb 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Versana Essential is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Essential clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, Vascular/Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Thoracic/Pleural, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD, B Flow/B Flow Color. Versana Essential is intended to be used in a hospital or medical clinic.

## Device Story

Versana Essential is a mobile console diagnostic ultrasound system for use in hospitals or clinics by trained healthcare professionals. It utilizes various transducers (convex, linear, sector, mechanical 4D) to acquire ultrasound signals from the human body. The system processes these signals to generate B-mode, M-mode, Doppler, 3D/4D, and elastography images. It features an operator control panel with a trackball and keyboard for image manipulation and analysis. The device supports external ECG gating/triggering and connectivity via USB, LAN, WiFi, or HDMI for data export and printing. Software tools assist in clinical productivity, including automated measurements (e.g., Auto Bladder, Auto EF) and clinical reporting (e.g., LI-RADS, BI-RADS, TI-RADS). The system provides real-time visualization of anatomy and interventional guidance, aiding clinicians in diagnostic assessment and procedural accuracy. It is powered by AC or an internal battery.

## Clinical Evidence

No clinical data was required to support substantial equivalence. The device evaluation relied on bench testing, including acoustic output, biocompatibility, thermal, electrical, electromagnetic, and mechanical safety testing, as well as system verification and validation.

## Technological Characteristics

Mobile console ultrasound system with 48 scan channels. Supports convex, linear, sector, and mechanical 4D transducers. Imaging modes: B, M, PW/CW/Color Doppler, Power Doppler, Harmonic, Coded Pulse, 3D/4D, Elastography, B-Flow. Connectivity: USB, LAN, WiFi, HDMI. Power: AC or internal battery. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, IEC 62359, ISO 14971, and DICOM standards.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Versana Balance ([K220446](/device/K220446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 15, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Lee Bush Regulatory Affairs Director GE Healthcare 9900 W. Innovation Drive WAUWATOSA WI 53226

Re: K223407

Trade/Device Name: Versana Essential Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 19, 2023 Received: January 19, 2023

Dear Lee Bush:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplets around the edge of the circle. The logo is simple and recognizable.

Expiration Date: 06/30/2023

See PRA Statement below.

Form Approved: OMB No. 0910-0120

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K223407

Device Name

Versana Essential

Indications for Use (Describe)

The Versana Essential is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Essential clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, VascularPeripheral Vascular, Musculoskeletal Conventional, Musculosketal Superficial, Thoracic/Pleural, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access).

Modes of operation include: B. M. PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD, B Flow/B Flow Color.

Versana Essential is intended to be used in a hospital or medical clinic.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are small white water droplet shapes around the inside of the circle. 

# K223407

# GE Healthcare 510(k) Premarket Notification Submission

## 510(k) Summary

| Date:                     | Jan. 19, 2022                                                                                                                                                                                    |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                | GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226                                                                                       |
| Primary Contact Person:   | Lee Bush<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)309-9429                                                                                                                      |
| Secondary Contact Person: | Jian Xie<br>Regulatory Affairs Leader<br>GE Healthcare                                                                                                                                           |
| Device Trade Name:        | Versana Essential                                                                                                                                                                                |
| Common/Usual Name:        | Diagnostic Ultrasound System                                                                                                                                                                     |
| Classification Names:     | Class II                                                                                                                                                                                         |
| Product Code(s):          | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | Versana Balance (K220446), Diagnostic Ultrasound System                                                                                                                                          |
| Classification Names:     | Class II                                                                                                                                                                                         |
| Product Code(s):          | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular border. The border has four droplet-like shapes extending outwards at the cardinal directions, giving it a dynamic and fluid appearance. The logo is presented in a blue color.

### Device Description:

The Versana Essential is a general purpose, Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals. The system is a mobile console that includes an operator control panel, display monitor and transducers.

The console provides digital acquisition, processing and display capability. The system has an internal battery to allow for acquisition while the system is not plugged into a power source. Acquisition can also be done while the system is connected to an AC power source.

The operator control panel includes function keys, trackball, an alfa-numeric keyboard.

The variety of transducers include convex, linear, sector, and mechanical 4D transducers. The access types include trans- body surface, transrectal, transvaginal and transcranial.

Data can be imported or exported by DVD, USB, LAN or WiFi if the USB wireless adapter is connected to the system. An external ECG module has been verified to use as input for gating/triggering during scanning. The system has a HDMI port. The Video adapter is optional part which used to extend the S-Video/Composite Out interface by connecting with HDMI port. The system has a DC outlet to allow connection of a DC printer which sits in the printer box of the console and an option for external Printer USB Isolator for other commercial printers to connect.

The system allows for electronic instructions for use and software updates to be download from a GE Healthcare website.

#### Intended Use/Indication for Use:

Versana Essential and predicate device Versana Balance (K220446) have the same intended use:

The Versana Essential is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Essential clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, Vascular/Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Thoracic/Pleura, Transcranial, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access).

Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD, B Flow/B Flow Color.

Versana Essential is intended to be used in a hospital or medical clinic.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle is surrounded by three white, curved shapes that resemble water droplets or stylized flames.

# GE Healthcare

510(k) Premarket Notification Submission

## Technology:

The Versana Essential employs the same fundamental scientific technology as its predicate device Versana Balance device (K220446).

## Determination of Substantial Equivalence:

The proposed Versana Essential system is substantially equivalent to the predicate device Versana Balance with regard to intended use, imaging capabilities, technological characteristics, imaging modes, hardware, and safety effectiveness.

## Transducers and Modes:

- The proposed Versana Essential transducers are same as predicate Versana Balance (K220446) systems . transducers
	- l 4C-RS
	- E8C-RS
	- 35c-RS
	- L6-12-RS
	- ર્દ-૪૮
	- 8C-KS
	- I E8CS-RS
	- RAB2-6-RS I
	- LK760-RS
- Versana Essential and predicate device Versana Balance (K220446) have the same imaging modes ●

## Software:

The proposed Versana Essential has the same features and functionality as the predicate device Versana Balance (K220446). The design of all features is same as predicate device Versana Balance, the workflow is same as predicate device Versana Balance non-touchscreen version.

The key feature list as below:

- LI-RADS:
- Probe Check (transducer element check) ●
- Whizz Color Flow ●
- Lateral Gain Compensation (LGC) ●
- VOCAL ●
- V-Live
- Standby mode (Fast boot up)
- Follow up Tool
- Breast Productivity (BI-RADS) ●
- Thyroid productivity (TI-RADS)
- Auto Bladder
- . B-Flow&B-Flow Color

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplet shapes around the circle. The logo is simple and recognizable.

- Breast Care ●
- . Needle recognition
- Auto EF
- TUI ●
- Stress Echo ●
- . Advance 3D
- Tricefy
- . Elastography imaging
- e-Delivery

Note: The predicate device Versana Balance (K220446) has two additional features, whizz label and multi-touch, that are not available in the proposed Versana Essential.

## Hardware:

The proposed Versana Essential and predicate Versana Balance (K2220446) non-touchscreen version have the same power input/output and similar design of internal circuits. Both of them have been designed in compliance with approved electrical and physical safety standards.

The proposed Versana Essential and predicate Versana Balance (K2220446) support the same accessories such printers, DVD, USB disk and barcode reader.

| Characteristic        | Proposed Device<br>Versana Essential                                                                             | Predicate Device<br>Versana Balance<br>(K220446)                                                                        | Comment on comparison                                                                                                                                                                                                                                      |
|-----------------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Scan channels         | 48 Channels                                                                                                      | 64 Channels                                                                                                             | The number of channels on Versana<br>Essential is less than on predicate<br>Versana Balance, however all safety and<br>performance requirements are met<br>according to the device evaluation and<br>verification. No impact to safety or<br>effectiveness |
| Touch panel           | Not available                                                                                                    | Option<br>(Configuration<br>dependent)                                                                                  | Identical with predicate device Versana<br>Balance non-touchscreen version.                                                                                                                                                                                |
| Network /<br>Archive: | USB, thermal b&w<br>image printer,<br>memory stick and<br>Encrypt patient<br>information store to<br>a removable | USB, thermal b&w<br>image printer,<br>memory stick and<br>Encrypt patient<br>information store to a<br>removable memory | The proposed device accessories for<br>network/archive were previously<br>available in predicate Versana Balance.<br>An additional Bluetooth adapter was<br>available in the predicate device but not<br>Versana Essential. However, there is no           |

Comparison of technological characteristics with Predicate device:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are also some water droplet shapes around the circle. The logo is simple and recognizable, representing the company's brand identity.

| Characteristic | Proposed Device<br>Versana Essential                                                   | Predicate Device<br>Versana Balance<br>(K220446)                                                                 | Comment on comparison                                                                                                                             |
|----------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
|                | memory device,<br>Ethernet/DICOM/LAN via hardware,<br>wireless adapter, Tricefy Uplink | device,<br>Ethernet/DICOM/LAN via hardware, wireless<br>adapter, <b>Bluetooth</b> USB adaptor, Tricefy<br>Uplink | impact to safety or effectiveness because<br>the memory stick, LAN, or wireless<br>adapter are alternative means to share<br>patient images/data. |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. The circle is surrounded by a series of curved lines that resemble water droplets or stylized flames. The logo is simple and recognizable, and it is often used to represent the company's brand.

### Summary of Non-Clinical Tests:

Versana Essential has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to comply with applicable medical device safety standards. The Versana Essential complies with voluntary standards:

- AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance - 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
- . IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - Edition 4.1, 2020
- . IEC 60601-2-37, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment -Edition 2.1, 2015
- . ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process - Fourth edition, 2009
- IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields - Edition 2.1, 2017
- ISO 14971, Medical devices - Application of risk management to medical devices, 2019
- NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set, 2020e

The following quality assurance measures are applied to the development of the system:

- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification)
- . Safety testing (Verification)

Transducer material and other patient contact materials are biocompatible.

#### Summary of Clinical Tests:

The subject of this premarket submission, Versana Essential, did not require clinical studies to support substantial equivalence.

#### Conclusion:

Based on the equipment design similarities, conformance to recognized performance

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative flourishes resembling water droplets or stylized leaves around the perimeter, giving it a distinctive and recognizable appearance.

standards, and performance testing, GE Healthcare considers the Versana Essential to be as safe, effective, and performs in a substantially equivalent manner as the predicate device Versana Balance (K220446).

---

**Source:** [https://fda-staging.innolitics.com/device/K223407](https://fda-staging.innolitics.com/device/K223407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
