K223379 · Qingdao Kingon Medical Science and Technology Co., Ltd. · CAW · Jun 22, 2023 · Anesthesiology
Device Facts
Record ID
K223379
Device Name
Portable Oxygen Concentrator (Model: P2-E7, P2-E)
Applicant
Qingdao Kingon Medical Science and Technology Co., Ltd.
Product Code
CAW · Anesthesiology
Decision Date
Jun 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Portable Oxygen Concentrator (Model: P2-E7, P2-E) is intended to provide supplemental oxygen in a home, institutional, or travel environment.
Device Story
Portable oxygen concentrator (POC) provides supplemental oxygen via pulse-dose delivery. Device uses pressure swing adsorption (PSA) with molecular sieves to extract oxygen from ambient air; nitrogen is adsorbed under pressure and released during exhaust cycles. Input: ambient air; Output: pulsed high-concentration oxygen delivered via nasal cannula. Operated by patient or caregiver in home, institutional, or travel settings. Device includes compressor, battery, solenoid valves, molecular sieves, and control electronics. Integrated sensors detect inhalation to trigger pulse delivery; volume depends on user setting. Provides visual/audible alarms for system status (e.g., low battery, no breath detected, high temperature). Benefits patients by providing portable, on-demand oxygen therapy.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2), performance (ISO 80601-2-69, ISO 80601-2-67), and biocompatibility (ISO 10993-5, ISO 10993-10, ISO 18562-2/3).
Indicated for adult patients requiring supplemental oxygen in home, institutional, or travel environments.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2023
Qingdao Kingon Medical Science and Technology Co., Ltd. Zhang Benrong Technical Supervisor RM.1711, Building K, NO.101 Science Ave International Creative Valley Qingdao Free Trade Zone, Shandong 266555 China
Re: K223379
Trade/Device Name: Portable Oxygen Concentrator (Model: P2-E7, P2-E) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: May 24, 2023 Received: May 24, 2023
Dear Zhang Benrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
For James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223379
Device Name
Portable Oxygen Concentrator (Model: P2-E7, P2-E)
Indications for Use (Describe)
The Portable Oxygen Concentrator (Model: P2-E7, P2-E) is intended to provide supplemental oxygen in a home, institutional, or travel environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> <path d="M3,3 L13,13" stroke="black"></path> <path d="M13,3 L3,13" stroke="black"></path> </svg> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16" x="0" y="0"></rect> </svg> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary as required by section 807.92(c)
### Date of the summary prepared: 10/30/2022 l.
### Administrative Information ll.
| Manufacturer<br>information | Establishment<br>registration<br>number | 3014777423 |
|-----------------------------|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Owner/Operator<br>Number | 10061814 |
| | Name | Qingdao Kingon Medical Science and Technology Co., Ltd. |
| | Address | Room 301-302, No.15 HanchengRoad, Qingdao Free Trade Zone,<br>Shandong, China, 266555 |
| | Contact Person | Name: Benrong Zhang<br>Address: Room 301-302, No.15 HanchengRoad, Qingdao Free Trade Zone, Shandong, China, 266555<br>TEL: +86-18565833539<br>FAX: +86 532 58792324<br>Email: augus@kingonmed.com |
| Submission<br>Correspondent | Contact Person | Name: Tracy Nanan<br>Address: Room 301-302, No.15 HanchengRoad, Qingdao Free Trade Zone, Shandong, China, 266555<br>TEL: +86-18565833539<br>FAX: +86 532 58792324<br>Email: ptg2022ptg@163.com |
# III. Device Information
| Type of 510(k) | Special 510(k): Device Modification |
|---------------------|---------------------------------------------------|
| Prior submission | No prior submission |
| Common Name | Generator, Oxygen, Portable |
| Classification name | Portable oxygen generator |
| Trade Name | Portable Oxygen Concentrator (Model: P2-E7, P2-E) |
| Review panel | Anesthesiology |
{4}------------------------------------------------
| Product code | CAW |
|-------------------|----------|
| Regulation Number | 868.5540 |
| Regulation Class | 2 |
# IV. Predicate Device Information
| Common Name | Generator, Oxygen, Portable |
|---------------------|--------------------------------------------|
| Classification name | Portable oxygen generator |
| Trade Name | Portable Oxygen Concentrator, model: P2-E6 |
| Review panel | Anesthesiology |
| Product code | CAW |
| Regulation Number | 868.5540 |
| Regulation Class | 2 |
# V. Device description
Portable Oxygen Concentrator (Model: P2-E7, P2-E) is a portable oxygen generator that is intended to release oxygen for respiratory by means of physical means (a molecular sieve). It supplies a pulsed high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Portable Oxygen Concentrator is small, portable and may be used in home, institutional, or travel environment.
The portable oxygen concentrator consists of two parts: an oxygen concentrator and accessories. The oxygen concentrator is composed of compressor, battery, solenoid valve, molecular sieve, circuit control system, heat dissipation system, and a flow control device. Accessories include power adapters.
Model difference: The only different between model P2-E7 and P2-E is that P2-E7 has biggest output oxygen flow of 1.4L/min (at 7 gears) and P2-E has biggest output oxygen flow of 1.0L/min (at 5 gears).
{5}------------------------------------------------
# VI. Principle of operation
Both model P2-E7 and P2-E has the same principle of operation. The portable oxygen concentrator works by getting use of the molecular sieves character that the internal pressure of a sealed container containing of molecular sieve will increase when injecting air into it. At this time, the molecular sieve will absorb a lot of nitrogen in the increasing of ambient pressure, while the oxygen in the air is still existed in gaseous form, then the oxygen are collected through some pipelines. When the nitrogen absorption process in the container reaches a certain level, then exhaust of the vacuum container and nitrogen will be released from molecular sieve with the ambient pressure decreases. It will detect when the user begins to take a breath and then delivers a pulsed volume of oxygen during the inhalation period. The volume of the oxygen pulse is dependent on the setting value.
# VII. Indications for Use
The Portable Oxygen Concentrator (Model: P2-E7, P2-E) is intended to provide supplemental oxygen in a home, institutional, or travel environment.
| ID | Comparison<br>Items | Subject device | Predicate device | Comparison |
|-----|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| 1. | 510K Number | / | K210371 | / |
| 2. | Manufacturer | Qingdao Kingon Medical Science and<br>Technology Co., Ltd. | Qingdao Kingon Medical Science and<br>Technology Co., Ltd. | Same |
| 3. | Device name | Portable Oxygen Concentrator | Portable Oxygen Concentrator | / |
| 4. | Model | P2-E7, P2-E | P2-E6 | / |
| 5. | Classification | 21CFR 868.5440 | 21CFR 868.5440 | Same |
| 6. | Product Code | CAW | CAW | Same |
| 7. | FDA Class | II | II | Same |
| 8. | Indications for<br>Use | The Portable Oxygen Concentrator<br>(Model: P2-E7, P2-E) is intended to<br>provide supplemental oxygen in a<br>home, institutional, or travel<br>environment. | The Portable Oxygen Concentrator,<br>model: P2-E6 is intended to provide<br>supplemental oxygen in a home,<br>institutional, or travel environment. | Same |
| 9. | Environment of<br>Use | Home, institutional, or travel<br>environment. | Home, institutional, or travel<br>environment. | Same |
| 10. | Design | table type | table type | Same |
| 11. | Prescriptive | Yes | Yes | Same |
| 12. | Patient<br>Population | Adult | Adult | Same |
| | | | | |
| 13. | Material of<br>Patient contact<br>components | Wiring cover: PC+ABS<br>Intake hood: PC+ABS<br>Nozzle fitting: Aluminum alloy<br>Button panel: PET<br>Main housing: PC+ABS | Wiring cover: PC+ABS<br>Intake hood: PC+ABS<br>Nozzle fitting: Aluminum alloy<br>Button panel: PET<br>Main housing: PC+ABS | Same |
| 14. | Duration and<br>type of contact | Type of contact: surface device;<br>Duration: permanent (> 30 d); | Type of contact: surface device;<br>Duration: permanent (> 30 d); | Same |
| 15. | Complete list of<br>all the<br>biocompatibility<br>tests performed | ISO 10993- 5 tested for Cytotoxicity;<br><br>ISO 10993-10 tested for Sensitization<br>and Irritation;<br><br>ISO 18562-2 tested for Particulate<br>matter;<br><br>ISO 18562-3 tested for Volatile<br>organic Compounds; | ISO 10993- 5 tested for Cytotoxicity;<br><br>ISO 10993-10 tested for Sensitization<br>and Irritation;<br><br>ISO 18562-2 tested for Particulate<br>matter;<br><br>ISO 18562-3 tested for Volatile<br>organic Compounds; | Same |
| 16. | Single Patient,<br>multi-use | Yes | Yes | Same |
| 17. | Patient<br>Interface | Cannula Port | Cannula Port | Same |
| 18. | Technology | Pressure Swing Adsorption with<br>molecular sieve | Pressure Swing Adsorption with<br>molecular sieve | Same |
| 19. | Dimensions | 6.30"H*3.35"W*8.70"L | 6.30"H*3.35"W*8.70"L | Same |
| 20. | Weight | 4.34lbs±0.07lbs (with standard<br>battery) | 4.34lbs±0.07lbs (with standard<br>battery) | Same |
| 21. | Oxygen<br>Concentration | 90%-3%/+6% at all settings | 90%-3%/+6% at all settings | Same |
| 22. | Setting | P2-E7: adjustable in 1 increments<br>from 1 to 7.<br>P2-E: adjustable in 1 increments from<br>1 to 5. | adjustable in 1 increments from 1 to 6 | Different<br>(See below note<br>ID_22) |
| 23. | Pulse mode<br>bolus size | P2-E7: 50mL per breath at setting 5<br>with 20BPM<br>P2-E: 50mL per breath at setting 5<br>with 20BPM | 50mL per breath at setting 5 with<br>20BPM | Same |
| 24. | Principle of<br>operation | by means of molecular sieve | by means of molecular sieve | Same |
| 25. | Filters | Input Filter, Patient Filter | Input Filter, Patient Filter | Same |
| 26. | Breath rate | 10 - 40 Breath per minute | 10 - 40 Breath per minute | Same |
| 27. | User Interface | Buttons, LCD Display | Buttons, LCD Display | Same |
| 28. | Power<br>requirements | AC adaptor: 100-240VAC ;50-60 Hz | AC adaptor: 100-240VAC ;50-60 Hz | Same |
| 29. | Maximum<br>oxygen<br>discharge<br>pressure | P2-E7: 20.6 PSI (142KPa)<br>P2-E: 20.9 PSI (144KPa) | 18.3 PSI (126KPa) | Different<br>(See below note<br>ID_29) |
| 30. | Inspiratory<br>trigger<br>sensitivity | -0.12cm/H20 | -0.12cm/H20 | Same |
| 31. | Software | Embedded | Embedded | Same |
| 32. | Acoustic Noise | P2-E7: 58.8 dBA at 1.4 LPM<br>P2-E: 52.0 dBA at 1.0 LPM | 58.2 dBA at 1.2 LPM | Different<br>(See below note<br>ID_32) |
| 33. | | Battery empty | Battery empty | Same |
| 34. | | Low pressure | Low pressure | Same |
| 35. | | No pulse | No pulse | Same |
| 36. | | High Temp | High Temp | Same |
| 37. | | Compressor Failure | Compressor Failure | Same |
| 38. | Alarms | Fan Failure | Fan Failure | Same |
| 39. | | Low Flow | Low Flow | Same |
| 40. | | Low Battery | Low Battery | Same |
| 41. | | No Breath Detected | No Breath Detected | Same |
| 42. | | EEPROM Failure | EEPROM Failure | Same |
| 43. | | Flow rates | Flow rates | Same |
| 44. | Status Indicator | Battery Condition | Battery Condition | Same |
| 45. | | Alarms | Alarms | Same |
| 46. | Battery<br>Duration | P2-E7: Up to 4.5 hours at 0.21 LPM<br>P2-E: Up to 4.5 hours at 0.21 LPM | Up to 4.5 hours at 0.21 LPM | Same |
| 47. | Operating<br>Environment | Temperature: 41 to 104°F (5 to 40°C)<br>Humidity: 10% to 90%, non-<br>condensing<br>Altitude: 0 to 10,000 ft. (0 to 3048<br>meters) | Temperature: 41 to 104°F (5 to 40°C)<br>Humidity: 10% to 90%, non-<br>condensing<br>Altitude: 0 to 10,000 ft. (0 to 3048<br>meters) | Same |
| 48. | Shipping<br>Storage | Temperature: -4 to 158°F (-20 to<br>70°C)<br>Humidity: 5% to 90%, non-condensing<br>Store in a dry environment | Temperature: -4 to 158°F (-20 to<br>70°C)<br>Humidity: 5% to 90%, non-condensing<br>Store in a dry environment | Same |
| 49. | Electrical Safety | AAMI ANSI ES 60601-1 | AAMI ANSI ES 60601-1 | Same |
| 50. | electromagnetic<br>compatibility | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| 51. | Biocompatibility | 4 VOC's less than ambient | 4 VOC's less than ambient | Same |
| 52. | Standards Met | ANSI AAMI ES 60601- 1: 2005 / (R)<br>2012 and A1: 2012 | ANSI AAMI ES 60601- 1: 2005 / (R)<br>2012 and A1: 2012 | Same |
| | | IEC 60601-1-2: 2014 | IEC 60601-1-2: 2014 | |
| | | IEC 60601-1-11: 2015 | IEC 60601-1-11: 2015 | |
| | | IEC 60601-1-8: 2006+ A1:2012 | IEC 60601-1-8: 2006+ A1:2012 | |
| | | ISO 80601-2-69: 2014 | ISO 80601-2-69: 2014 | |
| | | ISO 80601-2-67: 2014 | ISO 80601-2-67: 2014 | |
| | | ISO 18562-2: 2017 | ISO 18562-2: 2017 | |
| | | ISO 18562-3: 2017 | ISO 18562-3: 2017 | |
| | | IEC 62133: 2012 | IEC 62133: 2012 | |
| | | ISO 10993-5:2009<br>ISO 10993-10:2010 | ISO 10993-5:2009<br>ISO 10993-10:2010 | |
# VIII. Comparison with predicate device
{6}------------------------------------------------
# kıngon
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "kingon" in a sans-serif font. The word is written in a dark blue color. The letters are evenly spaced and the word is centered in the image. The background is white.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "kingon" in a dark blue sans-serif font. The letters are all lowercase, and there is a small circle above the "o" in "kingon". The word is centered and takes up most of the frame. The background is white.
### Note:
- ID 22: The max setting of subject device P2-E7 is 7 and of subject device P2-E is 5 while the predicate device is 6, the extra setting 7 of P2-E7 introduce risks of increased output flow and increased Emission of particulate matter of gas pass way, those risks are mitigated by tests tested according to ISO 80601-2-69: 2014, ISO 80601-2-67: 2014 and ISO 18562-2: 2017, therefore the difference does not raise new questions of safety and effectiveness.
- ID_29: The Maximum oxygen discharge pressure of subject device P2-E is different with predicate device P2-E6 (K210371), since the subject device has been tested against ISO 80601-2-69: 2014 with positive result, the difference of subject device with predicate device RS-00500 (K162433) do not raise new questions of safety and effectiveness.
- ID 32: The Acoustic Noise of subject device P2-E7 is litter bigger than predicate device P2-E6 (K210371) and subject device P2-E is litter smaller than predicate device P2-E6 (K210371), since the subject device P2-E7 has been tested against ISO 80601-2-69: 2014 and ANSI AAMI ES60601-1 with positive results, the difference of subject device with predicate device P2-E6 (K210371) do not raise new questions of safety and effectiveness.
## IX. Discussion of Non-Clinical Tests Performed for Safety and effectiveness are as
# follows
The recognized consensus standards for safety of medical equipment: ANSI AAMI ES60601 -1, IEC 60601 - 1 - 1 for safety, IEC 60601 - 1 -2 for electromagnetic compatibility, ISO 80601 - 2 - 69: 2014 and ISO 80601-2-67: 2014 for performance and IEC 62304 for software verification are complied. See below table for details:
{9}------------------------------------------------
# kıngo'n
| Standards | Standards Name |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI ES60601-<br>1:2005/(R)2012 and A1:2012 | Medical Electrical Equipment - Part 1: General Requirements For<br>Basic Safety And Essential Performance |
| IEC 60601-1-2: 2014 | Medical Electrical Equipment - Part 1-2: General Requirements For<br>Basic Safety And Essential Performance - Collateral Standard:<br>Electromagnetic Disturbances - Requirements And Tests |
| IEC 60601-1-11: 2015 | Medical Electrical Equipment - Part 1-2: General Requirements For<br>Basic Safety And Essential Performance - Collateral Standard:<br>Electromagnetic Disturbances - Requirements And Tests |
| IEC 60601-1-8: 2006+ A1:2012 | Medical electrical equipment - Part 1-8: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>General requirements, tests and guidance for alarm systems in<br>medical electrical equipment and medical electrical systems |
| ISO 80601-2-69: 2014 | Medical electrical equipment. Particular requirements for the basic<br>safety and essential performance of oxygen concentrator<br>equipment |
| ISO 80601-2-67: 2014 | Medical electrical equipment. Particular requirements for basic<br>safety and essential performance of oxygen-conserving equipment |
| ISO 18562-2: 2017 | Biocompatibility evaluation of breathing gas pathways in healthcare<br>applications - Part 3: Tests for emissions of volatile organic<br>compounds |
| ISO 18562-3: 2017 | Biocompatibility evaluation of breathing gas pathways in healthcare<br>applications. Tests for emissions of volatile organic compounds<br>(VOCs) |
| IEC 62133: 2012 | Secondary cells and batteries containing alkaline or other non-acid<br>electrolytes - Safety requirements for portable sealed secondary<br>cells, and for batteries made from them, for use in portable<br>applications |
| ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity |
| ISO 10993-10:2010 | Biological evaluation of medical devices - Part 10: Tests for irritation<br>and skin sensitization |
| IEC 62304:2006+A1:2015 | Medical device software - Software life cycle processes |
#### ● Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device P2-E6. The system complies with the AAMI ANSI ES60601-1, IEC 60601-1, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-67, and ISO 80601-2-69 standards for electrical safety and the IEC 60601-1-2 standard for EMC.
#### ● Software Verification and Validation Testing
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff, May 2005.
{10}------------------------------------------------
## Software Description:
The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could result in Minor Injury, either to a user of the device. The software of the system, on the whole, is accountable for the system scheduler of the device, including fulfilling alarm task, beep task, display task, monitor task, startup task, breath task, as well as dealing with USART1-interrupt, USART2-interrupt, WDT-interrupt, and controlling the sensor module.
- A Programming language
STMicroelectronics and Texas Instruments.
- > Hardware platform
- a) Applicable Device Name: Portable Oxygen Concentrator
- b) Applicable model: P2-E7, P2-E
- Micro-controller used: STM32F103RCT6, STM32F103C8T6, MSP430G2755 C)
- ROM Size: STM32F103RCT6 256K, STM32F103C8T6 64K, MSP430G2755 32K ರು)
- RAM Size: STM32F103RCT6 48K, STM32F103C8T6 20K, MSP430G2755 -4K e)
- Software Release Version Number: Keil uVision5 V5.25.20, IRA for 430 V7.12.1 f)
# X. Discussion of Clinical Accuracy Testing Performed
There was no clinical testing performed.
# XI. Conclusions
The Portable Oxygen Concentrator (Model: P2-E7, P2-E) have the same intended use and similar characteristics as the cleared predicate device Portable Oxygen Concentrator, model: P2-E6. Moreover, bench testing contained in this submission supplied demonstrate that the differences existed between Portable Oxygen Concentrator (Model: P2-E7, P2-E) and Portable Oxygen Concentrator, model: P2-E6 (K210371) do not raise any new questions of safety or effectiveness.
The non-clinical tests support the safety of the device and software verification and validation demonstrate that the Portable Oxygen Concentrator (Model: P2-E7, P2-E) performs as intended in the specified use conditions are same with predicate device. The performance tests demonstrate that the Portable Oxygen Concentrator (Model: P2-E7, P2-E) performs comparably to the predicate device that is currently marketed for the same intended use. Thus, Portable Oxygen Concentrator (Model: P2-E7, P2-E) is Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.