← Product Code [FMK](/productcode/FMK) · K223313

# Disposable Safety Lancets (K223313)

_Huaian Hening Medical Instruments Co., Ltd. · FMK · Jan 23, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223313

## Device Facts

- **Applicant:** Huaian Hening Medical Instruments Co., Ltd.
- **Product Code:** [FMK](/productcode/FMK.md)
- **Decision Date:** Jan 23, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4850
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The product is used to obtain capillary blood samples from fingertip in a hospital or at home. The device contains a sharp injury protection feature.

## Device Story

Disposable safety lancet; single-use; manual activation via pressure trigger. Components: trigger, plastic handle, outer shell, back cover, spring, protective cap, 304 stainless steel needle. Upon activation, needle punctures skin; spring mechanism automatically retracts needle into device body to prevent needlestick injury. Used by patients or healthcare providers in home or hospital settings. Output is a capillary blood sample for diagnostic testing. Benefits include reduced risk of accidental sharps injury compared to unprotected lancets.

## Clinical Evidence

Bench testing only. Performance data included appearance, launch length, penetration force (≤1.00N), needle retraction functionality, single-use verification, cap removal force (30-35 N*cm), needle bond strength (≥10N/15s), and drop testing. Biocompatibility testing performed per ISO 10993-1:2018 (cytotoxicity, sensitization, intracutaneous, systemic toxicity, hemolysis, pyrogenicity). Sterilization validated per ISO 11137. Shelf life validated to 5 years via stability/aging studies.

## Technological Characteristics

Needle: 304 austenitic stainless steel. Body/Cap: ABS plastic. Sensing/Actuation: Spring-loaded mechanical trigger. Dimensions: 21G, 23G, 26G, 28G, 30G gauges; 1.8mm-2.2mm launch length. Sterilization: Gamma irradiation. Single-use. No software or electronics.

## Regulatory Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

## Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- SurgiLance® Safety Lancets ([K101145](/device/K101145.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

January 23, 2023

Huaian Hening Medical Instruments Co., Ltd Zhengcan Da General Manager No. 6 West Hongdou Road Economic Development Zone Huaian, Jiangsu China

Re: K223313

Trade/Device Name: Disposable Safety Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: December 29, 2022 Received: December 29, 2022

Dear Zhengcan Da:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

**Jessica Carr -S**

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223313

Device Name Disposable Safety Lancets

# Indications for Use (Describe)

Disposable Safety Lancets

The product is used to obtain capillary blood samples from fingertip in a home. The device contains a sharp injury protection feature.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K223313 510(k) summary

## l Submitter

Device submitter: HUAIAN HENING MEDICAL INSTRUMENTS CO., LTD.

NO.6 West Hongdou Road Economic Development Zone Huaian,

Jiangsu China

Contact person: Zhengcan Da General Manager Phone: +86-13952318668 Fax: +86-517-83800501 Email: hn-lancets@vip.163.com

Date: 10/12/2022

#### II Device

Trade Name of Device: Disposable Safety Lancets Common Name: Blood Lancets Regulation Number: 21 CFR 878.4850 Regulation Name: Single Use Only Blood Lancet with an Integral Sharps Injury Prevention Feature Regulatory Class: II Product code: FMK Review Panel: General & Plastic Surgery

#### III Predicate Devices

| Trade name:             | SurgiLance® Safety Lancets                              |
|-------------------------|---------------------------------------------------------|
| Common name:            | Blood Lancets                                           |
| Classification:         | I-Lancet with Sharps Prevention Feature, 21CFR 878.4800 |
| Product Code:           | FMK                                                     |
| Premarket Notification: | K101145                                                 |
| Manufacturer:           | MediPurpose Pte, Ltd.                                   |

#### IV Device description

The Disposable Safety Lancets consists 7 parts, include a trigger, plastic handle, out shell, back cover, spring, protective cap and needle. The models of the Disposable Safety Lancets are 21G; 23G; 26G; 28G; 30G. The product is used to obtain capillary blood samples from fingertip in a hospital or at home. The device contains a sharp injury feature. The lancet is hit by pressure, and once the device strikes, the lancet needle can

{4}------------------------------------------------

puncture the skin. And once activated, the needle retracts into the body of the device which reduces the risk of injury as the result if an exposed needle. Used Gamma sterilization, and are products for single use.

## V Indications for use

#### Disposable Safety Lancets

The product is used to obtain capillary blood samples from fingertip in a hospital or at home. The device contains a sharp injury protection feature.

## VI Comparison of technological characteristics with the predicate devices

The Disposable Safety Lancets have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Disposable Safety Lancets and predicate devices do not alter suitability of the proposed device for its intended use.

| Device<br>feature                                            | Subject Device<br>(Disposable Safety<br>Lancets)                                                                                                                     | Predicate Device K101145<br>( SurgiLance® Safety<br>Lancets )                                                                                                                                                       | Comment                   |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Indications<br>for use                                       | The product is used to<br>obtain capillary blood<br>samples from fingertip in a<br>hospital or at home. The<br>device contains a sharp<br>injury protection feature. | The SurgiLance® Safety<br>Lancet is a puncture device to<br>obtain micro blood samples.<br>The<br>SurgiLance® Safety Lancet<br>has a sharps prevention<br>feature to protect the user<br>from a needlestick injury. | Similar<br>Comment<br>1   |
| Product<br>code                                              | FMK                                                                                                                                                                  | FMK                                                                                                                                                                                                                 | Equivalent                |
| Safety<br>protection<br>features                             | Yes                                                                                                                                                                  | Yes                                                                                                                                                                                                                 | Equivalent                |
| Reuse<br>durability                                          | Single use                                                                                                                                                           | Single use                                                                                                                                                                                                          | Equivalent                |
| Sterilizatio<br>n                                            | Irradiation                                                                                                                                                          | Not available                                                                                                                                                                                                       | Different<br>Comment<br>2 |
| Device feature                                               | Subject Device<br>(Disposable Safety<br>Lancets)                                                                                                                     | Predicate Device K101145<br>(SurgiLance® Safety<br>Lancets)                                                                                                                                                         | Comment                   |
| Model                                                        | 21G; 23G; 26G; 28G; 30G                                                                                                                                              | SLN 100: 21G;<br>SLN 200: 21G;<br>SLN 240: 21G;<br>SLN 300: 21G;<br>SLB 200: 18G;<br>SLB 250: 18G;<br>(Information gathered from<br>MediPurpose Pte, Ltd. official<br>website)                                      | Different<br>Comment 3    |
| Launch<br>length                                             | 21G-2.2mm<br>23G-1.8mm<br>26G-1.8mm<br>28G-1.8mm<br>30G -1.8mm                                                                                                       | SLN 100: 21G-1.0mm<br>SLN 200: 21G-1.8mm<br>SLN 240: 21G-2.2mm<br>SLN 300: 21G-2.8mm<br>SLB 200: 18G-1.8mm<br>SLB 250: 18G-2.3mm                                                                                    |                           |
| Materials<br>of parts in<br>contact<br>with<br>human<br>body | Lancet needle: 304<br>austenitic stainless steel;<br>Body and cap: ABS;                                                                                              | Lancet needle: medical grade<br>stainless steel;<br>Housing and cap: plastics                                                                                                                                       | Similar<br>Comment 4      |

{5}------------------------------------------------

Discussion:

Comment 1

The subject device and the predicate device have the same intended use, to puncture the skin to obtain drops of blood for testing purposes. The same basic technology characteristics for a lancet with sharps injury prevention of Disposable Safety Lancets as compared with the predicate device.

{6}------------------------------------------------

#### Comment 2

The sterilization method of predicate device is not available. However, the subject device was ensured sterility by sterilization. Therefore, the differences on sterilization do not raise new questions about safety and effectiveness.

#### Comment 3

The models and Launch length of subject device are different from the predicate device. The model was more than as the predicated products, while the puncture depths are same. Different models are only different in the outer diameter of the needle, which allowed to choose to meet blood volume needs. Different needle specification will be selected by physician per patient's condition and this different were addressed by performance tests. Testing of performance shows no impact of launch length on the sharpness /penetration force or bond between the lancet body and needle, this difference does not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

#### Comment 4

The subject device utilizes some of the same materials, specifically the use of medical grade stainless steel for the lancet needles but may use different types plastics for the bodies, caps. All the materials are known biocompatible materials that have been used in lancets or other similar medical devices, and the materials of subject device were demonstrated by the biocompatibility tests done. The differences between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

#### VII Summary of non-clinical testing

The following performance data were provided in support of the substantial equivalence determination.

| No | Testing item  | Specification                                                                                                                            | Result |
|----|---------------|------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 01 | Appearance    | Disposable Safety Lancets the surface<br>should be smooth without edge, no dirt and<br>damage, deformation and other poor<br>appearance. | Pass   |
| 02 | Launch Length | The length of the needles in the Disposable<br>Safety Lancet is different in different<br>gauges.                                        | Pass   |

#### Performance Testing for Disposable Safety Lancets

{7}------------------------------------------------

|    |                                  | The launch length of the needle is<br>determined according to the length of the<br>purchase, and the general emission length<br>is 1.8mm-2.2mm. |      |
|----|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 03 | Sharpness/Penetration<br>testing | Penetration force $≤1.00N.$                                                                                                                     | Pass |
| 04 | Feature                          | The tip of the needle can shrink quickly after<br>firing, and the tip of the needle is not<br>exposed.                                          | Pass |
| 05 |                                  | Disposable Safety Lancets Only one launch,<br>not another.                                                                                      | Pass |
| 06 | Initial bioburden                | Initial bioburden of the device shall be less<br>than 100CFU/g                                                                                  | Pass |
| 07 | Sterile                          | The sterile blood lancet shall be sterile                                                                                                       | Pass |
| 08 | Cap removal force                | The moment for breaking the safe mode<br>should range from 30 N*cm to 35 N*cm.                                                                  | Pass |
| 09 | Needle removal force             | The bond between the lancet body and<br>needle should be greater than or equal to<br>10N/15s.                                                   | Pass |
| 10 | Drop testing                     | The carton box should have no puncture<br>after the drop test.                                                                                  | Pass |

## Biocompatibility testing

Biocompatibility of the Disposable Blood Lancets and Disposable Safety Lancets were evaluated in accordance with ISO 10993-1:2018 for the body contact category. The following tests were performed, as recommended:

| Cytotoxic test               | ISO 10993-5:2009  |
|------------------------------|-------------------|
| Skin sensitization test      | ISO 10993-23:2021 |
| Intracutaneous test          | ISO 10993-10:2021 |
| Acute systemic toxicity test | ISO 10993-11:2017 |
| Hemolysis test               | ISO 10993-4:2017  |
| Pyrogen Test                 | ISO 10993-11:2017 |

## Sterilization and shelf life testing

- Irradiation sterilization validation per ISO 11137-1 and ISO 11137-3. A
- Pyrogen testing per ISO 10993-11:2017 A
- Simulated shipping per ISTA 2A: 2011 A
- The 5 years shelf life of the device is determined based on stability study which A includes ageing test.

## VIII Conclusion

{8}------------------------------------------------

The Disposable Safety Lancets is substantially equivalent to its predicate device (SurgilLance® Safety Lancets). The differences between the predicate and subject device do not raise any new or different questions of safety or effectiveness. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K223313](https://fda-staging.innolitics.com/device/K223313)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
