K223286 · Becton, Dickinson and Company · FMI · Dec 28, 2022 · General Hospital
Device Facts
Record ID
K223286
Device Name
BD AutoShield Duo™ Pen Needle
Applicant
Becton, Dickinson and Company
Product Code
FMI · General Hospital
Decision Date
Dec 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The BD AutoShield Duo™ Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs. The product has two safety shields, which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connection-end). The locked shields help reduce the occurrence of needle sticks from both ends of the needle.
Device Story
BD AutoShield Duo™ Pen Needle is a single-use, sterile, disposable hypodermic needle for subcutaneous drug delivery via pen injectors. Device features double-ended cannula with safety shields on both patient and non-patient ends; shields lock automatically after use to prevent accidental needle sticks. User attaches needle to pen injector via internal threads; non-patient end penetrates pen cartridge septum. Manual operation by patient or clinician. Device provides physical barrier against sharps injury. Benefits include reduced risk of needle-stick injuries for users. No electronic or software components.
Clinical Evidence
No clinical data. Bench testing only. Testing included functional performance (ISO 11608-2, ISO 9626, ISO 23908), biocompatibility (ISO 10993-1), particulate matter (USP <788>), sterility (ISO 11137-2), and shipping/shelf-life validation (ASTM D4169).
Technological Characteristics
Materials: Stainless steel cannula/springs, polycarbonate/polyethylene shields/sleeves, polypropylene hub, paper label. Energy: None (manual). Connectivity: None. Sterilization: Gamma irradiation (Cobalt-60, SAL 10^-6). Form factor: Pen needle assembly with dual safety shields. Standards: ISO 11608-2, ISO 9626, ISO 23908, ISO 10993-1, ISO 11137-2, USP <788>, ASTM D4169.
Indications for Use
Indicated for subcutaneous injection of drugs using pen injector devices. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 28, 2022
Becton, Dickinson and Company Alec Paterno Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
# Re: K223286
Trade/Device Name: BD AutoShield Duo™ Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: October 25, 2022 Received: October 25, 2022
## Dear Alec Paterno:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Courtney Digitally signed by Date: 2022.12.28 Evans -S 14:45:36 -05'00'
For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223286
#### Device Name BD AutoShield Duo™ Pen Needle
#### Indications for Use (Describe)
The BD AutoShield Duo™ Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.
The product has two safety shields, which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connection-end). The locked shields help reduce the occurrence of needle sticks from both ends of the needle.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K223286 510(k) Summary
| Submitted By: | Alec Paterno | | | | | | | | |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------------------|--------------|------------|-----------------|--------------------------------------------------------------------|---------------|--------------------------------------|
| | Senior Regulatory Affairs Specialist | | | | | | | | |
| | Becton Dickinson and Company | | | | | | | | |
| | 1 Becton Drive | | | | | | | | |
| | Franklin Lakes, NJ 07417 | | | | | | | | |
| | Tel: 908 670 3376 | | | | | | | | |
| Date Prepared: | December 21, 2022 | | | | | | | | |
| Device Name: | <table><tr><td>Trade Name:</td><td>BD AutoShield Duo™ Pen Needle</td></tr><tr><td>Common Name:</td><td>Pen Needle</td></tr><tr><td>Classification:</td><td>Class II device; 21 CFR 880.5570, (hypodermic single lumen needle)</td></tr><tr><td>Product Code:</td><td>FMI (hypodermic single lumen needle)</td></tr></table> | Trade Name: | BD AutoShield Duo™ Pen Needle | Common Name: | Pen Needle | Classification: | Class II device; 21 CFR 880.5570, (hypodermic single lumen needle) | Product Code: | FMI (hypodermic single lumen needle) |
| Trade Name: | BD AutoShield Duo™ Pen Needle | | | | | | | | |
| Common Name: | Pen Needle | | | | | | | | |
| Classification: | Class II device; 21 CFR 880.5570, (hypodermic single lumen needle) | | | | | | | | |
| Product Code: | FMI (hypodermic single lumen needle) | | | | | | | | |
#### Legally marketed predicate device to which substantial equivalence is being claimed:
K110703: BD AutoShield™ Duo Pen Needle
#### Purpose of Submission:
The purpose of this submission is to provide a minor administrative update to the BD AutoShield Duo™ Pen Needle intended use statement in order to align the statement with that of other marketed BD Pen Needles (cleared under K213478, K213156, and K212015). The overall intended use is the same as the predicate device (K110703), with the clarification, adding the word "subcutaneous", solely intended to harmonize the statement with the rest of our pen needle portfolio. The proposed clarification has no impact on the use of the device, the user interface, technological characteristics, or overall safety and effectiveness of the subject device compared to its predicate.
#### Device Description:
The BD AutoShield Duo™ Pen Needle has the same intended use/indications for use, technological characteristics, and principles of operation as the predicate device cleared under K110703.
The pen needle assembly consists of a double-ended cannula that is assembled onto an injection molded hub. The internal threads allow the subject device to be screwed onto a pen injector device, further enabling the non-patient end of the cannula to penetrate through the septum of the pen injector cartridge.
The patient and non-patient ends of the cannula are visible prior to attachment to the pen injector device. The BD AutoShield Duo™ Pen Needle has safety mechanisms on both the patient and non-patient ends of the needle, allowing the needle to be shielded and locked after use, which is designed to reduce the
{4}------------------------------------------------
occurrence of accidental needle-stick injuries. The BD AutoShield Duo™ Pen Needle is a single use disposable device and is provided sterile. The subject device is non-toxic and non-pyrogenic.
#### Intended Use:
The BD AutoShield Duo™ Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.
The product has two safety shields, which lock in place after use (patient-end) and upon removal of the needle from the pen (pen connection-end). The locked shields help reduce the occurrence of needle sticks from both ends of the needle.
The intended use of the subject device remains the same as the predicate device.
#### Comparison with Predicate Devices:
The subject device has the same fundamental scientific technology and device performance as the predicate device (K110703). The purpose of this submission is to provide a minor update to the wording of the BD AutoShield Duo™ Pen needle intended use statement to clarify proper use of the device and align the statement with that of other marketed BD Pen Needles. The table below provides a side-by-side comparison of the subject device compared to its predicate.
| Feature | Subject Device:<br>BD AutoShield Duo™ Pen<br>Needle | Predicate Device:<br>BD AutoShield™ Duo Pen<br>Needle | Comparison |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K110703 | N/A |
| Product Code | FMI | FMI | Unchanged |
| Regulation Number | 21 CFR 880.5570 | 21 CFR 880.5570 | Unchanged |
| Class | Class II | Class II | Unchanged |
| Intended Use | The BD AutoShield Duo™ Pen<br>Needle is intended for use with<br>pen injector devices for the<br>subcutaneous injection of<br>drugs.<br>The product has two safety<br>shields, which lock in place after<br>use (patient-end) and upon<br>removal of the needle from the<br>pen (pen connection-end). The<br>locked shields help reduce the<br>occurrence of needle sticks<br>from both ends of the needle. | The BD AutoShield Duo™ Pen<br>Needle is intended for use with<br>pen injector devices for the<br>subcutaneous injection of<br>drugs.<br>The product has two safety<br>shields, which lock in place after<br>use (patient-end) and upon<br>removal of the needle from the<br>pen (pen connection-end). The<br>locked shields help reduce the<br>occurrence of needle sticks<br>from both ends of the needle. | Unchanged, with<br>addition of word<br>"subcutaneous" to<br>clarify proper<br>injection depth |
| Needle Gauge | 30G | 30G and 31G | Unchanged |
| Needle Length | 5mm and 8mm | 5mm and 8mm | Unchanged |
| | Outer Cover | Outer Cover | Unchanged |
| | Sleeve | Sleeve | Unchanged |
| | Outer Shield | Outer Shield | Unchanged |
| | Inner Shield | Inner Shield | Unchanged |
| Pen Needle<br>Components | Hub | Hub | Unchanged |
| | Cannula | Cannula | Unchanged |
| | Non-patient end Shield | Non-patient end Shield | Unchanged |
| | Tear Drop Label | Tear Drop Label | Unchanged |
| | Patient end &<br>Non-patient end Springs | Patient end &<br>Non-patient end Springs | Unchanged |
| Needle insertion<br>method | Manual | Manual | Unchanged |
| Provided Sterile | Yes (gamma irradiation) | Yes (gamma irradiation) | Unchanged |
| Component Materials | Outer Cover | Polyethylene | Unchanged |
| | Polyethylene | | Unchanged |
{5}------------------------------------------------
| Feature | Subject Device:<br>BD AutoShield Duo™ Pen<br>Needle | Predicate Device:<br>BD AutoShield™ Duo Pen<br>Needle | Comparison |
|---------------------------------------|-----------------------------------------------------|-------------------------------------------------------|------------------------------------|
| Sleeve | Polycarbonate | Polycarbonate | Introduction of<br>alternate resin |
| Outer Shield | Polycarbonate | Polycarbonate | Introduction of<br>alternate resin |
| Inner Shield | Polycarbonate | Polycarbonate | Introduction of<br>alternate resin |
| Hub | Polypropylene | Polypropylene | Unchanged |
| Cannula | Stainless Steel | Stainless Steel | Unchanged |
| Lubricant | Medical Grade Lubricant | Medical Grade Lubricant | Unchanged |
| Non-patient end Shield | Polycarbonate | Polycarbonate | Introduction of<br>alternate resin |
| Tear Drop Label | Paper | Paper | Unchanged |
| Patient end & Non-patient end Springs | Stainless Steel | Stainless Steel | Unchanged |
The differences between the predicate and subject device do not raise any new questions of safety or effectiveness.
#### Performance Testing:
#### Non-Clinical Test Summary
The subject device has the same technological characteristics as the predicate device cleared under K110703. BD has validated the design of the subject device as part of its design control process in accordance with the Quality System Regulation. This testing included functional performance per ISO 11608-2:2012, ISO 9626:2016, ISO 23908:2011, BD internal test requirements, as well as Biocompatibility testing per ISO 10993-1:2018.
#### Functional Performance Testing
- ISO 11608-2:2012 - Needle-based injection systems for medical use - Requirements and test methods - Part 2: Needles
- . ISO 9626:2016 - Stainless steel needle tubing for the manufacture of medical devices requirements and test methods
- ISO 23908:2011 Sharps injury protection Requirements and test methods Sharps protection . features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- . BD internal test requirements
#### Biocompatibility Testing:
The following testing was conducted according to ISO 10993-1:2018 – Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process:
- Cytotoxicity
- Intracutaneous Reactivity
- Skin Sensitization ●
- Acute Systemic Toxicity
- Subacute/Subchronic Toxicity
- Genotoxicity - Bacterial and Mammalian
- Material-Mediated Pyrogenicity ●
- Implantation ●
{6}------------------------------------------------
- . Bacterial Endotoxin
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
Sterility, Shipping and Shelf-Life:
- The BD AutoShield Duo™ Pen Needle is sterilized using a validated gamma irradiation . method by Cobalt-60. It has been validated via the method described in ISO 11137-2:2013 -Sterilization of Health Care Product – radiation – Part 2: Establishing the Sterilization Dose. The sterilization parameters were chosen to assure the sterilization dose provides a minimum SAL of 10-6.
- . Residuals are not applicable for the gamma irradiation sterilization method
- . Limulus Amebocyte Lysate (LAL) assay was used to measure the endotoxin limit and the requirement was met. The product is non-pyrogenic.
- . This device is packaged in an Outer Cover with a tamper evident peel-away tear drop label. The tear drop label is the sterility barrier of the medical device
- o Sterile barrier testing performed on the subject device:
- . Tear drop label removal force
- . Seal integrity
- . Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems. All packaging was deemed acceptable for protection of product and sterility maintenance.
- Accelerated stability testing has been conducted to validate the sterility and performance of the BD ● AutoShield Duo™ Pen Needle device to support a shelf-life of 3 years.
#### Clinical Test Summary
No clinical study was included in this submission.
### Conclusion:
The results of functional performance and biocompatibility testing passed and successfully demonstrated the subject device met requirements. The non-clinical testing has demonstrated the is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.