K223267 · Konica Minolta, Inc. · MQB · Nov 17, 2022 · Radiology
Device Facts
Record ID
K223267
Device Name
SKR 3000
Applicant
Konica Minolta, Inc.
Product Code
MQB · Radiology
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SKR 3000 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpose diagnostic procedures. The SKR 3000 is not indicated for use in mammography, fluoroscopy and angiography applications.
Device Story
SKR 3000 is a digital radiography system using X-ray planar detectors to capture human anatomy images. System inputs digital signals from detectors into Console CS-7 software for processing; outputs diagnostic images to filing systems, printers, or displays. Used in clinical settings for general-purpose diagnostics, including pediatric patients. Modifications introduce wireless serial radiography functionality using P-65/P-75 detectors. CS-7 software performs image processing (gradation, frequency, dynamic range compression, smoothing, rotation, zooming, grid removal/scattered radiation correction via Intelligent-Grid). Healthcare providers use processed images for diagnostic decision-making. Benefits include replacement of traditional film/screen systems with digital workflow.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing, verification, and validation activities, including risk management analysis, confirming compliance with AAMI/ANSI ES 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Stationary X-ray system; digital planar detectors (P-61, P-71, P-81, P-65, P-75, P-82, P-85, P-95). Connectivity: Wireless and wired. Software: Console CS-7 for image processing. Standards: AAMI/ANSI ES 60601-1 (Ed.3.1), IEC 60601-1-2 (Ed.4.0).
Indications for Use
Indicated for generating radiographic images of human anatomy in general-purpose diagnostic procedures. Includes pediatric patients. Not indicated for mammography, fluoroscopy, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress (K221803)
Submission Summary (Full Text)
{0}------------------------------------------------
November 17, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Konica Minolta, Inc. % Jan Maniscalco Executive Vice President QA/RA Konica Minolta Healthcare Americas, Inc. 411 Newark Pompton Turnpike WAYNE NJ 07470
#### Re: K223267
Trade/Device Name: SKR 3000 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, LLZ Dated: October 24, 2022 Received: October 24, 2022
### Dear Jan Maniscalco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Zhkke
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223267
Device Name SKR 3000
Indications for Use (Describe)
The SKR 3000 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpose diagnostic procedures.
The SKR 3000 is not indicated for use in mammography, fluoroscopy and angiography applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------|
| <span></span> | |
| <span></span> | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circular shape at the top with several horizontal white lines running through it. Below the circular shape, the words "KONICA MINOLTA" are written in a bold, sans-serif font.
# 510(k) Summary K223267
| Company: | KONICA MINOLTA, INC. | |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | 1 Sakura-machi, Hino-shi, Tokyo, 191-8511, Japan | |
| Contact: | Tsutomu Fukui<br>Senior Manager of Regulatory & QMS Division<br>Quality Assurance Operations<br>Healthcare Business Headquarters<br>Telephone: +81 42 589 8429<br>Email: tsutomu.fukui1@konicaminolta.com | |
| Date Prepared: | November 04, 2022 | |
| Device Name: | SKR 3000 | |
| Common Name: | Digital Radiography | |
| Regulation Number: | 21 CFR 892.1680 | |
| Regulation Name: | Stationary x-ray system | |
| Regulatory Class: | Class II | |
| Product Code(s): | MQB, LLZ | |
| Predicate Device: | K213908 - SKR 3000 (KONICA MINOLTA, INC.) | |
| | Regulation Number: 21 CFR 892.1680 | |
| | Regulation Name: Stationary x-ray system | |
| | Regulatory Class: Class II | |
| | Product Codes: MQB, LLZ | |
| Reference Device: | K221803 - PHOENIX/AeroDR TX m01 and<br>PHOENIX/mKDR Xpress (SEDECAL SA) | |
| | Regulation Number: 21 CFR 892.1720 | |
| | Regulation Name: Mobile x-ray system | |
| | Regulatory Class: Class II | |
| | Product Codes: IZI, MQB | |
ー Page 1 of 4 一
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circle with horizontal white lines running through the center. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.
# Device Description
The digital radiography SKR 3000 performs X-ray imaging of the human body using an Xray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.
- This device is not intended for use in mammography
- This device is also used for carrying out exposures on children.
The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. The CS-7 is a software with Moderate level of concern. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation correction - (Intelligent-Grid). The Intelligent-Grid is cleared in K151465. The CS-7 modifications have been made for a wireless serial radiography.
The SKR 3000 is distributed under a commercial name AeroDR 3.
This submission is to introduce a wireless serial radiography into the SKR 3000 system. The wireless serial radiography function of P-65 / P-75 used with Phoenix was cleared under K221803. These detectors are wireless and their serial radiography functions are not being controlled by the x-ray generator. Hence no detector integration testing is necessary.
# Indications for Use
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, and angiography applications.
# Comparison Table
The comparison to the predicate device was summarized in the table blow.
- Page 2 of 4 -
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue globe-like shape with horizontal white lines running across it. Below the globe is the text "KONICA MINOLTA" in a bold, sans-serif font.
| | Subject Device | Predicate Device |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SKR 3000 | SKR 3000 |
| 510(K) Number | K223267 | K213908 |
| Indications for Use | This device is indicated for use in<br>generating radiographic images of<br>human anatomy. It is intended to replace a<br>replace radiographic film/screen<br>system in general-purpose diagnostic<br>procedures.<br>This device is not indicated for use in<br>mammography, fluoroscopy, and<br>angiography applications. | This device is indicated for use in<br>generating radiographic images of<br>human anatomy. It is intended to replace a<br>replace radiographic film/screen<br>system in general-purpose diagnostic<br>procedures.<br>This device is not indicated for use in<br>mammography, fluoroscopy, and<br>angiography applications. |
| DR Detector | P-61 (14" x 17")<br>P-71 (17" x 17")<br>P-81 (10" x 12")<br>P-65 (14" x 17")<br>P-75 (17" x 17")<br>P-82 (14" x 17")<br>P-85 (14" x 17")<br>P-95 (17" x 17") | P-61 (14" x 17")<br>P-71 (17" x 17")<br>P-81 (10" x 12")<br>P-65 (14" x 17")<br>P-75 (17" x 17")<br>P-82 (14" x 17")<br>P-85 (14" x 17")<br>P-95 (17" x 17") |
| Serial Radiography | Wired (P-61, P-71, P-65, P-75)<br>Wireless (P-65, P-75) | Wired (P-61, P-71, P-65, P-75) |
| Components or<br>Accessories | AeroDR Interface Units, Detector<br>Interface Units, Power Supply Unit,<br>Generator Interface Units, GIU SZ,<br>Battery Charger, etc. | AeroDR Interface Units, Detector<br>Interface Units, Power Supply Unit,<br>Generator Interface Units, Battery<br>Charger, etc. |
| Operator Console | CS-7 | CS-7 |
| Image Processing | Auto-gradation process<br>Frequency processing (F process)<br>Equalization processing (E process)<br>Hybrid processing (HF, HE, HS)<br>REALISM processing (RF, RE, RS)<br>Intelligent-Grid<br>Automatic exposure field recognition<br>process | Auto-gradation process<br>Frequency processing (F process)<br>Equalization processing (E process)<br>Hybrid processing (HF, HE, HS)<br>REALISM processing (RF, RE, RS)<br>Intelligent-Grid<br>Automatic exposure field recognition<br>process |
# Performance Data
The SKR 3000 is designed to comply with the following standard; AAMI/ANSI ES 60601-1 (Ed.3.1) and IEC 60601-1-2 (Ed.4.0). The verification and validation including the items required by the risk analysis for the SKR 3000 were performed and the results demonstrated that the predetermined acceptance criteria were met. The results of risk
- Page 3 of 4 -
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue sphere at the top with several horizontal white lines across it. Below the sphere, the words "KONICA MINOLTA" are written in a bold, sans-serif font.
management did not require clinical studies to demonstrate the substantial equivalency of the subject device modifications.
# Conclusion
The SKR 3000 has the same intended use and indications for use, technological characteristics, and principal operations. The technological differences raised no new issues of safety or effectiveness as compared to its predicate device (K213908). Performance tests demonstrate that the SKR 3000 performs according to specifications and functions as intended. It is concluded that the subject device is to be as safe and effective as the predicate and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.