K223237 · Sofwave Medical, Ltd. · OHV · Dec 16, 2022 · General, Plastic Surgery
Device Facts
Record ID
K223237
Device Name
SofWave System
Applicant
Sofwave Medical, Ltd.
Product Code
OHV · General, Plastic Surgery
Decision Date
Dec 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SofWave System is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older. The SofWave System is also intended for short-term improvement in the appearance of cellulite.
Device Story
System uses high-frequency ultrasonic pulses to elevate dermal temperature, causing controlled thermal damage to improve skin appearance. Consists of console (power, cooling, UI) and applicator (ultrasonic transducer array, active cooling). Used in clinical settings by trained operators. Applicator delivers continuous acoustic waves to skin; cooling element protects epidermis. Healthcare providers use device to treat facial/neck laxity and cellulite. Output is thermal coagulation in focal zone (1-2 mm depth). Clinical benefit includes improved skin appearance and lifting effects. No hardware/software changes required for cellulite indication; performance verified via bench testing and clinical study.
Clinical Evidence
Prospective clinical study (n=69) evaluated cellulite improvement. Subjects received 2 treatments 2-4 weeks apart. Primary endpoints: blinded review of clinical photographs (89% correct identification of post-treatment images), Cellulite Severity Scale (CSS), Global Aesthetic Improvement Scale (GAIS), and Laxity Scale (LS). Results showed clinical improvement consistent with predicate. Safety: 1 unrelated AE; mean pain score 4.6 (moderate).
Technological Characteristics
High-intensity, non-focused ultrasound (10-12 MHz). Components: console and applicator with piezoelectric ceramic plates (7x5 mm²). Cooling: thermoelectric cooler (TEC) and heat exchanger. Thermal coagulation confined to focal zone (<3 mm depth). Electrical safety: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-62. Biocompatibility: ISO 10993.
Indications for Use
Indicated for subjects aged 22 and older for non-invasive aesthetic treatment of facial lines/wrinkles, eyebrow lifting, lifting of lax submental and neck tissue, and short-term improvement in cellulite appearance.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
December 16, 2022
Sofwave Medical Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street Floor 23 Philadelphia, Pennsylvania 19103
Re: K223237
Trade/Device Name: SofWave System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System For Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: October 19, 2022 Received: October 19, 2022
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Carr -S
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K223237
Device Name
SofWave System
Indications for Use (Describe)
The Softem is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older. The SoftWave System is also intended for short-term improvement in the appearance of cellulite.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Sofwave Medical's SofWave System
# K223237
#### Submitter's Name, Address, Telephone Number, Contact Person, and Date Prepared
Sofwave Medical Ltd. 1 Ha-Otsma St. Yokneam Ilit, Israel 2069200
Submission Correspondent:
Janice M. Hogan Hogan Lovells US LLP janice.hogan@hoganlovells.com (267) 675-4611
Date Prepared: December 15, 2022
#### Name of Device:
SofWave System
#### Common or Usual Name:
Focused Ultrasound Stimulator System for Aesthetic Use
#### Classification Name:
21 CFR 878.4590 (Ultrasound for Tissue Heat or Mechanical Cellular Disruption), Class II, product code OHV
#### Predicate Devices
Sofwave Medical's SofWave System (K211483) (Predicate Device)
Soliton Inc.'s Rapid Acoustic Pulse (RAP) Device (K201801) (Predicate device)
#### Intended Use / Indications for Use
The SofWave System is indicated for use as a non-invasive dermatological aesthetic treatment to improve facial lines and wrinkles, lift the eyebrow, and lift lax submental (beneath the chin) and neck tissue; which can also affect the appearance of lax tissue in the submental and neck regions for subjects aged 22 and older. The SofWave System is also intended for short-term improvement in the appearance of cellulite.
{4}------------------------------------------------
#### Technological Characteristics
The SofWave System is an ultrasound system intended for aesthetic purposes. The system generates high frequency ultrasonic pulses that elevate the temperature in the dermis layer and cause controlled isolated areas of thermal damage.
The SofWave System consists of two main functional components: 1) the console and 2) the applicator. The console includes the power sources, cooling unit, electrical components and the user interface. The applicator is comprised of an array of ultrasonic transducers that emit continuous acoustic waves and an active cooling element that is used to cool the skin area in contact with the applicator. The applicator is connected by a flexible cable to the console.
#### Comparison of Technological Characteristics with the Predicate Device
The SofWave System has similar technological characteristics compared to the predicate devices. The subject SofWave device is almost identical to the previously SofWave device that was cleared in K211483, except minor hardware and software changes, which do not significantly affect clinical functionality or performance specifications of the device, and have been verified and tested. The treatment parameter or energy specification of the device remain identical to that cleared in K211483. The purpose of this 510(k) notification is to expand the indications for use to include short-term improvement in the appearance of cellulite. This additional indication for use does not require any additional hardware or software changes to the device.
The subject SoftWave device is also similar to the Soliton RAP predicate device (K201801) that is indicated for short-term improvement in the appearance of cellulite. Both devices are comprised of a console and a treatment applicator. Both devices deliver acoustic energy into the skin to alter the structure of the tissue, resulting in improved appearance of cellulite. The minor differences in the technology do not raise different questions of safety or effectiveness. For both devices, the same types of safety and efficacy questions arise. Clinical testing further demonstrates the performance of the SofWave device, showing that the magnitude of the clinical effects with treatment is comparable between SofWave and the Soliton RAP device.
{5}------------------------------------------------
| | Sofwave Medical's<br>SofWave System<br>(Subject Device) | Sofwave Medical's<br>SofWave System<br>(K211483)<br>(Predicate Device) | Soliton RAP for Cellulite<br>(K201801)<br>(Predicate Device) |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Class | II | II | II |
| CFR Regulation | 21 CFR 878.4590 | 21 CFR 878.4590 | 21 CFR 878.4810 |
| Product Code | OHV | OHV | GEX |
| Intended Use | The SofWave System is<br>indicated for use as a non-<br>invasive dermatological<br>aesthetic treatment to<br>improve facial lines and<br>wrinkles, lift the eyebrow,<br>and lift lax submental<br>(beneath the chin) and neck<br>tissue; which can also affect<br>the appearance of lax tissue<br>in the submental and neck<br>regions for subjects aged<br>22 and older. The SofWave<br>System is also intended for<br>short-term improvement in<br>the appearance of cellulite. | The SofWave System is<br>indicated for use as a non-<br>invasive dermatological<br>aesthetic treatment to<br>improve facial lines and<br>wrinkles, lift the eyebrow,<br>and lift lax submental<br>(beneath the chin) and neck<br>tissue; which can also affect<br>the appearance of lax tissue<br>in the submental and neck<br>regions for subjects aged<br>22 and older. | The Rapid Acoustic Pulse<br>Device (RAP) is indicated for<br>use as an accessory to the<br>1064 nm Q-Switched laser<br>for black ink tattoo removal<br>in Fitzpatrick Skin Type I-III<br>patients. The RAP device is<br>also intended for short-term<br>improvement in the<br>appearance of cellulite. |
| Device<br>Technology | High intensity, non-focused<br>ultrasonic pulse that can be<br>delivered percutaneously to<br>tissues to reduce fibrous<br>septa's tendency to deform<br>the skin surface | High intensity, non-focused<br>ultrasonic pulse that can be<br>delivered percutaneously to<br>tissues | Acoustic pulse that partially<br>disrupts the fibrous septa to<br>improve cellulite<br>appearance. |
| Type of Energy | High intensity therapeutic<br>ultrasound | High intensity therapeutic<br>ultrasound | Acoustic shockwave |
| System<br>components | Console that includes<br>• the power sources,<br>electrical components<br>and user interface<br>(touchscreen)<br>• Handpiece | Console that includes<br>• the power sources,<br>electrical components<br>and user interface<br>(touchscreen)<br>• Handpiece | Console that includes<br>• the power sources,<br>electrical components<br>and user interface<br>(touchscreen)<br>• Handpiece |
| Energy Type | High Intensity non-focused<br>Ultrasound | High Intensity non-focused<br>Ultrasound | Broadband/Ultrasonic |
| Treatment Depth | 1-2 mm | 1-2 mm | 1-10 mm* |
| Tissue at Focal<br>Point Temperature | 60°C -70°C | 60°C -70°C | N/A |
| Thermal Energy<br>Output per<br>channel | < 5J per channel | < 5J per channel | N/A |
| Thermal<br>Coagulation Point | Confined to focal zone;<br>shallow (<3 mm); no<br>thermal coagulation below<br>focal zone | Confined to focal zone;<br>shallow (<3 mm); no<br>thermal coagulation below<br>focal zone | N/A |
| Epidermal Impact | Non-invasive; Cooling<br>required | Non-invasive; Cooling<br>required | Non-invasive |
| | Sofwave Medical's<br>SofWave System | Sofwave Medical's<br>SofWave System<br>(K211483) | Soliton RAP for Cellulite<br>(K201801) |
| | (Subject Device) | (Predicate Device) | (Predicate Device) |
| Transducer<br>Acoustic Core | Energizer comprises:<br>- Array of<br>piezoelectric<br>ceramic plates (7 x<br>5 mm²)<br>- Temperature<br>control unit<br>(thermistors,<br>Thermoelectric<br>cooler (TEC), Heat<br>Exchanger) | Energizer comprises:<br>- Array of<br>piezoelectric<br>ceramic plates (7 x<br>5 mm²)<br>- Temperature<br>control unit<br>(thermistors,<br>Thermoelectric<br>cooler (TEC), Heat<br>Exchanger) | Cartridge is comprised of<br>electrodes immersed in the<br>circulating saline |
| Frequency | 10 MHz – 12 MHz | 10 MHz – 12 MHz | N/A |
| Treatment Area<br>Width | 35mm | 35mm | 38mm |
| Peak Acoustic<br>Pressure | 1.2 MPa | 1.2 MPa | 0.25 to 12 MPa |
| User Interface | LCD Touch Screen Graphic<br>User Interface | LCD Touch Screen Graphic<br>User Interface | LCD Touch Screen Graphic<br>User Interface |
| Electrical<br>Safety/EMC | IEC 60601-1 Compliant<br>IEC 60601-1-2 Compliant | IEC 60601-1 Compliant<br>IEC 60601-1-2 Compliant | IEC 60601-1 Compliant<br>IEC 60601-1-2 Compliant |
| Input Power | 100-240VAC<br>60Hz<br>10A | 100-240VAC<br>60Hz<br>10A | 240VAC<br>60Hz<br>15A |
# SofWave System Substantial Equivalence Chart
{6}------------------------------------------------
* Based on publicly available information for Soliton
#### Performance Data
The following nonclinical performance testing has been conducted to support the substantial equivalence of the SofMave System to its predicate device, consistent with FDA's "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the SofWave System functioned as intended.
- Biocompatibility of the patient-contacting components of the device was established in accordance with ISO 10993
- . Software verification and validation was performed, and demonstrated that the software performs as intended
- . Electrical Safety and Electromagnetic Compatibility was established in accordance with IEC 60601-1-2, IEC 60601-1, IEC 60601-1-6, and IEC 60601-2-62
- . Functional bench testing was conducted to verify that the modifications to the device did not affect the device performance
To support the expansion of indications for cellulite, the company conducted a clinical study that evaluated the safety and effectiveness of the non-invasive dermatological aesthetic improvement in cellulite appearance. A total of 69 subjects were enrolled and 68 subjects were treated at 4 sites in the United States.
Eligible patients received 2 treatments (2-4 weeks apart) using the SofWave System on one side (right or left) of the lateral / posterior upper thigh or buttocks. Serial clinical photographs were
{7}------------------------------------------------
collected under standardized conditions before treatment (baseline) and at the 3-month follow-up visits. Photographs were assessed by blinded independent reviewers to identify pre-treatment images when compared to post treatment images and to grade the pre-treatment and posttreatment images using the Cellulite Severity Scale (CSS). Improvement was also evaluated using the Global Aesthetic Improvement Scale (GAIS) and Laxity Scale (LS). Safety assessments included evaluation of AEs via physician examination during and after the treatment.
The blinded reviewers correctly identified the post treatment images for 89% of the subjects. Throughout the study, there was only one adverse event reported, which was not related to the device. No serious or unanticipated adverse event was reported during the study. The mean pain level was 4.6 (moderate pain). Based on a literature review, the study results were consistent with the predicate device.
## Conclusion
SoftWave has the same general intended use and similar indications, technological characteristics, and principles of operation as the company's previously cleared SoftWave System (K211483) and Soliton's RAP device (K201801) ("predicate devices"). The minor technological differences between the subject and the predicate devices do not raise different questions of safety or effectiveness. Performance testing of the device has demonstrated that the device performs as intended and thus, is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.