K223217 · Zynex Medical, Inc. · DSB · Jun 16, 2023 · Cardiovascular
Device Facts
Record ID
K223217
Device Name
Zynex Monitoring System, Model CM-1600
Applicant
Zynex Medical, Inc.
Product Code
DSB · Cardiovascular
Decision Date
Jun 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The Zynex Monitoring System, Model CM-1600 continuously monitors a patient's physiological parameters. These parameters include Bioelectrical Impedance, Electrocardiogram (ECG) Amplitude, Photoplethysmography (PPG) Amplitude, and Skin Temperature. This device is intended to be used in a professional medical environment (e.g., hospitals, and research institutions). The CM-1600 is a standalone device intended for desktop or mounted use (e.g., mounted to an IV pole), where operation is to be performed as uninterrupted patient monitoring. The CM-1600 shall only be used by a qualified operator. The operator shall have knowledge of the system and data interpretation obtained via medical education, system documentation, and/or specific training. The device does not report any diagnosis but provides numerical values to aid the diagnosis by a physician; it is the physician's responsibility to make proper judgements based on these measurements. IMPORTANT: This device must be ordered or prescribed by a licensed physician.
Device Story
System comprises wearable sensor and monitor tablet; collects bioelectrical impedance, ECG amplitude, PPG amplitude, and skin temperature. Wearable transmits data wirelessly to monitor. Device calculates 'Relative Index' value representing combined changes in parameters. Used in professional medical environments (hospitals, research institutions) by qualified operators (clinicians, surgeons, anesthesiologists). Output displayed on tablet as trend graphs and parameter values; aids physician in clinical decision-making and post-operative monitoring. Benefits include continuous data monitoring to improve post-operative outcomes and provider confidence.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and wireless coexistence (AAMI-TIR63).
Technological Characteristics
System includes wearable sensor, monitor tablet, and charging station. Sensing: bioelectrical impedance, ECG, reflective PPG, and skin temperature. Connectivity: wireless transmission between wearable and tablet. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993, ISO 14971. Form factor: wearable sensor and desktop/pole-mounted tablet.
Indications for Use
Indicated for monitoring bioelectrical impedance, ECG amplitude, PPG amplitude, and skin temperature, and their relative changes in fluid volume in adult patients.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 16, 2023
Zynex Medical, Inc. Donald Gregg Zynex Monitoring Solutions 9555 Maroon Circle Englewood, Colorado 80112
Re: K223217
Trade/Device Name: Zynex Monitoring System, Model CM-1600 Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: DSB, DQA Dated: December 5, 2022 Received: December 6, 2022
# Dear Donald Gregg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K223217
#### Device Name
Zynex Monitoring System, Model CM-1600
Indications for Use (Describe)
Monitoring of the following parameters and their relative of relative of relative changes in fluid volume in adult patients:
- · Bioelectrical Impedance
- ECG Amplitude
- PPG Amplitude
- · Skin Temperature
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Zynex Monitoring Solutions. The logo features the word "Zynex" in a bold, sans-serif font, with the words "Monitoring Solutions" in a smaller font below. A blue swoosh arcs over the word "Zynex", adding a dynamic element to the design. The overall design is clean and professional.
# 510(k) Summary Zynex Monitoring System, Model CM-1600
This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.
# 1. Submitter's Information
| Submitted by: | Zynex Monitoring Solutions, Inc. |
|---------------|-----------------------------------------------|
| | 9555 Maroon Circle<br>Englewood, CO 80112 USA |
Phone number: +1-800-495-6670 Fax Number: +1-866-870-4089
- Contact person: Donald Gregg President, Zynex Monitoring Solutions Phone number: +1-800-495-6670 E-mail: dgregg@zynex.com
Date of Summary: 07-October-2022
#### 2. Device Information
| Trade name: | Zynex Monitoring System, Model CM-1600 |
|----------------------|--------------------------------------------------------------------------------------|
| Common name(s): | Impedance cardiograph (ICG)<br>Electrocardiograph (ECG)<br>Photoplethysmograph (PPG) |
| Classification name: | See Table 1 |
| Device class: | Class II |
#### Table 1. Classification Name
| Classification Name | 21 CFR Section | Product Code |
|-----------------------------------------------------------------------------|----------------|--------------|
| Plethysmograph, Impedance | 870.2770 | DSB |
| Oximeter | 870.2700 | DQA |
| Thermometer, Electronic, Clinical | 880.2910 | FLL |
| Monitor, Physiological, Patient (Without Arrhythmia Detection<br>Or Alarms) | 870.2300 | MWI |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Zynex Monitoring Solutions. The word "Zynex" is written in a bold, sans-serif font, with the words "Monitoring Solutions" written in a smaller font below it. A blue arc is above the word "Zynex", adding a design element to the logo. The logo is simple and professional, and it is likely used to represent the company's brand.
# 3. Predicate Devices
The Zynex Monitoring System, Model CM-1600, is substantially equivalent in design (methodology) and indications for use to the predicate device shown in Table 2.
#### Table 2. Predicate Device
| Device Name | Manufacturer | 510(k) |
|---------------------------------------|---------------------|---------|
| Cardiac Monitor, Model 1500 (CM-1500) | Zynex Medical, Inc. | K191697 |
# 4. Description
Patient monitoring with the CM-1600 is intended to provide data to clinicians, doctors, and postoperative care providers. Accurate diagnostic tools and continuous data monitoring systems are valuable additions for anesthesiologists and surgeons through improving post-operative outcomes and provider confidence. The CM-1600 System is comprised of two (2) subsystems, the Wearable and the Monitor. The CM-1600 Wearable is designed and developed by Zynex Monitoring Solutions, and it collects physiological parameters and transmits those parameters to the CM-1600 Monitor via wireless communication. The CM-1600 Monitor is a Zynex-branded, third-party all-in-one medical grade tablet.
The Zynex Monitoring System, Model CM-1600, simultaneously monitors various parameters of a patient's body. These parameters include Bioelectrical Impedance, Electrocardiogram (ECG) Amplitude, Photoplethysmography (PPG) Amplitude, and Skin Temperature. A Relative Index value is calculated as a combination of the changes in these parameters and is represented by a single number.
# 5. Indications for use
Monitoring of the following parameters and their relative of relative of relative changes in fluid volume in adult patients:
- Bioelectrical Impedance
- ECG Amplitude
- PPG Amplitude
- Skin Temperature
# 6. Intended Use
The Zynex Monitoring System, Model CM-1600 continuously monitors a patient's physiological parameters. These parameters include Bioelectrical Impedance, Electrocardiogram (ECG) Amplitude, Photoplethysmography (PPG) Amplitude, and Skin Temperature.
This device is intended to be used in a professional medical environment (e.g., hospitals, and
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Zynex Monitoring Solutions. The logo features the word "Zynex" in a bold, sans-serif font, with the words "Monitoring Solutions" in a smaller font below. A blue, curved line arches over the word "Zynex", adding a visual element to the logo.
research institutions). The CM-1600 is a standalone device intended for desktop or mounted use (e.g., mounted to an IV pole), where operation is to be performed as uninterrupted patient monitoring.
The CM-1600 shall only be used by a qualified operator. The operator shall have knowledge of the system and data interpretation obtained via medical education, system documentation, and/or specific training. The device does not report any diagnosis but provides numerical values to aid the diagnosis by a physician; it is the physician's responsibility to make proper judgements based on these measurements.
IMPORTANT: This device must be ordered or prescribed by a licensed physician.
# 7. Summary of Technical Comparison with Predicate Device(s)
The Zynex Monitoring System, Model CM-1600 is a wireless version of the CM-1500, which monitors the same physiological and is indicated for use as a noninvasive monitor.
The CM-1600 features similar technology and techniques of obtaining physiological data as the CM-1500. The impedance cardiography (ICG) and electrocardiography (ECG) sections of the CM-1600 consist of similar architecture as the Cardiac Monitor, Model 1500. Both devices capture these signals with identical surface electrodes. For skin temperature monitoring, the Zynex Monitoring System, Model CM-1600, has identical indications for use, methodology, and processing as the Cardiac Monitor, Model 1500. For plethysmography amplitude monitoring the Zynex Monitoring System, Model CM-1600, method of measuring plethysmograph signal is reflective photoplethysmography on the bottom of the Wearable, whereas the Cardiac Monitor, Model CM-1500 used transmissive technology. For plethysmography amplitude monitoring, the CM-1600 has identical indications for use and similar processing as that of the predicate.
The basic elements of the UI (e.g., Relative Index™ trend graph, PPG waveform, and parameter pane locations) are identical between the Zynex Monitoring System, Model CM-1600 and the predicate.
#### 8. Performance Data
The following performance data were provided in support of substantial equivalence determination:
| Test Name | Testing Criteria | Test Results (Pass/Fail) |
|---------------------------------------------------------------------------------|---------------------------|--------------------------|
| Biocompatibility | ISO 10993 | Pass |
| Device Safety, Electrical Safety,<br>and Electromagnetic Compatibility<br>(EMC) | IEC 60601-1 and 60601-1-2 | Pass |
| Wireless Coexistence Testing | AAAMI-TIR63 | Pass |
| Animal Testing | N/A | N/A |
| Clinical Testing | N/A | N/A |
| Performance Testing | V&V/Predicate Testing | Pass |
#### Table 3. Performance test names, criteria, and results for the CM-1600.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Zynex Monitoring Solutions. The logo features the word "Zynex" in a bold, sans-serif font, with the words "Monitoring Solutions" in a smaller font below. A blue swoosh arcs over the word "Zynex", adding a dynamic element to the design.
Zynex Monitoring Solutions Inc. 9555 Maroon Circle, Englewood, CO 80112 Main: 800-495-6670
#### Biocompatibility Testing
The biocompatibility evaluation of the CM-1600 was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization ●
- . Irritation
#### Device Safety, Electrical safety, and electromagnetic compatibility (EMC)
Mechanical safety, Electrical safety and EMC testing were conducted on the CM-1600, consisting of the Wearable, Monitor, and Charging Station. The system complies with the IEC 60601-1 standard and the IEC 60601-1-2 standard for EMC.
Animal Testing: No Testing Required.
Clinical Testing: No Testing Required.
| Standard Number | Standard Name |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005+AMD1:2012<br>CSV | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| IEC 60601-1-2:2014 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Electromagnetic disturbances - Requirements and<br>tests |
| ISO 14971:2019 | Medical devices - Risk management - Application of risk<br>management to medical devices |
| ISO 10993-1:2009 | Biological evaluation of medical devices - Part 1: Evaluation<br>and testing |
#### Table 4. Medical Device Standards Compliance (full list in Section 9)
The data provided from the performance testing demonstrates that the Zynex Monitoring System, Model CM-1600 meets all applicable requirements and is substantially equivalent to the predicate device, the Zynex Monitoring Solutions Cardiac Monitor, Model 1500 (CM-1500).
# 9. Conclusion
The CM-1600 device is as safe and as effective as the predicate device, the CM-1500, when considering device indication for use, performance, safety, EMC, biocompatibility, and other data presented above. Therefore, the CM-1600 is substantially equivalent to the predicate device. Any minor differences between the subject and predicate devices do not raise any different safety or effectiveness concerns with the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.