The ClotTriever Thrombectomy System is indicated for: • The non-surgical removal of thrombi and emboli from blood vessels. • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).
Device Story
ClotTriever XL Catheter is a single-use, sterile embolectomy catheter for peripheral vasculature (10-28 mm vessels). Device consists of reinforced polymeric coaxial shafts with an expandable nitinol coring element and thrombus collection bag. Operated by physician under fluoroscopic guidance; catheter advanced beyond clot; self-expanding net deployed to core, separate, and entrap thrombus; net collapsed and retracted into ClotTriever/Protrieve sheath for removal. Syringe provided for aspiration and fluid infusion. Benefits include non-surgical thrombus removal, reducing need for invasive procedures.
Clinical Evidence
No clinical data required. Substantial equivalence supported by non-clinical bench testing (packaging, mechanical, simulated use, vacuum, leakage) and GLP animal safety/functional testing. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, reactivity, systemic toxicity, and hemocompatibility.
Technological Characteristics
Reinforced polymeric coaxial shafts; expandable nitinol wire net coring element; thrombus collection bag. Materials include Pebax, PTFE, and radiopaque markers. Dimensions: 16 Fr OD, 105 cm length. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6.
Indications for Use
Indicated for non-surgical removal of thrombi and emboli from peripheral blood vessels, including deep vein thrombosis (DVT), and for injection, infusion, or aspiration of fluids/contrast media.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 27, 2023
Inari Medical, Inc. Kaitlyn Weinkauf Senior Regulatory Affairs Specialist 6001 Oak Canyon, Suite 100 Irvine, California 92618
Re: K223210
Trade/Device Name: ClotTriever XL Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: October 14, 2022 Received: October 17, 2022
### Dear Kaitlyn Weinkauf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by
Gregory W. Gregory W. O'connell
-S
O'connell -S Date: 2023.04.27
12:05:46 -04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223210
Device Name ClotTriever XL Catheter
Indications for Use (Describe) The ClotTriever Thrombectomy System is indicated for:
- · The non-surgical removal of thrombi and emboli from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The ClotTriever Thrombectorny System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Page 1 of 5
## 510(K) SUMMARY
| Date prepared | April 27, 2023 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>949-600-8433 |
| Contact person | Kaitlyn Weinkauf<br>Sr. Regulatory Affairs Specialist |
| Name of Device | ClotTriever® Thrombectomy System |
| Device Trade Name | ClotTriever® XL Catheter |
| Common name | Embolectomy catheter |
| Regulation name | Embolectomy catheter |
| Classification number | 21 CFR 870.5150 |
| Primary product code | QEW |
| Secondary product<br>code | KRA |
| Regulatory class | II |
| Predicate device | ClotTriever® Thrombectomy System (K212632) |
| Description | The ClotTriever Thrombectomy System is a single-use, sterile medical device designed to<br>remove thrombi and emboli from the peripheral vasculature. The ClotTriever<br>Thrombectomy System consists of ClotTriever 13 Fr and 16 Fr Sheaths, Protrieve Sheath,<br>ClotTriever Catheter, ClotTriever BOLD Catheter, and the ClotTriever XL Catheter, each<br>packaged separately.<br>The ClotTriever Catheter, ClotTriever BOLD Catheter, and the ClotTriever XL Catheter are<br>comprised of reinforced polymeric coaxial shafts terminating in an expandable coring<br>element and thrombus collection bag. Three ports terminating in two stopcocks and a luer<br>lock connection are provided for de-airing the catheter shafts. To aid in fluoroscopic<br>visualization, the ClotTriever Sheaths, Protrieve Sheath, and ClotTriever Catheters have<br>radiopaque distal tips. The ClotTriever XL Catheter is intended for vessels 10 to 28 mm in<br>the peripheral vasculature. |
| Indications for Use | The introduction of the ClotTriever XL Catheter variant does not change the indications for<br>use of the ClotTriever Thrombectomy System.<br>The ClotTriever Thrombectomy System is indicated for:<br>● The non-surgical removal of thrombi and emboli from blood vessels. |
{4}------------------------------------------------
| | • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.<br>The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT). | | |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device modifications | The purpose of this submission is to introduce a ClotTriever BOLD Catheter variant, the ClotTriever XL Catheter, that is larger in diameter and longer in length.<br>There have been no changes to the ClotTriever 13 Fr and 16 Fr Sheaths, the Protrieve Sheath, the ClotTriever Catheter, or the ClotTriever BOLD Catheter. | | |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device | The proposed device and predicate device have a similar design and materials of construction. With the exception of the modifications to the catheter OD, catheter length, bag length, coring element length and diameter, and a minor material change, the predicate and proposed devices are the same. These modifications do not change the basic design or the principles of operation from the predicate device.<br>There are no new or different questions of safety or efficacy.<br>There is no change of intended use or fundamental scientific technology between the proposed device and predicate device. The changes made for the ClotTriever XL Catheter do not change the technological characteristics of the ClotTriever Thrombectomy System. | | |
| | Device | ClotTriever XL Catheter<br>(Proposed) | ClotTriever Thrombectomy<br>System<br>ClotTriever BOLD Catheter<br>Predicate (K212632) |
| | Manufacturer | Inari Medical | Inari Medical |
| | Product Code | QEW, KRA | QEW |
| | Intended Use | The ClotTriever Thrombectomy<br>System is intended for use in the<br>peripheral vasculature including<br>deep vein thrombosis (DVT). | The ClotTriever Thrombectomy<br>System is intended for use in the<br>peripheral vasculature including<br>deep vein thrombosis (DVT). |
| | Indications for Use | The ClotTriever Thrombectomy<br>System is indicated for:<br>• The non-surgical removal of<br>thrombi and emboli from blood<br>vessels.<br>• Injection, infusion, and/or<br>aspiration of contrast media and<br>other fluids into or from a blood<br>vessel.<br>The ClotTriever Thrombectomy<br>System is intended for use in the<br>peripheral vasculature including<br>deep vein thrombosis (DVT). | The ClotTriever Thrombectomy<br>System is indicated for:<br>• The non-surgical removal of<br>thrombi and emboli from blood<br>vessels.<br>• Injection, infusion, and/or<br>aspiration of contrast media and<br>other fluids into or from a blood<br>vessel.<br>The ClotTriever Thrombectomy<br>System is intended for use in the<br>peripheral vasculature including<br>deep vein thrombosis (DVT). |
| | Device Description | The ClotTriever Thrombectomy<br>System consists of the<br>ClotTriever Sheaths, the | The ClotTriever Thrombectomy<br>System consists of the<br>ClotTriever Sheaths, the |
| | Catheter, the ClotTriever BOLD<br>Catheter, and the ClotTriever XL<br>Catheter. The<br>ClotTriever/Protrieve Sheaths<br>are comprised of a reinforced<br>polymeric sheath equipped with<br>a self-expanding distal mesh<br>funnel, a flush/aspiration port,<br>and a proximal hemostasis valve.<br>The ClotTriever Catheter,<br>ClotTriever BOLD Catheter, and<br>ClotTriever XL Catheter are<br>comprised of reinforced<br>polymeric coaxial shafts<br>terminating in an expandable<br>coring element and thrombus<br>collection bag. Other accessories<br>provided include a pre-dilator,<br>the funnel loading tool, and a<br>large bore 60 cc syringe. | ClotTriever Catheter, and the<br>ClotTriever BOLD Catheter.<br>The ClotTriever/Protrieve<br>Sheaths are comprised of a<br>reinforced polymeric sheath<br>equipped with a self-expanding<br>distal mesh funnel, a<br>flush/aspiration port, and a<br>proximal hemostasis valve. The<br>ClotTriever and ClotTriever<br>BOLD Catheters are comprised<br>of reinforced polymeric coaxial<br>shafts terminating in an<br>expandable coring element and<br>thrombus collection bag. Other<br>accessories provided include a<br>pre-dilator, the funnel loading<br>tool, and a large bore 60 cc<br>syringe. | |
| Principles of Operation | The ClotTriever XL Catheter is<br>advanced into the vessel and<br>beyond the clot. The self-<br>expanding braided nitinol wire<br>net is deployed. The expanded<br>net cores, separates, and entraps<br>thrombus from the vessel as it is<br>being drawn to the funnel<br>opening of the<br>ClotTriever/Protrieve Sheath.<br>The net is collapsed and pulled<br>into and through the<br>ClotTriever/Protrieve Sheath<br>with the entrapped clot. A 60 cc<br>syringe is provided for the<br>aspiration of clot in the sheath<br>and the infusion of contrast<br>media and other fluids. | The ClotTriever BOLD<br>Catheter is advanced into the<br>vessel and beyond the clot. The<br>self-expanding braided nitinol<br>wire net is deployed. The<br>expanded net cores, separates,<br>and entraps thrombus from the<br>vessel as it is being drawn to the<br>funnel opening of the<br>ClotTriever/Protrieve Sheath.<br>The net is collapsed and pulled<br>into and through the<br>ClotTriever/Protrieve Sheath<br>with the entrapped clot. A 60 cc<br>syringe is provided for the<br>aspiration of clot in the sheath<br>and the infusion of contrast<br>media and other fluids. | |
| Target Vessel | Peripheral vessels 10-28 mm | Peripheral vessels 6-16 mm | |
| Catheter OD/ID | OD: 16 Fr<br>ID: 0.178" | OD: 12 Fr<br>ID: 0.124" | |
| Catheter Length | 105 cm | 80 cm | |
| Collection Bag | Length: 10 cm<br>Wire Diameter: 0.0040" | Length: 18 cm<br>Wire Diameter: 0.0045" | |
| Catheter Deployed Length | 126 cm | 111cm | |
| Coring Element OD | 30 mm | 16 mm | |
| Delivery Catheter Materials | Liner:<br>Pebax 72D, Propell | Liner:<br>PTFE | |
| | | Proximal segment: | Proximal segment: |
| | | Pebax 63D, Cool Grey 4C | Pebax 63D, Cool Grey 4C |
| | | Intermediate segment:<br>Pebax 63D, Violet C, fluoro-safe<br>marker | Intermediate segment:<br>Pebax 63D, Violet C, fluoro-<br>safe marker |
| | | Distal segment:<br>Pebax 63D, Cool Grey 4C<br>Pebax 55D, Violet C (distal tip) | Distal segment:<br>Pebax 55D, Violet C |
| | Guidewire compatibility | 0.035" | 0.035" |
| | Shelf-Life | 6 months | 2 years |
| | Sterile | SAL 10-6, EO | SAL 10-6, EO |
| | How provided | Sterile, single use | Sterile, single use |
| Summary of<br>substantial<br>equivalence | The proposed device, the ClotTriever XL Catheter, and the predicate device, the ClotTriever<br>BOLD Catheter, have the same indications for use, intended use, principles of operation,<br>and fundamental scientific technology. | | |
| | Non-Clinical Testing | | |
| | In accordance with the design failure modes and effects analysis, verification and validation<br>tests were identified to support the substantial equivalence of the ClotTriever XL Catheter<br>to the predicate device. This testing demonstrated compliance with relevant product<br>specifications. | | |
| | The following tests were performed on the proposed device to establish substantial<br>equivalence: | | |
| | Pouch Seal Visual Inspection and Dye Penetration Packaging Usability Visual and Dimensional Inspection Guidewire Compatibility Protrieve and ClotTriever Sheath Compatibility Radial Force Retraction Force into ClotTriever Sheath Deployment/Retraction Force Retraction force to expand the collection bag Kink Radius Fluid Leakage - Protrieve, ClotTriever Sheath, Catheter Vacuum Testing - Protrieve, ClotTriever Sheath Simulated Use, Track and Rotation (Torque) Simulated Use, Track and Tensile Particulate Matter Conical Fitting with 6% Luer Taper Pre-clinical/GLP Animal Safety Evaluation (including In Vivo Functional<br>Testing/Radiopacity Verification) | | |
{5}------------------------------------------------
# K223210
{6}------------------------------------------------
{7}------------------------------------------------
| The following testing was leveraged from the predicate ClotTriever BOLD Catheter (K212632):<br>Pouch peel and seal strength Corrosion Resistance Handle Torque Sterilization Validation |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.