K223193 · Airos Medical, Inc. · JOW · Dec 9, 2022 · Cardiovascular
Device Facts
Record ID
K223193
Device Name
AIROS 8P Sequential Compression Device
Applicant
Airos Medical, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Dec 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AIROS 8P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use.
Device Story
AIROS 8P is a digitally controlled pneumatic compression device for treating venous or lymphatic disorders. It consists of a main unit that supplies air cycles to compression garments worn on upper/lower extremities or the trunk. The device operates via three modes: Gradient, Pressure, and Peristaltic. It inflates garments from distal to proximal areas to increase blood flow and clear extracellular fluid. Used in home or hospital settings, the device is operated by patients or clinicians based on a physician's prescription. The provider sets treatment parameters; the device delivers cyclic pressure accordingly. Benefits include improved fluid clearance and management of edema/ulcers.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included error indicator, noise, cycle time accuracy, pressure time accuracy, therapeutic performance, therapy time accuracy, and garment integrity.
Technological Characteristics
Digitally controlled pneumatic compression system. Components: electrically generated air source, tubing, and compression garments. Modes: Gradient, Pressure, Peristaltic. Standards: ISO 14971:2019, ISTA 3A:2018, ES 60601-1, IEC 60601-1-2, IEC 61000-3-2, IEC 61000-3-3, IEC 60601-1-6, IEC 60601-1-11.
Indications for Use
Indicated for patients with lymphedema, venous stasis ulcers, venous insufficiency, or peripheral edema requiring gradient pneumatic compression therapy.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 9, 2022
AIROS Medical, Inc Darren Behuniak Vice President, Operations and Marketing 2501 Monroe Blvd. Suite 1200 Audubon, Pennsylvania 19403
Re: K223193
Trade/Device Name: AIROS 8P Sequential Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: October 12, 2022 Received: October 13, 2022
Dear Darren Behuniak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eric E. Richardson -5 D2022.12.09 16:00:05 -05'00'
Nicole Gillette for Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223193
Device Name
AIROS 8P Sequential Compression Device
Indications for Use (Describe)
The AIROS 8P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions:
- Lymphedema
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a green "A" inside, followed by the text "AIROS" in a simple, sans-serif font. Below "AIROS" is the word "MEDICAL" in a smaller font size.
# 510(k) Summary
October 12, 2022
### Submitter:
AIROS Medical, Inc. 2501 Monroe Blvd, Suite 1200 Audubon, PA 19403
# Contact Person:
Darren Behuniak, VP Operations & Marketing Email: dbehuniak(@airosmedical.com Phone: 866-991-6956
#### Common Classification & Proprietary Names
Trade Name: AIROS 8P Sequential Compression Device Common Name: Sequential Compression Device
#### Classification
| Classification Name | 21 CFR Regulation | Product Code | Class |
|-----------------------------|-------------------|--------------|-------|
| Compressible Limb<br>Sleeve | 870.5800 | JOW | II |
# Primary Predicate Device and Reference Device
The AIROS 8P Sequential Compression Device is substantially equivalent to the primary predicate. AIROS 8 Sequential Device, and the reference device, Mego Afek Lympha Press Optimal.
| | Primary Predicate | Reference Device |
|--------------------------|--------------------|------------------------|
| | AIROS 8 Sequential | Mego Afek Lympha Press |
| | Compression Device | Optimal |
| 510(k) | K193068 | K182003 |
| 21 CFR Regulation Number | 870.5800 | 870.5800 |
| Product Code | JOW | JOW |
| Classification | II | II |
# Device Description
AIROS 8P Sequential Compression Device is a pneumatic compression device used for treatment and management of venous or lymphatic disorders. The application of compression is effective by increasing blood flow and encouraging extracellular fluid clearance. The device consists of the mechanical device that is used to set the treatment options and supplies cycles of air to the compression garments. There are three primary treatment modes, Gradient Mode, Pressure Mode, and Peristaltic Mode. The air is supplied at adjustable pressures and sequences and inflates the compression garments from the distal to proximal areas of the body. The
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a gray hexagon with a green "A" inside of it on the left, and the text "AIROS" in gray on the right. Below the word "AIROS" is the word "MEDICAL" in a smaller font, also in gray.
compression garments are supplied in various sizes for the upper and lower extremity areas of the body.
# Indication for Use
The AIROS 8P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions:
- Lymphedema
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
#### Technological Characteristics
The manufacturer believes that the technological characteristics of the AIROS 8P are substantially equivalent to those of the primary predicate device and reference device. The AIROS 8P has similar components to its primary predicate and reference devices and has similar operating principles to the primary predicate and reference devices. The digitally controlled device consists of an electrically generated source of compressed air, tubing to convey the pressurized air to the sleeve, and like the primary predicate and reference devices, pressure is applied cyclically for a specified period of time, according to the physician's prescription. Three primary operating modes, Gradient, Pressure, and Peristaltic are available to the user and multiple size garments are available for the upper extremities, and trunk of the body. The AIROS 8P is identical to the AIROS 8 except for minor component changes that do not affect safety and efficacy of the device and the addition of new models of compression garments. The addition of the Peristaltic Mode is substantially equivalent to the reference device, the Mego Afek Lympha Press Optimal.
#### Functional Performance Testing
Testing was performed and to ensure that the system meets its specifications. The manufacturer believes that the technological characteristics of the AIROS 8P are substantially equivalent to those of the primary predicate and reference devices. The functional performance testing includes the following tests:
- Error Indicator Testing ●
- Noise Testing
- Cycle Time Accuracy
- Pressure Time Accuracy ●
- Therapeutic Performance ●
- Therapy Time Accuracy
- Garment Integrity
www.airosmedical.com 1.866.991.6956
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a stylized "A" inside, and the words "AIROS MEDICAL" to the right of the hexagon. The hexagon is gray, the "A" is green, and the words "AIROS MEDICAL" are gray.
### Standards
- ISO 14971:2019 Medical Devices Application of Risk Management to Medical . Devices
- ISTA 3A:2018 Packaged - Products for Parcel Delivery System Shipment
- ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009((R)2012 and A2:2010/(R)2012 -Medical electrical equipment- Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014 + AMD1:2020 CSV - Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances- Requirements and tests
- IEC 61000-3-2: 2018 A1: 2020 EMC Part 3-2: Limits Limits for harmonic current ● emissions (Equipment input current 16A per phase)
- IEC 61000-3-3: 2013 A1: 2017 EMC Part 3-3: Limits Limitation of voltage changes, ● voltage fluctuations and flicker in public low-voltage supply systems for equipment with rated current ≤ 16 A per phase and not subject to conditional connections
- IEC 60601-1-6: 2010/AMD1: 2013 Medical electrical equipment- Part 1-6: General . requirements for basic safety and essential performance- Collateral Standard: Usability
- IEC 60601-1-11: 2015 Medical electrical equipment- Part 1-11: General requirements ● for basic safety and essential performance- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
# Statement of Substantial Equivalence
The AIROS 8P is substantially equivalent in technology, function, operating parameters, and indications to the AIROS 8 primary predicate device and the Lympha Press Optimal reference device. There are no new risks introduced with the AIROS 8P.
#### Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, AIROS Medical, Inc., believes that the AIROS 8P is substantially equivalent to the primary predicate device and the reference device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.