K223179 · Alere San Diego, Inc. · CGA · Sep 20, 2023 · Clinical Chemistry
Device Facts
Record ID
K223179
Device Name
Cholestech LDX ™ System
Applicant
Alere San Diego, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Sep 20, 2023
Decision
SESE
Submission Type
Dual Track
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Cholestech LDX™ System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes: The Lipid Profile•GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported. The TC•HDL•GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood. The TC•GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood. The Lipid Profile cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported. The TC•HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood. The TC cassette is for the quantitative determination of total cholesterol in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Device Story
Portable analyzer and test cassette system; uses reflectance photometry to measure color changes on reagent pads; determines concentration of total cholesterol, HDL, triglycerides, and glucose in venous or capillary whole blood; calculates LDL, non-HDL, and TC/HDL ratio. Operated by professionals in clinical settings. Modification adds software-enforced limitations for hyperbilirubinemia and pediatric use (under 2 years); analyzer displays warning on screen, printed results, and EMR output. Benefits include rapid point-of-care lipid and glucose assessment for diagnosis/treatment of metabolic disorders.
Clinical Evidence
Bench testing only. Verification studies performed under design control to validate device modifications, including interference testing for conjugated and unconjugated bilirubin. Acceptance criteria met.
Technological Characteristics
Reflectance photometry; enzymatic assays; portable analyzer with LCD; whole blood sample input. Software-based warning system for specific patient populations. Sterilization/materials not specified. Complies with CLSI EP07-A3 for interference testing.
Indications for Use
Indicated for quantitative determination of total cholesterol, HDL cholesterol, triglycerides, and glucose in whole blood (fingerstick or venipuncture) for patients aged 2 years and older. Used for diagnosis/treatment of lipid, lipoprotein, and carbohydrate metabolism disorders, atherosclerosis, and liver/renal diseases. Contraindicated for children under 2 years of age.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K223179
B Applicant
Alere San Diego, Inc.
C Proprietary and Established Names
Cholestech LDX™ System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGA | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
| CHH | Class I | 21 CFR 862.1175 - Cholesterol (total) test system | CH - Clinical Chemistry |
| JGY | Class I | 21 CFR 862.1705 - Triglyceride test system | CH - Clinical Chemistry |
| JJE | Class I | 21 CFR 862.2160 - Discrete photometric chemistry analyzer for clinical use | CH - Clinical Chemistry |
| LBS | Class I | 21 CFR 862.1475 - Lipoprotein test system | CH - Clinical Chemistry |
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K223179 - Page 2 of 7
## II Submission/Device Overview:
### A Purpose for Submission:
Modification of a previously cleared device
### B Measurand:
- Total cholesterol (TC)
- High-density lipoprotein (HDL)
- Triglycerides (TRG)
- Glucose (GLU)
### C Type of Test:
Quantitative, enzymatic photometric
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The Cholestech LDX™ System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes:
The Lipid Profile•GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.
The TC•HDL•GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood.
The TC•GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood.
The Lipid Profile cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.
The TC•HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood.
The TC cassette is for the quantitative determination of total cholesterol in whole blood.
{2}
- Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
- HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
- Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
- Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Limitation – Cholestech LDX™ is not for use in children under the age of 2 years or patients with hyperbilirubinemia (Elevated bilirubin levels).
## D Special Instrument Requirements:
Cholestech LDX™ Analyzer.
## IV Device/System Characteristics:
### A Device Description:
The Cholestech LDX™ System is a small, portable analyzer and test cassette system. The system is for in vitro diagnostic use only. The Analyzer uses reflectance photometry (the amount of light reflected from a solid surface) to determine the amount of the analyte in blood. The analyzer measures color changes of the four reagent pads. The amount of color formed is converted by the analyzer to a concentration and the results are shown on the liquid crystal display (LCD) screen. The assays for use on the Analyzer include total cholesterol (TC), high-density lipoprotein cholesterol (HDL), triglycerides (TRG), glucose (GLU), and by calculation, low-density lipoprotein (LDL), non-HDL cholesterol, and total cholesterol/HDL cholesterol ratio.
The modification that is the subject of this submission addresses revised bilirubin interference claims for all analytes (see Section VII.A.3.) and includes a new limitation and a software upgrade to mitigate the modified risks. The revised limitations consist of the statement "Cholestech LDX™ is not for use in children under the age of 2 years or patients with hyperbilirubinemia (Elevated bilirubin levels)". The software upgrade incorporates the revised limitations, displaying it on the Cholestech LDX™ analyzer screen, printing it on all paper test results, and presenting it on all results sent to the electronic medical record.
K223179 - Page 3 of 7
{3}
K223179 - Page 4 of 7
B Principle of Operation:
The Cholestech LDX™ System principle of operation has not changed since the review of K120615.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Alere Cholestech LDX Analyzer, Alere Cholestech LDX Lipid Profile - GLU Cassette
B Predicate 510(k) Number(s):
K120615
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K223179 | K120615 |
| --- | --- | --- |
| Device Trade Name | Cholestech LDX™ System (Modified) | Cholestech LDX Analyzer and Cholestech LDX Lipid Profile•GLU Cassette |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the quantitative determination of total cholesterol, HDL cholesterol, triglycerides, and glucose in whole blood. | Same |
| Sample Type | Venous and Capillary Whole Blood | Same |
| Type of Use | For professional in vitro diagnostic use only | Same |
| General Device Characteristic Differences | | |
| Limitations | To address revised interference claims from conjugated and unconjugated bilirubin, the modified device includes the following limitation “Cholestech LDX™ is not for use in children under the age of 2 years or patients | Does not include the limitation |
{4}
K223179 - Page 5 of 7
| Device & Predicate Device(s): | K223179 | K120615 |
| --- | --- | --- |
| | with hyperbilirubinemia (Elevated bilirubin levels)". | |
## VI Standards/Guidance Documents Referenced:
CLSI – EP07-A3 – Interference Testing in Clinical Chemistry – 3rd edition
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. **Precision/Reproducibility:**
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
2. **Linearity:**
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
3. **Analytical Specificity/Interference:**
Interference testing was performed based on guidance from CLSI EP07-A3 and a study evaluating two interferents*, conjugated and unconjugated bilirubin was conducted. For the study, target analytes at different concentrations; triglycerides (100 mg/dL and 250 mg/dL), HDL (30 mg/dL and 80 mg/dL), glucose (50-70 mg/dL, 110-130 mg/dL and 225-270 mg/dL), total cholesterol (150 mg/dL and 220 mg/dL) were created in pooled samples that were spiked with the interferents, conjugated and unconjugated bilirubin, at a range of concentrations. Samples were tested in replicates of sixteen (16) using three (3) Cholestech LDX™ Lipid Profile Glucose test cassettes, on sixteen (16) Cholestech LDX™ Analyzers. The following table list the concentration of each substance at which no significant interference was detected; defined by the sponsor as a difference of less than or equal to ±10% between the test sample and control:
| Analyte | Analyte Target Values | Highest concentration with no observed Unconjugated Bilirubin Interference (mg/dL) | Highest concentration with no observed Conjugated Bilirubin Interference (mg/dL) |
| --- | --- | --- | --- |
| Total Cholesterol | 150 mg/dL ± 10% | 7.5 | 10 |
| | 220 mg/dL ± 10% | 7.5 | 12.5 |
| Triglycerides | 100 mg/dL ± 10% | 5 | 3.75 |
| | 250 mg/dL ± 10% | 5 | 7.5 |
| | 30 mg/dL ± 10% | 2.5 | 2.5 |
{5}
| Analyte | Analyte Target Values | Highest concentration with no observed Unconjugated Bilirubin Interference (mg/dL) | Highest concentration with no observed Conjugated Bilirubin Interference (mg/dL) |
| --- | --- | --- | --- |
| HDL | 80 mg/dL ± 10% | 5 | 12.5 |
| Glucose | 50-70 mg/dL | 10 | 12.5 |
| | 110-130 mg/dL | 20 | 15 |
| | 225-270 mg/dL | 27.1 | 35 |
Potential interference from conjugated and unconjugated bilirubin was also estimated for the calculated parameters.
Based on the data, the sponsor updated the interference claims in their labeling for the test system for conjugated and unconjugated bilirubin to indicate that concentrations up to 2.5 mg/dL do not interfere with the test system.
*Other aspects of this performance characteristic (including additional interferent testing) were established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296, K904082 and K120615.
4. Assay Reportable Range:
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
6. Detection Limit:
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
7. Assay Cut-Off:
Not applicable.
K223179 - Page 6 of 7
{6}
K223179 - Page 7 of 7
## B Comparison Studies:
1. **Method Comparison with Predicate Device:**
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
2. **Matrix Comparison:**
This performance characteristic was established in the previous 510(k) submissions, reviewed under K901900, K912136, K914469, K922612, K932727, K914033, K905296 and K904082.
## C Clinical Studies:
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not applicable.
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
The Cholestech LDX™ System’s expected values/reference range has not changed since the review of K120615.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.