Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface

K223168 · Medtronic, Inc. · QWF · Apr 19, 2023 · Cardiovascular

Device Facts

Record IDK223168
Device NameTubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface
ApplicantMedtronic, Inc.
Product CodeQWF · Cardiovascular
Decision DateApr 19, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4100
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medtronic tubing and accessories sets for extracorporeal membrane oxygenation (ECMO) with Balance biosurface are indicated for use in ECMO and extracorporeal life support (ECLS) procedures for adult patients with acute respiratory failure or acute cardiopulmonary failure when other available treatment options have failed, and when continued clinical deterioration is expected or the risk of death is imminent.

Device Story

Medtronic Tubing and Accessories Sets (eSets) provide components for assembling extracorporeal circuits used in ECMO and ECLS procedures. The system includes a Base eSet (drainage/return loops, tubing) and an Accessory eSet (nonstandard components). Used in clinical settings by perfusionists or trained medical staff to circulate blood during long-term respiratory or cardiopulmonary support. The device facilitates the connection of perfusion components to maintain extracorporeal circulation. It is a single-use, sterile, nonpyrogenic system. The Balance™ Biosurface is applied to components to enhance biocompatibility. The device is intended to support patients through critical failure states where conventional therapies are insufficient.

Clinical Evidence

No clinical or animal studies were performed. Safety and performance were demonstrated through non-clinical bench testing, including 21-day simulated use durability, tensile strength, pressure testing, kink testing, blood trauma testing, and coating coverage verification after life conditioning.

Technological Characteristics

Extracorporeal circuit tubing and accessories featuring Balance™ Biosurface. Materials include medical-grade plastics with TOTM plasticizer for prolonged use components. Sterilized via ethylene oxide. Single-use, nonpyrogenic. Designed for compatibility with standard extracorporeal circuit components. Complies with ISO 10993-1:2018 for biocompatibility.

Indications for Use

Indicated for adult patients with acute respiratory failure or acute cardiopulmonary failure requiring ECMO or ECLS support when other treatments have failed and clinical deterioration is expected or death is imminent.

Regulatory Classification

Identification

An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).

Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible; (2) The devices and accessories in the circuit must be demonstrated to be biocompatible; (3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories; (4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability; (5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and (6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. April 19, 2023 Medtronic Inc. Anna Wetherille Sr. Regulatory Affairs Specialist 8200 Coral Sea St. NE Mounds View, Minnesota 55112 Re: K223168 Trade/Device Name: Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal Circuit And Accessories For Long-Term Respiratory/Cardiopulmonary Failure Regulatory Class: Class II Product Code: QWF Dated: March 15, 2023 Received: March 16, 2023 Dear Anna Wetherille: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Image /page/1/Picture/4 description: The image shows a signature block. The text reads "Sincerely, Kathleen M. Grunder -S". The text is in a simple sans-serif font and is left-aligned. The background is white. for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K223168 Device Name Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface Indications for Use (Describe) The Medtronic tubing and accessories sets for extracorporeal membrane oxygenation (ECMO) with Balance biosurface are indicated for use in ECMO and extracorporeal life support (ECLS) procedures for adult patients with acute respiratory failure or acute cardiopulmonary failure when other available treatment options have failed, and when continued clinical deterioration is expected or the risk of death is imminent. Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness | Date Prepared: April 18, 2023 | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant: Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009 | | Contact Person: Anna Wetherille<br>Senior Regulatory Affairs Specialist<br>Phone: (763) 514-9842<br>Fax: (763) 367-8361<br>E-mail: anna.wetherille@medtronic.com<br><br>Juli Rubin (Alternate)<br>Senior Regulatory Affairs Program Manager<br>Phone: (763) 526-2357<br>Email: juli.c.rubin@medtronic.com | | Trade Name: Tubing and Accessories Sets for Extracorporeal Membrane<br>Oxygenation (ECMO) Balance™ Biosurface | | Common Name: Extracorporeal Life Support Circuit and Accessories | | Classification Name: Extracorporeal circuit and accessories for long-term<br>respiratory/cardiopulmonary failure | | Classification: Class II (with special controls) | | Regulation Number: 21 CFR 870.4100 | | Product Code: QWF | | Name of Predicate Device: FDA Final Order 81 FR 7451, February 12, 2016 | | Name of Reference Device: Tubing Pack (K171979) | {4}------------------------------------------------ ### Device Description: The Medtronic Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface ("eSets") contain components used to prepare an extracorporeal circuit for extracorporeal membrane oxygenation (ECMO) and extracorporeal life support (ECLS) procedures. The Base eSet Model BB22LSB contains a preassembled drainage and return loop, tubing assemblies, and other components used to prepare a basic extracorporeal circuit. The Accessory eSet Model BB22LSA contains nonstandard components to supplement the basic extracorporeal circuit, as needed per case and hospital protocols. This product is nonpyrogenic, is intended for single use, and has been sterilized using ethylene oxide. Maximum transit temperature: 50°C (122°F). Standard storage conditions are sufficient to safeguard the device. Store the device in the original packaging at room temperature in a dry place. Bench studies were performed after device preconditioning including exposure (up to 21 days) to simulated in vivo use conditions to demonstrate safety and reliability. #### Indication for Use: The Medtronic tubing and accessories sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance biosurface are indicated for use in ECMO and extracorporeal life support (ECLS) procedures for adult patients with acute respiratory failure or acute cardiopulmonary failure, when other available treatment options have failed, and when continued clinical deterioration is expected or the risk of death is imminent. ### Comparison to Predicate (Special Controls): Substantial equivalence evaluation includes a comparison to requirements in the FDA Final Order 81 FR 7451, February 12, 2016, as well as a comparison to the Reference Device. Per FDA Final Order (81 FR 7451) membrane lung devices for long-term pulmonary support, a preamendment class III device, as extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure, were redesignated and reclassified to class II (special controls) in patients with acute respiratory failure or acute cardiopulmonary failure where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The Medtronic Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface device meets all special controls identified in 21 CFR 870.4100, as follows: - Technological Characteristics: The geometry and design parameters of the subject ● device are consistent with the device's intended use in extracorporeal support procedures, and the device is compatible with the other devices and accessories in the extracorporeal circuit. - Biocompatibility: The subject device is demonstrated to be biocompatible in ● accordance with ISO 10993-1:2018 and with FDA guidance document Use of {5}------------------------------------------------ International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (4 September 2020). - Sterility and Shelf-life: Sterilization adoption evaluation and shelf-life testing ● demonstrate that the subject device maintains its sterility, integrity, durability, and reliability over the stated shelf life of the device. - Non-clinical Performance: Substantial equivalence of the performance characteristics ● is demonstrated on bench, mechanical integrity, durability, and reliability testing. - Labeling: The Instructions for Use include a detailed summary of the non-clinical ● evaluations pertinent to use of the device in an extracorporeal circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure. ### Comparison to Reference Device: A comparison of the Medtronic Tubing and Accessories Sets for ECMO with Balance™ Biosurface to the reference device indicates the following similarities: - Similar intended use, the only exception is the extended duration of extracorporeal support. - Same technological characteristics ● - Same operating principle - Same design features ● - Similar materials, the only exception is the new TOTM plasticizer used for the ● prolonged use components - Same shelf life, the subject device has a 2-year shelf life ### Summary of Performance Data In vivo pre-clinical and clinical studies involving patients or animals were not necessary to demonstrate the safety and performance of the subject device. Pre-clinical bench testing was used to verify the performance characteristics of this device. The following preclinical bench studies were conducted: - 21-day Simulated use durability testing ● - Tensile strength after life conditioning (long term use) - Pressure test after life conditioning (long term use) ● - Functional testing - Kink testing ● - Blood trauma testing ● - Coating coverage ● {6}------------------------------------------------ ## Conclusion: The data included in this submission are sufficient to demonstrate that the Medtronic Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface meets the special controls in FDA's Final Order (81 FR 7451, Redesignation as Extracorporeal Circuit and Accessories For Long- term Respiratory/cardiopulmonary Failure (ECMO), February 12, 2016), to connect perfusion devices and circulate blood as part of an extracorporeal circuit for use in Extracorporeal Membrane Oxygenation (ECMO) and Extracorporeal Life Support (ECLS) procedures.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...