Coltolux Comfort LED Curing Light

K223142 · Dent4you AG · EBZ · Dec 22, 2022 · Dental

Device Facts

Record IDK223142
Device NameColtolux Comfort LED Curing Light
ApplicantDent4you AG
Product CodeEBZ · Dental
Decision DateDec 22, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Coltolux Comfort LED Curing Light is indicated for the polymerization of intra- and extraoral dental restorations fabricated from materials (commonly referred to as visible light cured dental restoratives and/or composites and/or dental adhesives) that contain Camphorquinone as the photoinitiator.

Device Story

Handheld dental curing light; utilizes high-power LED to emit deep blue light (440-470 nm); activates Camphorquinone photo-initiators in dental resins/composites. Operated by dental professionals in clinical settings; lens placed in close proximity to restorative material; activated via pushbutton switch. Output is focused light energy; facilitates hardening/polymerization of dental materials. Benefits include efficient, reliable curing of light-activated dental restorations.

Clinical Evidence

Bench testing only. Performance verified per FDA guidance 'Dental Curing Lights - Premarket Notification (510(k))' and IEC 60601-1-2. Tests included light output uniformity, peak wavelength (ISO 10650:2018), irradiance (ISO 10650:2018), depth of cure, endurance, barrier sleeve attenuation, light shield filtration, life testing, duty cycle (IEC 60601-1:2012), EMC (IEC 60601-1-2:2014), and photobiological safety (IEC 62471:2006).

Technological Characteristics

High-power LED light source; wavelength 440-470 nm. Handheld form factor with pushbutton activation. Complies with ISO 10650:2018, IEC 60601-1:2012, IEC 60601-1-2:2014, and IEC 62471:2006. Battery-powered.

Indications for Use

Indicated for polymerization of intra- and extraoral dental restorations (composites, adhesives) containing Camphorquinone photoinitiator in dental patients.

Regulatory Classification

Identification

An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 22, 2022 Dent4You AG Dr. Tricia Cregger Global Regulatory Affairs Manager Bahnhofstrasse 2 Heerbrugg, 9435 SWITZERLAND Re: K223142 Trade/Device Name: Coltolux Comfort LED Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: October 3, 2022 Received: October 4, 2022 Dear Dr. Tricia Cregger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Michael E. Adjodha -S Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) K223142 Device Name Coltolux Comfort LED Curing Light Indications for Use (Describe) The Coltolux Comfort LED Curing Light is indicated for the polymerization of intra- and extraoral dental restorations fabricated from materials (commonly referred to as visible light cured dental restoratives and/or composites and/or dental adhesives) that contain Camphorquinone as the photoinitiator. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k) Summary | | | Prepared on: 2022-10-03 | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------|-------------------------|---------------------| | Contact Details | | K223142 | | 21 CFR 807.92(a)(1) | | Applicant Name | | Dent4You AG | | | | Applicant Address | | Bahnhofstrasse 2 Heerbrugg 9435 Switzerland | | | | Applicant Contact Telephone | | 330-916-8904 | | | | Applicant Contact | | Dr. Tricia Cregger | | | | Applicant Contact Email | | tricia.cregger@coltene.com | | | | Device Name | | | | 21 CFR 807.92(a)(2) | | Device Trade Name | | Coltolux Comfort LED Curing Light | | | | Common Name | | Ultraviolet activator for polymerization | | | | Classification Name | | Activator, Ultraviolet, For Polymerization | | | | Regulation Number | | 872.6070 | | | | Product Code | | EBZ | | | | Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | | | | Predicate # | | Predicate Trade Name (Primary Predicate is listed first)<br>Product Code | | | | K040551 | | Coltolux LED Curing Light | | EBZ | | Device Description Summary | | | | 21 CFR 807.92(a)(4) | | The proposed Coltolux Comfort LED Curing Light (Coltolux Comfort) is used for the polymerization of light-cured resin materials used in<br>dental restorations. The Coltolux Comfort incorporates a focused high-power LED to produce light in the "deep blue" spectrum<br>(wavelength 440 - 470 nm) for activating Camphorquinone (CQ) photo-initiators. The device is utilized by placement of the lens within<br>close proximity to the surface of the material to be cured, activation of the curing light via the pushbutton switch, and retention of the<br>tip near the dental material until the composite is cured. | | | | | | Intended Use/Indications for Use | | | | 21 CFR 807.92(a)(5) | | The Coltolux Comfort LED Curing Light is indicated for the polymerization of intra- and extraoral dental rom<br>materials (commonly referred to as visible light cured dental restoratives and/or dental adhesives) that contain<br>Camphorquinone as the photoinitiator. | | | | | | Indications for Use Comparison | | | | 21 CFR 807.92(a)(5) | | The indications for use of the Coltolux Comfort LED Curing Light are the indications for use of the predicate device, the<br>Coltolux LED Curing Light (K040551). | | | | | | Technological Comparison | | | | 21 CFR 807.92(a)(6 | The Coltolux Comfort LED Curing Light (subject device) and the Coltolux LED Curing Light (predicate device) are both LED curing lights for dental composite material. The indications for use and technological characteristics are substantially equivalent with only minor {4}------------------------------------------------ differences, including the mechanism of battery charging and the light intensity. These minor differences do not constitute a new intended use or raise different questions of safety and effectiveness compared with the predicate device. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Performance testing was conducted to verify that the proposed Coltolux Comfort Curing Light meets the requirements as defined in FDA Guidance Document "Dental Curing Lights - Premarket Notification [510(k)" and in accordance with IEC 60601-1-2. Software verification and validation of the device were conducted in accordance with the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" Performance Testing - Bench, of this Premarket Submission, contains detailed information reqarding the proposed Coltolux Comfort Curing testing protocols, test objectives, test articles, test methods and procedures, and acceptance criteria. Non-clinical tests performed to establish substantial equivalence to the identified predicate device included light output uniformity and dropoff testing, peak wavelength testing (ISO 10650:2018), irradiance output (ISO 10650:2018), depth of cure, endurance testing, light attenuation through barrier sleeve, light shield filtration, life testing, duty cycle (JEC 60601-1:2012), EMC (JEC 60601-1-2:2014), electromagnetic disturbances (IEC 60601-1-2:2014), and photobiological safety (IEC 62471:2006). Not Applicable Based on the non-clinical performance data the proposed Coltolux Comfort Curing Light is as safe, as effective, and performs as well as or better than the predicate device Coltolux LED Curing Light (K040551, 21 CRF 872.6070, product code EBZ).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...