The STR8 Clear Aligner is indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.
Device Story
STR8 Clear Aligners are custom-fabricated, removable, thermoformed plastic trays used for orthodontic tooth movement. Input consists of digital dental scans; specialized CAD/CAM software (ArchForm Treatment Planning System) processes these scans to generate a series of 3D models representing incremental tooth positions. A dental professional reviews and approves the treatment plan. The device is manufactured by thermoforming thermoplastic sheets (Zendura FLX) over the 3D-printed models. Aligners are worn by the patient 20-22 hours/day, removed for eating and cleaning, and replaced in sequence to apply gentle, continuous force to the dentition. The device is used in a clinical setting under the direction of a dental professional. Output is the physical movement of teeth toward a desired final position, facilitating orthodontic correction.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including dimensional accuracy validation of the manufacturing process (3D printing through thermoforming) and biocompatibility testing of the Zendura FLX material per ISO 10993-1 and ISO 10993-5:2009.
Technological Characteristics
Thermoformed clear aligners made of Zendura FLX (copolyester and polyurethane composite). Principle of operation: sequential application of gentle, continuous mechanical force. Dimensions: patient-specific. Connectivity: utilizes external CAD/CAM software (ArchForm) for treatment planning. Sterilization: non-sterile. Software: CAD/CAM-based treatment planning system.
Indications for Use
Indicated for the alignment of teeth during orthodontic treatment of malocclusions in patients requiring tooth movement via continuous gentle forces.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 24, 2023
STR8 Oral Care % Elisabeth Miller Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043
Re: K223141
Trade/Device Name: STR8 Clear Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: February 21, 2023 Received: February 21, 2023
Dear Elisabeth Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K223141
Device Name STR8 Clear Aligners
Indications for Use (Describe)
The STR8 Clear Aligner is indicated for the aligment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5. 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | STR8 Oral Care |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Address:<br>Phone:<br>Email: | Misee Harris<br>217 E Cherry Ave Jonesboro, AR 72401<br>931-797-1821<br>miseeharris@cdcjonesboro.com |
| Submission Correspondent:<br>Address:<br>Phone:<br>Email: | Elisabeth Miller, Regulatory Affairs Consultant<br>1321 Upland Dr. Suite 6792 Houston, TX 77043<br>(586) 242-7718<br>emiller@primepathmedtech.com |
| Date Prepared: | September 2022 |
| Proprietary Name: | STR8 Clear Aligners |
| Common Name: | Sequential Clear Aligner |
| Product Code: | NXC – Orthodontic plastic bracket |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Predicate Device: | uLab Systems Dental Aligner (K211510) |
| Classification Name: | Aligner, Sequential |
{4}------------------------------------------------
#### Device Description:
STR8 Clear Aligners are thermoformed plastic aligners designed to be worn in sequence to facilitate the movement to the teeth to the final desired position. The sequential aligners introduce incremental movements that move teeth by way of gentle continuous force. The aligners are to be worn 20 to 22 hours a day and are to be removed for eating and for cleaning.
STR8 Clear Aligners are designed from digital scans of a patient's dentition. Using the scan, specialized orthodontic CAD/CAM software will be used to develop the treatment plans that consist of sequential dental models wherein the teeth are gradually realigned with each step. For this 510(k), ArchForm Treatment Planning System (K213916) will be used for this application. ArchForm is approved for use in the management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options. The specialized orthodontic treatment planning software has a 510k clearance for the intended use under FDA Classification Product Code PNN, regulation 872.5470.
Once the treatment plan is reviewed and approved by a STR8 Oral Care dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The aligners are then provided to the patient by the prescribing dental professional group.
#### Indications for Use:
The STR8 Clear Aligner is indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.
{5}------------------------------------------------
## Comparison to Predicate Devices:
STR8 Clear Aligners are functionally equivalent to the following predicate device: uLab Systems Dental Aligner Kit (K211510).
| Specification | Subject Device: | Predicate Device: | Comparison Result |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | STR8 Clear Aligners | uLab Systems Dental Aligner Kit (K211510) | |
| Regulation Number | 21 CRF 872.5470 | 21 CRF 872.5470 | Same |
| Classification Name | Aligner, Sequential | Aligner, Sequential | Same |
| Product Code | NXC | NXC | Same |
| Classification | Class II | Class II | Same |
| OTC or Rx | Rx | Rx | Same |
| Material | Zendura FLX (copolyester and polyurethane composite) | Zendura A (thin thermoformed polyurethane) or Zendura FLX (copolyester and polyurethane composite) | Similar (Zendura A) or Same (Zendura FLX) |
| Biocompatible | Yes | Yes | Same |
| Sterile | No | No | Same |
| Device Description | A series of customized removable clear plastic orthodontic aligners that sequentially position teeth by way of continuous gentle force. | A series of customized removable clear plastic orthodontic aligners that sequentially position teeth by way of continuous gentle force. | Same |
| Patient Removable? | Yes | Yes | Same |
| Indication for Use | The STR8 Clear Aligner is indicated for the alignment of teeth during | The uLab Systems Dental Aligner is indicated for the alignment of teeth during | Same |
| | orthodontic treatment of<br>malocclusions by way of<br>continuous gentle forces. | orthodontic treatment of<br>malocclusions by way of<br>continuous gentle forces. | |
| Intended Use | Orthodontic Tooth<br>Movement | Orthodontic Tooth<br>Movement | Same |
| Principles of<br>Operation | Each preformed plastic<br>tray is worn in sequence<br>by the patient as<br>prescribed by the dental<br>practitioner. Orthodontic<br>movement occurs through<br>continuous gentle forces<br>applied to the dentition as<br>each tooth follows the<br>programmed<br>displacement based on<br>the doctor's prescription. | Each preformed plastic<br>tray is worn in sequence<br>by the patient as<br>prescribed by the dental<br>practitioner. Orthodontic<br>movement occurs through<br>continuous gentle forces<br>applied to the dentition as<br>each tooth follows the<br>programmed<br>displacement based on<br>the doctor's prescription. | Same |
| Aligner Design<br>Process | Standard dental software<br>for tooth alignment uses<br>digital scan (untreated<br>state) to generate the<br>image of a final,<br>provisional treated state<br>and then interprets a<br>series of images that<br>represent intermediate<br>teeth states. The dental<br>practitioner then reviews<br>these images and has the<br>option to reject or request<br>modifications to the set-<br>up prior to approving for<br>aligner fabrication. Once<br>the dental practitioner<br>approves the treatment<br>plan, the software<br>converts the files to<br>produce a series 3D<br>models used to produce<br>thermoformed aligners. | Standard dental software,<br>including the uLab<br>Systems uDesign.<br>K171295, for tooth<br>alignment uses digital<br>scan (untreated state) to<br>generate the image of a<br>final, provisional treated<br>state and then interprets<br>a series of images that<br>represent intermediate<br>teeth states. The dental<br>practitioner then reviews<br>these images and has the<br>option to reject, make or<br>request modifications to<br>the set-up prior to<br>approving it for aligner<br>fabrication. Once the<br>dental practitioner<br>approves the treatment<br>plan, the software<br>converts the files to<br>produce the series of 3D<br>models used to produce<br>thermoformed aligners. | Same |
Table 1. Predicate Device Comparison
{6}------------------------------------------------
{7}------------------------------------------------
#### Comparison for Indications for the Use to Predicate Device:
Based on the above comparison, the indications for use of the STR8 Clear Aligners in this submission is the same as the uLab Systems Dental Aligner Kit (K211510) as they are both indicated for the alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces. Thus, the STR8 Clear Aligners can be considered substantially equivalent to its predicate device.
## Comparison of Technical Characteristics to Predicate Device
Based on the above comparison, the design, construction, and performance characteristics of the STR8 Clear Aligners is similar to that of the uLAB Systems Dental Aligner Kit (K211510).
The only difference between the subject device (STR8 Clear Aligner) and predicate device (uLab Systems Dental Aligner Kit) is the material differences come from the additional material (Zendura A) used for the predicate device. The expansion of material does not increase the risk associated with manufacturing, and the subject device and predicate device do share an exact material, Zendura FLX.
Thus, the STR8 Clear Aligners can be considered substantially equivalent to its predicate device.
#### Non-Clinical Performance Testing
The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces. However, durability testing was completed on these aligners. Real world use was simulated to ensure that the aligner material and manufacturing process produced aligners that were suitable for their prescribed period of use.
An internal manufacturing validation was performed to establish the dimensional accuracy of the manufacturing process for the STR8 Clear Aligners. The submitted intraoral scans, digital dentition models, from treatment planning, 3D printed molds and the final thermoformed aligners were all assessed quantitatively in the validation. Thus, the robustness of the process was demonstrated from 3D printing through thermoforming.
All measurements were within 0.3 mm of the target input value, the predefined tolerance of the manufacturing process. Furthermore, throughout the qualitative assessment of the aligners no performance, cosmetic, or other detectable issues were identified. This validation has met the pre-established acceptance criteria to demonstrate that the STR8 manufacturing process yields dimensional accurate products that meet product specifications.
The dental thermoplastic material used for STR8 Clear Aligners is a well-establish and commonly used thermoplastic, Zendura FLX. Biocompatibility testing for the aligner
{8}------------------------------------------------
material, the only patient contacting material, was conducted by the manufacturer in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Additional cytotoxicity testing according to ISO 10993-5:2009 was performed on final manufactured STR8 Clear Aligners.
#### Clinical Performance Testing
The technical characteristics, indications for use, material, manufacturing, and sterilization processes are the same to the predicate device and therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.
#### Conclusion
Based on the similarities in indications for use, technological characteristics, and materials the STR8 Clear Aligners are substantially equivalent to the uLab Systems Dental Aligner Kit (K211510)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.