K223135 · Compal Electronics, Inc. · GEI · Jun 30, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223135
Device Name
AblatePal Radiofrequency Ablation System
Applicant
Compal Electronics, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AblatePal Radiofrequency Ablation System is designed to provide monopolar radiofrequency (RF) energy to be used for coagulation and ablation of soft tissue. The device is a prescription device.
Device Story
Monopolar RF generator system for soft tissue coagulation and ablation; utilizes active electrodes and neutral electrodes. Input: user-defined settings (time, temperature, impedance) via touchscreen; RF current delivered from generator through active electrode to patient tissue; neutral electrode returns current to generator. Heat generated by tissue resistance enables ablation. Operated by clinicians in clinical settings. Includes footswitches for activation and power adjustment. Output: RF energy delivery; informational signals and alerts on touchscreen. Benefits: controlled tissue ablation. System includes generator, footswitches, power cord, and single-use electrode kits (active electrode, tubing, neutral electrodes).
Clinical Evidence
No clinical data. Evidence consists of bench testing and ex-vivo animal tissue studies (liver, kidney, muscle). Bench testing verified electrical safety, EMC, and ablation zone dimensions compared to predicate. Ex-vivo testing demonstrated equivalent histological, thermal, and ablative performance.
Technological Characteristics
Monopolar RF generator; 480kHz ±10% output frequency; 0-200W power output. Impedance monitor (25-1000 Ohm); temperature continuity monitor (10-99°C). Materials: Al alloy neutral electrodes. Connectivity: touchscreen interface, footswitch inputs. Sterilization: EtO (active electrodes). Software: Moderate level of concern. Complies with IEC 60601-1, 60601-1-2, 60601-2-2.
Indications for Use
Indicated for coagulation and ablation of soft tissue. Prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2023
Compal Electronics, Inc Casper Chen Senior Director No. 581, Ruiguang Rd., Neihu District Taipei City, 11492 Taiwan
## Re: K223135
Trade/Device Name: AblatePal Radiofrequency Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 29, 2023 Received: June 1, 2023
## Dear Casper Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.06.30 15:38:18 -04'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223135
Device Name AblatePal Radiofrequency Ablation System
Indications for Use (Describe)
The AblatePal Radiofrequency Ablation System is designed to provide monopolar radiofrequency (RF) energy to be used for coagulation and ablation of soft tissue.
The device is a prescription device.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### 1. Date of Summary Preparation
June 28, 2023
#### 2. 510(k) Submitter/Holder
Compal Electronics, Inc. No. 581, Ruiguang Rd., Neihu Dist. Taipei, 11492, Taiwan (R.O.C.)
#### 3. Contact: Casper Chen
Senior Director of Medical Product Division
Casper_Chen@compal.com Email:
Address: No. 581, Ruiguang Rd., Neihu Dist. Taipei, 11492, Taiwan (R.O.C.)
#### 4. Device
| Device Trade Name: | AblatePal Radiofrequency Ablation System |
|----------------------|----------------------------------------------------------------|
| Device Common Name: | RF Ablation System |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation: | 21 CFR 878.4400 |
| Regulatory Class: | II |
| Product Code: | GEI |
#### 5. Predicate Device
The AblatePal Radiofrequency Ablation System is substantially equivalent to the following, commercially available predicate devices:
| Device Name | Classification Name and Class | 510(k) Number(s) |
|---------------------------------|------------------------------------------------------|---------------------------------------|
| Cool-tip™ RF Ablation<br>System | Electrosurgical, Cutting & Coagulation & Accessories | K042216<br>(Primary Predicate Device) |
#### 6 Intended Use / Indications for Use
The AblatePal Radiofrequency Ablation System is designed to provide monopolar radiofrequency (RF) energy to be used for coagulation and ablation of soft tissue.
The device is a prescription device.
{4}------------------------------------------------
## 7. Device Description
The AblatePal Radiofrequency Ablation System (subject device) (HFM01-MD2202) is a monopolar radiofrequency (RF) generator with dedicated components and accessories. The generator can be operated in only impedance mode that use various combinations of time, tissue temperature, tissue impedance, and manual control. An interactive touchscreen display allows user input to the generator and displays informational signals and alert messages. Active electrodes are placed in the tissue to deliver RF energy to the tissue to be ablated. Only one electrode can be attached to the generator and can be used individually. During the ablation of tissue, RF current flows from the generator to an active electrode which delivers the current to the patient. The RF current flows from the active electrode through the patient's body tissue to the Neutral Electrode, which recovers the current and returns it to the qenerator. The resistance to the current, provided by the patient's tissue, produces the heat that is necessary for ablation of tissue.
The dedicated system components include:
- o RF Generator, radiofrequency generator
- . Ablation Foot Switch (HFM01-AF0002): Single Foot Switch. Pressing the foot switch that RF Generator starts or stops ablating.
- . Power Control Foot Switch (HFM01-AF0001): Double Foot Switch. Pressing the foot switch to adjust power in whole units of Watts that the RF Generator will deliver during the ablation. The blue foot switch can increase the power, and the yellow foot switch can decrease the power.
- . Power Cord (WS-001H+WS-002): A line cord that provides AC power to the RFA system.
- Electrode Kit (HFM01-AC0001~ HFM01-AC0006), including radiofrequency electrode, inflowo outflow tubing sets and Neutral Electrodes (HFM01-AE0001).
The generator is packaged individually and is always supplied with the accessory items: footswitch and cables. The electrode kit HFM01-AC0001~ HFM01-AC0006 is separately packaged.
The following AblatePal Radiofrequency Ablation System compatible accessories are single use:
- . Electrode Kit (HFM01-AC0001~ HFM01-AC0006), including a radiofrequency electrode, inflow-outflow tubing sets and Neutral Electrodes.
#### 8. Comparison of Technological Characteristics with the Predicate Device
The AblatePal Radiofrequency Ablation System (subject device) relies on the same principles of operation and has similar performance characteristics, with respect to its intended use, as the predicate devices. The AblatePal Radiofrequency Ablation System combines the functional characteristics of the predicate Cool-tip™ generator (K042216).
{5}------------------------------------------------
There are three main differences between the AblatePal Radiofrequency Ablation System and the predicate Cool-tip™ RF Ablation System
1. New larger patient Neutral Electrode compared to the predicate device.
2. The generator combines the functional characteristics of the predicate Cool- tip™ generator into a single RF generator that includes a different user interface and software controls.
3. The addition of several optional accessories, specifically, footswitches.
Unlike the predicate device, the AblatePal Radiofrequency Ablation System has only one category of RF Ablation electrodes (monopolar). The AblatePal Radiofrequency Ablation System active electrodes deliver RF energy from the AblatePal Radiofrequency Ablation System generator for the ablation of tissue when used with an irrigation pump, which must comply with IEC 60601-1, and electrode kits. An electrode kit with a single active electrode (sterile), inflow-outflow tubing sets (sterile) and two Neutral Electrodes (non-sterile) are available. All kit items are single use.
| Description | Subject Device | Predicate Device | Comparison |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | AblatePal Radiofrequency<br>Ablation System<br>(K223135) | Cool-tip™ RF Ablation System<br>(K042216) | |
| Indications for Use | The AblatePal Radiofrequency Ablation System is designed to provide monopolar radiofrequency (RF) energy to be used for coagulation and ablation of soft tissue.<br><br>The device is a prescription device. | The Valleylab Cool-tip RF System (generator and accessories) is intended for the use in percutaneous, laparoscopic, and intraoperative coagulation and ablation of tissue, such as partial or complete ablation of non-resectable liver lesions. | The indication of "percutaneous, laparoscopic, and intraoperative coagulation and ablation of tissue, such as partial or complete ablation of non-resectable liver lesions" has been removed from the subject device. |
| Prescription or OTC | Prescription | Prescription | Same |
| ESU<br>• Major function | Monopolar<br>impedance monitor:<br>Range:25-1000 Ohm<br>Resolution:1 Ohm<br>Accuracy:50-200 Ohm (± 10%)<br><br>continuity monitor: temperature<br>Range: 10-99°C<br>Resolution: 0.1°C<br>Accuracy: ±4°C | Monopolar<br>impedance monitor:<br>Range:0-1000 Ohm<br>Resolution:1 Ohm<br>Accuracy:50-100 Ohm (± 10%), 500-1000 Ohm (± 20%)<br><br>continuity monitor: temperature<br>Range: 10-99°C<br>Resolution: 1°C<br>Accuracy: ±4°C | Same: Monopolar<br><br>Impedance monitoring of the subject device is in the range of that of the predicate device<br><br>Same: continuity monitor |
| ESU<br>• Performance specifications | Output frequency: 480kHz±10%<br>Waveform: AC sin wave (110 V) | Output frequency:<br>480kHz±10% | Accuracy of the subject device is in the range of that of the |
| | Power output: 0-200W Max 50 Ohm output load. Accuracy is ±20%. | Waveform: AC sin wave (100/120 V) | predicate device |
| | Voltage output: 100Vrms at 50Ω in 200W | Power output: 0-200W Max 50 Ohm output load. Accuracy is ±15%. | Others are same |
| | | Voltage output: 100Vrms at 50Ω in 200W | |
| ESU<br>• Physical<br>specifications | 12.0 kg<br>39 x 32.5 x 22.5 cm | 8.0 kg<br>41.8 x 28.0 x 14.0 cm | ESU of the subjective<br>device is heavier than<br>that of the predicate<br>device |
| Active accessory<br>• Monopolar or<br>bipolar<br>• Physical<br>dimensions<br>and design<br>(e.g., size, | - Monopolar<br>- Length: 10/15/20cm<br>- Exposure: 0.5/1/2/3 cm<br>- Diameter: 17/18 G<br>- EO sterilization<br>- Specification: 100V Vrms,2A | - Monopolar<br>- Length: 10/14.4/15/20/25 cm<br>- Exposure: 0.7/1/2/3/ cm<br>- Diameter: 17G<br>- EO sterilization<br>- Specification: 100V Vrms,2A | Length, exposure,<br>diameter of the subject<br>device is in the range<br>of that of the predicate<br>device |
| length,<br>connector<br>type)<br>• Rated voltage<br>• Materials<br>(electrode,<br>insulation,<br>coating, etc.) | | | Others are same |
| Neutral<br>electrodes<br>• Conductive or<br>capacitive<br>• Physical<br>specifications<br>• Materials | - Conductive<br>- Material: Al alloy<br>- Specification: 23x14 cm | - Conductive<br>- Material: Al alloy<br>- Specification: 22x13 cm | Neutral electrodes of<br>the subjective device<br>are larger than those<br>of the predicate device |
| Miscellaneous<br>accessories<br>• Functions | Ablation Foot Switch (Single Foot<br>Switch): Pressing the foot switch that<br>RF Generator starts or stops<br>ablating.<br>Power Control Foot Switch (Double<br>Foot Switch): Pressing the foot switch<br>to adjust power in whole units of<br>Watts that the RF Generator will<br>deliver during the ablation. The blue<br>foot switch can increase the power,<br>and the yellow foot switch can<br>decrease the power.<br>- Specification: 25VAC,5A | N/A | Addition of Ablation<br>Foot Switch and<br>Power Control Foot<br>Switch. |
| • Performance<br>specifications<br>• Physical<br>specifications<br>• Materials | | | Performance data is<br>provided in HFM01-F-<br>24 System test report<br>(section 20-22) and<br>IEC 60601-2-2 report<br>to support that, even<br>with significant<br>differences, the device<br>is as safe and effective<br>as the predicate. |
{6}------------------------------------------------
{7}------------------------------------------------
## 9. Performance Testing
Non-clinical performance testing was performed, and reports are provided in support of substantial equivalence with the predicate device.
#### 9.1 Summary of Bench Testing
Electrical Safety and Electromagnetic Compatibility (EMC):
Bench testing and verification were conducted to ensure proper device function. At the system level, electrical safety and electromagnetic compatibility were evaluated to establish conformity to the applicable IEC 60601 safety standards, as identified below.
- o IEC 60601-1:2005/AMD1:2020 Medical Electrical Equipment Part 1 General Requirements for Basic Safety and Essential Performance
- . IEC 60601-1-2:2021 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
- IEC 60601-2-2:2017 Medical Electrical Equipment Part 2-2: Particular Requirements for the . Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories
The performance and safety of the AblatePal Radiofrequency Ablation System was characterized using laboratory (bench) verification and simulated-use validation testing consisting of soft tissue performance providing a characterization of the AblatePal Radiofrequency Ablation System ablation performance across the complete range of modes, electrodes, and in three tissue types: liver, kidney, and muscle. Ablation performance verifying the equivalence between the predicate Cool-tip™ RFA System and AblatePal Radiofrequency Ablation System when used with the AblatePal Radiofrequency Ablation System was performed at the system level to demonstrate the following:
{8}------------------------------------------------
- . AblatePal Radiofrequency Ablation System RF Ablation Generator meets all design and performance requirements necessary to prove substantial equivalence with the predicate device,
- . AblatePal Radiofrequency Ablation System RF Ablation Generator meets all ablation performance design requirements and to demonstrate the ablations zone dimensions created with the AblatePal Radiofrequency Ablation System Ablation Generator (subject) and Cooltip™ RF Ablation Generator (predicate) are equivalent,
- The AblatePal Radiofrequency Ablation System Neutral Electrode, and RF Ablation cable ● meet design and performance requirements,
- o The AblatePal Radiofrequency Ablation System with Ablation Generator meets IEC, EMC. requlatory and safety requirements,
- Validate the usability of AblatePal Radiofrequency Ablation System Ablation Generator (HFM01-MD2202) with the intended users,
- The packaging of AblatePal Radiofrequency Ablation System, including RF Ablation Generator, Neutral Electrode, RF Ablation Electrode Kits, can provide protection during simulated transportation.
#### 9.2 Animal Testing
Ex-vivo testing conducted demonstrated the histological, thermal and ablative performance , as well as procedural safety of the subject device is equivalent to the predicate for the AblatePal Radiofrequency Ablation System intended use.
#### 10. Software
Software for the AblatePal Radiofrequency Ablation System was developed in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, for devices with a Moderate Level of Concern. System level software testing was completed successfully.
#### 11. Biocompatibility
The direct patient contacting devices of the AblatePal Radiofrequency Ablation System are the Neutral Electrode and active electrode. These devices are disposable and intended for single use. These devices were assessed for biocompatibility and potential adverse health effects associated with patient contacting disposable devices and were determined to be acceptable.
{9}------------------------------------------------
## Direct Patient Contacting Devices:
#### Neutral Electrode:
The Neutral Electrode is used to collect the RF energy delivered at the active electrode and return it to thegenerator.
The patient contacting materials in the Neutral Electrode were tested to thebiocompatibility requirements shown below.
- Cytotoxicity
- Sensitization
- o Skin Irritation
#### Active Electrode:
Active electrodes are placed in the tissue to deliver RF energy to the tissue being ablated.
The patient contacting materials in the Active Electrode were tested to the biocompatibility requirements shown below:
- . Cytotoxicity
- Sensitization ●
- Intracutaneous reactivity
- Acute Systemic Toxicity ●
- Pyrogen ●
#### 12. Sterilization and Shelf life
The active electrodes are provided sterile with SAL of 10°, are intended for use within the sterile field, and are designed for single use only.
#### Electrode kit:
The AblatePal Radiofrequency Ablation System electrode kits, including active electrode and tubing sets, are sterilized with EtO to a sterility level of 10 °. The sterile packaging is labeled as sterile for 3 years immediately following the date of manufacturing.
The single use neutral electrode are not provided sterile and have a shelf life of 1.5 years.
{10}------------------------------------------------
#### 13. Conclusion
The AblatePal Radiofrequency Ablation System (subject device) and Cool-tip™ RF Ablation System (predicate device) have equivalent power and ablation performance characteristics based on verification and validation testing, and pre-clinical studies. Verification testing of the optional accessories consisting of the footswitch was conducted for their use with the subject device. As a result, the AblatePal Radiofrequency Ablation System has been shown to be substantially equivalent to the predicate device and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.