The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
Digital flat-panel X-ray detector; captures projection radiographic images. Indirect conversion: scintillator (CsI or Gadox) converts X-rays to visible light; optical sensor/TFT array converts light to electrical charge; analog-to-digital conversion produces electronic image data. Used in clinical settings; operated by healthcare professionals. Integrates with X-ray generators via AED/AWC mode or Sync Trigger mode (TTL communication). Output transmitted to console PC for viewing, storage, or network distribution. Replaces film/screen systems; provides digital imaging for general radiography; facilitates faster image acquisition and electronic storage.
Clinical Evidence
Bench testing only. Performance evaluated via DQE and MTF measurements compared to predicate devices. Compliance with international standards (IEC 62220-1-1, IEC 60601-1, IEC 60601-1-2, IEC 62304) confirmed. No clinical data presented.
Technological Characteristics
Flat-panel digital X-ray detector; indirect conversion using CsI or Gadox scintillator and amorphous Silicon (a-Si) TFT array. Pixel pitch 140μm. Connectivity via Gigabit Ethernet (PoE). Power 100-240V, 50/60 Hz. Standards: ANSI AAMI ES60601-1, IEC 62220-1-1, IEC 60601-1-2, IEC 62304, ISO 14971.
Indications for Use
Indicated for digital radiographic diagnosis of human anatomy in general purpose diagnostic procedures. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 18, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
DRTECH Corporation % Ella An Regulatory Affairs Suite No. 1, 2 Floor/Suite No. 2, 3 Floor, 29, Dunchon-daero 541beon-gil Jungwon-gu, Seongnam-si, Gyeonggi-do 13216 KOREA
Re: K223124
Trade/Device Name: EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MOB Dated: September 29, 2022 Received: October 3, 2022
Dear Ella An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Zhkke
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223124
Device Name EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG
Indications for Use (Describe)
The EXPD 86P, EXPD 86PG, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Number: K223124
DRTECH Corporation Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, South Korea
## 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
### 1. Date Prepared [21 CFR 807.92(a) (1)]
September/29/2022
#### 2. Submitter's Information [21 CFR 807.92(a) (1)]
| • Name of Sponsor: | DRTECH Corporation |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------|
| • Address: | Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil<br>Jungwon-gu, Seongnam-si, Gyeonggi-do, South Korea |
- Contact Name: Ella An ●
- Telephone No.: + 82-31-779-7787
- Fax No .: + 82-31-779-7790
- Email Address : ● ysan96@drtech.co.kr
- Registration Number: ● 3005172103
- Name of Manufacturer: Same as Sponsor ●
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| • Trade Name: | EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG |
|---------------|--------------------------------------------|
|---------------|--------------------------------------------|
- . Common Name: Digital Flat Panel X-ray Detector
- Classification Name: Stationary X-ray System ●
- Classification Panel: Radiology
- Classification Regulation: 21 CFR 892.1680 ●
- Product Code: MOB
- Device Class: II ●
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
- 510(k) Number: ● K192400
- Applicant: DRTECH Corporation
- Trade Name: EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG ●
- Classification Name: Stationary X-ray System ●
- Classification Panel: Radiology ●
- Classification Regulation: 21 CFR 892.1680
- Product Code: MQB ●
- Device Class: II ●
{4}------------------------------------------------
#### 5. Description of the Modified Device [21 CFR 807.92(a) (4)]
<Modification>
Addition of EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG: The differences between the subject devices and the predicate devices are Dimension and Configuration of Components(Optional). The Dimension of Subject devices is longer than predicate devices.
#### 6. Indication for Use [21 CFR 807.92(a)(5)]
The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is a flat-panel type digital X-ray detector that captures projection radiographic images in digital format within seconds, eliminating the need for an entire x-ray film or an image plate as an image capture medium. EXPD 86P, EXPD 129P, EXPD 129PG differs from traditional X-ray systems in that, instead of exposing a film and chemically processing it to create a hard copy image, a device called a Detector is used to capture the image in electronic form.
The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG Detector is an indirect conversion device in the form of a square plate in which converts the incoming X-rays into visible light. This visible light is then collected by an optical sensor, which generates an electric charges representation of the spatial distribution of the incoming X-ray quanta.
The charges are converted to a modulated electrical signal thin film transistors. The amplified signal is converted to a voltage signal and is then converted from an analog to digital signal which can be transmitted to a viewed image print out, transmitted to remote viewing or stored as an electronic data file for later viewing.
Comparisons with the predicate, devices show the technological characteristics of the EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG to be same to the predicate devices. EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is functionally identical to the predicate devices.
{5}------------------------------------------------
# DRTECH
## 8. Intergrated Configuration for X-Ray system
The EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG is connected with General diagnostic X-ray system.
Image /page/5/Figure/4 description: This image shows a diagram of an X-ray system setup. The diagram includes components such as an X-ray System Generator, EXPD129P, Console PC, and Hospital Network. Cables labeled as LLD POE Cable, Ethernet Cable, and LLD SYNC Cable connect these components, with annotations indicating voltage, Gigabit Ethernet, and bidirectional TTL signal specifications.
| No. | Cable | Communication |
|-----|----------------|------------------------------------------------------------------------------------------------------------|
| ① | LLD POE Cable | 1) DC 24V voltage<br>2) It is supplied from Adaptor to Detector. |
| ② | LLD POE Cable | 1) Gigabit Ethernet Communications<br>2) It is transmitted from the detector to the CONSOLE PC. |
| ③ | LLD SYNC Cable | 1) two-way TTL communication<br>2) It is communicated between the detector and the X-ray System Generator. |
| ④ | LAN Cable | 1) 1) DICOM communication<br>2) It is transmitted from the CONSOLE PC to the hospital network. |
{6}------------------------------------------------
DRTECH
For successful integration of detector with X-ray system, we designed the hardware of EXPD 86P, EXPD 86PG, EXPD 129PG as below function. The Energy of generator is controlled by Radiography imaging software(Econsole1, K152172), not Detector.
| Mode | Description |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AED/AWC Mode | The detector detects actual amount of X-rays without any connection to the X-ray<br>generator.<br>No signal used (No need of connector interface cable) |
| Sync. Trigger Mode | The detector receives EXP_REQ signal that X-ray generator is prepared to<br>generate X-rays.<br>The detector prepares image acquiring and then responds EXP_OK signal to the<br>X-ray generator.<br>The X-ray generator confirms EXP_OK signal and generates X-ray, then the<br>detector performs image acquiring, according to image acquisition time and<br>transmits the image data.<br>EXP_REQ (Generator→ Detector), EXP_OK (Detector →Generator) |
### 9. Substantial Equivalence [21 CFR 807.92(b)]
| Parameter | Subject Device | Predicate Device |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | Unknown | K192400 |
| Manufacturer | DRTECH Corporation | DRTECH Corporation |
| Model Name | EXPD 129P/ EXPD 129PG/<br>EXPD 86P/ EXPD 86PG | EVS 4343A / EVS 4343AG<br>EVS 3643A / EVS 3643AG |
| Classification Name | Stationary X-ray System | |
| Classification Panel | Radiology | |
| Classification<br>Regulation | 21 CFR 892.1680 | |
| Product Code | MQB | |
| Device Class | Class II | |
| Intended Use | The EXPD 86P, EXPD 86PG, EXPD<br>129P, EXPD 129PG Digital X-ray<br>detector is indicated for digital imaging<br>solution designed for providing general<br>radiographic diagnosis of human<br>anatomy. This device is intended to<br>replace film or screen based<br>radiographic systems in all general<br>purpose diagnostic procedures. This<br>device is not intended for<br>mammography applications | The EVS 4343A / EVS 4343AG / EVS<br>3643A / EVS 3643AG Digital X-ray<br>detector is indicated for digital imaging<br>solution designed for providing general<br>radiographic diagnosis of human<br>anatomy. This device is intended to<br>replace film or screen based<br>radiographic systems in all general<br>purpose diagnostic procedures. This<br>device is not intended for<br>mammography applications. |
{7}------------------------------------------------
| Parameter | Subject Device | Predicate Device | | |
|---------------------------------|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Detector<br>Size | EXPD 86P/ EXPD 86PG<br>: Rectangular Panel | EVS 3643A: Rectangular Panel<br>EVS 3643AG: Rectangular Panel | | |
| | EXPD 129P/ EXPD 129PG<br>: 17" X 51" | EVS 4343A: 17" X 17"<br>EVS 4343AG: 17" X 17" | | |
| | EXPD 86P/ EXPD 86PG<br>: 17" X 34" | EVS 3643A: 13" X 17"<br>EVS 3643AG: 13" X 17" | | |
| Dimensions | EXPD 129P/EXPD 129PG<br>: 460mm (W) × 1324mm (L) × 20mm (H) [±0.5mm] | EVS 4343A:<br>460(W) x 483(L) x 15.5(H)<br>EVS 4343AG:<br>460(W) x 483(L) x 15.5(H) | | |
| | PD 86P/EXPD 86PG<br>: 460mm (W) × 894.14mm (L) × 15.5mm (H) [±0.5mm] | EVS 3643A:<br>460(W) × 409(L) × 15.5(H)<br>EVS 3643AG:<br>460(W) × 409(L) × 15.5(H) | | |
| Pixel Pitch | 140μm | 140μm | | |
| Image Size | EXPD 129P/EXPD 129PG<br>: 3,072 x 9,216 | EVS 4343A: 3,072 x 3,072<br>EVS 4343AG: 3,072 x 3,072 | | |
| | EXPD 86P/EXPD 86PG<br>: 3,072 x 6,144 | EVS 3643A: 2,560 x 3,072<br>EVS 3643AG: 2,560 x 3,072 | | |
| Scintillator & TFT<br>Materials | TFT -amorphous Silicon | TFT -amorphous Silicon | | |
| | EXPD 129P/ EXPD 86P: CsI | EVS 4343A/EVS 3643A: CsI | | |
| | EXPD 129PG/ EXPD 86PG: Gadox | EVS 4343AG/ EVS 3643AG: GOS | | |
| Performance | DQE | EXPD 129P: 50.0 % at 0.5lp/mm | EVS 4343A: 52.9% at 1.0 lp/mm | |
| | | EXPD 129PG: 25.0 % at 0.5lp/mm | EVS 4343AG: 27.2 % at 1.0 lp/mm | |
| | | EXPD 86P: 50.0 % at 0.5lp/mm<br>EXPD 86PG: 25.0 % at 0.5lp/mm | EVS 3643A: 50.5 % at 1.0 lp/mm<br>EVS 3643AG: 26.3 % at 1.0 lp/mm | |
| | MTF | EXPD 129P: 45.0 % at 2.0 lp/mm<br>EXPD 129PG: 45.0 % at 2.0 lp/mm<br>EXPD 86P: 45.0 % at 2.0 lp/mm<br>EXPD 86PG: 45.0 % at 2.0 lp/mm | EVS 4343A: 44.1 % at 2.0 lp/mm<br>EVS 4343AG: 49.2 % at 2.0 lp/mm<br>EVS 3643A: 44.5% at 2.0 lp/mm<br>EVS 3643AG: 46.3 % at 2.0 lp/mm | |
| | Resolution | 3.5 lp/mm | 3.5 lp/mm | |
| Anatomical Sites | | General Radiography | General Radiography | |
| Power Supply | | 100~240V~, 50/60 Hz | 100~240V~, 50/60 Hz | |
| Component | | Adaptor & AC Power Cord<br>LLD POE Cable<br>LLD Sync Cable | PoE Adaptor<br>Lan cables(2 ea)<br>AC Power Cable<br>Ethernet Connector Bracket | |
| Communication | | | Wired: Ethernet | Wired: Ethernet |
| Parameter | Modified Device | Predicate Device | Remark | |
| DOE | EXPD 129P: 50.0 % at 0.5lp/mm | EVS 4343A: 52.9 % at 1.0 lp/mm | | |
| | EXPD 129PG: 25.0 % at 0.5lp/mm | EVS 4343AG: 27.2 % at 1.0 lp/mm | | |
| | EXPD 86P: 50.0 % at 0.5lp/mm | EVS 3643A: 50.5 % at 1.0 lp/mm | | |
| | EXPD 86PG: 25.0 % at 0.5lp/mm | EVS 3643AG: 26.3 % at 1.0 lp/mm | | |
| MTF | EXPD 129P: 45.0 % at 2.0 lp/mm | EVS 4343A: 44.1 % at 2.0 lp/mm | | |
| | EXPD 129PG: 45.0 % at 2.0 lp/mm | EVS 4343AG: 49.2 % at 2.0 lp/mm | | |
| | EXPD 86P: 45.0 % at 2.0 lp/mm | EVS 3643A: 44.5 % at 2.0 lp/mm | | |
| | EXPD 86PG: 45.0 % at 2.0 lp/mm | EVS 3643AG: 46.3 % at 2.0 lp/mm | | |
DRTECH
{8}------------------------------------------------
When compared to the predicate devices (K192400), the EXPD 86P, EXPD 86PG, EXPD 129P, EXPD 129PG presented in this submission have similar:
- . Intended Use
- . Technological characteristics
- . Operating principle
- Performance (Resolution)
- Communication Method
- . Performance (DQE and MTF)
- . Panel Material
- . Software(Firmware)
A few differences are as follows
- . Size
- Configuration of Components(Optional)
There are no significant differences between the EXPD 86PG, EXPD 129P, EXPD 129PG and the predicate device(s) that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
According to bench test report, it is proved that the DQE and MTF of predicated device and subject device are basically equal or worth than the predicate device. As a result, subject devices performance is equal or worth than the predicate device.
#### 10. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The non-clinical performance testing constrains that the main physical values for comparison of Xray devices like DQE and MTF are basically equal or worth than the predicate device as following table:
The EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG detector complies with the following international and FDA-recognized consensus standards:
| Recognition No. | Standard No. | Title of Standard | Remark |
|-----------------|--------------|-------------------|--------|
|-----------------|--------------|-------------------|--------|
{9}------------------------------------------------
| Recognition No. | Standard No. | Title of Standard | Remark |
|-----------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 19-46 | ANSI AAMI ES60601-<br>1:2005/(R)2012 & A1:2012,<br>C1:2009/(R)2012 &<br>A2:2010/(R)2012 (Cons. Text)<br>[Incl. AMD2:2021] | Medical electrical equipment - Part 1:<br>General requirements for basic safety<br>and essential performance (IEC 60601-<br>1:2005, MOD) [Including Amendment<br>2 (2021)] | |
| 12-289 | IEC 62220-1-1 Edition 1.0 2015-03 | Medical electrical equipment-<br>Characteristics of digital X-ray imaging<br>devices Part 1-1: Determination of the<br>detective quantum efficiency Detectors<br>used in radiographic imaging | |
| 5-125 | ISO 14971 Third Edition 2019-12 | Medical devices - Application of risk<br>management to medical devices | |
| 5-132 | IEC 60601-1-6 Edition 3.2 2020-07<br>CONSOLIDATED VERSION | Medical electrical equipment - Part 1-6:<br>General requirements for basic safety<br>and essential performance - Collateral<br>standard: Usability | |
| 12-342 | PS 3.1 - 3.20 2021e | Digital Imaging and Communications in<br>Medicine (DICOM) Set | |
| 19-36 | IEC 60601-1-2 Edition 4.1 2020-09<br>CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety<br>and essential performance - Collateral<br>Standard: Electromagnetic disturbances<br>- Requirements and tests | |
| 13-79 | IEC 62304 Edition 1.1 2015-06<br>CONSOLIDATED VERSION | Medical device software - Software life<br>cycle processes | |
DRTECH
### 11. Related FDA Guidance documents
We reviewed the below FDA guidance documents and it was reflected to development of the modified EXPD 86P/ EXPD 86PG/ EXPD 129P/ EXPD 129PG detector.
| Title of Guidance | Remark |
|-------------------------------------------------------------------------------------|--------|
| Content of Premarket Submissions for Management of Cybersecurity in Medical Devices | |
| Pediatric Information for X-ray Imaging Device Premarket Notifications | |
{10}------------------------------------------------
#### 13. Conclusion [21 CFR 807.92(b)(3)]
The modified EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG detector is substantially equivalent to the currently marketed and predicate device (EVS 4343A / EVS 4343AG / EVS 3643A / EVS 3643AG - K192400) in terms of fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, ANSI AAMI ES60601-1:2005/AMD2:2021, IEC 60601-1-2:2020, IEC 60601-1-6 Edition 3.2, IEC 62304:2015 and IEC 6220-1-1 which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.
The results of these tests demonstrate that The EXPD 86P / EXPD 86PG/ EXPD 129P/ EXPD 129PG meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data and safety testing demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.