← Product Code [JDW](/productcode/JDW) · K223112

# MAVERICK Mini External Fixation System (K223112)

_Smith & Nephew, Inc. · JDW · Dec 8, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223112

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JDW](/productcode/JDW.md)
- **Decision Date:** Dec 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Smith & Nephew's MAVERICK and MAVERICK Mini External Fixation System is intended to be used for temporary stabilization of the long bones, feet, pelvis, and wrist. The indications for the Smith & Nephew MAVERICK Mini External Fixation System are the following: - Open and closed fracture fixation - Bony or soft tissue deformities

## Device Story

MAVERICK Mini External Fixation System provides temporary bone stabilization; used in orthopedic surgery for fracture management and deformity correction. System comprises stainless steel bone pins, composite/titanium bars, and titanium clamps; components assembled into custom frames tailored to specific anatomical applications. Operated by surgeons in clinical/OR settings. Principle of operation involves securing bone pins to an external frame via clamps to maintain alignment and stability. Device output is a rigid construct providing mechanical support to the skeletal site. Benefits include stabilization of fractures and deformities to facilitate healing.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance supported by mechanical testing including torsional/axial grip strength of clamps and bars, and load-to-failure testing of full constructs. Biocompatibility assessment completed via Declaration of Conformity; MRI safety addressed via justification.

## Technological Characteristics

External fixation system; materials include stainless steel (pins), titanium (clamps, bars), and composite (bars). Principle of operation: mechanical stabilization via pin-to-bar clamping. Modular assembly allows frame customization. Non-powered, manual device. Biocompatibility confirmed; MRI safety supported by justification.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- MAVERICK External Fixation System ([K213874](/device/K213874.md))
- Smith & Nephew External Fixation System (Jet-X) ([K994143](/device/K994143.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 8, 2022

Smith & Nephew, Inc. Allison Francis Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016

Re: K223112

Trade/Device Name: MAVERICK Mini External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT. JDW Dated: September 30, 2022 Received: October 3, 2022

Dear Allison Francis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Ting Song -S

Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K223112

Device Name MAVERICK Mini External Fixation System

Indications for Use (Describe)

Smith & Nephew's MAVERICK and MAVERICK Mini External Fixation System is intended to be used for temporary stabilization of the long bones, feet, pelvis, and wrist.

The indications for the Smith & Nephew MAVERICK Mini External Fixation System are the following:

- * Open and closed fracture fixation
* Bony or soft tissue deformities

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Smith Nephew

| Submitted by:                                         | Smith & Nephew, Inc.             Orthopaedic Division             1450 East Brooks Road             Memphis, Tennessee 38116 |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary:                                      | September 30, 2022                                                                                                           |
|                                                       | Primary Contact:                                                                                                             |
|                                                       | Ms. Allison Francis             Regulatory Affairs Specialist             T 901-572-5959                                     |
|                                                       | Secondary Contact:                                                                                                           |
|                                                       | Mr. Brad Sheals, MS             Senior Regulatory Affairs Manager             T 901-288-7141             F 901-566-7911      |
| Name of Device:                                       | MAVERICK Mini External Fixation System                                                                                       |
| Common Name:                                          | Multi-lateral Fixators and Accessories                                                                                       |
| Device Classification Name and             Reference: | 21 CFR 888.3030 Single/multiple component metallic             bone fixation appliances and accessories                      |
|                                                       | 21 CFR 888.3040 Smooth or threaded metallic bone             fixation fastener                                               |
| Device Class:                                         | Class II                                                                                                                     |
| Panel Code:                                           | Orthopaedics/87                                                                                                              |
| Product Code:                                         | KTT, JDW                                                                                                                     |

### Predicates

| Manufacturer            | Description                                              | Submission<br>Number | Clearance Date       |
|-------------------------|----------------------------------------------------------|----------------------|----------------------|
| Smith & Nephew,<br>Inc. | MAVERICK External Fixation<br>System (Primary Predicate) | K213874              | February 11,<br>2022 |
| Smith & Nephew,<br>Inc. | Smith & Nephew External<br>Fixation System (Jet-X)       | K994143              | February 18,<br>2000 |

### Device Description

The subject of this premarket notification is the MAVERICK Mini External Fixation System. The system is an external fixation system that consists of various components used in the management of pelvic, feet, wrist, and long bone fractures and reconstructive and corrective orthopedic surgery. The system consists of stainless steel bone pins, composite and titanium bars, and titanium clamps. These components can be combined to build a frame which is appropriate for each specific application.

{4}------------------------------------------------

## Smith Nephew

### Indications for Use

Smith & Nephew's MAVERICK and MAVERICK Mini External Fixation System is intended to be used for temporary stabilization of the long bones, feet, pelvis, and wrist.

The indications for the Smith & Nephew MAVERICK Mini External Fixation System are the following:

- Open and closed fracture fixation .
- Bony or soft tissue deformities .

### Technological Characteristics

The principle of operation for the subject and the predicate devices is stabilization of bones through fixation with pins clamped to an external fixation frame built with composite bars. The MAVERICK Mini External Fixation System is similar to legally marketed predicate devices in that they are manufactured from similar materials, are of structurally similar size, strength and stiffness, and incorporate similar methods of assembly and adjustment.

### Performance Testing

To further support a determination of substantial equivalence, Mechanical testing and/or assessments were utilized. A review of the data indicates that the subject MAVERICK Mini External Fixation System is substantially equivalent to one or more of the previously cleared predicate devices listed above. The following overall bench performance strategy was used as a basis for the determination of substantial equivalence.

| Device Testing                                 |                                         |
|------------------------------------------------|-----------------------------------------|
| Subject Clamps Testing                         | Torsional Grip Strength with Pins       |
|                                                | Axial Grip Strength with Pins           |
|                                                | Resistance to Slip at Internal Swivel   |
|                                                | Junction                                |
|                                                | Axial Grip Strength with Bars           |
|                                                | Torsional Grip Strength of Bars         |
| Rigidity Testing to Failure of Full Constructs | Load to Failure of Wrist Spanning Style |
|                                                | Frames – double pin clamps              |
|                                                | Load to Failure of Wrist Spanning Style |
|                                                | Frames - bar to pin clamps              |
| Mating to MAVERICK Large System                | Large Clamp Grip on 6mm Bars            |

### Overall Bench Performance Testing Strategy

A Declaration of Conformity for biocompatibility assessment was completed. A MRI justification was utilized in lieu of testing.

### Conclusion

Based on similarities to the predicate multi-lateral external fixation systems and a review of testing performed, the subject MAVERICK Mini External Fixation System is substantially equivalent to the commercially available predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K223112](https://fda-staging.innolitics.com/device/K223112)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).