← Product Code [FPA](/productcode/FPA) · K223101

# BD Secondary Infusion Set (K223101)

_Care Fusion · FPA · May 12, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223101

## Device Facts

- **Applicant:** Care Fusion
- **Product Code:** [FPA](/productcode/FPA.md)
- **Decision Date:** May 12, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The BD Secondary Infusion Set is intended for the intravascular administration of fluids from a container to a patient's vascular system.

## Device Story

BD Secondary Infusion Set is a gravity-fed administration set for delivering IV fluids and medications (including lipids) to a patient's vascular system. Components include a bag spike, drip chamber, tubing, roller clamp, and micro luer lock connector. The device is used in clinical settings by healthcare providers to facilitate fluid delivery. The provider inserts the bag spike into a fluid container and connects the distal end to the patient's vascular access device. The roller clamp allows for manual flow regulation. The device is single-use, sterile, and supports use for up to 7 days (168 hours). It benefits patients by providing a reliable, sterile pathway for continuous or intermittent intravenous therapy.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), sterility, and performance testing per ISO 594-1, ISO 594-2, ISO 8536-4, and ISO 8536-14.

## Technological Characteristics

Gravity-fed infusion set. Materials: PVC, ABS, Polyethylene, MABS. Connectivity: None. Sterilization: Gamma irradiation (SAL 10^-6). Standards: ISO 594-1/2 (luer fittings), ISO 8536-4 (gravity infusion), ISO 8536-14 (clamps), ISO 11137 (sterilization), ISO 11607 (packaging), ISO 10993 (biocompatibility).

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Intravascular Administration and Extension Sets ([K051499](/device/K051499.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a sans-serif font.

May 12, 2023

CareFusion Paulina Davis Staff Regulatory Affairs Specialist 10020 Pacific Mesa Blvd San Diego, California 92121

Re: K223101

Trade/Device Name: BD Secondary Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: April 12, 2023 Received: April 13, 2023

# Dear Paulina Davis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223101

Device Name BD Secondary Infusion Set

#### Indications for Use (Describe)

The BD Secondary Set is indicated for continuous or intermittent delivery of IV fluids, medications, including lipids, through intravascular routes of administration. The BD Secondary Set may be used with any patient with consideration given to the procedure being performed and fluids being infused.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="color: #000000;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: #000000;"> </span> </span> </span> <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="color: #000000;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> | <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="color: #000000;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: #000000;"> </span> </span> </span> <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="color: #000000;">Over The Counter Use (21 CFR 8</span> </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| X | Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".

# 510(k) SUMMARY K223101

|                          | Submitter Name:           | CareFusion                                                                                                         |
|--------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------|
|                          | Submitter Address:        | 10020 Pacific Mesa Blvd<br>San Diego, CA 92121, USA                                                                |
| Submitter<br>Information | Contact Person:           | Paulina Davis<br>Staff Regulatory Affairs Specialist<br>Phone: 714-330-6037<br>Email Address: Paulina.Davis@BD.com |
|                          | Date of Preparation:      | May 8, 2023                                                                                                        |
|                          | Trade Name:               | BD Secondary Infusion Set                                                                                          |
|                          | Common Name:              | Intravascular Administration Set and Extension Set                                                                 |
|                          | Regulation Name:          | Intravascular Administration Set                                                                                   |
| Subject Device           | Classification Panel:     | General Hospital                                                                                                   |
|                          | Regulation Number:        | 21 CFR 880.5440                                                                                                    |
|                          | Regulatory Class:         | Class II                                                                                                           |
|                          | Product Code:             | FPA                                                                                                                |
|                          | Trade Name:               | Intravascular Administration Set and Extension Set                                                                 |
|                          | Premarket Notification #: | K051499                                                                                                            |
|                          | Regulation Name:          | Intravascular Administration Set                                                                                   |
|                          | Classification Panel:     | General Hospital                                                                                                   |
| Predicate Device         | Regulation Number:        | 21 CFR 880.5440                                                                                                    |
|                          | Regulatory Class:         | Class II                                                                                                           |
|                          | Product Code:             | FPA                                                                                                                |
|                          | Manufacturer:             | Medegen Medical Manufacturing Services<br>(Now owned by Becton, Dickinson and Company)                             |

| Reason for<br>Submission | CareFusion is submitting this traditional premarket notification for the BD<br>Secondary Infusion Set in support of a new 510(k) baseline for the<br>Secondary Infusion Set.                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description       | The BD Secondary Infusion Set is an administration set consisting<br>primarily of a bag spike, drip chamber, tubing, roller clamp, micro luer<br>lock connector, and a hanger. The drip chamber cap is located at the<br>proximal end of the administration set, and the micro luer lock cap is<br>located at the distal end of the administration set. The bag spike, located<br>on the proximal end of the drip chamber is inserted into a prepared fluid<br>container. The BD Secondary Infusion Set is supplied as fluid-path sterile<br>using gamma irradiation in a perforated pouch, is non-pyrogenic, and is for |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield inside. A portion of the shield is colored yellow. To the right of the logo is the text "CareFusion" in gray. Below the word "CareFusion" is the text "has joined BD" in a smaller, lighter gray font.

|                                  | single-use only. The BD Secondary Infusion Set is for gravity<br>administration. The device is intended for prescription use (RX) only.<br>DEHP or natural rubber latex are not part of the material formulation. The<br>set can be used for up to 7 days (168 hours) provided the set is<br>continuously attached to the primary administration set.                                                             |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                     | The BD Secondary Infusion Set is intended for the intravascular<br>administration of fluids from a container to a patient's vascular system.                                                                                                                                                                                                                                                                      |
| Indications for Use              | The BD Secondary Set is indicated for continuous or intermittent delivery of<br>IV fluids, medications, including lipids, through intravascular routes of<br>administration. The BD Secondary Set may be used with any patient<br>population with consideration given to the procedure being performed and<br>fluids being infused.                                                                               |
| Technological<br>Characteristics | The BD Secondary Infusion Set was shown to be substantially equivalent to<br>the predicate device cleared per K051499. Any differences in technological<br>characteristics such as the material were addressed through<br>biocompatibility and performance testing. The biocompatibility and<br>performance data demonstrated substantial equivalence. There were no<br>new questions of safety or effectiveness. |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, it says "has joined BD" in a smaller font.

The following table provides a comparison between the subject and the predicate device.

| Characteristic               | Subject Device:<br>BD Secondary Infusion Sets                                                                                                                                                                              | Predicate Device:<br>Intravascular Administration and<br>Extension Sets                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                                                                |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Status                | Subject of 510(k)                                                                                                                                                                                                          | K051499<br>Clearance date June 22, 2005                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                   |
| Applicant                    | CareFusion (Now BD)                                                                                                                                                                                                        | Medegen Medical Manufacturing<br>Services (Now BD)                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                   |
| Model Numbers                | 72215N<br>10016073                                                                                                                                                                                                         | Various                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                   |
| Classification<br>Regulation | Same as predicate                                                                                                                                                                                                          | 21 CFR 880.5440                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                   |
| Regulation Name              | Same as predicate                                                                                                                                                                                                          | Intravascular Administration Set                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                   |
| Device Class                 | Same as predicate                                                                                                                                                                                                          | ll                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                   |
| Product Code                 | Same as predicate                                                                                                                                                                                                          | FPA                                                                                                                                                                                                                                                    | N/A                                                                                                                                                                                                                   |
| Review Branch                | Same as predicate                                                                                                                                                                                                          | General Hospital                                                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                   |
| Indications for Use          | The BD Secondary Set is indicated<br>for continuous or intermittent<br>delivery of IV fluids, medications,<br>including lipids through intravascular<br>routes of administration. The BD<br>Secondary Set may be used with | The Intravascular Administration Set and<br>Extension Set is a device used to<br>administer fluids from a container to a<br>patient's vascular system through a<br>needle or catheter inserted into the<br>patient's artery or vein. The Intravascular | The indications are essentially the<br>same with the exception being the<br>subject device lists specific fluid<br>types and a general patient<br>population. The predicate device<br>does not specify fluid types or |
| Characteristic               | Subject Device:<br>BD Secondary Infusion Sets                                                                                                                                                                              | Predicate Device:<br>Intravascular Administration and<br>Extension Sets                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                                                                |
|                              | any patient population with<br>consideration given to the<br>procedure being performed and<br>fluids being infused.                                                                                                        | Administration Set and Extension Set<br>may incorporate componentry that aid in<br>the prevention of accidental needle<br>sticks.                                                                                                                      | patient population. The subject<br>device fluids were assessed based<br>on clinical use. Functional testing<br>on the subject device was<br>conducted to demonstrate<br>compatibility with fluid types.               |
| Intended Use                 | The BD Secondary Infusion Set is<br>intended for the intravascular<br>administration of fluids from a<br>container to a patient's vascular<br>system.                                                                      | The Intravascular Administration Set and<br>Extension Set is a device used to<br>administer fluids from a container to a<br>patient's vascular system through a<br>needle or catheter inserted into the<br>patient's artery or vein.                   | The subject device intended use is<br>the same; intended for fluid<br>administration to the patient's<br>vascular system through a needle<br>or catheter.                                                             |
| Drip Chamber                 | PVC, Acrylonitrile<br>Butadiene Styrene (ABS),<br>Polyethylene                                                                                                                                                             | Butadiene Styrene (ABS),<br>Polypropylene, PTFE/PES, PVC                                                                                                                                                                                               | The safety of materials that<br>comprise the finished device was<br>demonstrated by the results of<br>biocompatibility.                                                                                               |
| Tubing (Non<br>DEHP)         | Same as predicate                                                                                                                                                                                                          | PVC                                                                                                                                                                                                                                                    | The safety of materials that<br>comprise the finished device was<br>demonstrated by the results of<br>biocompatibility.                                                                                               |
| Characteristic               | Subject Device:<br>BD Secondary Infusion Sets                                                                                                                                                                              | Predicate Device:<br>Intravascular Administration and<br>Extension Sets                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                                                                |
| Roller Clamp                 | Acrylonitrile<br>Butadiene Styrene (ABS)                                                                                                                                                                                   | Material type not listed in predicate<br>510(k)                                                                                                                                                                                                        | Roller clamp is not patient<br>contacting. Verification testing was<br>completed to demonstrate the<br>performance of the roller clamp.                                                                               |
| Male Luer Lock               | Modified Acrylonitrile<br>Butadiene Styrene (MABS)                                                                                                                                                                         | Acrylic                                                                                                                                                                                                                                                | The safety of materials that<br>comprise the finished device was<br>demonstrated by the results of<br>biocompatibility.                                                                                               |
| Male Luer Lock<br>Cap        | Polyethylene                                                                                                                                                                                                               | Polypropylene                                                                                                                                                                                                                                          | The safety of materials that<br>comprise the finished device was<br>demonstrated by the results of<br>biocompatibility.                                                                                               |
| Hanger                       | Same as predicate                                                                                                                                                                                                          | HDPE                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                   |
| Set Length                   | 10016073: 31 inches<br>72215N: 40 inches                                                                                                                                                                                   | 40 inches                                                                                                                                                                                                                                              | Substantially equivalent; the subject<br>and predicate devices have a<br>similar set length.                                                                                                                          |
| Priming Volume               | 10016073:  11 mL<br>72215N:   13 mL                                                                                                                                                                                        | Not listed in predicate 510(k)                                                                                                                                                                                                                         | Verification testing was completed<br>for the priming volume.                                                                                                                                                         |
| Drip Rate                    | 10016073: 20/mL<br>72215N: 15/mL                                                                                                                                                                                           | 15/mL and 20/mL                                                                                                                                                                                                                                        | Verification testing was conducted<br>to confirm performance of drip rate.                                                                                                                                            |
| Characteristic               | Subject Device:<br>BD Secondary Infusion Sets                                                                                                                                                                              | Predicate Device:<br>Intravascular Administration and<br>Extension Sets                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                                                                |
| Packaging Type               | Perforated Pouch                                                                                                                                                                                                           | Peelable Tyvek/Film Pouch                                                                                                                                                                                                                              | The packaging is not the sterile<br>barrier. The sterile barrier consists<br>of the drip chamber cap and male<br>luer cap. Microbial barrier testing<br>was conducted as part of shelf-life<br>performance testing.   |
| Sterilization<br>Method      | Same as predicate                                                                                                                                                                                                          | Irradiation                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                   |
| Sterility Assurance<br>Level | Same as predicate                                                                                                                                                                                                          | 10-6                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                   |
| Device Utility               | Same as predicate                                                                                                                                                                                                          | Disposable, Single use only, Fluid path<br>sterile                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                   |
| Shelf Life                   | Same as predicate                                                                                                                                                                                                          | 3 years                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                   |
| Device Duration              | Up to 7 days (168 hours)                                                                                                                                                                                                   | Up to 96 hours                                                                                                                                                                                                                                         | Verification testing was conducted<br>to show performance of the subject<br>device for the intended duration of 7<br>days (168 hours).                                                                                |
| Mode of Fluid<br>Delivery    | Same as predicate                                                                                                                                                                                                          | Gravity                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                   |
| Biocompatibility             | Same as predicate                                                                                                                                                                                                          | Meets ISO 10993                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                   |

3 | Раде

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the CareFusion logo, which consists of an orange circle with a white shield inside. A yellow arrow is pointing to the right inside the shield. To the right of the logo is the text "CareFusion" in gray. Below the text is the phrase "has joined BD" in a smaller, lighter gray font.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, it says "has joined BD" in a smaller font.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the CareFusion logo, which includes an orange circle with a white shield and a yellow accent on the upper right. To the right of the logo is the word "CareFusion" in gray. Below the word "CareFusion" are the words "has joined BD" in a smaller, lighter gray font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".

# Safety and Performance Testing

## Nonclinical Testing

The BD Secondary Infusion Set was evaluated via non-clinical safety and performance testing to demonstrate that the subject device is substantially equivalent to the predicate device.

The bench and nonclinical testing on the BD Secondary Infusion Set was conducted according to the following FDA recognized consensus standards listed below. The subject device met the applicable test specifications and acceptance criteria as described in the submission.

- . FDA Recognition Number 6-11, ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
- . FDA Recognition Number 6-129, ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- . FDA Recognition Number 6-447, ISO 8536-4:2019, Infusion equipment for medical use -Part 4: Infusion sets for single use, gravity feed
- . ISO 8536-14:2016, Infusion equipment for medical use – Part 14: Clamps and flow regulators for transfusion and infusion equipment without fluid contact
- . USP 41, National Formulary 36 (USP), General Chapter <788> Particulate Matter In Injections
- FDA Recognition Number 14-528, ISO 11137-1:2006, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
- FDA Recognition Number 14-409, ISO 11137-2:2013, Sterilization of health care . products - Radiation - Part 2: Establishing the sterilization dose
- FDA Recognition Number 14- 514, ISO 11737-1:2018, Sterilization of health care . products -- Microbiological methods -- Part 1: Determination of a population of microorganisms on products
- . FDA Recognition Number 14-530. ISO 11607-1:2019. Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems

## Biocompatibility

Biocompatibility tests for the subject device were performed in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller font, it says "has joined BD".

Management Process as recognized by the FDA. The BD Secondary Infusion Set is categorized as an externally communicating, indirect blood path, prolonged (> 24 hours to 30 days) device in accordance with ISO 10993-1 and FDA Guidance, Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." The battery of tests included the following:

- Cytotoxicity (ISO 10993-5, FDA recognition number 2-245) ●
- Sensitization (ISO 10993-10, FDA recognition number 2-296) .
- Irritation or Intracutaneous Activity (ISO 10993-10, FDA recognition number 2-296) ●
- Acute Systemic Toxicity (ISO 10993-11, FDA recognition number 2-255) ●
- Sub-acute/ Sub-chronic Toxicity (ISO 10993-11, FDA recognition number 2-255) ●
- Material Mediated Pyrogenicity (ISO 10993-11, FDA recognition number 2-255) ●
- Hemocompatibility (ASTM F756, FDA Recognition number 2-250 and ISO 10993-4, ● FDA recognition number 2-248)

## Clinical Data

Not Applicable. There are no clinical data included in this submission.

## Conclusion

The information provided in this 510(k) submission, including the non-clinical safety and performing testing, is sufficient to demonstrate substantial equivalence of the BD Secondary Infusion Set to the predicate device. The BD Secondary Infusion Set and its predicate have similar indications for use, the same intended use, and similar technological characteristics. Any differences between the subject and predicate devices are supported by performance data and do not raise new questions of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K223101](https://fda-staging.innolitics.com/device/K223101)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).