Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14)

K223100 · Medline Industries, Inc. · CAF · Jun 27, 2023 · Anesthesiology

Device Facts

Record IDK223100
Device NameHudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14)
ApplicantMedline Industries, Inc.
Product CodeCAF · Anesthesiology
Decision DateJun 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Hudson RCI AQUATHERM® III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AQUAPAK® system to provide continuous heated aerosol for respiratory applications.

Device Story

The AquaTherm III Plus is an adjustable electronic heater placed between a nebulizer adaptor and a sterile solution reservoir. When gas flow initiates, sterile solution is drawn through a stainless-steel puncture pin in the heater center; the solution is heated as it flows through the pin. The heated solution enters the nebulizer adaptor, where it is aerosolized into the patient gas stream. Used in hospital environments, the device is operated by healthcare professionals. It provides continuous heated aerosol to support respiratory therapy. The device is mains-powered and features adjustable temperature settings (25°C–37°C).

Clinical Evidence

Bench testing only. No clinical data was provided. Non-clinical testing included biocompatibility (ISO 10993-1, -5, -10, -11), gas pathway evaluation (ISO 18562-1, -2, -3), EMC testing (IEC 60601-1-2), and electrical safety (IEC 60601-1-11).

Technological Characteristics

Mains-powered (115 VAC, 60Hz) external adjustable electronic heater. Materials include stainless steel (puncture pin). Dimensions/form factor: external heater unit. Connectivity: standalone. Sterilization: non-sterile. Standards: ISO 10993, ISO 18562, IEC 60601-1-2, IEC 60601-1-11. Temperature range: 25°C–37°C.

Indications for Use

Indicated for use in hospital settings to provide continuous heated aerosol for respiratory applications via the Hudson RCI Nebulizer Adaptor and AQUAPAK system.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 27, 2023 Medline Industries, Inc % Joy Gutermuth Senior Specialist (Consultant) Ram+ 2790 Mosside Blvd. Suite 800 Monroeville, Pennsylvania 15146 Re: K223100 Trade/Device Name: Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14) Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: June 13, 2023 Received: June 13, 2023 Dear Joy Gutermuth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/5 description: The image shows the name "Ethan L. Nyberg -S" in a large, sans-serif font. The text is black against a white background. The letters are evenly spaced and clearly legible. Ethan Nyberg, Ph.D Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. Submission Number (if known) K223100 Device Name Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater (050-14) Indications for Use (Describe) The Hudson RCI AQUATHERM® III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AQUAPAK® system to provide continuous heated aerosol for respiratory applications. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # DATE PREPARED June 26, 2023 ## MANUFACTURER AND 510(k) OWNER Medline Industries 1 Three Lakes Drive, Northfield, IL, 60093 Telephone: Official Contact: Nicole Schaffer, ### REPRESENTATIVE/CONSULTANT Joy Gutermuth RQM+ Telephone: Email: iqutermuth@rgmplus.com ### DEVICE INFORMATION Proprietary Name/Trade Name: Hudson RCI AquaTherm III Plus External Adjustable | | Electronic Heather ("AquaTherm Heater") | |--------------------|-----------------------------------------------------------------------| | Common Name: | Nebulizer Heater | | Regulation Number: | 868.5630 | | Class: | II | | Product Code: | CAF | | Premarket Review: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental<br>Devices (OHT1) | | Review Panel: | Anesthesiology | # PREDICATE DEVICE IDENTIFICATION The AquaTherm Heater is substantially equivalent to the following predicates: | 510(k)<br>Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | |------------------|-------------------------------------------------------------|----------------------| | K880473 | Aero-Mist Nebulizer Heater/ Pegasus Research<br>Corporation | ✓ | The predicate devices have not been subject to a design related recall. # DEVICE DESCRIPTION The AquaTherm devices are adjustable output temperature heaters. The AquaTherm Heater is placed between a nebulizer adaptor and a reservoir of sterile solution. When a gas flow is initiated to the nebulizer, sterile solution is drawn up from the reservoir and through the stainless-steel puncture pin in the center of the AquaTherm Heater. The solution is heated as it flows through the pin. The heated solution is then drawn into the nebulizer adaptor where it is aerosolized into the patient gas stream. {4}------------------------------------------------ # INDICATIONS FOR USE The Hudson RCI AquaTherm III Plus External Adjustable Electronic Heater is designed for use with the Hudson RCI Nebulizer Adaptor and the AquaPak system to provide continuous heated aerosol for respiratory applications. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS Medline Industries believes that the AquaTherm Heater is substantially equivalent to the predicate devices based on the information summarized here: The subject device has a similar design and dimensions, and uses similar or identical materials as the device cleared in K880473. The subject device has the same intended use and similar technological characteristics to the device cleared in K880473. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicate. | Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater | |-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Description | The AquaTherm devices are<br>adjustable output temperature<br>heaters. The AquaTherm Heater is<br>placed between a nebulizer<br>adaptor and a reservoir of sterile<br>solution. When a gas flow is<br>initiated to the nebulizer, sterile<br>solution is drawn up from the<br>reservoir and through the<br>stainless-steel puncture pin in the<br>center of the AquaTherm Heater.<br>The solution is heated as it flows<br>through the pin. The heated<br>solution is then drawn into the<br>nebulizer adaptor where it is<br>aerosolized into the patient gas<br>stream. | The heater warms and evaporates<br>particles impinging on the heater<br>disc which is disposed below the<br>nebulizer. Return water drains<br>through the center tube into the<br>sterile water supply container<br>below the heater. The water is<br>drawn up from the sterile water<br>supply by a draw tube which is part<br>of the nebulizer itself. The draw<br>tube passes through the center<br>tube of the heater in use. | | Intended Use | To provide continuous heated<br>aerosol for respiratory<br>applications. | "… to provide warm aerosol of<br>water particles into the respiratory<br>tract."*<br>*Intended use not clearly stated in<br>510(k) summary | | Indications for Use | The Hudson RCI AquaTherm III<br>Plus External Adjustable<br>Electronic Heater is designed for | "Respiratory therapy, in hospital, to<br>provide warm aerosol of water<br>particles into the respiratory tract."* | | Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater | | | use with the Hudson RCI<br>Nebulizer Adaptor and the<br>AquaPak system to provide<br>continuous heated aerosol for<br>respiratory applications. | *Indications for use not clearly<br>stated in 510(k) summary | | Representative<br>Image*<br>*images feature<br>nebulizer adaptors<br>and tubing which<br>are not included<br>within this<br>submission. | Image: Hudson RCI AquaTherm III Plus Nebulizer Heater | Image: Pegasus Research Corporation Aero-Mist Nebulizer Heater | | Classification | Class II | Class II | | Product Code | CAF | CAF | | Regulation<br>Number | §868.5630 | §868.5630 | | Regulation Name | Nebulizer | Nebulizer | | Accessories | Aquapak Nebulizer adapter Aquapak prefilled nebulizer reservoir Sold separately and not part of<br>this submission | Nebulizer adapter Prefilled nebulizer reservoir Sold separately | | Environment | Hospital | Hospital | | Sterilization<br>Method | Non-sterile | Non-sterile | | Patient Contacting<br>Materials | Indirect patient contact only:<br>Stainless steel | Unknown- 510(k) does not<br>describe patient-contacting<br>materials | | Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater | | Principles of<br>Operation | Adjustable temperature. The<br>AquaTherm Heater is placed<br>between a nebulizer adaptor (not<br>included) and a reservoir of sterile<br>solution (not included). When a<br>gas flow is initiated to the<br>nebulizer, sterile solution is drawn<br>up from the reservoir and through<br>the stainless-steel puncture pin in<br>the center of the AquaTherm<br>Heater. The solution is heated as<br>it flows through the pin. The<br>heated solution is then drawn into<br>the nebulizer adaptor where it is<br>aerosolized into the patient gas<br>stream. Rainout is returned to the<br>reservoir via the nebulizer adapter<br>return tube (rainout return tube is<br>part of the nebulizer adapter, not<br>the subject device). | Adjustable temperature. Warms<br>and evaporates particles on the<br>heater disc below the nebulizer.<br>Return water drains through the<br>center tube into the container<br>below the heater. The water is<br>drawn up from the water supply by<br>a draw tube; the nebulizer draw<br>tube passes through the center<br>tube of the heater. | | Power Supply | Mains powered | Mains powered | | Voltage and Power | 115 VAC at 60Hz, 1.5 A | 115 VAC (+/- 10 volts) 1.7 A | | Storage and<br>Transportation<br>Environment | Temperature: -20 °C to 50 °C<br>Relative Humidity: 0% to 94%<br>(non-condensing) | Temperature: -40 °C to 70 °C<br>Relative Humidity: 10% to 100%<br>(including condensate) | | Output<br>Temperature<br>Range | 25°C -37°C | 25°C -40°C | | Heat up Time | 25 minutes (about 20 minutes<br>after each subsequent<br>temperature adjustment) | Approximately 20 minutes | | Operating<br>Environment | • Temperature: 18°C to 26°C<br>• Relative Humidity: 30% to 75%<br>(non-condensing)<br>• Pressure: 700 to 1060 hPa | • Temperature: 10°C to 40°C<br>• Relative Humidity: 30% to 75%<br>(non-condensing)<br>• Pressure: 700 to 1060 hPa | | Product Features | Proposed<br>Hudson RCI AquaTherm III Plus<br>External Adjustable Electronic<br>Heater | Predicate<br>Pegasus Research Corporation<br>Aero-Mist Nebulizer Heater | | Biocompatibility | External communicating, limited contact device that indirectly contacts tissue/bone/dentin. Indirect gas pathway | Unknown- 510(k) does not<br>describe any biocompatibility. | | Standards Utilized | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3<br>IEC 60601-1-2<br>IEC 60601-1-11 | UL544 | | Non-Clinical<br>Testing | Packaging<br>Cleaning process<br>EMC testing | Unknown- 510(k) does not state<br>testing. | {5}------------------------------------------------ {6}------------------------------------------------ {7}------------------------------------------------ # SUMMARY OF NON-CLINICAL TESTING No FDA performance standards have been established for the AquaTherm Heater. The following tests were performed to demonstrate safety based on current industry standards: - Biocompatibility ● - Packaging ● - Cleaning - EMC Testing ● The results of these tests indicate that the AquaTherm Heater is substantially equivalent to the predicate devices. # CONCLUSION Based on the testing performed, including gas pathway biocompatibility testing, and compliance to the latest edition of IEC 60601-1-2, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed AquaTherm Heater are assessed to be substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...