← Product Code [DQK](/productcode/DQK) · K223094

# EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade; EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™, Software License (K223094)

_ABBOTT MEDICAL · DQK · Dec 27, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K223094

## Device Facts

- **Applicant:** ABBOTT MEDICAL
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Dec 27, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures. EnSite™ X EP System Contact Force Software License: When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters. EnSite™ X EP System Surface Electrode Kit: The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use. The EnSite™ X EP System TactiFlex™ Ablation Catheter Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use. EnSite™ X EP System, TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade and EnSite™ X EP System, TactiFlexTM Ablation Catheter, Sensor EnabledTM, Software License: The EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.

## Device Story

EnSite™ X EP System is a catheter navigation and mapping system used in clinical EP labs by physicians. It processes signals from electrophysiology catheters to display 3D catheter position, cardiac electrical activity (waveform traces), and 3D isopotential/isochronal maps of cardiac chambers. The system constructs anatomical models by collecting and labeling points as the catheter moves within the heart. The subject device (v2.0.1) adds compatibility with the TactiFlex™ Ablation Catheter, Sensor Enabled™, providing user-selectable Force Number Refresh Rates and a Force Direction Indicator (arrows at catheter tip). These features assist clinicians in visualizing contact force and orientation during ablation procedures, potentially improving procedural precision and patient outcomes. The system integrates with TactiSys™ Quartz Equipment for force visualization.

## Clinical Evidence

Bench testing only. Design verification activities, including software/system-level verification, preclinical validation, human factors evaluations, and electrical safety testing, confirmed the system meets user requirements and safety standards (IEC 62304, ISO 14971, ANSI AAMI IEC 62366-1, IEC 60601-1).

## Technological Characteristics

Catheter navigation and mapping system; 3D cardiac modeling; software-based signal processing; connectivity with TactiSys™ Quartz Equipment; compatible with Sensor Enabled™ catheters. Standards: IEC 62304, ISO 14971, ANSI AAMI IEC 62366-1, IEC 60601-1.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- EnSite™ X EP System ([K221213](/device/K221213.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 27, 2022

Abbott Medical Alyssa Timmers Senior Regulatory Affairs Specialist One St. Jude Medical Device St. Paul, Minnesota 55117

Re: K223094

Trade/Device Name: EnSite™ X EP System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK Dated: September 29, 2022 Received: September 30, 2022

Dear Alyssa Timmers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223094

Device Name EnSite™ X EP System

#### Indications for Use (Describe) EnSite™ X EP System

The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.

The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.

EnSite™ X EP System Contact Force Software License:

When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.

EnSite™ X EP System Surface Electrode Kit:

The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.The EnSite™ X EP System TactiFlex™ Ablation Catheter Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.

EnSite™ X EP System, TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade and EnSite™ X EP System, TactiFlexTM Ablation Catheter, Sensor EnabledTM, Software License:

The EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;">         ☑ Prescription Use (Part 21 CFR 801 Subpart D)       </span> | <span style="white-space: nowrap;">         ☐ Over-The-Counter Use (21 CFR 801 Subpart C)       </span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

The 510(k) Summary was drafted in accordance with 21 CFR 807.92 and is included below.

| 510(k) Information                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                          | K223094                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 510(k) Type                            | Traditional 510(k)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared                          | 29 September 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Submitter Information                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Manufacturer Name & Address            | Abbott Medical<br>One St. Jude Medical Drive,<br>St. Paul, Minnesota, 55117, USA<br>Manufacturer of the EnSite™ X EP System<br><br>Abbott Medical Costa Rica Ltda.<br>Edificio #44 Calle 0, Ave. 2<br>Zona Franca Coyol<br>El Coyol Alajuela, Costa Rica 1897-4050<br>Manufacturer of the EnSite™ X EP System Surface Electrode Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Contact Person                         | Alyssa Timmers<br>Senior Regulatory Affairs Specialist<br>651-756-3706<br>alyssa.timmers@abbott.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| EnSite™ X EP System Device Information |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Trade Name                             | EnSite™ X EP System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Common Name                            | Programmable Diagnostic Computer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Class                                  | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name                    | 870.1425, computer, diagnostic, programmable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code                           | DQK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device                       | EnSite™ X EP System (K221213)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description                     | The EnSite™ X EP System is a catheter navigation and mapping system. A catheter<br>navigation and mapping system is capable of displaying the 3-dimensional (3-D)<br>position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as<br>well as displaying cardiac electrical activity as waveform traces and as three-<br>dimensional (3D) isopotential and isochronal maps of the cardiac chamber.<br><br>The contoured surfaces of the 3D maps are based on the anatomy of the patient's<br>own cardiac chamber. The system creates a model by collecting and labeling the<br>anatomic locations within the chamber. A surface is created by moving a selected<br>catheter to locations within a cardiac structure. As the catheter moves, points are<br>collected at and between all electrodes on the catheter. A surface is wrapped around<br>the outermost points. |
| Indications for Use                    | EnSite™ X EP System<br>The EnSite™ X EP System is a suggested diagnostic tool in patients for whom<br>electrophysiology studies have been indicated.<br>The EnSite™ X EP System provides information about the electrical activity of the<br>heart and displays catheter location during conventional electrophysiological (EP)<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                        | EnSite™ X EP System Contact Force Software License                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                        | When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System<br>Contact Force Module is intended to provide visualization of force information from<br>compatible catheters.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                        | EnSite™ X EP System Surface Electrode Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                        | The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP<br>System in accordance with the EnSite™ X EP System indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                        | EnSite™ X EP System, TactiFlex™ Ablation Catheter, Sensor Enabled™,<br>Software Upgrade and Software License                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                        | The EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™<br>Software Module is indicated for use with the EnSite™ X EP System in accordance<br>with the EnSite™ X EP System indications for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Comparison                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Comparison                             | The subject devices are:<br>EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor<br>Enabled™, Software Upgrade, v2.0 andEnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor<br>Enabled™, Software License.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                        | The subject devices are intended to be installed on the EnSite™ X EP System with<br>version 2.0.1 software. The predicate device is EnSite X EP System v2.0 with the<br>Contact Force License, catalog number ENSITE-CF-01 (included in the predicate<br>submission).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                        | The subject device adds a new indication for use specific for the EnSite™ X EP<br>System TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade. The<br>predicate intended use and indications for use have not changed with this addition.<br>The addition of this intended use is not critical to the intended diagnostic use of the<br>device as it refers to use of the subject device in accordance with the indications for<br>use of the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                        | The subject device adds additional user selectable settings for Force Number<br>Refresh Rate as compared with the predicate which has a fixed setting for this. This<br>feature is available when either a TactiCath™ Contact Force Ablation Catheter,<br>Sensor Enabled™ or TactiFlex™ Ablation Catheter, Sensor Enabled™ are<br>connected.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                        | The predicate device displays force direction by a graphical representation and the<br>Force Cone feature, while the subject device uses those features and adds Force<br>Direction Indicator arrows displayed at the catheter tip at the 3D location of the<br>catheter. This feature is only available when a TactiFlex™ Ablation Catheter, Sensor<br>Enabled™ is connected.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                        | The subject and predicate device use the same fundamental scientific technology to<br>facilitate catheter position and orientation, as well as cardiac mapping and model<br>creation. There were no changes to the hardware. The device software was modified<br>to enable compatibility with Abbott's ablation catheter, TactiFlex™ Ablation Catheter,<br>Sensor Enabled™. All risks associated with these modifications were mitigated to<br>acceptable levels. No new questions of safety or effectiveness were raised.                                                                                                                                                                                                                                                                                                                                                                  |
| Non-Clinical Testing<br>Summary        | Design verification activities were performed and met their respective acceptance<br>criteria to ensure that the devices in scope of this submission are safe and effective.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                        | Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                        | The EnSite™ X EP System v2.0.1 used with the EnSite™ X EP System TactiFlex™<br>Ablation Catheter, Sensor Enabled™, Software Upgrade and License was developed<br>and tested in accordance with the following industry guidance documents and<br>standards:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                        | Guidance for the Content of Premarket Submissions for Software<br>Contained in Medical Devices IEC 62304:2015-06 Edition 1.1, Medical Device Software - Software<br>Life Cycle Processes ISO 14971:2019 Medical Devices - Application of Risk Management to<br>Medical Devices ANSI AAMI IEC 62366-1:2015, Medical devices - Part 1: Application of<br>usability engineering to medical devices IEC 60601-1: 2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 +<br>A1:2012, Medical electrical equipment - Part 1: General requirements<br>for basic safety and essential performance.                                                                                                                                                                                                                                                                                                    |
| Types of Testing Performed             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                        | Software Verification at a software and system level to test the new<br>software features added for the display of the TactiFlex™ Ablation<br>Catheter, Sensor Enabled™ Force Number Refresh Rate Force Direction Indicator Preclinical Validation Testing to confirm the system could meet user<br>requirements and its intended use after modifications Human Factors Evaluations to confirm the user interface of the subject<br>device can be used as intended by the defined user groups Electrical Safety Testing to ensure the system when used with the<br>TactiFlex™ Ablation Catheter, Sensor Enabled™ continues to comply<br>with safety standards                                                                                                                                                                                                                               |
| Statement of Equivalence               | The subject and predicate devices have the same intended use, and similar<br>indications for use. The devices operate using the same fundamental scientific<br>technology to facilitate catheter position and orientation, as well as cardiac mapping<br>and model creation. The non-clinical and clinical testing completed and submitted in<br>this Traditional 510(k) provides objective evidence the subject device is as safe, as<br>effective, and performs as well as or better than the legally marketed device<br>predicate.                                                                                                                                                                                                                                                                                                                                                       |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K223094](https://fda-staging.innolitics.com/device/K223094)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
