K223088 · Care Fusion · FPA · Apr 7, 2023 · General Hospital
Device Facts
Record ID
K223088
Device Name
BD SmartSite™ Needle-Free Connector
Applicant
Care Fusion
Product Code
FPA · General Hospital
Decision Date
Apr 7, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The BD SmartSite™ Needle-Free Connector (NFC) is a sterile, single patient use connector for needle free access to the IV line and/or IV catheter during IV therapy. The BD SmartSite™ NFC can be used for direction, intermittent infusion and/or the continuous infusion of fluids, drugs, IV nutrition, lipids, and/or blood/blood products or the aspiration of blood. The BD SmartSite™ NFC may be used with power injector procedures to a maximum pressure of 325 psi up to a flow rate of 10 mL per second.
Device Story
Needle-free connector for IV access; consists of female luer, male luer, and internal silicone valve. Operation: male luer attachment compresses internal valve/piston, opening fluid pathway for infusion or aspiration. Used in clinical settings for IV therapy; operated by healthcare professionals. Benefits: enables needle-free medication administration and fluid access; reduces needle-stick risk. Output: provides open fluid path for delivery/aspiration. Device supports power injection up to 325 psi.
Clinical Evidence
No clinical data.
Technological Characteristics
Materials: Acrylic Multi-polymer (GS90) body/male luer, Isoplast 2530 polyurethane female luer, silicone piston (Elastosil LR3003/80), silicone lubricants. Principle: Mechanical valve compression. Sterilization: Radiation (SAL 10-6). Standards: ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-23, ISO 8536-4, ISO 594-1, ISO 594-2, USP <788>, USP <85>, USP <161>, ANSI/AAMI ST72, ISO 11137-1/2, ISO 11607-1/2.
Indications for Use
Indicated for needle-free access to IV lines or catheters for direct injection, intermittent/continuous infusion of fluids, drugs, nutrition, lipids, blood/blood products, or blood aspiration. Compatible with power injectors up to 325 psi and 10 mL/s flow rate.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
April 7, 2023
CareFusion Jacob Lee Sr. Manager Regulatory Affairs 10020 Pacific Mesa Blvd San Diego, California 92121
Re: K223088
Trade/Device Name: BD SmartSite™ Needle-Free Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: March 8, 2023 Received: March 8, 2023
Dear Jacob Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223088
Device Name BD SmartSite™ Needle-Free Connector
#### Indications for Use (Describe)
The BD SmartSite™ Needle-Free Connector (NFC) is a sterile, single patient use connector for needle free access to the IV line and/or IV catheter during IV therapy. The BD SmartSite™ NFC can be used for direction, intermittent infusion and/or the continuous infusion of fluids, drugs, IV nutrition, lipids, and/or blood/blood products or the aspiration of blood. The BD SmartSite™ NFC may be used with power injector procedures to a maximum pressure of 325 psi up to a flow rate of 10 mL per second.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" are the words "has joined BD" in a smaller, lighter font.
# K223088 -0(k) Summarv
#### Submitter Information
Submitter:
Contact Person: Phone: Email: Date Prepared:
CareFusion 10020 Pacific Mesa Blvd. San Diego, CA 92121, USA Jacob Lee 385-255-0184 iacob.lee@bd.com April 7, 2023
#### Subject Device Identification
| Trade Name: |
|-----------------------|
| Common Name: |
| Classification Name: |
| Classification Panel: |
| Regulation Number: |
| Regulatory Class: |
| Product Code: |
BD SmartSite™ Needle-Free Connector Intravascular Administration Set Intravascular Administration Set General Hospital 21 CFR 880.5440 Class II FPA
#### Predicate Device Identification
SmartSite™ Needle-Free Valve Trade Name: Common Name: Intravascular Administration Set Classification Name: Intravascular Administration Set Classification Panel: General Hospital Regulation Number: 21 CFR 880.5440 Regulatory Class: Class II Product Code: FPA CareFusion, Inc. Manufacturer: 510k Number: K061285 510K Clearance Date: July 26, 2006
#### Reason for Submission
The objective of this submission is to introduce new labeling claims. Specifically, the following claims are proposed to be added to the labeling:
- Duration of use to 7 days (168 hours) or 200 activations, whichever occurs first ●
- Sterility claim from fluid path sterile to content sterile ●
- . Update disinfection swab time
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for CareFusion. The logo features an orange circle with a white shield-like shape inside, with a small orange accent in the upper right corner. To the right of the circle is the name "CareFusion" in a dark gray, sans-serif font. Below the name, in smaller, lighter gray font, is the phrase "has joined BD".
# Device Description
The BD SmartSite™ NFC allows the user to add medication into the primary line without the use of a needle. It consists of a female luer on one side and a male luer connection on the other side. Both connections have locking threads. The connector has a silicone valve inside which is in an expanded position in the free state. When the male luer end of a compatible vascular access device is attached securely to the female luer of the SmartSite connector, the valve/piston is compressed which opens the fluid pathway. This open pathway allows administration of fluids as well as aspiration through the connector without the use of a needle.
# Indication for Use
The BD SmartSite™ Needle-Free Connector (NFC) is a sterile, single patient use connector for needle free access to the IV line and/or IV catheter during IV therapy. The BD SmartSite™ NFC can be used for direct injection, intermittent infusion and/or the continuous infusion of fluids, drugs, IV nutrition, lipids, and/or blood/blood products or the aspiration of blood. The BD SmartSite™ NFC may be used with power injector procedures to a maximum pressure of 325 psi up to a flow rate of 10 mL per second.
#### Technological Characteristics
The following table presents an overview of comparisons between the subject device and the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for CareFusion. The logo features an orange circle with a white shield inside on the left. To the right of the logo is the name "CareFusion" in a sans-serif font. Below the name, in smaller letters, is the phrase "has joined BD".
| Attributes | Subject:<br>BD SmartSite™ Needle-Free<br>Connector | SmartSite Needle Free Valve<br>(K061285) | Equivalence Discussion |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Reg.<br>Number | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| FDA Regulation<br>Name | Intravascular Administration Set | Intravascular Administration Set | Same |
| FDA Class | Class II | Class II | Same |
| FDA Product<br>Code | FPA | FPA | Same |
| Indication for<br>use | The BD SmartSite™ Needle-Free<br>Connector (NFC) is a sterile, single<br>patient use connector for needle free<br>access to the IV line and/or IV<br>catheter during IV therapy. The BD<br>SmartSite™ NFC can be used for<br>direct injection, intermittent infusion<br>and/or the continuous infusion of<br>fluids, drugs, IV nutrition, lipids,<br>and/or blood/blood products or the<br>aspiration of blood. The BD<br>SmartSite™ NFC may be used with<br>power injector procedures to a<br>maximum pressure of 325 psi up to a<br>flow rate of 10 mL per second. | The SmartSite Needle Free Valve<br>Administration Sets are intended for<br>use in today's growing professional<br>healthcare environment including<br>healthcare facilities, home care, and<br>medical transport that utilize infusion<br>systems for the delivery of fluids,<br>medication, blood and blood products.<br>The SmartSite valve allows the user to<br>add medication into the primary line<br>without the use of a needle. The<br>SmartSite valve also be used with low<br>pressure power injectors rated for a<br>maximum setting of 325 psi. | Different – mechanical<br>hemolysis and blood<br>compatibility was<br>conducted to verify blood<br>aspiration claim |
| Attributes | Subject:<br>BD SmartSite™ Needle-Free<br>Connector | SmartSite Needle Free Valve<br>(K061285) | Equivalence Discussion |
| Device<br>Components /<br>Materials | SmartSite Body and Male Luer:<br>Acrylic Multi-polymer (GS90) | SmartSite Body and Male Luer:<br>Acrylic Multi-polymer (GS90) | Same |
| | SmartSite Female Luer:<br>Isoplast (2530 Polyeurethane) | SmartSite Female Luer:<br>Isoplast (2530 Polyeurethane) | |
| | Piston: Silicone (Elastosil LR3003/80) | Piston: Silicone (Elastosil LR3003/80) | |
| | Lubricant (piston opening): Silicone<br>fluid (FS-1265 Fluorosilicone Fluid) | Lubricant (piston opening): Silicone<br>fluid (FS-1265 Fluorosilicone Fluid) | |
| | Lubricant: Silicone fluid (550 Silicone<br>Fluid) | Lubricant: Silicone fluid (550 Silicone<br>Fluid) | |
| Packaging<br>Configuration | Each device is individually packaged in<br>pouches. Fifty (50) pouches and one<br>(1) Directions for Use per dispenser<br>box. Two (2) dispenser boxes per<br>shipper box. | Each device is individually packaged in<br>pouches. Fifty (50) pouches and one<br>(1) Directions for Use per dispenser<br>box. Two (2) dispenser boxes per<br>shipper box. | Same |
| Sterilization<br>Method | Radiation (SAL 10-6) | Radiation (SAL 10-6) | Same |
| Sterilization<br>Claim | Content Sterile | Fluid Path Sterile | Different - Package<br>integrity testing was<br>conducted to verify sterile<br>barrier claim. |
| Biocompatibility | Biocompatible for the intended use<br>per ISO 10993-1 | Biocompatible for the intended use<br>per ISO 10993-1 | Same |
| Non-Pyrogenic | Yes | Yes | Same |
| Non-DEHP | Yes | Yes | Same |
| No Natural<br>Rubber Latex | Yes | Yes | Same |
| Attributes | Subject:<br>BD SmartSite™ Needle-Free<br>Connector | SmartSite Needle Free Valve<br>(K061285) | Equivalence Discussion |
| Duration of Use | 7 days (168 hours) | 72 hours; 24 hours for infusions of<br>blood, blood products or lipids<br>emulsions | Different - All intended<br>fluids per indications<br>extended to 7 day<br>duration. verification<br>testing was conducted to<br>extend the device<br>duration. See Section 18<br>for Harsh Infusate<br>testing. |
| Priming Volume | 0.1 mL | 0.11 mL | Equivalent - the priming<br>volume is within the<br>specification of the<br>predicate and has no<br>impact on functional or<br>performance<br>characteristics. |
| Residual<br>Volume | 0 mL | 0 mL | Same |
| Number of<br>Valve<br>Activations | 200 | 100 | Different – verification<br>conducted to extend<br>number of valve<br>activations. |
| Method of<br>Disinfection | Prior to every access, swab top of NFC<br>access surface for 2 - 5 seconds with<br>70% isopropyl alcohol and allow to<br>dry | Prior to every access, swab top of<br>Needle-Free Valve port with 70%<br>isopropyl alcohol (1 – 2 seconds) and<br>allow to dry (approximately 30<br>seconds) | Different - the longer<br>swab time does not raise<br>new or different<br>questions of safety or<br>effectiveness. |
| Power Infusion<br>Flow Rate | ≤ 325 psi; 10 mL/s | Maximum 325 psi | Equivalent |
| Use | Single Patient Use | Single Patient Use | Same |
| Shelf Life | 3 years | 3 years | Same |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in dark gray. Below the text, there is a smaller text that says "has joined BD".
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside on the left. To the right of the circle is the word "CareFusion" in gray text. Below the word "CareFusion" are the words "has joined BD" in a smaller, lighter gray font.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is smaller text that says "has joined BD".
### Substantial Equivalence Discussion:
Design verification testing was performed to demonstrate that the subject device is equivalent to the predicate device. All test results met their acceptance criteria and support that the BD SmartSite™ NFC is safe and effective and is substantially equivalent to the predicate SmartSite Needle Free Valve. The subject device and the predicate have the equivalent indications for use and intended use. Both devices are sterilized via irradiation and are single patient use devices.
Both the subject and predicate devices have the same principle of operation. The primary technological differences between the subject device and the predicate are similar. The BD SmartSite™ NFC has a longer duration of use (up to 7 days or 200 activations, whichever comes sooner). The BD SmartSite™ NFC is also claiming content sterile and the predicate claims fluid path sterile.
The BD SmartSite™ NFC contains minor differences in technological characteristics when compared to the predicate device, these differences do not change the intended use and do not raise new questions of safety and effectiveness as supported by verification testing.
#### Discussion of Non-Clinical Tests:
The BD SmartSite™ NFC, like the predicate device, was evaluated for biocompatibility appropriate to the contact characterization (externally communicating, blood path (indirect) for prolonged duration (> 24 hours to 30 days)). Testing is performed in accordance with the requirements of ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and the FDA Guidance for Industry - Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Specific testing included:
- . ISO 10993-4:2017 Biological evaluation of medical device - Part 4: Selection of tests for interactions with blood
- . ISO 10993-5:2009 Biological evaluation of medical device - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical device – Part 10: Tests for skin sensitization
- . ISO 10993-11:2017 Biological evaluation of medical device – Part 11: Tests for systemic toxicity
- . ISO 10993-23:2021 Biological evaluation of medical device - Part 23: Test for irritation
Other standards followed included:
- ISO 8536-4:2019 Infusion equipment for medical use Part 4: Infusion Sets for . Single Use Gravity Feed, Section 8 Chemical and Biological Requirements
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in a dark gray sans-serif font. Below the text, in a smaller font, it says "has joined BD".
- USP <788>:2021 Particulate Matters in Injection
# Particulate Testing:
The BD SmartSite™ NFC was tested to demonstrate the product meets particulate requirements of United States Pharmacopeia, National Formulary (USP), General Chapter <788>, Particulate Matter in Injections (Current Standard).
#### Sterilization and Shelf Life:
The subject device is radiation sterilized and data supports a shelf-life claim of 3 years. Sterilization and shelf-life testing were completed in accordance with the following FDA recognized guidelines:
Sterilization:
- . ISO 11137-1:2006 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
- ISO 11137-2:2013 Sterilization of health care products Radiation Part 2: ● Establishing the sterilization dose
- United States Pharmacopeia, National Formulary (USP), General Chapter <85>, Bacterial Endotoxins Test
- . United States Pharmacopeia, National Formulary (USP), General Chapter <161>, Medical Devices - Bacterial Endotoxin and Pyrogen Tests
- ANSI/AAMI ST72:2019 - Bacterial endotoxins - Test methods, routine monitoring and alternatives to batch testing
Shelf-Life:
- ISO 11607-1:2019 Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems
- ISO 11607-2:2019 Packaging for Terminally Sterilized Medical Devices - Part 2: Validation Requirements for Forming, Sealing and Assembly Processes
- Package testing included:
- Standard Test Method for Determining Integrity of seals for Flexible o Packaging by Visual Inspection: ASTM F1886
- Standard Test Method for Seal Strength of Flexible Barrier Materials: ASTM o F88/F88M-21
- Seal Transfer Width: Internal testing O
- Bubble Leak Detection Test: ASTM F2096 and ASTM F2096-11 O
- Ink Legibility: Internal testing O
- Standard Test Method for Thickness Measurement of Flexible Packaging o Material: ASTM F2251-13
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller font, it says "has joined BD".
- O Label Adhesion: Internal Testing
- Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging O by Dye Penetration: ASTM F1929
# Performance Testing:
The BD SmartSite™ NFC was tested to verify compliance with the relevant sections of the following standards:
- ISO 8536-4:2019 Infusion equipment for medical use Part 4: Infusion sets for . single use, gravity feed
- ISO 594-1: 1986 Conical fittings with a 6% (Luer) taper for syringes, needles and . certain other medical equipment - Part 1: General requirements
- ISO 594-2: 1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
# Microbial Ingress Testing:
Microbial ingress was performed based on the following FDA quidance document:
- Guidance for Industry and FDA staff; Intravascular Administration Sets Premarket . Notification Submissions [510(k)], July 11, 2008
Additional performance testing was conducted to simulate use with fluids:
- Harsh Infusates testing: Device tests for multiple days with worst case infusates ●
# Clinical Data:
There are no clinical data included in this submission.
# Conclusion:
The information in this submission supports the safety and efficacy of the subject device for its intended use and demonstrates substantial equivalence with the predicate device. The BD SmartSite™ NFC differences in geometry, materials and sterilization claim do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.