K223019 · ZetrOZ Systems, LLC · PFW · Jun 23, 2023 · Physical Medicine
Device Facts
Record ID
K223019
Device Name
sam 2.0 Long Ultrasound Device
Applicant
ZetrOZ Systems, LLC
Product Code
PFW · Physical Medicine
Decision Date
Jun 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5300
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223019 · Jun 23, 2023
sam 2.0 Long Ultrasound Device
ZetrOZ Systems, LLC
Systematic literature review; ClinicalTrials.gov; PubMed; EBSCOhost; Academic Search Complete; Google Scholar
A systematic literature review and meta-analysis were conducted to evaluate the clinical effects of Sustained Acoustic Medicine (SAM) on musculoskeletal injuries, including functional outcomes, quality of life, pain reduction, and safety, to support the expansion of indications.
Systematic review; Meta-analysis; Musculoskeletal; Pain relief; Clinical literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic literature review and meta-analysis of SAM therapy; Systematic review and meta-analysis
Patients with musculoskeletal injuries (e.g., knee osteoarthritis, chronic myofascial pain, shoulder, elbow and ankle tendinopathy)
Not applicable for this study
Functional outcomes (strength and range of motion), quality of life, pain reduction, and safety profile
Indications for Use
The sam 2.0 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, the local increase in circulation, and the relief of pain associated with limited mobility and function related to soft tissue injuries (e.g., knee osteoarthritis, chronic myofascial pain, and shoulder, elbow and ankle tendinopathy).
Device Story
Device is a stationary, long-duration ultrasound system for home use; prescribed by healthcare professionals. Inputs: electrical power from lithium battery or AC adapter. Transformation: piezoelectric crystal converts electrical energy into 3 MHz ultrasonic energy; delivered via applicators and coupling patches to generate deep heat in tissues. Output: continuous wave ultrasound (0.65 W single applicator; 1.3 W dual). Operation: patient applies applicators to skin for up to 4 hours. Clinical use: relief of pain, muscle spasms, joint contractures, increased circulation, and improved mobility in soft tissue injuries. Benefit: sustained deep heat therapy for chronic musculoskeletal conditions.
Clinical Evidence
Systematic literature review and meta-analysis using PubMed, EBSCOhost, Academic Search Complete, Google Scholar, and ClinicalTrials.gov. Evaluated Sustained Acoustic Medicine (SAM) effects on musculoskeletal functional outcomes (strength, range of motion), quality of life, pain reduction, and safety. Evidence supports efficacy for deep heat generation, pain relief, and treatment of knee osteoarthritis, tendinopathy, and chronic myofascial pain.
Indicated for home use in patients requiring deep heat therapy for pain relief, muscle spasms, joint contractures, increased local circulation, and pain associated with limited mobility/function from soft tissue injuries including knee osteoarthritis, chronic myofascial pain, and shoulder, elbow, or ankle tendinopathy.
Regulatory Classification
Identification
An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 23, 2023
ZetrOZ Systems, LLC Sabrina Lewis QAR Director 56 Quarry Road Trumbull, Connecticut 06611
Re: K223019
Trade/Device Name: sam 2.0 Long Ultrasound Device Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Regulatory Class: Class II Product Code: PFW Dated: May 17, 2023 Received: May 17, 2023
Dear Sabrina Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Amber T. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223019
Device Name sam 2.0 Long Duration Ultrasound Device
#### Indications for Use (Describe)
The sam 2.0 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, the local increase in circulation, and the relief of pain associated with limited mobility and function related to soft tissue injuries (e.g., knee osteoarthritis, chronic myofascial pain, and shoulder, elbow and ankle tendinopathy).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | sam 2.0 Long Duration Ultrasound Device |
|--------------------|---------------------------------------------------------------------------------------|
| Manufacturer: | ZetrOZ Systems, LLC<br>56 Quarry Road<br>Trumbull, CT 06611 |
| Contact: | Ms. Sabrina Lewis<br>QAR Director<br>Phone: 888-202-9831<br>Email: sabrina@zertoz.com |
| Date Prepared: | June 23, 2023 |
| 510(k) Number: | K223019 |
| Classifications: | 21 CFR §890.5300; Ultrasonic diathermy. |
| Class: | II |
| Product Code: | PFW |
| Predicate Device: | sam 2.0 Long Duration Ultrasound Device (K191568) |
#### Indications For Use:
The sam 2.0 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, the local increase in circulation, and the relief of pain associated with limited mobility and function related to soft tissue injuries (e.g., knee osteoarthritis, chronic myofascial pain, and shoulder, elbow and ankle tendinopathy).
#### Device Description:
The device was previously cleared in K191568. The sam 2.0 Long Duration Ultrasound Device consists of ultrasound:
- Power Controller ●
- Cables ●
- Applicators ●
- Coupling Patches
The sam 2.0 Long Duration Ultrasound Device contains firmware for logging usage time. There is no control function of this firmware.
The Power Controller and cables can be used to power one or two Applicators simultaneously to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per
{4}------------------------------------------------
Applicator. In single Applicator mode, the patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with a onetime use Coupling Patches.
The system is intended to apply ultrasonic energy for a long duration (4 hours) to generate deep heat within body tissues. The system is intended for prescription home use after proper instruction from a healthcare professional.
## Predicate Device:
ZetrOZ submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the sam 2.0 Long Duration Ultrasound Device is substantially equivalent in intended use, design principles, and performance to the following predicate device:
Primary Predicate: sam 2.0 Long Duration Ultrasound Device (K191568)
## Performance Testing Summary:
Clinical performance testing was utilized to support the substantial equivalence of this device. A systematic literature review was conducted using PubMed, EBSCOhost, Academic Search Complete, Google Scholar and ClinicalTrials.gov to identify studies evaluating the effects of the subject device, Sustained Acoustic Medicine (SAM), on the musculoskeletal system of humans. This systematic review and meta-analysis aim to summarize the clinical effects of SAM treatment on musculoskeletal injuries functional outcomes (strength and range of motion), quality of life, pain reduction, and safety profile of the intervention. Cumulatively, these studies demonstrate the efficacy of SAM therapy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. As evidenced by the study data summarized above, the relief of pain associated with limited mobility and function related to the soft tissue injuries that have been clinically validated include knee osteoarthritis, shoulder, elbow and ankle tendinopathy, and chronic myofascial pain.
No non-clinical performance testing was performed to support the expansion of indications.
## Substantial Equivalence:
Overall, the subject device is nearly identical in design and intended use to the primary predicate K191568. The notable difference is in that of the indications. The subject device presents an expansion of indications to better align with the clinical literature on SAM. The SAM is a stationary ultrasound device that is capable of creating (>4℃) into the tissue. The subject device and predicate K191568 are identical with respect to design, technology, and application.
| | ZetrOZ sam 2.0<br>Long Duration<br>Ultrasound Device<br>(Subject Device) | ZetrOZ sam 2.0 Long<br>Duration Ultrasound<br>Device (K191568)<br>(Primary<br>Predicate) | Comparison |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Information | | | |
| Classification Name | Ultrasonic<br>Diathermy Device | Ultrasonic<br>Diathermy Device | Identical |
| Service Type | Physical<br>Medicine | Physical<br>Medicine | Identical |
| Classification | 21 CFR<br>890.5300 | 21 CFR<br>890.5300 | Identical |
| Class | II | II | Identical |
| Product Code | PFW | PFW | Identical |
| Indications | The sam 2.0 Long Duration<br>Ultrasound Device is<br>intended for home use to<br>apply ultrasonic energy to<br>generate deep heat within<br>body tissues for the<br>treatment of selected<br>medical conditions such as<br>the relief of pain, the relief<br>of muscle spasms, the<br>treatment of joint<br>contractures, the local<br>increase in circulation, and<br>the relief of pain associated<br>with limited mobility and<br>function related to soft<br>tissue injuries (e.g., knee<br>osteoarthritis, chronic<br>myofascial pain, and<br>shoulder, elbow and ankle<br>tendinopathy). | The sam 2.0 Long Duration<br>Ultrasound Device is<br>intended for home use to<br>apply ultrasonic energy to<br>generate deep heat within<br>body tissues for the<br>treatment of selected<br>medical conditions such as<br>the relief of pain, the relief<br>of muscle spasms, the<br>treatment of joint<br>contractures, and the local<br>increase in circulation. | Similar to the Primary<br>Predicate.<br>The device is intended to<br>treat medical conditions<br>such as the relief of pain,<br>the relief of muscle spasms,<br>the treatment of joint<br>contractures, and the local<br>increase in<br>circulation. It also includes<br>medical<br>conditions relief of pain<br>associated with limited<br>mobility and function<br>related to soft tissue<br>injuries including knee<br>osteoarthritis, shoulder,<br>elbow and ankle<br>tendinopathy, and chronic<br>myofascial pain as<br>supported by clinical<br>evidence. |
| Manufacturer | ZetrOZ | ZetrOZ | Identical |
| Console/Generator<br>Dimensions (L x W x H cm) | 6.10 cm L x 7.09 cm H x<br>1.88cm W | 6.10 cm L x 7.09 cm H x<br>1.88cm W | Identical |
| Treatment Head<br>Dimensions (L x W x H cm) | 3.81 cm L x 3.30 cm W x<br>1.14 cm H | 3.81 cm L x 3.30 cm W x<br>1.14 cm H | Identical |
| Console/Generator Weight<br>(kg) | 0.10 kg | 0.10 kg | Identical |
| Treatment Head Weight<br>(kg) | 0.01 kg | 0.01 kg | Identical |
| Power Supply | 120/240 VAC with 5V<br>DC Input Power Jack and<br>LithiumBattery Powered | 120/240 VAC with 5V<br>DC Input Power Jack and<br>Lithium<br>Battery Powered | Identical |
| Leakage Current | 0.3 mA | 0.3 mA | Identical |
| Crystal Material | Lead Zirconate-<br>Titanate | Lead Zirconate-<br>Titanate | Identical |
| Technology of ultrasound<br>generation (e.g., | Piezoelectric | Piezoelectric | Identical |
| Treatment Mode(s) | Two discrete settings of<br>power at same Frequency | Two discrete settings of<br>power at same Frequency | Identical |
| Beam Type (collimated | | | |
| or<br>divergent) | Divergent | Divergent | Identical |
| Transducer Diameter<br>(cm) | 5 cm | 5 cm | Identical |
| Acoustic Working<br>Frequency and<br>Accuracy (MHz) | 3MHz ± 20% | 3MHz ± 20% | Identical |
| Effective Radiating Area | One: 6 cm² | One: 6 cm² | Identical |
{5}------------------------------------------------
{6}------------------------------------------------
| | ± 20% | ± 20% | | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------|-------------|------------------------------|-----------|-----------|
| Beam Nonuniformity<br>Ratio and Accuracy | BNR: <5:1 ±<br>20% | BNR: <5:1 ±<br>20% | Identical | | | | |
| Output Mode:<br>(Continuous<br>Wave/Amplitude -<br>Modulated Wave) | Continuous<br>Wave - 100%<br>duty cycle | Continuous<br>Wave - 100%<br>duty cycle | Identical | | | | |
| Maximum Timer Setting<br>and Accuracy | 4 Hours +/- 1 minute | 4 Hours +/- 1 minute | identical | | | | |
| Beam Maximum Intensity<br>and Accuracy (W/cm²) | 0.132 W/cm² ± 20% | 0.132 W/cm² ± 20% | Identical | | | | |
| Maximum Value of the<br>Output Power (Rated<br>Output Power) and<br>Accuracy (W) | Single Applicator:<br>0.65W ± 20%<br><br>Dual Applicator:<br>1.3W ± 20% | Single Applicator:<br>0.65W ± 20%<br><br>Dual Applicator:<br>1.3W ± 20% | Identical | | | | |
| Maximum Value of the<br>Effective Intensity and<br>Accuracy (Not to exceed 3<br>W/cm2 *) | 0.264 W/cm² ± 20% | 0.264 W/cm² ± 20% | Identical | | | | |
| For Amplitude<br>Modulated Waves | Not Amplitude Modulated | Not Amplitude Modulated | identical | | | | |
| Peak Temperature Rise<br>vs. Time and Tissue<br>Depth to Maximum<br>Treatment Time (for<br>fixed Treatment Head<br>Placement) (deg C) | 8°C at 1 cm<br>6°C at 3 cm<br>3°C at 5 cm<br>Max treatment time: 4 hours | 8°C at 1 cm<br>6°C at 3 cm<br>3°C at 5 cm<br>Max treatment time: 4 hours | Identical | | | | |
| Maximum Patient Contact<br>Surface Temperature of<br>Treatment Head under<br>Simulated or Actual Use<br>Conditions for all Operating<br>Conditions (Continually<br>operated for maximum<br>treatment time) (deg C)<br>Therapeutically Applied | 44 °C | 44 °C | Identical | | | | |
| Applicator Type | Ultrasound Coupling Patch<br>Up to two circular<br>Applicators with 3 MHz<br>output | Ultrasound Coupling Patch<br>Up to two circular<br>Applicators with 3 MHz<br>output | identical | | | | |
| Applicator Type Applicator<br>Emitting Surface Areas<br>(cm2) | Up to two circular<br>Applicators<br>One Applicator : 5 cm²<br>Two Applicators :10<br>cm² | Up to two circular<br>Applicators<br>One Applicator : 5<br>cm²<br>Two Applicators :10<br>cm² | Identical | | | | |
| Coupling Bandage | ABS Plastic with<br>integrated coupling<br>medium | ABS Plastic with<br>integrated coupling<br>medium | Identical | | | | |
| Applicator Lens Material | TPX | TPX | identical | | | | |
| Environmental - Operating<br>Temperature Range | 0°C to +50°C (32°F to<br>+122°F) | 0°C to +50°C (32°F to<br>+122°F) | Identical | | | | |
| Performance Standards | 21 CFR 1050.10 | 21 CFR 1050.10 | Identical | | | | |
| Sterility | Designed to meet Electrical<br>Safety Standards | Non Sterile | IEC 60601-1<br>IEC 60601-1-2 | Non Sterile | IEC 60601-1<br>IEC 60601-1-2 | identical | Identical |
{7}------------------------------------------------
| | IEC 60601-1-11 | IEC 60601-1-11 | |
|--------------------------------------------------------|-----------------------------------|-----------------------------------|-----------|
| Biocompatibility | Yes | Yes | Identical |
| Mechanical safety | Yes | Yes | Identical |
| Radiation safety (if<br>not radioactive state as such) | Not Radioactive | Not Radioactive | identical |
| Software/Firmware | Yes | Yes | Identical |
| Output Channels | Two Independent Power<br>Channels | Two Independent Power<br>Channels | Identical |
#### Conclusion:
The subject device and the predicate device have the same intended use, technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in similar materials. The data included in this submission demonstrate substantial equivalence to the predicate device listed above. The sam 2.0 Long Duration Ultrasound Device is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.