K223000 · Inari Medical · QEW · Jan 11, 2023 · Cardiovascular
Device Facts
Record ID
K223000
Device Name
Artix BG
Applicant
Inari Medical
Product Code
QEW · Cardiovascular
Decision Date
Jan 11, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Artix BG balloon thrombectomy sheath is indicated for: · Use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures. · Use as a conduit for retrieval devices. · The non-surgical aspiration of emboli and thrombi from blood vessels. · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Artix BG balloon thrombectomy sheath is intended for use in the peripheral vasculature.
Device Story
Artix BG is a single-use, over-the-wire balloon thrombectomy sheath; used in peripheral vasculature. Device facilitates intravascular catheter insertion/guidance; provides temporary vascular occlusion via distal compliant balloon; acts as conduit for retrieval devices; enables non-surgical aspiration of emboli/thrombi; allows infusion/aspiration of fluids/contrast media. Operated by physicians in clinical settings. Input: guidewire access. Transformation: mechanical insertion; balloon inflation via 1 mL syringe; aspiration via 30 mL syringe. Output: vascular access; thrombus removal; flow occlusion. Benefits: minimally invasive removal of clots; procedural support for angiography/interventions.
Clinical Evidence
Bench testing only. No clinical or animal data required. Verification and validation testing included sheath visual/dimensional analysis, simulated use cycling, sheath aspirational flowrate, and clot burden removal. All acceptance criteria met.
Technological Characteristics
Single-use, over-the-wire sheath with four embedded lumens (inflation/deflation), proximal hemostasis valve, stopcock, and flush port. Features hydrophilic coating on distal shaft and two radiopaque marker bands (tip and balloon location). Compliant balloon for occlusion. 8 Fr diameter, 65 cm or 105 cm lengths. Sterilized via EtO (ISO 11135:2014/Amd 1:2018, AAMI TIR 28:2016). Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for adults undergoing interventional procedures in the peripheral vasculature requiring intravascular catheter guidance, temporary vascular occlusion, conduit for retrieval devices, non-surgical aspiration of emboli/thrombi, or fluid/contrast media infusion/aspiration.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 11, 2023
Inari Medical Ellen Nguyen Regulatory Affairs Specialist 6001 Oak Canyon Suite 100 Irvine, California 92618
Re: K223000
Trade/Device Name: Artix BG Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, DQY Dated: December 9, 2022 Received: December 9, 2022
Dear Ellen Nguyen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S ○'connell -S Date: 2023.01.11
05:00 nell -S 09:19:48 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223000
Device Name Artix BG
Indications for Use (Describe)
The Artix BG balloon thrombectomy sheath is indicated for:
· Use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures.
· Use as a conduit for retrieval devices.
· The non-surgical aspiration of emboli and thrombi from blood vessels.
· Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Artix BG balloon thrombectomy sheath is intended for use in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## PAGE 1 OF 5
# 510(K) SUMMARY
| Date prepared | January 10, 2023 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>877.923.4747 |
| Contact person | Ellen Nguyen<br>Regulatory Affairs Specialist |
| Trade name | Artix BG |
| Common name | Balloon thrombectomy sheath |
| Regulation name | Percutaneous catheter |
| Classification number | 21 CFR 870.1250 |
| Product code | QEW |
| Secondary product<br>code | DQY |
| Regulatory class | II |
| Predicate device | Inari Medical, ClotTriever Thrombectomy System (K212632) |
| Reference device | Inari Medical, Artix Balloon Guiding Sheath (K221846) |
| Description | The Artix BG is a single-use, over-the-wire system designed to facilitate the<br>insertion and guidance of an intravascular catheter into a selected peripheral<br>blood vessel and act as a conduit for retrieval devices. A compliant balloon<br>mounted at the sheath's distal tip provides temporary vascular occlusion<br>during angiographic and interventional procedures. The Artix BG also non-<br>surgically aspirates thromboemboli from selected vessels and is capable of<br>infusion/aspiration of fluids into or from a selected vessel. The Artix BG is<br>packaged with the following components:<br>• Artix BG balloon thrombectomy sheath (8 Fr, 65 cm or 105 cm)<br>• 8 Fr Introducer Dilator (0.014" and 0.035" guidewire compatibility)<br>• Balloon Inflation Syringe, 1 mL<br>• Large Bore Syringe, 30 mL<br>• 3-way Stopcock |
| Indications for Use | The Artix BG balloon thrombectomy sheath is indicated for:<br>• Use in facilitating the insertion and guidance of an intravascular<br>catheter into a selected blood vessel. The balloon provides temporary<br>vascular occlusion during these and other angiographic procedures.<br>• Use as a conduit for retrieval devices.<br>• The non-surgical aspiration of emboli and thrombi from blood<br>vessels. |
{4}------------------------------------------------
#### PAGE 2 OF 5
- Injection, infusion, and/or aspiration of contrast media and other ● fluids into or from a blood vessel.
The Artix BG balloon thrombectomy sheath is intended for use in the peripheral vasculature.
The predicate ClotTriever Thrombectomy System is indicated for "The nonsurgical removal of emboli and thrombi from blood vessels and injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel." The Artix BG's expanded indications for use are identical. Both devices are intended for use in the peripheral vasculature, with the only difference being that the predicate device is also intended for the treatment of deep vein thrombosis (DVT).
A tabular comparison of specific technological characteristics between the predicate and subject devices is provided below:
| Feature | Artix BG - Subject Device | ClotTriever<br>Thrombectomy System -<br>Predicate (K212632) | Artix Balloon Guiding<br>Sheath - Reference<br>Device (K221846) |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Inari Medical | Inari Medical | Inari Medical |
| Product code | QEW, DQY | QEW | DQY |
| Intended<br>use/Indications for<br>use | The Artix BG balloon<br>thrombectomy sheath is<br>indicated for:<br>Use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures. Use as a conduit for retrieval devices. The non-surgical aspiration of thrombi and emboli from blood vessels. Infusion, injection, and/or aspiration of contrast media or other fluid into or from a blood vessel. The Artix BG balloon thrombectomy sheath is intended for use in the peripheral vasculature. | The ClotTriever<br>Thrombectomy System is<br>indicated for:<br>The non-surgical removal of thrombi and emboli from blood vessels. Infusion, injection, and/or aspiration of contrast media or other fluid into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT). | The Artix Balloon<br>Guiding Sheath is<br>indicated for use in<br>facilitating the insertion<br>and guidance of an<br>intravascular catheter<br>into a selected blood<br>vessel. The balloon<br>provides temporary<br>vascular occlusion<br>during these and other<br>angiographic procedures.<br>The Artix Balloon<br>Guiding Sheath is also<br>indicated for use as a<br>conduit for retrieval<br>devices.<br>The Artix Balloon<br>Guiding Sheath is<br>intended for use in the<br>peripheral vasculature. |
| Feature | Artix BG - Subject Device | ClotTriever<br>Thrombectomy System -<br>Predicate (K212632) | Artix Balloon Guiding<br>Sheath - Reference<br>Device (K221846) |
| Device description | The Artix BG is comprised<br>of a single through lumen<br>shaft with four (4)<br>embedded lumens for<br>balloon inflation and<br>deflation, a proximal<br>integrated hemostasis valve<br>to minimize blood loss, a<br>stopcock with flush port,<br>and a balloon port. To assist<br>with insertion of the sheath<br>into the vasculature, a<br>hydrophilic coating covers a<br>portion of the distal sheath<br>shaft. The distal tip of the<br>sheath shaft contains two<br>(2) radiopaque marker<br>bands to aid with<br>angiographic visualization:<br>The distal marker indicates<br>the location of the sheath's<br>tip, while the proximal<br>marker band marks the<br>location of a compliant<br>balloon mounted at the<br>distal end. The balloon<br>provides temporary vascular<br>occlusion during<br>angiographic and<br>interventional procedures. | The ClotTriever<br>Thrombectomy System<br>consists of ClotTriever 13<br>Fr and 16 Fr Sheaths<br>("Sheath") and the<br>ClotTriever/ClotTriever<br>Bold Catheter<br>("Catheter"), each<br>packaged separately.<br>The sheath is an introducer<br>sheath with a distal self-<br>expanding funnel,<br>aspiration port, and<br>proximal hub. A dilator is<br>provided to aid insertion<br>and positioning of the<br>sheath. Other provided<br>accessories include a 60 cc<br>large bore syringe that<br>provides a vacuum source<br>and collects aspirated<br>contents. Radiopaque<br>markers aid sheath<br>positioning under<br>fluoroscopic visualization.<br>The dilator tip is<br>radiopaque, and there is a<br>radiopaque marker band<br>near the distal end of the<br>sheath. | The Artix Balloon<br>Guiding Sheath is<br>comprised of a single<br>through lumen shaft with<br>four (4) embedded<br>lumens for balloon<br>inflation and deflation, a<br>proximal integrated<br>hemostasis valve to<br>minimize blood loss, a<br>stopcock with flush port,<br>and a balloon port. To<br>assist with insertion of<br>the sheath into the<br>vasculature, a<br>hydrophilic coating<br>covers a portion of the<br>distal sheath shaft. The<br>distal tip of the sheath<br>shaft contains two (2)<br>radiopaque marker bands<br>to aid with angiographic<br>visualization: The distal<br>marker indicates the<br>location of the sheath's<br>tip, while the proximal<br>marker band marks the<br>location of a compliant<br>balloon mounted at the<br>distal end. The balloon<br>provides temporary<br>vascular occlusion<br>during angiographic and<br>interventional<br>procedures. |
| Principles of<br>operation | After the target vessel is<br>accessed and dilated, the<br>Artix BG is inserted into the<br>vessel over a guidewire.<br>Once the Artix BG is in<br>position, the dilator is<br>removed, and all compatible<br>catheter devices, such as the<br>Artix MT (K220600), can<br>be inserted through the<br>Artix BG for access into the<br>peripheral vasculature. The<br>sheath's balloon can also be<br>inflated using a 1 mL<br>syringe to provide flow | After the target vessel is<br>accessed and dilated, the<br>ClotTriever Sheath is<br>inserted into the vessel<br>over a guidewire. The<br>ClotTriever/ClotTriever<br>Bold Catheter can then be<br>advanced through the<br>sheath and beyond the clot.<br>The self-expanding braided<br>nitinol wire net is<br>deployed. The expanded<br>net cores, separates, and<br>entraps thrombus from the<br>vessel as it is being drawn | After the target vessel is<br>accessed and dilated, the<br>Artix Balloon Guiding<br>Sheath is inserted into<br>the vessel over a<br>guidewire. Once the<br>Artix Sheath is in<br>position, the dilator is<br>removed, and all<br>compatible catheter<br>devices can be inserted<br>through the Artix Sheath<br>for access into the<br>peripheral vasculature.<br>The sheath's balloon can |
| | | | |
| Feature | Artix BG - Subject Device | ClotTriever<br>Thrombectomy System -<br>Predicate (K212632) | Artix Balloon Guiding<br>Sheath - Reference<br>Device (K221846) |
| | occlusion of the vessel<br>during the procedure. The<br>provided 30 mL syringe can<br>be used to aspirate clot in<br>the vessel or sheath and<br>infuse contrast media and<br>other fluids as required. | to the funnel opening of<br>the ClotTriever Sheath.<br>The net is collapsed and<br>pulled into and through the<br>ClotTriever Sheath with<br>the entrapped clot. A 60 cc<br>syringe is provided for the<br>aspiration of clot in the<br>sheath and the infusion of<br>contrast media and other<br>fluids. | also be inflated using a 1<br>mL syringe to provide<br>flow occlusion of the<br>vessel during the<br>procedure. |
| Intended patient<br>population | Adults undergoing<br>interventional procedures | Adults undergoing<br>interventional procedures | Adults undergoing<br>interventional procedures |
| Target vessel | Peripheral vasculature | Peripheral vasculature | Peripheral vasculature |
| Placement duration | < 24 hours | < 24 hours | < 24 hours |
| Guidewire<br>compatibility | 0.014" or 0.035" | 0.035" | 0.014" or 0.035" |
| Sterilization | EtO | EtO | EtO |
| Single-use | Yes | Yes | Yes |
Summary of substantial equivalence
{5}------------------------------------------------
{6}------------------------------------------------
Summary of substantial equivalence
### Biocompatibility
This submission proposes no changes to the device materials. Therefore, the previous passing results demonstrating that the Artix BG (K221846) and accessories meet biological safety requirements per ISO 10993-1 are still applicable.
### Sterilization
The subject device, including its accessories, is sterilized using EtO to achieve a sterility assurance level (SAL) of 10 6 using a validated sterilization process in accordance with the principles of ISO 11135:2014/Amd 1:2018 (Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices - Amendment 1: Revision of Annex E, Single batch release) and AAMI TIR 28:2016 (Product adoption and process equivalence for ethylene oxide sterilization). This submission proposes no changes to the device design or materials. Therefore, the previous sterilization process per K221846 remains applicable.
### Non-Clinical Testing
In accordance with the Design Failure Modes and Effects Analysis, verification and validation testing were identified to support the substantial
{7}------------------------------------------------
### PAGE 5 OF 5
equivalence of the Artix BG to the ClotTriever Thrombectomy System. These tests included:
- Sheath Visual & Dimensional
- . Simulated Use Cycling
- Sheath Aspirational Flowrate ●
- Clot Burden Removal ●
Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.
Neither animal testing nor clinical testing were required for the determination of substantial equivalence.
### Conclusion
The nonclinical tests demonstrate the Artix BG's substantial equivalence to the ClotTriever Thrombectomy System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.