K222999 · Hubei Woozon Healthcare Co., Ltd. · FYA · Jun 16, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222999
Device Name
Surgical Gown
Applicant
Hubei Woozon Healthcare Co., Ltd.
Product Code
FYA · General, Plastic Surgery
Decision Date
Jun 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided non-sterile. Before the use, the Surgical Gowns must be sterilized by EO sterilization based on ISO 11135:2014.
Device Story
Surgical gown; single-use, disposable, non-sterile apparel. Composed of SMS nonwoven fabric and polyester fiber; blue color; available in sizes S-XXL. Features laced neck/waist, polyester elastic sleeve closures, and heat-sealed seams. Used by operating room personnel during surgical procedures to provide AAMI Level 3 barrier protection. Protects patient and staff from transfer of microorganisms, body fluids, and particulates. Requires EO sterilization by end-user prior to use.
Clinical Evidence
No clinical data. Bench testing only. Device conforms to AAMI PB70:2012, ASTM F2407-20, and ISO 10993 standards. Testing included impact penetration (AATCC 42), hydrostatic resistance (AATCC 127), tensile/tear strength (ASTM D5034/D5587), seam strength (ASTM D1683), flammability (16 CFR 1610), and biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity).
Technological Characteristics
SMS nonwoven fabric and polyester fiber construction. AAMI Level 3 barrier protection. EO sterilization (ISO 11135). Mechanical properties tested per ASTM standards (D5034, D5587, D1683). Biocompatibility per ISO 10993. Flammability per 16 CFR 1610.
Indications for Use
Indicated for operating room personnel during surgical procedures to provide AAMI Level 3 barrier protection against microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 16, 2023
Hubei Woozon Healthcare Co.,Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K222999
Trade/Device Name: Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: May 10, 2023 Received: May 22, 2023
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bifeng Qian -S
Bifeng Qian, MD, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222999
Device Name Surgical Gown
#### Indications for Use (Describe)
Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requrements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided non-sterile. Before the use, the Surgical Gowns must be sterilized by EO sterilization based on ISO 11135:2014.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K222999
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
#### 1.0 Submitter's information
Name: Hubei Woozon Healthcare Co.,Ltd. Address: Nongfeng Road, Nonwoven Fabric Industrial Park, Wangzhou Avenue, Pengchang Town, Xiantao City, Hubei, China. Contact: Li Chunlin Phone Number: 86-728-3280667 Fax number: 86-728-3280667 Date of Preparation: June 12, 2023
#### Designated Submission Correspondent
Contact: Mr. Boyle Wang Address: Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device information
| Trade name: | Surgical Gown |
|----------------------|------------------|
| Common name: | Surgical gown |
| Classification name: | Gown, Surgical |
| Model(s): | S, M, L, XL, XXL |
#### 3.0 Classification
| Production code: | FYA - Gown, Surgical |
|--------------------|-----------------------------------|
| Regulation number: | 21CFR 878.4040 - Surgical apparel |
| Classification: | Class II |
| Panel: | Surgical apparel |
#### 4.0 Predicate device information
Manufacturer: Hubei Wanli Protective Products Co. Ltd
{4}------------------------------------------------
Surgical Gown Device: K211509 510(k) number:
#### 5.0 Intended Use/Indication for Use Statement
Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided non-sterile. Before the use, the Surgical Gowns must be sterilized by EO sterilization based on ISO 11135:2014.
#### 6.0 Device description
The Surgical Gown is composed of collar, body, sleeve and tie, in a blue color, offered in sizes S, M, L, XL and XXL. The neck and waist are laced, the sleeves are made of polyester elastic closure by sewing, and the rest are made of heat sealing. It has been tested according to AAMI PB70:2012 and meet AAMI Level 3 barrier level protection for a surgical gown.
| Table 3 - General Comparison | | | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Item | Proposed device<br>(K222999) | Predicated device<br>(K211509) | Remark |
| Product Code | FYA | FYA | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Product name | Surgical Gown | Surgical Gown | Same |
| 510(k) No. | K222999 | K211509 | / |
| Models | S,M,L,XL,XXL | S,M,L,XL,XXL | Same |
| Intended<br>Use/Indications for<br>Use | Surgical Gowns are intended to be worn by operating room personnel<br>during surgical procedures to protect the surgical patient and<br>operating room personnel from the transfer of microorganisms, body<br>fluids and particulate material. This surgical gown meets the<br>requirements of AAMI Level 3 barrier protection for a surgical<br>gown per ANSI/AAMI PB70:2012 Liquid barrier performance and<br>classification of protective apparel and drapes intended for use in<br>health care facilities (AAMI PB70). The Surgical Gowns are single | | Same |
| | | Surgical Gowns are intended to be worn by operating room personnel<br>during surgical procedures to protect the surgical patient and operating<br>room personnel from the transfer of microorganisms, body fluids and<br>particulate material. This surgical gown meets the requirements of<br>AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI<br>PB70:2012 Liquid barrier performance and classification of<br>protective apparel and drapes intended for use in health care<br>facilities (AAMI PB70). The Surgical Gowns are single use, disposable | |
| | use, disposable medical devices;<br>provided non-sterile. Before the<br>use, the Surgical Gowns must be<br>sterilized by EO sterilization based<br>on ISO 11135:2014. | medical devices; provided non-<br>sterile. | |
| Material | SMS nonwoven fabric<br>polyester fiber | Sleeve/body<br>(polyethylene SMS Nonwoven) | Similar |
| Color | Blue | Blue | Same |
| Sterility | Sterile for end-user | Sterile for end user | Same |
| Single Use | Yes | Yes | Same |
| Impact Penetration | ≤1.0 g | ≤1.0 g | Same |
| Hydrostatic<br>Resistance | >50 cm | >50 cm | Same |
| Tensile strength | Machine direction mean≥32.9 lbf;<br>Cross direction mean≥26.3 lbf. | Machine direction mean≥27.8 lbf;<br>Cross direction mean≥19.1 lbf | * Gap 1 |
| Tear resistance | Machine direction mean≥9.3 lbf;<br>Cross direction mean≥18.2 lbf | Machine direction mean≥10.7 lbf;<br>Cross direction mean≥6.2 lbf | * Gap 2 |
| Steam strength | Shoulder: 48.7N<br>Armhole:33.9N<br>Sleeve: 49.6N | Shoulder: 71.8N<br>Armhole: 66.0N<br>Sleeve: 77.0N | * Gap 3 |
| 16 CFR Part 1610<br>Flammability | Meets requirements of Flame<br>Resistant CPSC 1610 Class 1 | Meets requirements of Flame<br>Resistant CPSC 1610 Class 1 | Same |
| Linting | Log10<4 | Log10<4 | Same |
| Liquid barrier<br>performance | Level 3 AAMI PB70 | Level 3 AAMI PB70 | Same |
| Cytotoxicity | Under the condition of the test, no<br>potential cytotoxicity | Under the condition of the test, no<br>potential cytotoxicity | Same |
| Irritation<br>Sensitization | Under the condition of the test, no<br>irritation and sensitization | Under the condition of the test, no<br>irritation and sensitization | Same |
| Acute systemic<br>toxicity | Under the condition of acute<br>systemic toxicity test, the test<br>article did not show acute systemic<br>toxicity in vivo. | / | / |
| Sterilization<br>Modality | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Same |
#### tah
7.0Technological Characteristic Comparison Table
{5}------------------------------------------------
* Gap analysis:
Gap 1-3, the two devices have some little deviation in product performance, but the difference in the performance test result does not raise additional questions for safety and effectiveness.
#### 8.0 Non-Clinical Test Conclusion
The proposed device was tested and conformed to the related recognized standards and the requirements stated in the "Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes" dated August, 1993.
Based on ASTM F2407-20 Standard Specification for Surgical Gowns Intended for Use in Healthcare Facilities, the subject device has been conducted the test as following:
AATCC 42-2013 Water Resistance: Impact Penetration Test
{6}------------------------------------------------
AATCC 127-2014 Water Resistance: Hydrostatic Pressure Test
ASTM D5034-09(2017) Standard Test Method for Breaking Strength and Elongation of Texile Fabrics (Grab Test)
ASTM D5587-15(2019) Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure
ASTM D1683-17(2018) Standard Test Method for Failure in Sewn Seams of Woven Apparel Fabrics
ASTM F1868-17(2017) Standard Test Method for Thermal and Evaporative Resistance of Clothing Materials Using a Sweating Hot Plate
ISO 9073-10:2003 Textiles - Test methods for nonwovens - Part 10: Lint and other particles generation in the dry state
16 CFR Part 1610(a) Standard for The Flammability of Clothing Textiles
ISO 11737-2:2019 Sterilization of health care products - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
| Test<br>Methodology | Purpose | Acceptance Criteria | Results |
|----------------------------|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| AATCC 42 | Impact Penetration | Level 3, ≤1.0g | <1.0g |
| AATCC 127 | Hydrostatic<br>Resistance | Level 3, ≥50cm | ≥50cm |
| ASTM D5034 | Tensile strength | Machine direction mean:<br>≥18 lbf;<br>Cross direction mean: ≥12<br>lbf. | Machine direction mean≥32.9 lbf;<br>Cross direction mean≥26.3 lbf. |
| ASTM D5587 | Tear resistance | Farbic direction A mean:<br>≥4.0 lbf;<br>Farbic direction B mean:<br>≥2.5 lbf; | Machine direction mean≥9.3 lbf;<br>Cross direction mean≥18.2 lbf. |
| ASTM D1683 | Seam strength | Shoulder: ≥8.5 lbf;<br>Arm opening: ≥5.5 lbf;<br>Sleeve: ≥8.5 lbf; | Shoulder: ≥10.1 lbf;<br>Arm opening: ≥7.2 lbf;<br>Sleeve: ≥10.1 lbf; |
| ASTM F1868 | Evaporative<br>Resistance of<br>fabrics | Mean Evaporative<br>Resistance<br>(Ref)≥0.06(kPa·m²/W). | Mean Evaporative<br>Resistance<br>(Ref)≥0.06(kPa·m²/W). |
| ISO 9073-10 | Lint and Other<br>particles<br>generation in the<br>dry state | Log 10 < 4 | Index for Particulate Matter<br>(IPM)< 2.89 |
| 16 CFR Pat<br>1610 | Flammability<br>testing | Class 1, Non Flammable | Class 1, Non Flammable |
| Cytotoxicity | Biocompatibility-<br>cytotoxicity | Under the condition of the<br>test, no potential<br>cytotoxicity | Under the condition of the test,<br>no potential cytotoxicity |
| Irritation | Biocompatibility-<br>irritation | Under the condition of the<br>test, no irritation | Under the condition of the test,<br>no irritation |
| Sensitization | Biocompatibility-<br>sensitization | Under the condition of the<br>test, no sensitization | Under the condition of the test,<br>no sensitization |
| Acute systemic<br>toxicity | Biocompatibility-<br>Acute systemic<br>toxicity | Under the condition of acute<br>systemic toxicity test, the<br>test article did not show<br>acute systemic | Under the condition of acute<br>systemic toxicity test, the test<br>article did not show acute<br>systemic |
{7}------------------------------------------------
| | | toxicity in vivo. | toxicity in vivo. |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------------------------|
| Ethylene oxide<br>residues | The test was<br>performed in<br>accordance with ISO<br>10993-7:2008<br>Ethylene oxide<br>sterilization residuals<br>to determine the EO<br>and ECH residuals<br>of the test sample. | EO ≤ 4mg/device<br>ECH ≤ 9mg/device | EO: 3.68 mg/device<br>ECH: 0.94 mg/device |
#### 9.0 Clinical Test Conclusion
No clinical study implemented for the Surgical Gown.
#### 10.0 Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Surgical Gown is as safe, as effective, and performs as well as or better than the legally marketed predicate device K211509.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.