Arm Blood Pressure Monitor, models AOJ-33A and AOJ-33B
Applicant
Shenzhen AOJ Medical Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jul 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
Device Story
Battery-powered, automatic non-invasive blood pressure monitor; uses oscillometric technique to measure systolic/diastolic pressure and pulse rate via upper arm cuff. Device performs automated inflation/deflation; calculates measurements; displays results (mmHg/Kpa) and WHO BP classification. Includes data storage for review of measurements and low battery indication. Used in medical facilities or home settings by adults. AOJ-33B model includes Bluetooth for data transfer to external applications; AOJ-33A lacks wireless. Healthcare providers or patients review stored data to monitor BP trends; aids in hypertension management and cardiovascular health assessment.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018 using the Same Arm Sequential Method. Study included 100 adult subjects (49 female, 51 male, ages 12-60). Results demonstrated mean error and standard deviation of differences for systolic and diastolic pressure within ISO 81060-2:2018 limits. No adverse effects reported.
Technological Characteristics
Oscillometric measurement principle. Powered by 3.7V Li-ion battery. Cuff size 22-42 cm. Connectivity: Bluetooth (AOJ-33B only). Biocompatibility per ISO 10993-1, -5, -10. Electrical safety per IEC 60601-1, IEC 60601-1-11; EMC per IEC 60601-1-2. Performance per ISO 80601-2-30.
Indications for Use
Indicated for adult patients for non-invasive measurement of systolic/diastolic blood pressure and pulse rate via oscillometric technique. Not established for use in pregnant or pre-eclamptic patients.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
July 26, 2023
Shenzhen AOJ Medical Technology Co., Ltd. Jack Wang Deputy Chief Room 301 & 4F, Blk A, Building A, Jingfa IM Park, Xiaweiyuan Gushu Community, Xixiang, Baoan Shenzhen, Guangdong 518126 China
Re: K222994
Trade/Device Name: Arm Blood Pressure Monitor, models AOJ-33A and AOJ-33B Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 12, 2023 Received: June 13, 2023
Dear Jack Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Robert T. Kazmierski - S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222994
Device Name
Arm Blood Pressure Monitor, models AOJ-33A, AOJ-33B
Indications for Use (Describe)
The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Shenzhen AOJ Medical Technology Co., Ltd.<br>Room 301&4F, Block A, Building A,<br>Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang<br>Street, Bao'an District, 518126, Shenzhen, China<br>TEL: 86 755-27786026 | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Contact Person: | Jack Wang | | | |
| Prepare date: | September 16, 2022 | | | |
| 2. Device name and<br>classification: | Device Name: Arm Blood Pressure Monitor<br>Models: AOJ-33A and AOJ-33B<br>Regulation No .: 21 CFR 870.1130<br>Review Panel: Cardiovascular<br>Classification Name: Cardiovascular Diagnostic Devices<br>Product code: DXN<br>Regulatory Class: Class II | | | |
| 3. Reason for<br>Submission: | New Application. No prior submission for this device before. | | | |
| 3. Class III device<br>statement | Not applicable, the subject device is a Class II device. | | | |
| 4. Predicate Devices: | Reference predicate: Shenzhen AOJ Medical Technology Co., Ltd., AOJ-30B<br>Electronic Blood Pressure Monitor cleared under K191180. | | | |
| 5. Device Description: | The predicates have not been subject to any recall before.<br>AOJ-33series arm blood pressure monitor is designed as a battery driven automatic<br>non-invasive blood pressure monitor. It can automatically complete the inflation,<br>deflation and measurement, which can measure systolic and diastolic blood<br>pressure as well as the pulse rate of adult person at upper arm within its claimed<br>range and accuracy via the oscillometric technique. The result will be displayed in<br>the international unit mmHg or Kpa. | | | |
| | The device has the data storage function in order for data reviewing, including the<br>systolic pressure, diastolic pressure, pulse rate and measurement time. The device<br>also has low voltage indication, which will be triggered when the battery is low. | | | |
| | The proposed device is intended to be used in medical facilities or at home. And the<br>effectiveness of this sphygmomanometer has not been established in pregnant<br>(including pre-eclamptic) patients. | | | |
| | The product is provided non-sterile, and not to be sterilized by the user prior to use. | | | |
| | All the models included in this submission follow the same software, same<br>measurement principle and same specifications. The main differences are color of<br>the face shell and keys, which will not affect the safety and effectiveness of the<br>device. | | | |
{4}------------------------------------------------
#### K222994
AOJ-33B include the Bluetooth transmission functionality which can transfer data for Application in the external instruments, and the measuring data, including systolic diastolic pressures and pulse rate can be displayed, stored and reviewed by the Application in the external instruments without any control feature, therefore, no interoperability happened. This function is not available for AOJ-33A.
6. Indications for The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive Use: oscillometric technique at medical facilities or at home.
## 7. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate devices.
Please refer to following table to find differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
| ITEM | Proposed Device<br>AOJ-33 series | Predicate Device<br>AOJ-30B/K191180 | Comparison<br>Result |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer | Shenzhen AOJ Medical Technology<br>Co., Ltd. | Shenzhen AOJ Medical Technology<br>Co., Ltd. | Same |
| Intended Use/Indications<br>for Use | The Arm Blood Pressure Monitor is<br>intended to measure the systolic and<br>diastolic blood pressure as well as the<br>pulse rate of adult person via<br>non-invasive oscillometric technique<br>in which an inflatable CUFF is<br>wrapped around the upper arm at<br>medical facilities or at home. | The Electronic Blood Pressure<br>Monitor is intended to measure the<br>systolic and diastolic blood pressure<br>as well as the pulse rate of adult<br>person via non-invasive oscillometric<br>technique in which an inflatable<br>CUFF is wrapped around the upper<br>arm at medical facilities or at home. | Same |
| Contraindications | Not Known | Not Known | Same |
| Application scenario | Medical Facilities and Home Use | Medical Facilities and Home Use | Same |
| Operational Specifications | | | |
| Principle | Oscillometric | Oscillometric | Same |
| Measurement Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same |
| Patient population | Adult | Adult | Same |
| Measurement site | Upper arm | Upper arm | Same |
| Blood pressure<br>measurement range | 30-255 mmHg | 30-255 mmHg | Same |
| Accuracy | $\pm$ 3 mmHg | $\pm$ 3 mmHg | Same |
| Heart rate measurement<br>range | 40-199 bpm | 40-199 bpm | Same |
| Accuracy | $\pm$ 5% of reading | $\pm$ 5% of reading | Same |
| Cuff size | 22 - 42 cm | 22 - 36 cm | Different |
| Display | Blood Pressure (Systolic and<br>Diastolic), Pulse rate, Date, Time,<br>WHO BP Classification | Blood Pressure (Systolic and<br>Diastolic), Pulse rate, Date, Time,<br>WHO BP Classification | Same |
| | | | |
| | Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | Indicating Bar, Low Battery Icon, Heart Icon, Memory Record Number | |
| Auto shutdown | YES | YES | Same |
| Operating environment | Temperature: 5°-40°C Humidity:<br>15%-90% RH, Atmospheric pressure: 70 kPa -106 kPa | Temperature: 5°-40°C Humidity:<br>15%-90% RH, Atmospheric pressure:<br>70 kPa -106 kPa | Same |
| Storage environment | Ambient Temperature: -20°C to 55°C<br>Relative Humidity: 10-93% RH,<br>Atmospheric pressure:<br>70 kPa -106 kPa | Ambient Temperature: -20°C to 55°C<br>Relative Humidity: 10-93% RH,<br>Atmospheric pressure:<br>70 kPa -106 kPa | |
| Battery type | Lithium-ion battery, D.C. 3.7V | 6Vdc (4 * 1.5V AAA batteries) | Different |
| Weight | AOJ-33A: About 225g<br>AOJ-33B: About 225g | Approx. 483.8 g without battery | Different |
| Dimensions | AOJ-33A:123 mm (Length) x 59 mm<br>(Width) x 28mm (Height) (48.43<br>inches x 23.23 inches x11.02<br>inches )<br>AOJ-33B: 125 mm (Length) x 48<br>mm (Width) x 28 mm (Height)<br>(49.21 inches x 18.9 inches x 11.02<br>inches) | 138 mm * 120 mm * 59 mm | |
| Patient Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same |
| Biocompatibility<br>evaluation | Cytotoxicity, skin sensitization and<br>irritation | Cytotoxicity, skin sensitization and<br>irritation | Same |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Data transmission | Has wireless function with Bluetooth<br>and available for AOJ-33B only | Not available | Different |
Table 1 Comparison between predicate AOJ-30B and the subject device
{5}------------------------------------------------
As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. The differences between the subject and predicate devices will not raise different questions of safety or effectiveness.
## 8. Performance Testing:
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
## Non-Clinical Data:
The following performance data were provided in support of the substantial equivalence determination.
{6}------------------------------------------------
## Biocompatibility testing
The biocompatibility evaluation for the device were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the device complies with the IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety, IEC 60601-1-11 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements far basic safety and essential performance -Collateral Standard: Requirements for medical equipment and medical electrical systems used in the home healthcare environment^ and the IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC.
## Bench Testing
Bench testing was conducted on the device, consisting of all the accessories in the system. The system complies with the ISO 80601-2-30 Medical electrical equipment — Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers for performance effectiveness.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software fbr this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
## Clinical data:
AOI-33A was tested to ISO 81060-2: 2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. the clinical validation data on AOJ-33A can cover AOJ-33B. The Same Arm Sequential Method was chosen and performed on AOJ-33A. This study included 100 adult subjects (49 female, 51 male) with an age range of 12 to 60 years. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018. No adverse effect and/or complication is found in this study.
## 9. Conclusion:
Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the AOJ-33 series Arm Blood Pressure Monitor is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.