Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Screws, Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Screws
K222967 · Avanti Orthopaedics, Inc. · HWC · Nov 23, 2022 · Orthopedic
Device Facts
Record ID
K222967
Device Name
Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Screws, Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Screws
Applicant
Avanti Orthopaedics, Inc.
Product Code
HWC · Orthopedic
Decision Date
Nov 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Compression Screws: The Avanti Orthopedics 2.2 mm and 2.7mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Compression Screws: The Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones and small bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.
Device Story
Metallic cannulated bone screws for internal fixation of fractures, fusions, osteotomies, non-unions, and malunions. Screws are cannulated to allow placement over guidewires for precise positioning in bone. Used by surgeons in clinical settings. Provides mechanical stabilization of bone fragments to facilitate healing. Manufactured from stainless steel via subtractive techniques. Available in multiple diameters (2.2mm, 2.7mm, 3.2mm, 4.0mm) and lengths (14-60mm).
Clinical Evidence
Bench testing only. Performance testing conducted per ASTM F543-17 and FDA guidance 'Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway'. All predetermined acceptance criteria were met.
Technological Characteristics
Stainless steel (ASTM F138); cannulated design for guidewire placement; self-tapping; subtractive manufacturing; headless form factor; end-user steam sterilized (132°C, 4 min exposure).
Indications for Use
Indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones and small bone fragments in adults, children (2-12 years), and adolescents (12-21 years) where growth plates are fused or will not be crossed by fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
November 23, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Avanti Orthopaedics, Inc. % Linda Braddon, CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock. Georgia 30188
Re: K222967
Trade/Device Name: Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Screws, Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: September 26, 2022 Received: September 27, 2022
Dear Linda Braddon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Shumava Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K222967
Device Name
Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Compression Screws
Indications for Use (Describe)
The Avanti Orthopaedics 2.2mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions of small bones in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K222967
Device Name
Avanti Orthopaedics 3.2mm, and 4.0mm Cannulated Compression Screws
Indications for Use (Describe)
The Avanti Orthopaedics 3.2mm, and 4.0mm Cannulated Compression Screws are intended for fixatures, fusions, osteotomies, non-unions of small bones and small bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------|
| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY:
Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Screws Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Screws
| Date Prepared | November 18, 2022 | | |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|--|
| Sponsor | Avanti Orthopaedics, Inc<br>1814 Gilpin Avenue<br>Wilmington, DE 19806<br>Phone 302-530-6369<br>Fax 302-351-4896 | | |
| 510(k) Contact | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681<br>Regulatory@SecureBME.com | | |
| Trade Name | Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Screws<br>Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Screws | | |
| Common Name | Prosthetic Device | | |
| Code –<br>Classification | Product Code: HWC<br>Regulation: 21 CFR 888.3040<br>Device Class: II | | |
| | Subject Device | Primary Predicate | |
| Primary Predicate | Avanti Orthopaedics 2.2mm and<br>2.7mm Cannulated Compression<br>Screws | K161616 DePuy Synthes 2.4 mm<br>Cannulated Screws | |
| | Avanti Orthopaedics 3.2mm and<br>4.0mm Cannulated Compression<br>Screws | K161616 DePuy Synthes 3.5 mm and<br>4.0 mm Cannulated Screws | |
| | Subject Device | Additional Predicate(s) | |
| Additional Predicate | Avanti Orthopaedics 2.2mm and<br>2.7mm Cannulated Compression<br>Screws | K180541 DePuy Synthes 2.0 mm<br>Quick Insertion Screws | |
| | Avanti Orthopaedics 3.2mm and<br>4.0mm Cannulated Compression<br>Screws | K162825 CAPTIVATE™<br>Compression Screws | |
| | Subject Device | Reference Device(s) | |
| Reference Device | Avanti Orthopaedics 2.2mm and<br>2.7mm Cannulated Compression<br>Screws | K191118 Avanti Distal Radius and<br>Forearm System | |
| | Avanti Orthopaedics 3.2mm and<br>4.0mm Cannulated Compression<br>Screws | K191118 Avanti Distal Radius and<br>Forearm System | |
{5}------------------------------------------------
| Device Description | The Avanti Orthopaedics Cannulated Compression Screw System are metallic bone screws manufactured from Stainless Steel (ASTM F138). The screws are available in multiple lengths and diameters, and are intended to be used as stand-alone bone screws for internal bone fixation of fractures, fusions, osteotomies, non-unions, and malunions in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by the screw fixation. The Avanti Orthopaedics Cannulated screws are cannulated for use with guidewires for precise placement in bone. |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use Statement | Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Compression Screws:<br>The Avanti Orthopedics 2.2 mm and 2.7mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.<br><br>Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Compression Screws:<br>The Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bones and small bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation. |
{6}------------------------------------------------
# Comparison of Technological Characteristics: Avanti Orthopaedics 2.2mm and 2.7mm Cannulated Compression Screws
| Characteristic | Subject Device<br>Avanti Orthopaedics 2.2mm and<br>2.7mm Cannulated Compression Screws | Primary Predicate<br>Synthes USA Products, LLC DePuy<br>Synthes 2.4 mm Cannulated Screws<br>K161616 |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Classification | Class II | Class II |
| Product Code | HWC | HWC |
| Purpose | Subject device | Same indications |
| Diameter (mm) | 2.2 – 2.7mm | 2.4mm |
| Lengths (mm) | 14 – 60mm | 10 – 50mm |
| Partially / Fully<br>Threaded | Partially Threaded<br>Fully Threaded | Partially Threaded<br>Fully Threaded |
| Head Type | Headless | Low-Profile |
| Driver Shape | Star | Star |
| Cannulated | Yes | Yes |
| Self-Tapping | Yes | Yes |
| Sterility | End-User Sterilized, Steam<br>Sterilization<br>Cycle: Pre-Vacuum<br>Temperature: 132°C (270°F)<br>Exposure Time: 4 minutes<br>Drying Time: 60 minutes | End-User Sterilized, Steam<br>Sterilization |
| Material | Stainless Steel per ASTM F138 | Stainless Steel per ASTM F138 |
| Manufacturing | Subtractive manufacturing | Subtractive manufacturing |
| Biocompatibility | Implant: Permanent Duration -<br>Implant for Tissue / Bone (>30 days)<br>Instrument: Limited Duration -<br>Implant for Tissue / Bone (< 24 hours) | Implant: Permanent Duration -<br>Implant for Tissue / Bone (>30 days)<br>Instrument: Limited Duration -<br>Implant for Tissue / Bone (< 24 hours) |
| Indicated Patient<br>Population | Adults and children (2-12 years) and<br>adolescents (12-21 years) in which<br>growth plates have fused or in which<br>growth plates will not be crossed by<br>screw fixation | Adults and children (2-12 years) and<br>adolescents (12-21 years) in which<br>growth plates have fused or in which<br>growth plates will not be crossed by<br>screw fixation |
| Characteristic | Subject Device<br>Avanti Orthopaedics 3.2mm<br>and 4.0mm Cannulated<br>Compression Screws | Primary Predicate<br>Synthes USA Products, LLC<br>DePuy Synthes 3.5 mm and<br>4.0 mm Cannulated Screws<br>K161616 |
| Device Classification | Class II | Class II |
| Product Code | HWC | HWC |
| Purpose | Subject device | Same indications |
| Diameter (mm) | 3.2 - 4.0mm | 3.5 - 4.0mm |
| Lengths (mm) | 14 - 60mm | 10 - 50mm |
| Partially / Fully Threaded | Partially Threaded<br>Fully Threaded | Partially Threaded<br>Fully Threaded |
| Head Type | Headless | Low-Profile |
| Driver Shape | Star | Hex |
| Cannulated | Yes | Yes |
| Self-Tapping | Yes | Yes |
| Sterility | End-User Sterilized, Steam<br>Sterilization<br>Cycle: Pre-Vacuum<br>Temperature: 132°C (270°F)<br>Exposure Time: 4 minutes<br>Drying Time: 60 minutes | Terminally Sterilized, Gamma<br>Irradiation |
| Material | Stainless Steel per ASTM F138 | Stainless Steel per ASTM F138 |
| Manufacturing | Subtractive manufacturing | Subtractive manufacturing |
| Biocompatibility | Implant: Permanent Duration -<br>Implant for Tissue / Bone (>30<br>days)<br>Instrument: Limited Duration -<br>Implant for Tissue / Bone (< 24<br>hours) | Implant: Permanent Duration -<br>Implant for Tissue / Bone (>30<br>days)<br>Instrument: Limited Duration -<br>Implant for Tissue / Bone (< 24<br>hours) |
| Indicated Patient Population | Adults and children (2-12 years)<br>and adolescents (12-21 years) in<br>which growth plates have fused<br>or in which growth plates will<br>not be crossed by screw<br>fixation. | Adults and children (2-12 years)<br>and adolescents (12-21 years) in<br>which growth plates have fused<br>or in which growth plates will<br>not be crossed by screw<br>fixation. |
{7}------------------------------------------------
# Comparison of Technological Characteristics: Avanti Orthopaedics 3.2mm and 4.0mm Cannulated Compression Screws
{8}------------------------------------------------
## Technological Characteristics
There are no significant technological differences between the subject and predicate devices. The subject device uses the same material as the additional predicate and reference device and all devices are manufactured using subtractive techniques. Any differences in available sizes between the subject and predicate devices were addressed via the additional predicates (K180541, K162825).
## Non-Clinical Performance Testing Summary
Testing was performed per ASTM F543-17 and the device met all predetermined acceptance criteria as defined in the FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers -Performance Criteria for Safety and Performance Based Pathway".
## Conclusions
Based on the similarities of the intended use/indications for use, technological and functional characteristic, and the results of the non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.