K222958 · Sciton, Inc · GEI · Apr 23, 2024 · General, Plastic Surgery
Device Facts
Record ID
K222958
Device Name
mJOULE RF System and Accessories
Applicant
Sciton, Inc
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The mJoule RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
The mJOULE RF System is an electrosurgical device used in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The system comprises a control console and RF delivery accessories, including single-use needle arrays. The console houses a power supply, cooling system, and microprocessor-based control system. The clinician enters treatment parameters via a touchscreen and activates RF emission using a footswitch. The device delivers bipolar RF energy to tissue through needle arrays at depths of 0.5–4.0 mm. The system monitors energy delivery to the tissue. The device is intended for use by healthcare professionals in a clinical setting. By creating controlled thermal coagulative zones, the device enables hemostasis and coagulation, providing a therapeutic benefit for dermatologic and surgical applications.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and high-frequency surgical equipment standards (IEC 60601-2-2). Biocompatibility was assessed per ISO 10993. Sterilization was validated to a SAL of 10^-6. Thermal testing on ex vivo porcine tissue demonstrated that coagulation zones are proportional to energy delivered, confirming the device's ability to achieve electrocoagulation and hemostasis.
Technological Characteristics
Bipolar RF electrosurgical system. Operates at 1 MHz. Features include microprocessor control, touchscreen interface, and footswitch activation. Needle arrays (7x7, 6x6) with 0.2 mm diameter and 0.5-4.0 mm depth. Power output max 50 W. Sterilization via EO. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-1-6. Connectivity includes remote interlock connector.
Indications for Use
Indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. No specific patient population age or gender restrictions are stated.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
April 23, 2024
Sciton, Inc. Jay Patel VP of Regulatory Affairs 925 Commercial Street Palo Alto, California 94303
Re: K222958
Trade/Device Name: mJOULE RF System and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: March 22, 2024 Received: March 25, 2024
Dear Jay Patel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.04.23 08:18:06 -04'00'
Mark Trumbore, Ph.D. Acting Division Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222958
Device Name mJoule RF System
Indications for Use (Describe)
The mJoule RF System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Sciton, Inc. |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 925 Commercial Street, Palo Alto, CA 94303 |
| Phone: | (650) 493-9155 |
| Fax: | (650) 493-9146 |
| Contact Person: | Jay M. Patel, VP of Regulatory Affairs |
| Date Prepared: | April 19, 2024 |
| Device Trade Name: | mJOULE RF System |
| Common Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories, 21 CFR 878.4400 |
| Classification Product Code: | GEI |
| Legally Marketed Predicate<br>Device: | K192545 Potenza |
| Description of the mJOULE<br>RF System: | The mJOULE RF System consists of console and radio frequency delivery<br>accessories. The control console houses the power supply, cooling system and RF<br>delivery system. The treatment parameters are entered via a touchscreen on the<br>console. The user activates RF emission by means of a footswitch. |
| Indications for Use: | The mJOULE RF system is intended for use in dermatologic and general surgical<br>procedures for electrocoagulation and hemostasis. |
| Technological<br>Characteristics: | The mJOULE System shares the same indications for use. Predicate Comparison<br>of this submission, shares similar design features (including frequency and delivery<br>systems, power supply, cooling and control system), functional features (including<br>power output and needle array size), and is therefore substantially equivalent to the<br>above legally marketed predicate device. |
{4}------------------------------------------------
Comparative representation of Potenza (K192545), which is the predicate for the subject device, mJOULE
RF System (K222958).
| Specification | Predicate Device | This Application |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K192545 | K222958 |
| Product | Potenza | mJOULE RF System |
| Regulation | GEI, 21 CFR 878.4400 | GEI, 21 CFR 878.4400 |
| Device Class | ll | ll |
| Intended Use | The POTENZA is intended for use<br>in dermatologic and general<br>surgical procedures for<br>electrocoagulation and<br>hemostasis. | mJOULE RF System is intended<br>for use in in dermatologic and<br>general surgical procedures for<br>electrocoagulation and<br>hemostasis. |
| Frequency | 1MHz, 2MHz | 1MHz |
| Needle Array Size | 7x7, 5x5, 4x4, 1 | 7x7, 6x6 |
| Needle Array Pitch | 1.4 mm, 2 mm | 1.4 mm, 1.7 mm |
| Needle Exposed<br>Length | 0.4 mm | 0.4 mm |
| Needle Depth | 0.5 - 4.0 mm | 0.5 - 4.0 mm |
| Needle Polarity | Bipolar | Bipolar |
| Needle Diameter | 0.2 mm | 0.2 mm |
| Treatment Time | 10 – 500 ms | 10 – 600 ms |
| Power Max | 50 W | 50 W |
| Power per Needle | 1.02 W for 7x7 array | 1.02 W for 7x7 & 6x6 array |
| Sterilization | EO | EO |
| Power Source<br>(Active Accessory) | 100 - 240 VAC, 50/60 Hz | 100 - 240 VAC, 50/60 Hz |
| Electrode - Single<br>Use (Yes/No) | Yes | Yes |
| Control System | Microprocessor | Microprocessor |
| Energy Monitor | Display Indicates Energy<br>Delivered to Tissue | Display Indicates Energy<br>Delivered to Tissue |
| Safety | NA | Safety Eyewear and Remote<br>Interlock Connector |
| Safety Compliance<br>Standards | IEC 60601-1, IEC 60601-1-2, IEC<br>60101-2-2, IEC 60601-1-6 | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-2, IEC 60601-1-6 |
| Console Dimension | 20" x 14" x 12" | 12" x 15" x 43" |
| User Interface | Touch screen control | Touch screen control |
{5}------------------------------------------------
Performance Testing – Bench:
The safety and efficacy of the mJoule RF System were established by a series of performance tests; lab performance tests, design validation and software verification and validation. Sciton conducted bench testing to ensure that the mJoule RF System operates safely and within predefined design specifications.
- 1. Tested parameters include:
- · Frequency: Bipolar at 1MHz
- Power measurements
- · Energy per pin and total energy
- · Impedance measurement accuracy and range
- · Safety test of various warnings / failsafe mechanisms
- · Needle depth
- Motor speed level
2. Biocompatibility testing The biocompatibility tests were performed according to ISO 10993 standards for acute systemic toxicity, cytotoxicity, irritation, sensitization, and pyrogenicity assessment of mJoule RF device as recommended in ISO 10993-1 and FDA's guidance document.
- 3. Sterilization testing The sterilization of single use disposable tips was tested to ensure the sterility assurance level (SAL) of at least 10-6.
| 4. | Electrical safety and electromagnetic compatibility (EMC) Electrical safety, EMC, device related electrical safety for higher frequency and usability were conducted on the mJoule RFMN system according to the following consensus standards: IEC 60601-1 Medical Electrical equipment Part1: General requirements for basic safety and essential performance IEC 60601-1-2 EMC testing- General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Tests IEC 60601-2-2 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5. | Software verification and validation testing Software verification and validation testing was conducted in accordance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for the subject device. |
| Performance Testing – Thermal Testing: | Thermal testing is conducted on ex vivo porcine tissue in accordance with FDA's Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, March 9, 2020 to quantitatively determine the depth, height and width of thermal coagulative zones. It is concluded from this study that coagulation zones are created below the surface of the tissue that are proportional to the amount of energy delivered to the tissue. The ex-vivo study results show that the RF microneedling tips are safe for use and effective in achieving the specified indications of dermatologic and general surgical procedures for electrocoagulation and hemostasis. |
| Safety and Effectiveness: | The mJOULE RF System is similar to predicate device with respect to principles of operation, technological characteristics, as well as performance characteristics. Results of design validation and verification activities, i.e., testing to designated |
Safet Effec Results of design validation and verification activities, i.e., testing to standards and performance testing of the device has demonstrated substantial equivalence of the subject device to the predicate device in terms of safety and effectiveness for requested intended use.
{6}------------------------------------------------
Conclusion:
The mJOULE RF System shares similar indications for use, design features, and similar functional features as, and therefore is substantially equivalent to the currently marketed predicate device. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.