← Product Code [IZL](/productcode/IZL) · K222951

# SM-V (K222951)

_Sedecal., Sa. · IZL · Nov 21, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222951

## Device Facts

- **Applicant:** Sedecal., Sa.
- **Product Code:** [IZL](/productcode/IZL.md)
- **Decision Date:** Nov 21, 2022
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1720
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

This is a mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

## Device Story

Mobile diagnostic x-ray system; capacitor-assisted; used by qualified doctors/technicians in clinical settings. Inputs: manual selection of kVp and mAs via two-point control. Operation: monoblock x-ray generator produces x-ray beams for imaging; manual collimation; optional ion chamber for DAP reading. Output: radiographic exposures for external imaging systems (not provided). Benefits: ergonomic, lightweight, maneuverable design; tube protection circuitry; closed-loop control of tube current/kVp/filaments minimizes errors. Healthcare providers use output to perform diagnostic imaging; device facilitates clinical decision-making by providing high-quality radiographic exposures.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and non-clinical performance evaluation. System verified for proper operation and compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-2-28, 60601-1-6, and 62304 standards.

## Technological Characteristics

Mobile x-ray system; monoblock generator (16, 20, 32 kW); 40-150 kV range; 0.32-200 mAs range. Manual collimation (Varex Optica 10). Features closed-loop control of tube current, kVp, and filaments. Power: 100-240V AC single-phase. Compliance: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-2-28, 60601-1-6, 62304. No imaging software or digital detectors included.

## Regulatory Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- PHOENIX ([K212291](/device/K212291.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 21, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

SEDECAL SA % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Court NAPLES FL 34114

Re: K222951

Trade/Device Name: SM-V Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: September 20, 2022 Received: September 27, 2022

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

2022.11.21
*Lu Jiang* 14:55:12 -05'00'

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222951

Device Name SM-V.

#### Indications for Use (Describe)

This is a mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: 510(k) Number K222951

Image /page/3/Picture/1 description: The image is a logo for SEDECAL. The logo is a teal rounded rectangle with a white border. Inside the rectangle is a white icon of a desk with a lamp on top, followed by the word "SEDECAL" in white, sans-serif font.

SEDECAL SA C/ Pelaya, 9 - 13, Pol. Ind. Río de Janeiro 28110 Algete, Madrid España (Spain) Tel.- +34 91 6280544 Fax.- +34 91 6280574 Date Prepared: October 12, 2022 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager

- 1) Identification of the Device: Trade/Device Name: SM-V Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL. Common/Usual Name: Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K212291 Trade/Device Name: PHOENIX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Indications for Use: This is a mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
- 4) Description of the Device: This is a capacitor assisted mobile diagnostic x-ray system which is provided without the digital x-ray acquisition systems used on our predicate model. The unit is provided with a battery to supply the internal PC up to 3 hours without mains connection. This system has been made lighter and more maneuverable, such that battery assisted motorized movement is no longer necessary. The removal of the movement motorization and the required rechargeable batteries resulted in a much lighter system. The generator is also smaller resulting in further weight reduction. The indication for use remains the same as the predicate. The system uses a different xray generator than the one used by the predicate but both are monoblock x-ray generators (same type of generator). The collimator has been changed from the Ralco R108F to the Varex Optica 10. They are functionally identical, having just two manual controls. Although the available generator

{4}------------------------------------------------

options have lower peak outputs, the full range of examinations can still be made. The main features of this equipment are:

- a) A solid and ergonomic design. Easy operation, security and precision of all positioning movements relative to the position of the patient.
- b) Standard electrical outlet operation with single-phase lines at 100 -- 240 V*. Automatic line voltage compensation.
- c) Head-Assembly rotation in relation to its transversal axis (±±180°) and horizontal axis (+90°, --30°). Collimator rotation in relation to its vertical axis (±90°).
- d) Controls for lock release of Rotating Column and Telescopic Arm.
- e) Column rotation in relation to its vertical axis (±317°), telescopic and vertical motion of the Arm.
- Two Point control by selecting kVp and mAs. t)
- g) X-ray Handswitch for X-ray exposures and Collimator Light.
- h) Infrared Remote X-ray Handswitch (optional) for X-ray exposures and Collimator Light.
- i) Dosimetry (optional):
	- i) DAP reading with an Ion Chamber.
	- ii) Calculated DAP (eDAP).
- j) Manual Collimation.
- k) Tube protection circuitry prolongs Tube life and increases system performance.
- I) Equipped with closed loop control of X-ray Tube current, kVp and filaments, which minimize potential errors and the need for readjustments.

There is no imaging hardware or software supplied with the unit. That is the responsibility of the purchaser.

| Characteristic                      | Predicate: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                              | SM-V                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use:             | This is a digital mobile diagnostic x-ray system<br>intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the prone<br>or supine position. Not for mammography | This is a mobile diagnostic x-ray system<br>intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the prone<br>or supine position. Not for mammography.<br>(Same) |
| Configuration                       | Mobile System with digital x-ray panel and<br>image acquisition computer                                                                                                                                                                                                                                                                                                                                                                | SAME but no image acquisition software is<br>supplied.                                                                                                                                                                                                                                                                                                                                                                                     |
| X-ray<br>Generator(s)               | 20, 32, 40 kW and 50 kW Monoblock<br>generator<br>kV range: from 40 kV to 150 kV<br>mAs range: from 0.1 mAs to 500 mAs                                                                                                                                                                                                                                                                                                                  | 16, 20, 32 kW Monoblock generator<br>kV range: from 40 kV to 150 kV<br>mAs range: from 0.32 mAs to 200 mAs<br>Although the peak power is lower, the full<br>range of examinations from hand to chest can<br>be effectively imaged.                                                                                                                                                                                                         |
| Collimator                          | Ralco R108F                                                                                                                                                                                                                                                                                                                                                                                                                             | Varex Optica 10. Controls are the same.                                                                                                                                                                                                                                                                                                                                                                                                    |
| Characteristic                      | Predicate: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                              | SM-V                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Photos                              | Image: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                                  | Image: SM-V                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Touch screen<br>control             | Yes, 19"                                                                                                                                                                                                                                                                                                                                                                                                                                | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Motorized<br>movement               | Included                                                                                                                                                                                                                                                                                                                                                                                                                                | Not included                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Meets US<br>Performance<br>Standard | YES 21 CFR 1020.30                                                                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Power Source                        | Universal power supply, from 100 V~ to 240<br>V~. 1 phase                                                                                                                                                                                                                                                                                                                                                                               | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Imaging<br>Software                 | Konica-Minolta control software CS-7<br>(K151465 or K172793) OR Konica-Minolta<br>control software Ultra.                                                                                                                                                                                                                                                                                                                               | Not included                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Panel Interface                     | Ethernet or Wi-Fi wireless                                                                                                                                                                                                                                                                                                                                                                                                              | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Size                                | 122 x 54 x 223 cm                                                                                                                                                                                                                                                                                                                                                                                                                       | 123 x 59 x 191 cm (including column) /<br>123 x 59 x 126 cm. Smaller, more compact.                                                                                                                                                                                                                                                                                                                                                        |
| Weight                              | 520 kg                                                                                                                                                                                                                                                                                                                                                                                                                                  | 275kg Much lighter. Weight reduction due to<br>elimination of motion motors and their heavy<br>rechargeable batteries as well as using a<br>smaller generator.                                                                                                                                                                                                                                                                             |
| Digital<br>Detectors                | AeroDR Series                                                                                                                                                                                                                                                                                                                                                                                                                           | Not included.                                                                                                                                                                                                                                                                                                                                                                                                                              |

## 5) Substantial Equivalence Chart

{5}------------------------------------------------

- 6) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. The results of bench testing indicate that the new device is as safe and effective as the predicate device. Here is a summary of the changes we made to the device: We removed the imaging system hardware and software, and we removed the motorized motion system, including motors and rechargeable batteries.

{6}------------------------------------------------

- 7) Summary of non-clinical testing: A system was assembled and found to be operating properly. Firmware was essentially unchanged except for new technique stations. Because the system has an Ethernet interface, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff. The internal computer is available for the installation of user supplied imaging software.
The SM-V Battery Mobile X-Ray Unit has been tested to be in compliance with the following International Standards:

IEC 60601-1:2005+A1:2012 (Edition 3.1) IEC 60601-1-2:2020 + A1:2020 (Edition 4.1) IEC 60601-1-3:2008+A1:2013 (Edition 2.1) IEC 60601-2-54:2009+A1:2015 + A2:2018 (Edition 1.2) IEC 60601-2-28:2017 (Edition 3.0) IEC 60601-1-6:2010 + A1:2013 (Edition 3.1) IEC 62304:2006 + A1:2015 (Edition 1.1)

- 8) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
- 9) Conclusion: After analyzing bench and non-clinical tests, it is the conclusion of Sedecal SA. that the SM-V Diagnostic Mobile X-Ray System is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K222951](https://fda-staging.innolitics.com/device/K222951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
