A clinical sample evaluation was performed to assess the diagnostic quality of the Nanox.ARC tomosynthesis images in human subjects as an adjunct to conventional radiography.
Clinical sample evaluation at Shamir Medical Center; Clinical sample evaluation
Healthy adult human subjects; Sample Size: 9; Number of Sites: 1
Standard of care radiographies
Diagnostic quality of tomosynthesis images
Indications for Use
Nanox.ARC is a stationary X-ray system intended to produce tomographic images of the human musculoskeletal system adjunctive to conventional radiography, on adult patients. This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists. Digital Tomosynthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions. This device is not intended for mammographic, cardiac, pulmonary, intra-abdominal, intra-cranial, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
Device Story
Nanox.ARC is a floor-mounted, stationary tomographic X-ray system for musculoskeletal imaging. It utilizes a tilting arc gantry with five independently, electronically switched cold-cathode X-ray tubes (MEMS-based quantum tunneling) operated sequentially to acquire multiple low-dose projections during a single non-linear spherical sweep. Input data from a flat-panel scintillator-photodetector is processed by a reconstruction engine to generate tomographic slices parallel to the table plane. The system is operated by radiographers, radiologists, or physicists in professional healthcare settings. Image reconstruction, protocol management, and DICOMization services can be performed locally or via Nanox.CLOUD. The output provides depth information on structures of interest, reducing overlying structure interference to assist clinicians in diagnostic decision-making. The device features a motorized patient table and is controlled via a tablet interface.
Clinical Evidence
No clinical trials performed. Evidence consists of non-clinical bench testing, phantom performance evaluations, and a clinical sample evaluation. Nine digital tomosynthesis cases from healthy adult subjects and twelve phantom exams were used to confirm diagnostic quality and performance against standard-of-care radiography.
Technological Characteristics
Stationary tomographic X-ray system; cold-cathode X-ray source (MEMS chip, quantum tunneling); five X-ray tubes; flat-panel scintillator-photodetector; 43x43cm fixed field size; 1.8kW generator; 40-110 kVp range; PMMA tabletop; DICOM connectivity; local or cloud-based processing; Ubuntu 20.04 LTS OS; tablet-based control console. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 60601-2-28, ISO 10993-1, ISO 14971, IEC 62366.
Indications for Use
Indicated for adult patients requiring tomographic imaging of the musculoskeletal system as an adjunct to conventional radiography. Contraindicated for pediatric/neonatal patients and mammographic, cardiac, pulmonary, intra-abdominal, intra-cranial, interventional, or fluoroscopic applications.
Regulatory Classification
Identification
A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
April 28, 2023
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Nano-X Imaging Ltd. % Odelia Maron VP QA & RA Communications Center, Bldg. C. Entrance 1 Neve Ilan. 9085000 ISRAEL
# Re: K222934
Trade/Device Name: Nanox.ARC Regulation Number: 21 CFR 892.1740 Regulation Name: Tomographic x-ray system Regulatory Class: Class II Product Code: IZF, MQB Dated: March 27, 2023 Received: March 27, 2023
Dear Odelia Maron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22934
Device Name Nanox.ARC
## Indications for Use (Describe)
Nanox.ARC is a stationary X-ray system intended to produce tomographic images of the human musculoskeletal system adjunctive to conventional radiography, on adult patients. This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists. Digital Tomosynthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions. This device is not intended for mammographic, cardiac, pulmonary, intra-abdominal, intra-cranial, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
Type of Use (*Select **one** or **both**, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Nanox. The logo features a stylized, abstract design composed of intersecting blue lines that form a spherical or globe-like shape. Below the graphic is the word "NANOX" in a simple, sans-serif font, also in blue.
# Nanox.ARC 510(k) Summary
K222934
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the alphanumeric string "K222934" in a simple, sans-serif font. The characters are black against a white background. The text appears to be a code or identifier, possibly a serial number or product code.
Image /page/5/Picture/1 description: The image shows the logo for Nanox. The logo consists of a blue, abstract design that resembles a stylized globe or sphere, composed of interconnected lines. Below the design, the word "NANOX" is written in a bold, blue sans-serif font.
#### Submission Sponsor 1
Nano-x Imaging Ltd. Communications Center, Bldg. C, Entrance 1, Neve Ilan, Israel 9085000 Establishment Registration Number: The Company will register following FDA clearance
#### 2 Submission Correspondent
Odelia Maron, Ph.D. VP QA & RA Phone: +972-50-5733837 E-mail: Odelia.m@nanox.vision
#### 3 Date Prepared
March 24, 2023
#### 4 Device identification
| Name of Device: | Nanox.ARC |
|------------------------------|--------------------------|
| Classification Name: | Tomographic X-ray System |
| Regulation: | 21 CFR §892.1740 |
| Regulatory Classification: | Class II |
| Product Classification Code: | IZF and MQB |
| Classification panel: | Radiology |
#### 5 Legally Marketed Predicate and Reference Device
| Predicate Manufacturer: | AGFA NV. |
|--------------------------------|---------------------------|
| Predicate Trade Name: | DR 600 with Tomosynthesis |
| Predicate Classification Code: | IZF and MQB |
| Predicate 510(k): | K193262 |
| Reference Manufacturer: | AGFA NV. |
| Reference Trade Name: | DR 800 with Tomosynthesis |
| Reference Classification Code: | IZF and JAA |
| Reference 510(k): | K183275 |
#### 6 Device description
Nanox.ARC is a tomographic and solid-state X-ray system (product codes IZF and MQB) intended to produce tomographic images of the human musculoskeletal system from a single tomographic sweep, as an adjunct to conventional radiography, on adult patients.
Nanox.ARC is a floor-mounted tomographic system that consists of a user control console, a multisource, tiltable arc gantry with five alternately-switched tubes, a motorized patient table, a flatpanel detector of a scintillator-photodetector type, and Protocols database and Image processing software packages.
Nanox.ARC utilizes several small-sized X-ray tubes that are independently and electronically switched, thereby dividing the overall power requirements over multiple tubes. Nanox.ARC utilizes a tilting imaging ring with five X-ray tubes, operated sequentially, one at a time, used to generate multiple low-dose X-ray projection images acquired from different angles during a single spherical (non-linear) sweep. The sweep is performed over a motorized patient table. Patients can be placed in prone, supine, and lateral positions.
The acquired projection imaging data is automatically reconstructed to form tomographic slices of the imaged object, with each slice parallel to the table plane. The Tomosynthesis image result reduces the effect of overlying structures and provides depth information on structures of interest. The image reconstruction service, as well as the system's protocol database and DICOMization
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo for Nanox, a company specializing in medical imaging technology. The logo consists of a stylized, abstract design in blue, resembling a symmetrical, curved pattern. Below the graphic is the word "NANOX" in a bold, sans-serif font, also in blue, clearly identifying the company.
services, can be hosted either locally or as part of the Nanox.CLOUD, according to customer preference. The resultant images are sent using the DICOM protocol.
#### 7 Indications for Use
Nanox.ARC is a stationary X-ray system intended to produce tomographic images of the human musculoskeletal system adjunctive to conventional radiography, on adult patients.
This device is intended to be used in professional healthcare facilities or radiological environments, such as hospitals, clinics, imaging centers, and other medical practices by trained radiographers, radiologists, and physicists.
Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep. Applications can be performed with the patient in prone, supine, and lateral positions.
This device is not intended for mammographic, angiographic, cardiac, pulmonary, intraabdominal, intra-cranial, interventional, or fluoroscopic applications. This device is not intended for imaging pediatric or neonatal patients.
#### 8 Technological Characteristics
Nanox.ARC is a stationary X-ray tomographic system utilizes predefined protocols with predefined optimal acquisition parameters depending on the anatomy being imaged. The System allows the user to further configure these parameters for each examination type.
The acquired projection imaging data is automatically reconstructed to form tomographic imaged object.
A reconstruction engine converts the 2D images from the detector into Tomographical layers data. This reconstruction engine, as well as the system's protocol database and DICOMization services, can be hosted either locally, or as part of the Nanox.CLOUD.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo for Nanox. The logo consists of a stylized blue graphic above the word "NANOX" in a simple sans-serif font. The graphic is composed of curved lines that intersect to form a pattern resembling a stylized eye or a network of interconnected pathways.
#### 9 Substantial Equivalence
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence with respect to indications for use, principles of operation, technological characteristics, materials, and performance.
The comparison of the devices in Table 1 below provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new or different questions of safety or effectiveness based on the similarities to the predicate device.
| | Table 1 – Comparison of Subject Device with Predicate Device | |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Subject Device<br>Nanox.ARC | Predicate Device<br>DR 600 with Tomosynthesis |
| Name | Nanox.ARC | DR 600 with Tomosynthesis |
| Manufacturer | Nano-x Imaging Ltd. | AGFA NV |
| 510(k) | K222934 | K193262 |
| Date of clearance | To be assigned | March 9, 2020 |
| Indications for use | Nanox.ARC is a stationary X-ray<br>system intended to produce<br>tomographic images of the human<br>musculoskeletal system adjunctive<br>to conventional radiography, on<br>adult patients.<br><br>This device is intended to be used<br>in professional healthcare facilities<br>or radiological environments, such<br>as hospitals, clinics, imaging<br>centers, and other medical<br>practices by trained radiographers,<br>radiologists, and physicists.<br><br>Digital Tomosynthesis is used to<br>synthesize tomographic slices from<br>a single tomographic sweep.<br><br>Applications can be performed<br>with the patient in prone, supine,<br>and lateral positions.<br><br>This device is not intended for<br>mammographic, angiographic,<br>cardiac, pulmonary, intra-<br>abdominal, intra-cranial,<br>interventional, or fluoroscopic<br>applications. This device is not<br>intended for imaging pediatric or<br>neonatal patients. | The DR 600 System is a general<br>radiography X-ray imaging system used<br>in hospitals, clinics and medical<br>practices by radiographers, radiologists<br>and physicists to make, process and<br>view static X-ray radiographic images<br>of the skeleton (including skull, spinal<br>column and extremities), chest,<br>abdomen and other body parts on adult,<br>pediatric or neonatal patients. In<br>addition, the System provides the<br>AGFA tomosynthesis option, intended<br>to acquire tomographic slices of human<br>anatomy and to be used with AGFA DR<br>X-ray systems. Digital Tomosynthesis<br>is used to synthesize tomographic slices<br>from a single tomographic sweep.<br>Applications can be performed with the<br>patient in a sitting, standing or lying<br>position. This System is not intended<br>for mammography applications. |
| Intended users | Radiographers, radiologists, and<br>physicists | Radiographers, radiologists, and<br>physicists |
| Intended use environment | Hospitals, clinics, imaging centers,<br>and other healthcare facilities | Hospitals, clinics, imaging centers, and<br>other healthcare facilities |
| Classification Name | Tomographic X-ray system | Tomographic X-ray system |
| Regulation number | 21 CFR § 892.1740 | 21 CFR § 892.1740 |
| Regulatory class | II | II |
| Product code | IZF, MQB | IZF, MQB |
| Item | Subject Device | Predicate Device |
| | Nanox.ARC | DR 600 with Tomosynthesis |
| Communications with | DICOM | DICOM |
| external storage devices | | |
| Patient positioning options | | |
| Vertical stand | Not included | Included as an Option |
| Table stand | Included | Included as an Option |
| Patient Table | | |
| Patient table height | 48cm<br>(motorized w/fixed height) | 55-90 cm<br>(motorized w/ height adjustment) |
| Patient weight allowance<br>(kg/lbs.) | 150kg / 330 lbs. | 320 Kg / 705 lbs. |
| Table length (mm) | 2600 | 2200 |
| Table width (mm) | 720 | 810 |
| Tabletop material | PMMA | Resopal HPL (DIN EN438) |
| Generator | | |
| Tube voltage range (kVp) | 40-110 | 40-150 |
| Nominal electric power<br>(kW) | 1.8 | Selectable 50, 65 or 80 |
| Total mAs range per<br>Tomosynthesis study | 12-90 mAs | 12-90 mAs |
| X-ray accuracy | Fulfills the X-ray radiation<br>accuracy according to EN IEC<br>60601-2-54 with a variation of<br>max. 0.05 (5%). | Fulfills the X-ray radiation accuracy<br>according to EN IEC 60601-2-54 with a<br>variation of max. 0.05 (5%). |
| X-ray source | | |
| Electron stream source | MEMS chip | Tungsten filament |
| Physical phenomena<br>generating electron<br>stream | Quantum tunneling via an applied<br>External electrical field (cold<br>cathode) | Thermionic excitation (Hot Cathode) |
| Focal spot size (nominal<br>FS value large/small<br>focus) | 0.7 | E7884X: 0.6/1.2<br>E7252X: 0.6/1.2<br>E7254FX: 0.6/1.2<br>E7869XX: 0.6/1.2 |
| Anode design | Stationary Anode…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.