Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure Monitor
Applicant
Famidoc Technology Company Limited
Product Code
DXN · Cardiovascular
Decision Date
Apr 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure is intended to measure the blood pressure and pulse rate of adults and children at least 12 years of age, at household or medical center. with the cuff around the left upper arm according to the instruction in the user's guide manual(Cuff size 220 ~ 420 mm).
Device Story
Device is an automatic, wireless, upper-arm blood pressure monitor. It uses an inflatable cuff and an oscillometric method to measure systolic/diastolic blood pressure and pulse rate. The device consists of modular components including a pneumatic flow control module, pressure detection module, and a single-chip microcomputer. The microcomputer controls inflation/deflation, processes amplified/filtered pulse pressure signals via a proprietary algorithm, and calculates measurements. Results are displayed on an LED or LCD screen and stored in memory. Some models include Bluetooth/Wi-Fi for data transmission. The device is intended for use by adults and children (12+ years) in home or clinical settings. Healthcare providers or patients use the output to monitor blood pressure trends, aiding in hypertension management and cardiovascular health assessment.
Clinical Evidence
Clinical validation conducted per ISO 81060-2:2018 using a same-arm sequential method. Study included 85 participants (43 males). A manual mercury sphygmomanometer served as the reference. Results confirmed accuracy within acceptable limits specified by ISO 81060-2. No adverse events or side effects reported.
Technological Characteristics
Oscillometric measurement principle. Modular design with single-chip microcomputer. Power: 3.7V lithium battery or 4xAAA batteries/USB 5V. Display: LED or LCD. Connectivity: Bluetooth/Wi-Fi (select models). Biocompatibility per ISO 10993-5, -10. Safety/EMC compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30.
Indications for Use
Indicated for adults and children at least 12 years of age for measurement of blood pressure and pulse rate at home or medical centers using the left upper arm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
April 10, 2023
Famidoc Technology Company Limited Amos Zou Management Representative No. 212 Yilong Road, Hexi Industrial Zone Jinxia. Changan Town Dongguan, Guangdong 523853 China
Re: K222887
Trade/Device Name: Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 6, 2023 Received: April 7, 2023
Dear Amos Zou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Aneesh S. Deoras -S
for
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K22887
#### Device Name
Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure Monitor
Indications for Use (Describe)
Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure is intended to measure the blood pressure and pulse rate of adults and children at least 12 years of age, at household or medical center. with the left upper arm according to the instruction in the user's guide manual(Cuff size 220 ~ 420 mm).
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(K) safety and effective information is being submitted in accordance with the requirement SMDA and 21 CFR 807.92.
## 1. Submitter of 510(K):
| Date of Prepared: | 17/1/2023 |
|-------------------|------------------------------------------------------------------------------------------------|
| Submitter's Name: | Famidoc Technology Company Limited |
| Address: | No.212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan Town<br>Dongguan Guangdong 523853 CN |
| Contact person: | Amos zou |
| TEL: | +(86) -769-8927-2488 |
| FAX: | +(86) -769-8927-2498 |
| E-mail: | qa@famidoc.com |
## 2.Proposed Device and code:
| Device Trade Name: | Arm Blood Pressure Monitor/Automatic Wireless Smart Blood<br>Pressure monitor<br>(Model:FDBP-A7B、FDBP-A7BL、FDBP-A7BT、FDBP-A7BLT、<br>FDBP-A8B、FDBP-A8BL、FDBP-A8BT、FDBP-A8BLT、<br>FDBP-A9B、FDBP-A9BL、FDBP-A9BT、FDBP-A9BLT、<br>FDBP-A10B、FDBP-A10BL、FDBP-A10BT、FDBP-A10BLT) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Medical<br>Specialty | Noninvasive blood pressure measurement system. |
| Product Code: | DXN |
| Regulation number | 21 CRF 870.1130 |
| Device Class | 2 |
## 3. Predicate Device:
| 510(K) | Trade or Proprietary or Model Name | Manufacturer |
|---------|------------------------------------------------------|-----------------------|
| K210770 | Fully Automatic Electronic<br>Blood Pressure Monitor | Andon Health Co,.Ltd. |
### 4. Description of Proposed Device:
Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure Monitor(Model:FDBP-A7B、FDBP-A7BL、FDBP-A7BT、FDBP-A7BLT、FDBP-A8B、 FDBP-A8BL、FDBP-A8BT、FDBP-A8BLT、FDBP-A9B、FDBP-A9BL、FDBP-A9BT、
{4}------------------------------------------------
FDBP-A9BLT、FDBP-A10B、FDBP-A10BL、FDBP-A10BT、FDBP-A10BLT)includes utilize modular design method. It consists of nine main modules:
□ power-on self-test module, system initialization module,sampling data processing and pressure, pulse rate calculation module, display processing module, power detection processing module, data storage module, key scanning processing module, sampling processing module, , wireless function module, and each module communicates through a message queue.
□ The blood pressure monitor controls the pneumatic flow control module through singlechipped microcomputer to pressurize the cuff module in order to exceed the lower pressure of patients, the blood being pushed against the artery walls;
□ Pneumatic Flow Control Module being directed to release the pressure, while the pressure detection module collect pulse pressure signal and amplify filter;
[ amplified filtersignal being read by single-chipped microcomputer for pressure and pulse signal,through unique algorithm to obtain the systolic and diastolic pressure with pulse;
□ Single-chipped microcomputer will control the inflation/deflation module to release the pressure after receive measurements;
□ in the meanwhile, display the measurements results then stored the values with memory module.
## 5. Intended for Use
Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure is intended to measure the blood pressure and pulse rate of adults and children at least 12 years of age, at household or medical center. with the cuff around the left upper arm according to the instruction in the user's guide manual(Cuff size 220 ~ 420 mm).
#### 6. Technical and Performance
The following table compares the device to the predicate device with basic technological characteristics.
{5}------------------------------------------------
7.Comparison of Technological Characteristics with Predicate Device
| Elements of<br>Comparison | Subject Devices | | | | Predicate Device | Comparison<br>Result |
|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Device Name | Arm Blood Pressure Monitor/Automatic Wireless Smart Blood Pressure Monitor | | | | Fully Automatic Electronic<br>Blood Pressure Monitor | / |
| Device Model | FDBP-A7BL | FDBP-A7BT | FDBP-A10BL | FDBP-A10BT | KD-5810, KD5810B,<br>KD-5811, KD-5920TL,<br>KD-552 | / |
| 510 (k) Number | K222887 | K222887 | K222887 | K222887 | K210770 | / |
| Product Code | DXN | DXN | DXN | DXN | DXN | SE |
| Regulation No. | 870.1130 | 870.1130 | 870.1130 | 870.1130 | 870.1130 | SE |
| Classification | II | II | II | II | II | SE |
| Intended Use and indications for Use | | | | | | |
| Intended Use | Arm Blood Pressure<br>Monitor/Automatic<br>Wireless Smart Blood<br>Pressure is intended to<br>measure the blood<br>pressure and pulse rate of<br>adults and children at<br>least 12 years of age, at<br>household or medical<br>center. with the cuff<br>around the left upper arm<br>according to the<br>instruction in the user's<br>guide manual(Cuff size<br>220 ~ 420 mm). | Arm Blood Pressure<br>Monitor/Automatic<br>Wireless Smart Blood<br>Pressure is intended to<br>measure the blood<br>pressure and pulse rate of<br>adults and children at<br>least 12 years of age, at<br>household or medical<br>center. with the cuff<br>around the left upper arm<br>according to the<br>instruction in the user's<br>guide manual(Cuff size<br>220 ~ 420 mm). | Arm Blood Pressure<br>Monitor/Automatic<br>Wireless Smart Blood<br>Pressure is intended to<br>measure the blood<br>pressure and pulse rate of<br>adults and children at<br>least 12 years of age, at<br>household or medical<br>center. with the cuff<br>around the left upper arm<br>according to the<br>instruction in the user's<br>guide manual(Cuff size<br>220 ~ 420 mm). | Arm Blood Pressure<br>Monitor/Automatic<br>Wireless Smart Blood<br>Pressure is intended to<br>measure the blood<br>pressure and pulse rate of<br>adults and children at<br>least 12 years of age, at<br>household or medical<br>center. with the cuff<br>around the left upper arm<br>according to the<br>instruction in the user's<br>guide manual(Cuff size<br>220 ~ 420 mm). | Fully Automatic Electronic<br>Blood Pressure Monitor is<br>for use by medical<br>professionals<br>or at home and is a<br>non-invasive blood<br>pressure measurement<br>system intended to<br>measure the diastolic and<br>systolic blood pressures<br>and pulse rate of an adult<br>individual<br>by using a non-invasive<br>technique in which an<br>inflatable cuff is wrapped | SE |
| upper arm. The cuff<br>circumference is limited to<br>22cm-48cm. | | | | | | |
| Performance Specification | | | | | | |
| Measuring<br>Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | Oscillometric Method | SE |
| Measuring<br>Range | Systolic: 60-250mmHg<br>Diastolic: 30-195mmHg<br>Pulse: 40-199beats/min | Systolic: 60-250mmHg<br>Diastolic: 30-195mmHg<br>Pulse: 40-199beats/min | Systolic: 60-250mmHg<br>Diastolic: 30-195mmHg<br>Pulse: 40-199beats/min | Systolic: 60-250mmHg<br>Diastolic: 30-195mmHg<br>Pulse: 40-199beats/min | Systolic: 60-260mmHg<br>Diastolic: 40-199mmHg<br>Pulse rate: 40-180<br>beats/min | Similar<br>Note 1 |
| Pressure<br>resolution | 1 mmHg or 0.1 kPa | 1 mmHg or 0.1 kPa | 1 mmHg or 0.1 kPa | 1 mmHg or 0.1 kPa | 1 mmHg or 0.1 kPa | SE |
| Accuracy | Pressure:<br>±3 mmHg (±0.4kPa)<br>Pulse:±5% | Pressure:<br>±3 mmHg (±0.4kPa)<br>Pulse:±5% | Pressure:<br>±3 mmHg (±0.4kPa)<br>Pulse:±5% | Pressure:<br>±3 mmHg (±0.4kPa)<br>Pulse:±5% | Pressure: ±3mmHg<br>Pulse rate:<br>Less than 60: ±3bpm<br>More than 60 (incl.) : ±5% | Similar<br>Note 2 |
| Patient<br>Population | Adult | Adult | Adult | Adult | Adult | SE |
| Measurement<br>Site of Body | Upper Arm | Upper Arm | Upper Arm | Upper Arm | Upper Arm | SE |
| Inflation and<br>Deflation | Automatic | Automatic | Automatic | Automatic | Automatic | SE |
| Memory Size | 2x60 sets record | 2x60 sets record | 2x60 sets record | 2x60 sets record | 4×60 times with time and<br>date stamp | Similar<br>Note 3 |
| Indicators | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory | SE |
| | | | | K222887 510(k) Summary | | |
| | Record<br>Number,Bluetooth<br>symbol | Record<br>Number,Bluetooth<br>symbol | Record<br>Number,Bluetooth<br>symbol | Record Number,Bluetooth<br>symbol | Record Number,Bluetooth<br>symbol | |
| Cuff<br>Circumference | 220mm ~420mm | 220mm ~420mm | 220mm ~420mm | 220mm ~420mm | 22cm-30cm,<br>30cm-42cm(Optional),<br>42cm-48cm(Optional),<br>22cm-42cm (Optional) | Similar<br>Note 4 |
| Power Battery | 3.7V lithium battery or<br>USB 5V adapter | 4xAAA 1.5V alkaline<br>batteries or USB 5V<br>adapter | 3.7V lithium battery or<br>USB 5V adapter | 4xAAA 1.5V alkaline<br>batteries or USB 5V<br>adapter | 4xAA 1.5V alkaline<br>batteries or DC 6V<br>adapter | Similar<br>Note 5 |
| Display | LED Digital Display | LED Digital Display | LCD Digital Display | LCD Digital Display | LCD Digital Display | Similar<br>Note 6 |
| | OPERATING&STORAGE CONDITIONS | | | | | |
| Operating<br>Environment | Temperature: 5°C~40°C<br>Humidity: 15% RH~90%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: 5°C~40°C<br>Humidity: 15% RH~90%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: 5°C~40°C<br>Humidity: 15% RH~90%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: 5°C~40°C<br>Humidity: 15% RH~90%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: 10°C~40°C<br>Humidity: ≤85%RH<br>Atmospheric pressure:<br>80kPa~105kPa | Similar<br>Note 7 |
| Storage<br>Environment | Temperature: -25°C~55°C<br>Humidity: 15% RH~95%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: -25°C~55°C<br>Humidity: 15% RH~95%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: -25°C~55°C<br>Humidity: 15% RH~95%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: -25°C~55°C<br>Humidity: 15% RH~95%<br>RH, No condensation<br>Atmospheric pressure:<br>70kPa~106kPa | Temperature: -20°C~50°C<br>Humidity: ≤85%RH<br>Atmospheric pressure:<br>80kPa~105kPa | Similar<br>Note 8 |
| | COMPLIANCE STANDARDS | | | | | |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Safety Engineering |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | Safety Engineering |
| Home Use | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | Safety Engineering |
| Performance | IEC 80601-2-30 | IEC 80601-2-30 | IEC 80601-2-30 | IEC 80601-2-30 | IEC 80601-2-30 | Safety Engineering |
| Biocompatibility | All the patient | All the patient…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.