K222871 · Sparrow Acoustics, Inc. · DQD · Mar 29, 2023 · Cardiovascular
Device Facts
Record ID
K222871
Device Name
Stethophone v1
Applicant
Sparrow Acoustics, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Mar 29, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Stethophone v1 is intended for medical diagnostic purposes only. It may be used for detection and amplification of sounds from the heart and lungs with the use of selective frequency ranges. It can be used on adults undergoing physical assessment.
Device Story
Stethophone v1 is an electronic stethoscope software application for smartphones; captures chest sounds via smartphone microphone; amplifies and filters sounds for real-time auscultation or recording; provides oscillogram and spectrogram visualizations; includes three selectable frequency filters (Bell, Diaphragm, Starling). Used by healthcare providers in clinical and non-clinical environments to assist in physical examinations; supports decision-making by allowing providers to hear, visualize, and store chest sounds for later analysis. Benefits include enhanced sound clarity and visual representation of heart/lung sounds for improved diagnostic assessment.
Clinical Evidence
Bench testing only. No clinical data required. Performance validated via repeatability and reproducibility tests, anechoic chamber testing, internal medical analyst evaluations, and an independent usability and performance study involving external medical specialists.
Technological Characteristics
Standalone software application operating on iOS smartphones. Uses smartphone microphone for sound acquisition. Features digital signal processing (DSP) for frequency filtering (Bell: 25-300 Hz; Diaphragm: 170-850 Hz; Starling: 60-1400 Hz) and visualization (oscillogram/spectrogram). Non-sterile, no electrical/EMC risks as it is standalone software.
Indications for Use
Indicated for detection and amplification of heart and lung sounds in adults undergoing physical assessment. Prescription use only.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Steth IO® Smartphone Stethoscope and Phonocardiogram, Model 1.0 (K160016)
3M™ Littmann® Electronic Stethoscope, Model 3200 (K083903)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
March 29, 2023
Sparrow Acoustics Inc. Nadezda Ivanova Chief Product Officer 2416 Natura Drive Lucasville, NS B4B 0X3 Canada
Re: K222871
Trade/Device Name: Stethophone v1 Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: September 21, 2022 Received: September 22, 2022
Dear Nadezda Ivanova:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222871
Device Name
Stethophone v 1
Indications for Use (Describe)
Stethophone v1 is intended for medical diagnostic purposes only. It may be used for detection and amplification of sounds from the heart and lungs with the use of selective frequency ranges. It can be used on adults undergoing physical assessment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div><span style="font-size:120%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|------------------------------------------------------------------------------------------------|
| <div><span style="font-size:120%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Sparrow Acoustics. The word "SPARROW" is written in a sans-serif font in a light blue color. To the right of the word "SPARROW" is a graphic of a bird inside of a circle with sound waves emanating from it. Below the word "SPARROW" is the word "ACOUSTICS" written in a smaller sans-serif font.
## 510(k) SUMMARY
- 1. Summary Date: September 21, 2022
- 2. Submitter: Sparrow Acoustics Inc. 2416 Natura Dr. Lucasville, NS Canada. B4B 0X3 Tel: +1 (902) 989-3908
- 3. Correspondent: Nadia Ivanova
- 4. Device Trade Name: Stethophone v1
- 5. Device Common Smartphone stethoscope Name:
- 6. Classification: Electronic Stethoscope 21 CFR 870.1875(b) Class II Product Code: DQD Panel: Cardiovascular
- 7. Intended Use/ Stethophone v1 is intended for medical diagnostic purposes only. It may be used for the detection and amplification of sounds from the heart and Indications for Use: lungs with the use of selective frequency ranges. It can be used on adults undergoing physical assessment.
- 8. Device Description: Stethophone v1 is an electronic stethoscope software application that operates on smartphones. Stethophone v1 allows for the capture and amplification of chest sounds
that are listened in real-time or recorded, cloud storage of sound records, sound filtering for selected frequency ranges, and visualization to assist in sound analysis.
Stethophone v1 is designed to assist healthcare providers to hear and visualize heart and lung sounds during a physical examination of a patient as well as storing recorded sounds for later analysis. Stethophone v1 is a decision support device used for the assessment of chest sounds of adult patients in clinical and non-clinical environments. Key product features:
- Acquiring sound through the smartphone microphone ●
- Real-time listening of chest sounds .
- Recording of chest sounds .
- Two modes of sound visualization: oscillogram and spectrogram ●
- Selecting among three sound filters for listening: ●
- Bell: Classic filter used for low frequency sounds O
- Diaphraqm: Classic filter used for higher frequency sounds of O heart and lungs
- Starling: Filter for listening to the full frequency of chest sounds O
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Sparrow Acoustics. The word "SPARROW" is written in a large, sans-serif font in a light blue color. To the right of the word "SPARROW" is a stylized image of a bird inside of a circle with sound waves emanating from it. Below the word "SPARROW" is the word "ACOUSTICS" written in a smaller, sans-serif font.
| 9. | Predicate Devices: | Steth IO® Smartphone Stethoscope<br>and Phonocardiogram, Model 1.0 | 3M™ Littmann® Electronic<br>Stethoscope, Model 3200 |
|-----|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| | | 510(k) Number: K160016<br>Manufacturer: StratoScientific, Inc.<br>Product Code: DQD<br>Classification: 870.1875(b) | 510(k) Number: K083903<br>Manufacturer: 3M Company<br>Product Code: DQD<br>Classification: 870.1875(b) |
| 10. | Comparison to<br>Predicates: | The comparison chart below provides evidence to facilitate the<br>substantial equivalence determination between Stethophone v1 to the<br>predicate devices with respect to intended use, technological<br>characteristics, and principles of operation. | |
| | Stethophone v1 | Steth IO®<br>(primary) | 3M™ Littmann®<br>Model 3200 | Comments |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | | | | |
| Classification | Class II device | Class II device | Class II device | Same |
| Regulation | 21 CFR<br>870.1875(b) | 21 CFR<br>870.1875(b) | 21 CFR<br>870.1875(b) | Same |
| Product code | DQD | DQD | DQD | Same |
| Device | Stethoscope,<br>Electronic | Stethoscope,<br>Electronic | Stethoscope,<br>Electronic | Same |
| FDA Clearance | Pending | k160016 | K083903 | |
| Intended Use/Indications for Use | | | | |
| Intended Use/<br>Indications for<br>Use | Stethophone v1 is<br>intended for<br>medical<br>diagnostic<br>purposes only. It<br>may be used for<br>detection and<br>amplification of<br>sounds from the<br>heart and lungs<br>with the use of<br>selective<br>frequency ranges.<br>It can be used on<br>adults undergoing<br>physical<br>assessment. | The StratoScientific<br>Steth IO<br>Stethoscope and<br>Phonocardiogram<br>Model 1.0 is<br>intended for<br>medical diagnostic<br>purposes only. It<br>may be used for the<br>detection and<br>amplification of<br>sounds from the<br>heart, and lungs<br>with the use of<br>selective frequency<br>ranges. It has been<br>tested for use on<br>adults undergoing a<br>physical<br>assessment. | The 3M™<br>Littmann®<br>Electronic<br>Stethoscope, Model<br>3200 is intended for<br>medical diagnostic<br>purposes only. It<br>may be used for the<br>detection and<br>amplification of<br>sounds from the<br>heart, lungs,<br>arteries, veins, and<br>other internal<br>organs with the use<br>of a selective<br>frequency. It can be<br>used on any person<br>undergoing a<br>physical<br>assessment. | Substantially<br>equivalent |
| Prescription/<br>OTS | Prescription | Prescription | Prescription | Same |
| Population Type | Adults | Adults | All types of patients | Same with Steth<br>IO, narrower<br>population than<br>Littmann 3200 |
| | | | | |
| | | | | which doesn't<br>affect safety and<br>performance for<br>the proposed<br>intended use |
| Principles of Operation | | | | |
| Form Factor | Device that is<br>held in the<br>Doctor's hand is<br>the form of the<br>smartphone | Device that is held<br>in the Doctor's<br>hand is the form of<br>the smartphone | Similar to a<br>traditional<br>stethoscope | Same as Steth IO.<br>Different than<br>Littmann 3200,<br>which doesn't<br>affect safety and<br>performance |
| Dedicated<br>Device vs.<br>iPhone | Operates on<br>iPhone<br>smartphone using<br>its hardware and<br>operating system | Operates on<br>iPhone smartphone<br>using its hardware<br>and operating<br>system | Dedicated<br>proprietary<br>hardware (chest<br>piece) | Same as Steth IO.<br>Different than<br>Littmann 3200,<br>which doesn't<br>affect safety and<br>performance |
| iPhone model<br>Compatibility | 6s, 6S Plus, SE<br>1st gen, 7, 7 Plus,<br>8, 8 Plus, X, XS,<br>XS MAX, XR, SE<br>2 generation, 11,<br>11 Pro, 11 Pro<br>Max, 12, 12 Pro,<br>12 Mini, 12 Pro<br>Max, 13 mini, 13,<br>13 Pro, 13 Pro<br>Max | iPhone 6, 7/8, 7+ or<br>8+, X | N/A | Stethophone v1 is<br>available for a<br>larger number of<br>iPhone models<br>than Steth IO,<br>which may affect<br>availability, but<br>doesn't affect<br>safety and<br>performance as<br>stethoscopes |
| Software<br>Platform | iOS | iOS | Windows, MacOS | Same as Steth IO.<br>Different than<br>Littmann 3200,<br>which may affect<br>availability, but<br>doesn't affect<br>safety and<br>performance as<br>stethoscopes |
| Sound Type | Heart, lungs | Heart, lungs | Heart, lungs,<br>arteries, veins, and<br>other internal<br>organs | Same with Steth<br>IO, narrower than<br>Littmann 3200,<br>which doesn't<br>affect safety and<br>performance for<br>the proposed<br>intended use |
| Signal Input<br>Method | Uses microphone<br>of the smartphone<br>to acquire a<br>sound | Uses microphone<br>of the smartphone<br>to acquire a sound | Uses microphone<br>of the stethoscope<br>to acquire a sound | Same as Steth IO,<br>equivalent to<br>Littmann 3200 |
| Audio Output<br>Method | Headphones | Headphones | Eartubes with<br>eartips | Same as Steth IO,<br>equivalent to<br>Littmann 3200 |
| Record and<br>Playback<br>Sounds | Yes | Yes | Yes | Same |
| Real-Time<br>Auscultation | Yes | Yes | Yes | Same |
| Filter Selection | Yes | Yes | Yes | Same |
| Digital Signal<br>Processing<br>(DSP) | Yes | Yes | Yes | Same |
| Display | Smartphone<br>screen | Smartphone screen | LCD display of the<br>device | Same as Steth IO,<br>equivalent to<br>Littmann 3200 |
| Sound<br>Visualization | Yes | Yes | Yes | Same |
| Technical Characteristics | | | | |
| Frequency<br>Response | Stethophone v1<br>picks up and<br>amplifies the<br>sound between<br>20 and 2000 Hz.<br>Based on the<br>selected audio<br>filter, specific<br>frequency ranges<br>are further<br>emphasized:<br>Bell filter<br>emphasizes<br>range from 25 to<br>300 Hz<br>Diaphragm filter<br>emphasizes<br>frequency<br>spectrum range<br>170-850 Hz,<br>Starling filter<br>works between 60<br>-1400 Hz | Steth IO has two<br>filters: Heart and<br>Lung. The specific<br>frequency filtration<br>of these filters<br>corresponds to<br>Stethophone's Bell<br>(high-cut filter) and<br>closely to<br>Stethophone's<br>Diaphragm and<br>Starling filters. This<br>correspondence is<br>based on frequency<br>response curves. | The Bell mode<br>amplifies sounds<br>from 20 - 1000 HZ,<br>but emphasizes<br>lower frequency<br>sounds between 20<br>- 200Hz.<br>The Diaphragm<br>mode amplifies<br>sounds from 20 -<br>2000 Hz but<br>emphasizes the<br>sounds between<br>100 - 500 Hz.<br>The Extended<br>Range mode<br>amplifies sounds<br>from 20 - 2000Hz<br>similar to the<br>Diaphragm Mode<br>but provides more<br>Low-frequency<br>response between<br>50 - 500Hz. | Substantially<br>equivalent |
| Volume Control | Yes | Yes | Yes | Same |
| Battery | Uses a built-in<br>battery of a<br>smartphone | Uses a built-in<br>battery of a<br>smartphone | Uses a built-in<br>battery of a<br>chestpiece | Same as Steth IO,<br>equivalent to<br>Littmann 3200 |
| Recording<br>Lengths | 20 sec | Up to 1 min | Maximum of 29<br>seconds for on-<br>board recording, up<br>to 1 min for<br>wireless recording<br>to computer<br>software | Substantially<br>Equivalent,<br>differences don't<br>affect safety and<br>effectiveness |
| Available Sound<br>Visualization | Spectrogram,<br>oscillogram | Spectrogram,<br>oscillogram | Spectrogram,<br>oscillogram | Same |
| Ability to Zoom | Yes | Yes | Yes | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sparrow Acoustics. The word "SPARROW" is written in a light blue sans-serif font, with a stylized image of a bird in the middle of the second "R". Above the bird are three curved lines, resembling sound waves. Below the word "SPARROW" is the word "ACOUSTICS" in a smaller, sans-serif font.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sparrow Acoustics. The word "SPARROW" is written in large, light blue letters, with a stylized image of a bird inside the "O". Below the word "SPARROW", the word "ACOUSTICS" is written in smaller, light blue letters.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Sparrow Acoustics. The logo features the word "SPARROW" in a light blue sans-serif font, with the "W" replaced by a stylized image of a bird inside of three concentric circles. Below the word "SPARROW" is the word "ACOUSTICS" in a smaller, sans-serif font.
The proposed device Stethophone v1 and the predicate devices Steth IO and Littmann 3200 share the same intended use, design, user functions and fundamental scientific operational technology. The devices are functionally the same: all three devices capture, amplify the sound, use frequency filters, visualize via two types of graphs (oscillogram and spectrogram), store records and allow auscultating heart and lung sound in real time. Software comparison and their validation support the equivalency claim. Performance and usability comparison testing was performed to support the equivalence claims.
- Sparrow Acoustics Inc. submitted performance testing information in this 11. Performance Data: 510(k) to demonstrate safety and efficacy of Stethophone v1, to validate that the device meets predetermined specifications and performs according to pre-specified acceptance criteria, and to support the substantial equivalence determination.
Testing includes repeatability and reproducibility tests, performance tests using an anechoic chamber, internal tests run by a medical analysts' team, tests involving external medical specialists with auscultation experience, as well as a usability and performance study conducted by an independent medical facility.
- 12. Biocompatibility Not Applicable (Standalone Software) Testing:
- Sterilization & Shelf-Not Applicable (Standalone Software). Therefore, it is a non-sterile 13. life Testing: device and shelf-life is not applicable to this device because of low likelihood of time-dependent product degradation.
- 14. Electrical safety and Not Applicable (Standalone Software). Therefore, there is no source of electromagnetic risk for electrical safety or electromagnetic compatibility associated compatibility (EMC): directly with the device.
- 15. Animal Study: Animal performance testing was not required to demonstrate safety and effectiveness of the device.
- 16. Human Clinical Clinical testing was not required to demonstrate the safety and Performance Testing: effectiveness of the device.
- 17. Statement of Stethophone v1 and the predicate devices Steth IO and Littmann 3200 Substantial share the same intended use, design, user functions and fundamental Equivalence: scientific operational technology. The devices are functionally the same. Software comparison and their validation support the equivalency claim. The differences that exist between the devices do not affect the relative safety and/or effectiveness.
Sparrow Acoustics believes the Stethophone v1 device and its predicates Steth IQ (Model 1.0) and 3M Littmann Electronic Stethoscope, Model 3200 used in conjunction with the Littmann StethAssist phonocardiogram software, are substantially equivalent.
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.