K222869 · Andon Health Co, Ltd. · DXN · Feb 15, 2023 · Cardiovascular
Device Facts
Record ID
K222869
Device Name
Arm Blood Pressure Monitor
Applicant
Andon Health Co, Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Feb 15, 2023
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
Device Story
Fully Automatic Electronic Blood Pressure Monitor (models KD-5811A, KD-5811V, KD-5815) measures systolic/diastolic blood pressure and pulse rate; utilizes oscillometric method with silicon integrated pressure sensor; operates by inflating cuff around upper arm and measuring pressure during deflation; displays results on LCD; includes irregular heartbeat (IHB) detection and blood pressure classification indicator; intended for home or clinical use by adults; provides data for monitoring cardiovascular health; results displayed to user for clinical decision-making support.
Clinical Evidence
No new clinical testing performed. Clinical performance relies on reference device (K083246) which shares the same algorithm and design principle. Accuracy verified against ISO 81060-2 criteria.
Technological Characteristics
Oscillometric measurement; silicon integrated pressure sensor; 4x1.5V AA battery or DC mains power; LCD display; upper arm cuff (22-48cm); compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 80601-2-30.
Indications for Use
Indicated for adult individuals requiring non-invasive measurement of systolic and diastolic blood pressure and pulse rate via an upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or at home.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
February 15, 2023
Andon Health Co., Ltd. Liu Yi. President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin, Tianjin 300190 China
Re: K222869
Trade/Device Name: Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 26, 2022 Received: January 4, 2023
Dear Liu Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222869
Device Name
Fully Automatic Electronic Blood Pressure Monitor
Indications for Use (Describe)
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K222869
# 510(K) SUMMARY
(In accordance with 21 CFR 807.92)
#### 1.0 Submitter's Information
| Name: | Andon Health Co., Ltd. |
|----------------------|-----------------------------------------------------------------------------------|
| Address: | No 3, Jinping Street, Ya An Road, Nankai District, Tianjin,<br>300190, P.R. China |
| Phone Number: | 86-22-87611660 |
| Fax Number: | 86-22-87612379 |
| Contact: | Mr. Liu Yi |
| Date of Preparation: | September 15, 2022 |
#### 2.0 Device Information
| Device Name: | Fully Automatic Electronic Blood Pressure Monitor |
|----------------------|---------------------------------------------------|
| Common Name: | Arm Blood Pressure Monitor |
| Classification Name: | Non-Invasive Blood Pressure Measurement System |
#### 3.0 Classification
| Product Code: | DXN |
|--------------------|--------------------|
| Regulation Number: | 21 CFR 870.1130 |
| Classification: | II |
| Review Panel: | 870 Cardiovascular |
#### 4.0 Predicate Device Information
| Manufacturer: | Andon Health Co., Ltd. |
|----------------|---------------------------------------------------|
| Device: | Fully Automatic Electronic Blood Pressure Monitor |
| 510(k) Number: | K210770 |
| Classification | II |
| Product Code | DXN |
#### 5.0 Intended Use
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.
{4}------------------------------------------------
## 6.0 Device Description
Fully Automatic Electronic Blood Pressure Monitor ( KD-5811A, KD-5811V and KD-5815) is designed and manufactured according to IEC 80601-2-30.
The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.
# 7.0 Comparison of Technological Characteristics with Predicate Device
The following table is the summary of the technological characteristics of the proposed subject device and predicate device.
| Item | Subject Device | Predicate Device<br>(KD-5811<br>K210770) | Comparison<br>Result |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and<br>mode | Fully Automatic Electronic<br>Blood Pressure Monitor | Fully Automatic Electronic<br>Blood Pressure Monitor | -- |
| Model | KD-5811A, KD-5811V,<br>KD-5815 | KD-5811 | -- |
| Rx or OTC | OTC | OTC | Same |
| Population | Adult | Adult | Same |
| Cuff<br>Location | Upper arm | Upper arm | Same |
| Physical Attributes | | | |
| Weight<br>(exclude<br>batteries and<br>cuff) | KD-5811A: About 223g<br>KD-5811V:About 239g<br>KD-5815:About 189g | About 239g | KD-5811V is<br>same with<br>predicate.<br>KD-5811A and<br>KD-5815 are<br>changed. |
| Dimensions | KD-5811A:<br>139.4mm×93.8mm×42mm<br>KD-5811V:<br>139.4mm×93.8mm×43.4mm<br>KD-5815:<br>138mm×98mm×48.9mm | 139.4mm×93.8mm×43.4mm | KD-5811V is<br>same with<br>predicate.<br>KD-5811A and<br>KD-5815 are<br>changed. |
| Memory | KD-5811A, KD-5811V:<br>2×120 times<br>2×60 times<br>2×30 times | 2×120 times<br>2×60 times<br>2×30 times | KD-5811A and<br>KD-5811V are<br>same with<br>predicate. |
| Item | Subject Device | Predicate Device<br>(KD-5811<br>K210770) | Comparison<br>Result |
| | KD-5815:<br>1×60 times | | KD-5815 is less<br>memory times<br>than predicate.. |
| Displayed<br>Calculated<br>Parameters | SYS | SYS | Same |
| | DIA | DIA | |
| | Pulse | Pulse | |
| | IHB | IHB | |
| Display<br>component | KD-5811A, KD-5811V:<br>LCD with backlight<br><br>KD-5815: LCD | LCD with backlight | KD-5811A and<br>KD-5811V are<br>same with<br>predicate.<br>KD-5815 is<br>without<br>backlight. |
| Average<br>function | 1.Average value of all results in<br>the current user memory zone.<br>2.Average the value of the latest<br>3 times. | 1.Average value of all results<br>in the current user memory<br>zone.<br>2.Average the value of the<br>latest 3 times | Same |
| Other<br>Displayed<br>Information | KD-5811A:<br>Date<br>Time<br>Memory<br>Battery usage<br>Blood pressure classification<br>Bodymove<br><br>KD-5811V:<br>Date<br>Time<br>Memory<br>Battery usage<br>Blood pressure classification<br><br>KD-5815<br>Memory<br>Battery usage<br>Blood pressure classification | Date<br>Time<br>Memory<br>Battery usage<br>Blood pressure classification | KD-5811A is<br>more function<br>than predicate.<br><br>KD-5811V is<br>same with<br>predicate.<br><br>KD-5815 is less<br>function than<br>predicate. |
| Electrical Power | | | |
| DC Mains | 5V | 6V | Changed |
| Battery | 4 ×1.5V SIZE AA | 4 ×1.5V SIZE AA | Same |
| Item | Subject Device | Predicate Device<br>(KD-5811<br>K210770) | Comparison<br>Result |
| Environmental Operation | | | |
| Temperature | 10~40°C | 10~40°C | Same |
| Humidity | <85% | <85% | Same |
| Environmental Storage | | | |
| Temperature | -20~50°C | -20~50°C | Same |
| Humidity | <85% | <85% | Same |
| Performance NIBP | | | |
| Pulse Rate<br>Range | 40 -180times/min | 40 -180times/min | Same |
| Pulse Rate<br>Accuracy | Within ±5% | Within ±5% | Same |
| Technique/<br>Method | Oscillometric | Oscillometric | Same |
| Measure<br>process | Measure during deflating | Measure during deflating | Same |
| Pressure<br>Accuracy | Within ±3mmHg | Within ±3mmHg | Same |
| Cuff | | | |
| Pressure<br>Range | 0-300mmHg | 0-300mmHg | Same |
| Overpressure<br>Limit | 300mmHg | 300mmHg | Same |
| Algorithm | Amplitude | Amplitude | Same |
{5}------------------------------------------------
{6}------------------------------------------------
There are no significant differences between the two products and they are identical in terms of intended use, materials, design, manufacturing methods.
## 8.0 Discussion of Non-Clinical Testing
Non-clinical tests conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- 60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And . IEC A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2014,Medical Electrical Equipment -- Part 1-2: General . Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests
{7}------------------------------------------------
- IEC 60601-1-11 Edition 2.0 2015-01 Medical electrical equipment Part 1-11: . General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 80601-2-30: Edition 2.0 2018-03 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
None of the test demonstrate that the new Blood Pressure Monitors raises new questions of safety and effectiveness as compared to the predicate.
# 9.0 Clinical Test
Comparison and evaluation are carried out between the subject device and cleared devices, and it is showed that:
The proposed devices KD-5811A, KD-5811V and KD-5815 has the same algorithm and design principle with cleared device KD-5961 (K083246), so the clinical test report of KD-5961 can be used as a reference when considering the clinical effect of KD-5811A, KD-5811V and KD-5815.
Accuracy of the blood pressure monitors for the clinical test report was verified by meeting criteria 1 and criteria 2 of ISO 81060-2.
## 10. Comparison to the Predicate Device and Conclusion
The conclusion drawn from the nonclinical tests and clinical test demonstrate that the subject device Fully Automatic Electronic Blood Pressure Monitor (KD-5811A , KD-5811V, and KD-5815) is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5811 (K210770).
The subject devices are very similar with its predicate device in the intended use, design principle, materials, performance and applicable standards. Their appearance, the memory capacity,displayed information, DC mains and MCU is different.
However, the tests in this submission demonstrates that these small differences do not raise any new questions of safety and effectiveness. And the subject device is as safe and effective as the legally marketed predicate K210770.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.