Digital Radiography CXDI-Pro, Digital Radiography D1
K222855 · Canon, Inc. · MQB · Oct 17, 2022 · Radiology
Device Facts
Record ID
K222855
Device Name
Digital Radiography CXDI-Pro, Digital Radiography D1
Applicant
Canon, Inc.
Product Code
MQB · Radiology
Decision Date
Oct 17, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DIGITAL RADIOGRAPHY CXDI-Pro / DIGITAL RADIOGRAPHY D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
Solid-state X-ray imager (flat panel detector) captures X-ray photons; scintillator (CsI(Tl)) converts photons to visible light; photodetector array converts light to electrical signals; signals digitized and transmitted via wired/wireless connection to PC. Used in clinical settings by healthcare professionals for general radiography. CXDI Control Software processes signals into images for display on external monitors. Output assists clinicians in diagnostic decision-making. Benefits include digital image capture replacing traditional film/screen systems. Device series includes CXDI-703C, CXDI-803C, and CXDI-403C wireless detectors.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation activities confirmed that modifications meet applicable safety and performance standards.
Technological Characteristics
Solid-state flat panel detector using TFT technology and CsI(Tl) scintillator. Magnesium alloy case. Pixel pitch 140μm. Wireless connectivity (IEEE 802.11n/ac/a/g/b, 2.4GHz/5GHz, Bluetooth LE). IP55 rating. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-54, ISO 10993-1.
Indications for Use
Indicated for digital image capture of human anatomy for general purpose diagnostic radiographic procedures. Not for mammography. Suitable for all patient populations.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
DIGITAL RADIOGRAPHY CXDI-Pro / DIGITAL RADIOGRAPHY D1 (K221876)
Submission Summary (Full Text)
{0}------------------------------------------------
October 17, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
Canon Inc. % Ms. Saori Sawaki Business Manager, Regulatory Consultant Ken Block Consulting LLC 800 E. Campbell Road, Suite 202 RICHARDSON TX 75081
## Re: K222855
Trade/Device Name: Digital Radiography CXDI-Pro Digital Radiography D1 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: September 21, 2022 Received: September 21, 2022
Dear Ms. Sawaki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222855
Device Name DIGITAL RADIOGRAPHY CXDI-Pro DIGITAL RADIOGRAPHY D1
#### Indications for Use (Describe)
The DIGITAL RADIOGRAPHY CXDI-Pro / D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the word "Canon" in a bold, red font. The letters are closely spaced together, creating a solid block of text. The font appears to be sans-serif, with rounded edges, giving the word a modern and clean look.
# 510(k) Summary
### K222855
| Applicant/ Sponsor: Canon Inc. | | |
|--------------------------------|--------------------------------------|--|
| | 9-1 Imaikami-cho | |
| | Nakahara-ku, Kanagawa 211-8501 JAPAN | |
- Contact Person: Mr. Akira Hirai General Manager TEL: 81-3-3758-2111 FAX: 044-739-6695 hirai.akira@mail.canon
Date Prepared: October 3, 2022
Submission Type: Special 510(k) Submission
| Proposed Device | Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code:<br>Regulation:<br>Class: | Canon Inc.<br>DIGITAL RADIOGRAPHY CXDI-Pro<br>DIGITAL RADIOGRAPHY D1<br>Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Stationary X-ray System<br>MQB<br>892.1680, Stationary X-ray System<br>II |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device: | Clearance:<br>Manufacturer:<br>Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code:<br>Regulation:<br>Class: | K221876<br>Canon Inc.<br>DIGITAL RADIOGRAPHY CXDI-Pro<br>DIGITAL RADIOGRAPHY D1<br>Solid State X-ray Imager (Flat Panel/Digital Imager)<br>Stationary X-ray System<br>MQB<br>892.1680, Stationary X-ray System<br>II |
| Device<br>Description: | The DIGITAL RADIOGRAPHY CXDI-Pro, also called the DIGITAL<br>RADIOGRAPHY D1, (hereinafter referred to as CXDI-Pro) is a solid-state x-ray<br>imager. The CXDI-Pro is a series of detectors, and in the predicate submission<br>(K221876) consists of the CXDI-703C Wireless and CXDI-403C Wireless<br>detectors, also called the AR-D3543W and AR-D4343W detectors respectively.<br>The detectors intercept x-ray photons, and the scintillator emits visible spectrum<br>photons that illuminate an array of photodetectors that create electrical signals.<br>After the electrical signals are generated, the signals are converted to digital values.<br>The digital values are sent to the PC via a wired or wireless connection, converted<br>to images with the CXDI Control Software, and then displayed on the PC/monitors.<br>The PC/monitors used with the CXDI-Pro are not a part of this submission.<br>The proposed changes to the predicate device, CXDI-Pro, includes the addition of<br>the new detector, CXDI-803C Wireless (also called the AR-D2735W) to the CXDI-<br>Pro series; a firmware update from 01.01.03.00 to 01.02.00.01; and a CXDI Contro<br>Software version update from 3.10.2.2 to 3.10.2.6.<br>The new detector, CXDI-803C Wireless, which differs in pixel count, imaging area,<br>external dimensions, and weight, has the same image performance as the predicate<br>detectors. None of the CXDI-Pro detectors have any dynamic functions (such as<br>fluoroscopy). | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are closely spaced together, creating a solid block of text. The font is sans-serif, and the overall impression is one of a strong, recognizable brand logo. The red color is vibrant and eye-catching.
510(k) Summary
# K222855
| Indications for Use: | The DIGITAL RADIOGRAPHY CXDI-Pro / DIGITAL RADIOGRAPHY D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Technological Characteristics: | Comparison with the predicate device shows the technological characteristics of the CXDI-Pro is substantially equivalent to the predicate device. The flat panel detector units are functionally same, using the same components. |
| | Proposed Device | Predicate Device | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Trade Name | DIGITAL RADIOGRAPHY CXDI-Pro / D1 | DIGITAL RADIOGRAPHY CXDI-Pro / D1 | |
| 510(k) Submitter<br>[Number] | Canon Inc.<br>[K222855] | Canon Inc.<br>[K221876] | IDENTICAL |
| Indication for Use | The DIGITAL RADIOGRAPHY CXDI-Pro /<br>DIGITAL RADIOGRAPHY D1 provides<br>digital image capture for conventional<br>film/screen radiographic examinations. This<br>device is intended to capture, for display,<br>radiographic images of human anatomy, and to<br>replace radiographic film/screen systems in all<br>general purpose diagnostic procedures. This<br>device is not intended for mammography<br>applications. | The DIGITAL RADIOGRAPHY CXDI-Pro /<br>DIGITAL RADIOGRAPHY D1 provides<br>digital image capture for conventional<br>film/screen radiographic examinations. This<br>device is intended to capture, for display,<br>radiographic images of human anatomy, and to<br>replace radiographic film/screen systems in all<br>general purpose diagnostic procedures. This<br>device is not intended for mammography<br>applications. | IDENTICAL |
| Application | General Radiography | General Radiography | IDENTICAL |
| Components | Detector (Sensor)<br>Documentation (Manuals)<br>Battery Charger (BC-1A/BC-01)<br>Battery Pack (LB-4A)<br>X-ray Interface Box (XB-1A)<br>Power Box (PB-01)<br>Wiring Cable, 1.5/7.5/10/15/25m (WC-01)<br>PC Connection Cable (CP-01)<br>Ready Indicator (RI-3A)<br>Detector Stand (DS-01) | Detector (Sensor)<br>Documentation (Manuals)<br>Battery Charger (BC-1A/BC-01)<br>Battery Pack (LB-4A)<br>X-ray Interface Box (XB-1A)<br>Power Box (PB-01)<br>Wiring Cable, 1.5/7.5/15/25m (WC-01)<br>PC Connection Cable (CP-01)<br>Ready Indicator (RI-3A)<br>Detector Stand (DS-01) | MODIFIED |
| Other Compatible<br>Components | Multi Box (MB-02)<br>Status Indicator (SI-01/SI-4A) | Multi Box (MB-02)<br>Status Indicator (SI-01/SI-4A) | IDENTICAL |
| Detector Sensor | CXDI-703C Wireless (AR-D3543W)<br>CXDI-803C Wireless (AR-D2735W)<br>CXDI-403C Wireless (AR-D4343W) | CXDI-703C Wireless (AR-D3543W)<br>CXDI-803C Wireless (AR-D2735W)<br>CXDI-403C Wireless (AR-D4343W) | MODIFIED |
| External Dimensions | CXDI-703C Wireless: 384 x 460 x 15.7 mm<br>CXDI-803C Wireless: 384 x 307.5 x 15.7 mm<br>CXDI-403C Wireless: 460 x 460 x 15.7 mm | CXDI-703C Wireless: 384 x 460 x 15.7 mm<br>CXDI-403C Wireless: 460 x 460 x 15.7 mm | MODIFIED |
| Case Material | Magnesium alloy | Magnesium alloy | IDENTICAL |
| Detector Technology | TFT | TFT | IDENTICAL |
| Pixel Pitch | 140μm | 140μm | IDENTICAL |
| Scintillator | CsI(Tl)<br>[Cesium Iodide doped with Thallium] | CsI(Tl)<br>[Cesium Iodide doped with Thallium] | IDENTICAL |
| Spatial<br>Resolution | 35% [MTF@2lp/mm] | 35% [MTF@2lp/mm] | IDENTICAL |
| DQE | 58% [@0.5 lp/mm, 3.5 uGy] | 58% [@0.5 lp/mm, 3.5 uGy] | IDENTICAL |
| IP Level | IP55 | IP55 | IDENTICAL |
| Software | CXDI Control Software V3.10.2.6 | CXDI Control Software V3.10.2.2 | MODIFIED |
| Detector Firmware | V01.02.00.01 | V01.01.03.00 | MODIFIED |
| Wireless<br>Communication | IEEE 802.11n/ac/a/g/b<br>2.4GHz/5GHz<br>Bluetooth Low Energy | IEEE 802.11n/ac/a/g/b<br>2.4GHz/5GHz<br>Bluetooth Low Energy | IDENTICAL |
| Photographing Mode | Standard Synchronization Mode,<br>Non Generator Connection Mode | Standard Synchronization Mode,<br>Non Generator Connection Mode | IDENTICAL |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are closely spaced together, creating a solid block of text. The font appears to be sans-serif, with rounded edges on the letters. The overall impression is a strong and recognizable brand logo.
### 510(k) Summary
### Summary of The fundamental scientific technology of the CXDI-Pro has not been modified. Non-Clinical / The major differences between the proposed CXDI-Pro and the predicate are Test Data: differences in the new detector and a newer version of the firmware and the CXDI control software. The proposed indications for use statement is identical to the indications for use statement of the predicate device. Evaluation of the addition of the new detector and optional accessories confirmed that the changes did not impact CXDI-Pro conformance with the U.S. Performance Standard for radiographic equipment and relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-6, and 60601-2-54. These verification/validation activities successfully demonstrated that the device continues to meet the standards for the areas impacted by the modifications to the predicate device and raises no new questions regarding either safety or effectiveness when compared to the predicate device. Clinical testing is not necessary for the current submission, based on the minor differences from the predicate device. Adequate detector bench testing should be sufficient to demonstrate that the subject detector. CXDI-Pro, works as intended. Therefore, the verification/validation activities conducted support a determination of substantial equivalence for the CXDI-Pro. The proposed device follows the applicable elements of the following FDA guidance documents: Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices, Guidance for the Content of Premarket Submissions for Software contained in Medical Devices, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Radio Frequency Wireless Technology in Medical Devices, and Pediatric Information for X-ray Imaging Device Premarket Notifications. Conclusion: Canon Inc. considers the DIGITAL RADIOGRAPHY CXDI-Pro / D1 to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
K22855
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.