Nipro SafeTouch Needle; Nipro Syringe with SafeTouch Needle
K222852 · Nipro Medical Corporation · FMI · May 25, 2023 · General Hospital
Device Facts
Record ID
K222852
Device Name
Nipro SafeTouch Needle; Nipro Syringe with SafeTouch Needle
Applicant
Nipro Medical Corporation
Product Code
FMI · General Hospital
Decision Date
May 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nipro SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries. The needle is compatible with standard Luer-lock syringes. The Nipro Syringe with SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries.
Device Story
Disposable hypodermic single lumen needle with integrated mechanical safety shield; provided individually or pre-attached to Luer-lock syringe (1-10mL). Used by healthcare professionals for intramuscular, intravascular, or subcutaneous injections. Safety shield manually activated by clinician post-treatment to cover needlepoint, preventing accidental needlesticks; mechanism is irreversible once activated. Device facilitates fluid delivery/aspiration; protects clinicians from sharps injuries during handling and disposal.
Clinical Evidence
Bench testing only; no clinical trials. Simulated clinical study performed per FDA guidance for sharps injury prevention features. 500 devices evaluated by 10 trained professionals; zero device failures or user-related adverse events reported.
Technological Characteristics
Stainless steel cannula; polypropylene hub, plunger, and needle shield; polyisoprene rubber stopper. EO gas sterilization (SAL 10^-6). 5-year shelf life. Luer-lock connectivity. ISO 6009 color-coding. Standards: ISO 7886-1, ISO 7864, ISO 80369-7, ISO 23908, ISO 9626, ISO 10993-1, ASTM F1980-07, ASTM D4169-16, ASTM F1929-15.
Indications for Use
Indicated for general purpose injection of fluid from vials and to/from parts of the body below the skin surface. Intended for use by healthcare professionals. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 25, 2023
Nipro Medical Corporation Jessica Oswald-Mcleaod Director, Quality and Regulatory 3150 NW 107th Ave. Doral, Florida 33172
Re: K222852
Trade/Device Name: Nipro SafeTouch Needle; Nipro SafeTouch Needle with Syringe Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: April 25, 2023 Received: April 28, 2023
Dear Jessica Oswald-Mcleaod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature of Courtney Evans -S. The signature includes the date 2023.05.25 and the time 15:33:54 -04'00'. The signature is likely used to verify the authenticity and integrity of a digital document.
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222852
#### Device Name
Nipro SafeTouch Needle; Nipro Syringe with SafeTouch Needle
### Indications for Use (Describe)
The Nipro SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries. The needle is compatible with standard Luer-lock syringes.
The Nipro Syringe with SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue abstract symbol on the left and the word "NIPRO" in blue block letters on the right. The abstract symbol appears to be two interlocking shapes, possibly representing two links in a chain.
# K222852 510(K) Summary
Prepared on: May 23, 2023
# 1. Contact Details - 21 CFR 807.92(a)(1)
| Applicant Name: | Nipro Medical Corporation, North America division |
|-----------------------------|---------------------------------------------------|
| Applicant Address: | 3150 NW 107th Ave. Doral FL 33172 USA |
| Establishment Registration: | 1056186 |
Contact Person:
| Contact Person: | Jessica Oswald-McLeod |
|-----------------|----------------------------------|
| | Director, Quality and Regulatory |
| Phone: | 1-305-432-6699 |
| Email | JessicaO@nipromed.com |
# 2. Device Name - 21 CFR 807.92(a)(2)
| Trade Name: | Nipro SafeTouch Needle Nipro Syringe with SafeTouch Needle |
|----------------------|-------------------------------------------------------------|
| Common Name: | Disposable Syringe with Safety Needle |
| Classification Name: | Piston syringe, Hypodermic single lumen needle |
| Regulation Number: | 880.5860, 880.5570 |
| Panel: | General Hospital (80) |
| Product Code: | FMF, FMI |
| Regulatory Class: | Class 2 |
# 3. Legally Marketed Predicate Device - 21 CFR 807.92(a)(3)
K162081: BD 1mL Luer-lok Hypodermic Syringe with BD Eclipse Hypodermic Needle
# 4. Device Description Summary - 21 CFR 807.92(a)(4)
The Nipro SafeTouch Needle is a disposable hypodermic single lumen needle with attached safety device. The safety device is a needle shield that covers the needle post treatment eliminating accidental needlestick injuries. It is provided with and without a piston syringe.
The Nipro SafeTouch Needle is available in gauges of 18-30, lengths of ½". The needles are available individually or attached to a luer lock syringe in volumes of 1, 3, 5 and 10mL.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue symbol that looks like two interlocking infinity symbols. To the right of the symbol is the word "NIPRO" in blue, block letters. The background is white.
#### 4.1 Nipro Syringe with SafeTouch Needle Configurations
| Syringe Volume (mL) | Needle Gauge | Needle Length (inches) |
|---------------------|--------------|------------------------|
| 1 | 25 | 5/8 |
| | 27 | 1/2 |
| | 30 | 1/2 |
| 3 | 25 | 1/2 |
| | 23 | 1 |
| | 22 | 1 |
| | 21 | |
| | 20 | |
| | 18 | |
| 5 | 22 | |
| | 21 | 1 |
| | 20 | 1-1/2 |
| 10 | 22 | |
| | 21 | |
| | 20 | |
#### 4.2 Nipro SafeTouch Needle Configurations
| Needle Gauge | Needle Length (inches) |
|--------------|------------------------|
| 18 | |
| 19 | 1 |
| 20 | 1-1/2 |
| 21 | |
| 22 | 3/4 |
| | 1 |
| | 1-1/2 |
| 23 | 1 |
| | 1-1/2 |
| 25 | 5/8 |
| | 1 |
| | 1-1/2 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue abstract symbol on the left and the word "NIPRO" in blue on the right. The abstract symbol appears to be two interlocking shapes, possibly representing the letter "N". The word "NIPRO" is written in a bold, sans-serif font.
| Needle Gauge | Needle Length (inches) |
|--------------|------------------------|
| 26 | 1/2 |
| 27 | 1/2 |
| 27 | 1 1/4 |
| 30 | 1/2 |
The device is sterile, single use only, non-toxic, and non-pyrogenic. It is sterilized by Ethylene Oxide gas and has a shelf-life of 5 years.
The Nipro Svringe with SafeTouch Needle is identical to the reference device. K944355 Nioro Disposable Syringes and Needles, in terms of manufacturing and sterilization methods and facilities and materials of construction. This submission is to add the safety shield to the hypodermic needle and market it independently as well as attached to a Nipro luer-lock syringe.
# 5. Intended Use/Indications for Use - 21 CFR 807.92(a)(5)
The Nipro SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries. The needle is compatible with standard Luer-lock syringes.
The Nipro Syringe with SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries.
### 6. Technology Comparison - 807.92(a)(6)
The subject device is identical to the predicate device in operational mode, physical construction, and intended use. Both devices contain a mechanism that requires physical action by the clinician to cover the needlepoint after use protecting against accidental needlesticks. Once activated, the safety cannot be deactivated.
They are provided sterile with a 5-year shelf-life and are for single use only. They have patient contact and duration.
| Item for Comparison | Subject device:<br>K222852 SafeTouch<br>Needle / Syringe<br>with SafeTouch<br>Needle | Predicate device:<br>K162081 BD Luer-<br>Lok™ Hypodermic<br>Syringe with<br>BD Eclipse™<br>Hypodermic Needle | Conclusion | |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------|
| 1. Device Description | | | | |
| 1.1. Indications for<br>Use | The Nipro SafeTouch<br>Needle with syringe is<br>indicated for general<br>purpose injection and<br>aspiration of fluid to | The BD 1mL Luer-<br>Lok™ Hypodermic<br>Syringe with BD<br>Hypodermic Needle or<br>BD Eclipse™ | Similar - no new<br>issues of safety and<br>effectiveness | |
| Item for Comparison | Subject device:<br>K222852 SafeTouch<br>Needle / Syringe<br>with SafeTouch<br>Needle | Predicate device:<br>K162081 BD Luer-<br>Lok™ Hypodermic<br>Syringe with<br>BD Eclipse™<br>Hypodermic Needle | Conclusion | |
| | and from parts of the<br>body below the surface<br>of the skin. It includes<br>a needle shield to<br>prevent against<br>accidental needlestick<br>injuries. | Hypodermic Needle is<br>intended for use by<br>health care<br>professionals for<br>general purpose<br>injection and<br>aspiration of fluid from<br>vials, ampoules, and<br>parts of the body<br>below the surface of<br>the skin. The BD<br>Eclipse™ Hypodermic<br>Needle contains a<br>mechanism that covers<br>the needlepoint after<br>use. In the activated<br>position the needle<br>cover guards against<br>accidental needle<br>sticks during normal<br>handling and disposal<br>of the used needle/<br>syringe combination. | | |
| 1.2. Mechanism of<br>Action | Mechanical/Manual | Mechanical/Manual | Same | |
| 1.3. Technology<br>Overview | Graduated disposable<br>piston syringe provided<br>with a safety<br>hypodermic needle<br>consisting of a<br>Stainless-Steel cannula<br>with plastic hub and<br>needle shield. | Graduated disposable<br>piston syringe<br>provided with a safety<br>hypodermic needle<br>consisting of a<br>Stainless-Steel cannula<br>with plastic hub and<br>needle shield. | Same | |
| 1.4. Anatomical<br>Location | intramuscular,<br>intravascular, or<br>subcutaneous injection | intramuscular,<br>intravascular, or<br>subcutaneous injection | Same | |
| Item for Comparison | Subject device:<br>K222852 SafeTouch<br>Needle / Syringe<br>with SafeTouch<br>Needle | Predicate device:<br>K162081 BD Luer-<br>Lok™ Hypodermic<br>Syringe with<br>BD Eclipse™<br>Hypodermic Needle | Conclusion | |
| 2.1. Syringe barrel | Polypropylene | Styrene acrylic<br>copolymer | Similar - no new<br>issues of safety and<br>effectiveness | |
| 2.2. Lubricant | Silicone | Silicone | Same | |
| 2.3. Plunger | Polypropylene | Polypropylene | Same | |
| 2.4. Stopper | Polyisoprene Rubber | Polyisoprene Rubber | Same | |
| 2.5. Needle Cannula | Stainless Steel | Stainless Steel | Same | |
| 2.6. Needle Hub with<br>integrated safety<br>mechanism<br>(needle shield) | Polypropylene | Polypropylene | Same | |
| 3. Physical Characteristics | | | | |
| 3.1. Needle Gauge | 18 - 30 | 20 - 30 | Nipro provides a larger<br>gauge needle. Testing<br>results demonstrate no<br>new issues of safety<br>and effectiveness | |
| 3.2. Needle hub Color | The needle hub is<br>color-coded to the<br>appropriate gauge<br>needle per ISO 6009 | The needle hub is<br>color-coded to the<br>appropriate gauge<br>needle per ISO 6009 | Same | |
| 3.3. Needle Length | 1/2" - 1 1/2" | 1/2" - 1 | Nipro provides a longer<br>length needle. Testing<br>results demonstrate no<br>new issues of safety<br>and effectiveness | |
| 3.4. Syringe Volume | 1 - 10 mL | 1-10mL | Same | |
| 4. Sterility | | | | |
| 4.1. Method | EO | Gamma | Both methods are<br>recognized as Category<br>A. | |
| 4.2. Single-Use | yes | yes | Same | |
| 4.3. SAL | 10-6 | 10-6 | Same | |
| | Item for Comparison | Subject device:<br>K222852 SafeTouch<br>Needle / Syringe<br>with SafeTouch<br>Needle | Predicate device:<br>K162081 BD Luer-<br>Lok™ Hypodermic<br>Syringe with<br>BD Eclipse™<br>Hypodermic Needle | Conclusion |
| | 4.4. Shelf Life | 5 years | 5 years | Same |
| 5. | Biological Specifications | | | |
| 5.1. | Systemic<br>Injection | No abnormalities or<br>mice death | No abnormalities or<br>mice death | Within Criteria; Same |
| | 5.2. Intracutaneous<br>reaction test | No erythema, edema,<br>or necrosis | No erythema, edema,<br>or necrosis | Within Criteria; Same |
| | 5.3. Hemolysis test | Hemolytic index shall<br>not exceed 2% | Hemolytic index shall<br>not exceed 2% | Within Criteria; Same |
| | 5.4. Bacterial<br>Endotoxin | <= 20 EU/device | <= 20 EU/device | Within Criteria; Same |
| 5.5. | Microbial barrier<br>test | No microbial growth<br>was observed | No microbial growth<br>was observed | Within Criteria; Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue symbol that looks like two interlocking loops, followed by the word "NIPRO" in blue, block letters. The symbol is on the left, and the word is on the right.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue symbol that looks like two interlocking letter "C" shapes. To the right of the symbol is the word "NIPRO" in blue, block letters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue abstract symbol on the left and the word "NIPRO" in blue capital letters on the right. The abstract symbol appears to be two interlocking shapes, possibly representing two links in a chain.
# 7. Non-Clinical and/or Clinical Tests Summary and Conclusions - 21 CFR 807.92(b)
The submitted performance testing is representative of the functional testing that is performed at final inspection and testing during routine manufacturing. Bench testing was performed on the subject device and the predicate device. The following standards were used. No deviations were noted.
- a. Visual Inspection (ISO 7886-1, ISO 7864, ISO 6009)
- b. Dimensional Specifications (ISO 7886-1, ISO 7864, ISO 80369-7)
- C. Mechanical and Performance Characteristic (80369-7, ISO 23908, ISO 9626, ISO 7886-1, ISO 7864, ISO 9626)
- d. Chemical Specifications (ISO 7886-1, ISO 7864)
- Sterilization (ISO 11607-1, ISO 11607-2, ISO 11135, USP <71>, USP <55>) e.
- Shelf-life testing (ASTM F1980-07) f.
- g. Packaging and Transportation tests (ASTM D4169-16)
- h. Dye Penetration packaging tests (ASTM F1929-15)
- Biocompatibility and Chemical Characterization (ISO 10993-1) i.
- Cytotoxicity: MEM Elution (ISO 10993-5) 트
- . Systemic Toxicity: Systemic Injection (ISO 10993-11)
- . Sensitization: Guinea Pig Maximization (ISO 10993-10)
- Irritation: Intracutaneous Reactivity (ISO 10993-23) "
- 트 Pyrogenicity:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Nipro. The logo consists of a blue abstract symbol on the left and the word "NIPRO" in blue capital letters on the right. The abstract symbol appears to be two interlocking shapes, possibly representing two links in a chain.
- Bacterial Endotoxin (BET) (ANSI/AAMI ST72:2011/(R)2016, USP-NF M98910 01 01 -<161>, USP-NF M98830 02 01 <85>,)
- Material mediated (USP-NF<151>) -
- I Hemocompatibility: Hemolysis (indirect) and Direct contact (ISO 10993-4 and ASTM F756)
- . Sterilization Residuals: EO/ECH Residuals (ISO 10993-7)
- . Particulates Testing: Sizing and counting Particulates: Light Obscuration Method (USP <788>)
A simulated clinical study was performed according to FDA Guidance, Medical Devices with Sharps Injury Prevention Features. Five hundred (500) subject devices were evaluated across 10 trained professionals. There were no device failures, no user-related adverse events or safety device related adverse events.
Internal verification and validation testing confirms that the SafeTouch Needle with or without Syringe meets specifications equivalent in design and technological characteristics to the predicate device.
# 8. Conclusion Statement
The SafeTouch Needle with or without Syringe complies with all applicable voluntary consensus for performance, biocompatibility, packaging, transportation. Through functional performance testing the subject devices have demonstrated substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.