K222832 · Endomagnetics Ltd., · NEU · Jan 21, 2023 · General, Plastic Surgery
Device Facts
Record ID
K222832
Device Name
Sentimag System
Applicant
Endomagnetics Ltd.,
Product Code
NEU · General, Plastic Surgery
Decision Date
Jan 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The Sentimag system is intended for intra-operative use, to help surgeons detect and locate magnetic marker in the body. The system is specifically intended to detect Magseed® implantable magnetic marker used to mark soft tissue lesions. The Endomag Magseed® Magnetic Marker (K173587) is placed within the target soft tissue prior to planned surgical removal. The currently cleared marker (K173587), when used in conjunction with the Sentimag Gen 3 System in Counts mode, can be used as a guide for the surgeon to follow in the excision of tissue.
Device Story
Sentimag Gen 3 is a non-imaging magnetic localization system for intra-operative detection of implanted Magseed magnetic markers. System comprises a mains-powered base unit, a hand-held probe (standard or thin), and a footswitch. Operation involves the probe emitting an alternating magnetic field to induce a magnetic response in the implanted marker; the base unit detects this response via susceptometry. Output is provided as a numerical 'counts' value on a display and an audible signal with varying pitch. Used in operating rooms by surgeons for breast lumpectomy or other soft tissue lesion excision. The device features a 'Counts' mode (equivalent to Gen 2 predicate) and a 'Measure' mode (currently dormant). Risk controls, including visual display differences and training, prevent accidental use of the dormant mode. The system aids surgeons in precisely locating non-palpable lesions, facilitating accurate excision and potentially improving surgical outcomes.
Clinical Evidence
Bench testing only. Testing included usability studies, sensitivity comparisons between Gen 3 (standard/thin probes) and Gen 2, and simulated use in various tissue models (liver, thyroid, breast, abdominal, lung) to confirm performance in 'Counts' mode.
Technological Characteristics
Mains-powered base unit with LCD; detachable hand-held probe (13.5mm or 18.5mm diameter) and footswitch. Sensing principle: magnetic susceptometry (alternating magnetic field emission). Connectivity: wired probe/footswitch. Software: embedded, includes dormant 'Measure' mode. Sterilization: single-use sterile sheath required for probe.
Indications for Use
Indicated for assisting in the localization of soft tissue lesions in patients undergoing surgical removal of such lesions.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
1/21/2023
Endomagnetics Ltd Mehryar Behizad Regulatory Director 330 Cambridge Science Park, Milton Road Cambridge, Cambridgeshire CB4 0WN United Kingdom
# Re: K222832
Trade/Device Name: Sentimag System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: December 19, 2022 Received: December 20, 2022
# Dear Mehryar Behizad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Deborah A. Fellhauer -S
Deborah Fellhauer RN. BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222832
Device Name Sentimag System
Indications for Use (Describe)
The Sentimag Magnetic Localization System when used with the Magseed® marker is indicated to assist in localizing soft tissue lesions.
The Endomag Sentimag System is the only non-imaging guidance system intended for use with the Magsed® magnetic marker.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTER INFORMATION
| Submitter's Name:<br>Address: | Endomagnetics Ltd.<br>330 Cambridge Science Park, Milton Road<br>Cambridge<br>CB4 0WN |
|-------------------------------|---------------------------------------------------------------------------------------|
| Contact Person: | Mehryar Behizad PhD |
| | Regulatory Director |
| Tel: | +44 1223 652540 |
| Email: | mbehizad@endomag.com |
| Date summary prepared: | 25th August 2022 |
# DEVICE INFORMATION
| Trade name: | Sentimag System |
|------------------------|----------------------------------|
| Common name: | Gen 3 |
| Classification name: | Marker, Radiographic Implantable |
| Regulation: | 21 CFR 878.4300 |
| Device Classification: | Class II |
| Product Code: | NEU |
# PREDICATE DEVICE
- Principal Predicate K153044 Sentimark Magnetic Marker System
- Secondary Predicate K173587 Magseed® Magnetic Marker System
-Tertiary Predicate is Special 510(k) K163541 Magseed® Magnetic Marker System
The rationale for the ranking of the predicates is as follows:
The Principal Predicate is K153044. This Traditional 510(k) submission describes the SentiMark magnetic marker in conjunction with Sentimag. This system is used to mark and locate soft tissue for surgical removal. It describes the indication and the magnetic mechanism of detection and that the Sentimag detector is the unit that induces and detects the magnetic response in the Magseed magnetic marker. Magseed marker in combination with Sentimag (Gen 2) has been legally marketed in the United States since its 510(k) clearance on 2nd March 2016.
{4}------------------------------------------------
The Secondary Predicate is K173587. This has been identified as the second most important predicate because this is a derivative predicate that describes the use of the same system (Sentimag/Magseed). However, in this predicate, data for implantation of Magseed for more than 30 days was presented, reviewed and cleared by the FDA (16 February 2018).
The target tissue, magnetic mode of detection and materials of construction are exactly the same as the primary predicate. The overall indication (marking, detection and removal of soft tissue) remained the same as the Principal Predicate, but the extended duration to more than 30 days for the Magseed marker implantation was approved. This has no bearing on the Sentimag Gen 2 detector.
The Tertiary Predicate is Special 510(k) K163541. This Premarket Submission is solely concerned with the change of name for the marker from SentiMark to Magseed. There is no revision to the implanted marker and it has no bearing on the Sentimag Gen 2 detector. This was cleared on 13th January 2017.
# DESCRIPTION OF DEVICE
The Sentimag Gen 3 System is the 3d generation of Sentimag detector, intended for the nonimaging detection and localization of Magseed® magnetic marker that has been implanted in a lumpectomy site intended for surgical removal. It consists of a base unit and a probe.
This generation of the device has two modes:
- a) Counts Mode: This mode is the exact equivalent of the current generation of the Sentimag detector (Sentimag Gen 2 - described in K153044). In this mode Magseed (K173587) and is detectable based on a counts signal displayed on the base unit of Sentimag device.
# b) Measure Mode: Please note that this mode has a different graphic display and is currently dormant in Sentimag Gen 3
Risk control measures in relation to users wrondly using Sentimag Gen 3 in Measure mode are outlined in Risk Management File: Use risk Analysis Appendix 1, line U 6.4.
Briefly these are:
- a) Magseed does not provide a clinically usable signal with Measure mode,
- b) The displays from the two modes are visually clearly distinguishable. Counts mode looks very similar to the Gen2 display with yellow graphics, while Measure mode is graphically very different and in a grey color (initially) as shown from the screen shots of the initial 'ready to use' display below:
Image /page/4/Figure/13 description: The image shows three different modes of a device. The first mode is "Gen 2" and displays a value of "0" along with a speaker icon and a probe icon. The second mode is "Gen3 Counts mode" and also displays a value of "0" along with a graph. The third mode is "Gen3 Measure mode" and displays a value of ">40mm" along with a visual representation of distance.
Initial 'Ready to use' screenshots from the Gen2 and Gen3 Sentimag displays
{5}------------------------------------------------
- c) The Gen 3 system remembers which mode was previously selected when it is next used, so once set in Counts mode, users would not have any need to change modes.
- d) Specific training will be provided to ensure end users use Sentimag Gen 3 system in the correct mode.
- e) The IFU will provide explicit direction in relation to not using the Measure mode in addition to the above.
Note: Other components of risk management file are enclosed as part of this submission.
The probability of end users utilising Sentimag Gen 3 in the inappropriate mode has been minimised through a series of risk control measures outlined in Use risk analysis, Appendix 1, lines U 6.4.
Whilst we recognize that there is the possibility that end users might accidentally switch to measure mode, we believe the visual difference, lack of signal and suitable training, would ensure the use of the appropriate mode. End users are trained to seek a signal before commencing surgery, so as to make sure the system is fully operational. Existing end users are familiar with counts mode through their use with Gen 2 and new users would be trained to use only this mode for the Magseed, and provided with instruction via the IFU.
The predicate for the Sentimaq Gen 3 device is the previous generation of the detector and was introduced in previous submissions to the FDA as Sentimaq but in fact was the second generation in this family of devices. So. throughout this submission in order to distinguish between the predicate detector and the detector which is the subject of this submission the predicate will be referred to as "Sentimag Gen 2" and the detector in this submission will be referred to as "Sentimag Gen 3".
Sentimag Gen 3 also supports a standard diameter and a thin probe. The functional performance of the two different probes is equivalent in the two different Gen 3 modes. In Counts mode both the probes are functionally equivalent to Sentimag (Gen 2) predicate.
The Sentimag System is designed for use in an operating room environment by suitably trained physicians who are experienced in diagnosis and treatment of breast lesions. The Sentimag System aids the surgeon to detect surgically invasively magnetic marker material that has been placed for the purpose of detecting a non-palpable lesion, and to locate target excision sites.
Prior to a lumpectomy procedure, the Magseed® Magnetic Marker which already has clearance (K173587) is placed percutaneously into the breast, using imaging guidance such as ultrasound or radiography, to temporarily mark a site intended for surgical removal. During a surgical procedure, the hand-held Sentimag probe emits an alternating magnetic field that detects the magnetic response of the Magseed® magnetic marker, this signal is converted by the base unit into a visual and audible response that is similar in use to the predicate device.
The Sentimag Gen 3 device is the successor to the Sentimag Gen 2 device (Principal Predicate K153044. Secondary Predicate K173587 and Tertiary Predicate 510(k) K163541). In common with the principal predicate, Endomag Sentimag Gen 3 device is able to detect the Magseed Magnetic Marker using proximity detection based on susceptometry.
{6}------------------------------------------------
# INTENDED USE
The Sentimag system is intended for intra-operative use, to help surgeons detect and locate magnetic marker in the body. The system is specifically intended to detect Magseed® implantable magnetic marker used to mark soft tissue lesions.
The Endomag Magseed® Magnetic Marker (K173587) is placed within the target soft tissue prior to planned surgical removal. The currently cleared marker (K173587), when used in conjunction with the Sentimag Gen 3 System in Counts mode, can be used as a guide for the surgeon to follow in the excision of tissue.
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
As the predicate Sentimag Gen 2 detector (K153044), the Sentimag Gen 3 System utilizes nonimaging technologies that are comprised of a console that incorporates electronics and a simple user interface, plus a probe handpiece. Like the predicate (Sentimag Gen 2), Sentimag Gen 3 is a proximity detector. The Sentimag Gen3 System comprises of a mains-powered base unit, a detachable hand-held probe available in 2 sizes (Standard Probe and Thinner Probe) that is connected to the base unit with a flexible cable, and a detachable electronic footswitch that is connected to the base unit with a flexible cable. A single use sterile sheath is required to be placed over the probe prior to the procedure, this is to be supplied by the hospital.
The hand-held probe which emits an alternating magnetic field that detects the magnetic response of the marker. When the marker is exposed to a magnetic field it becomes magnetic, and this magnetism can then be detected using the probe which gives an audible and visual signal of the level of magnetism. In count mode this is in the format of a numerical reading. In this mode the Sentimag Gen 3 device can measure Magseed® in exactly the same way as the predicate (Sentimag Gen 2).
A location marker is placed percutaneously in situ at the clinical target site by a delivery system using 5 cm to 12 cm delivery systems described in predicates K173587 and K163541 and then the detector handpiece is used for the intraoperative detection and localization of the implanted marker. The handpiece is connected by a flexible cable to a console unit that provides the user with a visual indication of the presence and proximity of the marker.
In Counts mode Sentimag Gen 3 can detect Magseed® (K173587) in exactly the same way and with exactly the same sensitivity as the predicate (Sentimag Gen 2 - described in K153044). In common with the Sentimag Gen 2 predicate, in Sentimag Gen 3 in Count mode a visible numerical representation of the detected signal level is simultaneously displayed on the base unit's liquid crystal display.
# Additionally, Sentimag Gen 3 supports a Measure Mode. Please note that this mode is currently dormant in Sentimag Gen 3.
In both Count mode and Measure mode the sensing of the magnetic signal is indicated by a change in pitch (frequency) of an audio output from the Sentimag base unit.
The Gen 3 Sentimaq also provides an option for a thinner (13.5 mm diameter) probe. This probe functions in exactly the same way as a normal diameter (18.5 mm) probe in each mode described above.
{7}------------------------------------------------
# DISCUSSION OF NON-CLINICAL TESTS SUBMITTED
Performance testing was conducted to evaluate and characterize the performance of the Sentimag System (Gen 3) with the previously approved Magseed®(K173587) Preclinical/Laboratory testing included:
- Usability testing of Sentimag (Gen 3)- Standard and thin probe .
- Sensitivity testing of Sentimag (Gen 3) standard and thin probes against Magseed . (K173587).
- . Comparison of Sentimag (Gen 3) and Sentimag (Gen 2) sensitivity in counts mode
- Simulated Use of Sentimag (Gen 3) with Magseed (K173587) in Counts mode using the . following simulated tissue:
- 1. Liver
- 2. Thyroid (CIRS Model 074)
- 3. Breast (CIRS Model 073)
- 4. Abdominal Tissue Plate (SynDaver 141720, 2N Std, 5mm Fat, 5mm muscle)
- 5. Lung Model (SynDaver)
# CONCLUSION
Endomagnetics believes that Sentimag System (K153044) is the predicate device because the Sentimag (Gen 3) is the same as the predicate Sentimag in Counts mode. The Measure mode in Sentimag Gen 3 is an additional function. The Sentimag Gen3 Measure mode is currently dormant.
Specific risk control measures have been adopted to prevent the end user from using this mode. These risk control measures are outlined in section 5.4 above. Sentimag Gen 3 has the same Intended Use and very similar technological characteristics to the predicate device.
The different technological characteristics do not raise any new questions of safety or effectiveness. The test, verification and validation data presented in this submission demonstrate substantial equivalence of the Sentimag Gen 3 to the predicate (Sentimag Gen 2).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.