VentStar Helix dual heated (N) Plus (MP02650), VentStar Helix heated (N) Plus (MP02608)
K222822 · Draegerwerk AG & CO Kgaa · BZE · Jun 14, 2023 · Anesthesiology
Device Facts
Record ID
K222822
Device Name
VentStar Helix dual heated (N) Plus (MP02650), VentStar Helix heated (N) Plus (MP02608)
Applicant
Draegerwerk AG & CO Kgaa
Product Code
BZE · Anesthesiology
Decision Date
Jun 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
VentStar Helix dual heated N Exten. Kit (MP02280): Inspiratory and expiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient. VentStar Helix heated (N) Plus (MP02608): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.
Device Story
VentStar Helix devices are disposable, heated breathing circuits designed for neonatal patients (tidal volume ≤100 mL). The system connects to a Fisher & Paykel MR850 humidifier to deliver humidified gas to the patient. The VentStar Helix Dual Heated N Exten. Kit provides inspiratory and expiratory heating, while the VentStar Helix Heated (N) Plus provides inspiratory heating. Used in hospital settings or during intrahospital transport, these circuits are compatible with Dräger Babylog VN-Series ventilators. The device functions as a conduit for breathing gas; heating wires integrated into the circuit walls maintain gas temperature and humidity levels to prevent condensation. Healthcare providers connect the circuit to the humidifier and ventilator; the system operates passively to maintain gas conditions. Benefits include consistent humidification and temperature control for neonates, reducing the risk of airway drying or thermal injury. The device is not for use during MRI or electrocautery.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including mechanical, thermal safety, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-1, ISO 18562).
Technological Characteristics
Disposable breathing circuits with integrated heating wires. Materials evaluated per ISO 10993-1 and ISO 18562. Connectors comply with ISO 5356-1. Electrical: 22V, 2.73A, 60W. Connectivity: Physical connection to MR850 humidifier and Dräger Babylog VN-Series ventilators. Sterilization: Not specified (disposable).
Indications for Use
Indicated for neonatal patients with tidal volume up to 100 mL requiring humidified breathing gas delivery via connection to a Fisher & Paykel MR850 humidifier. Contraindicated for use during MRI or electrocautery.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
VentStar heated (N) (MP00308), Drägerwerk AG & Co. KGaA (K102618)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 14, 2023
Dräegerwerk AG & Co. KGaA Jan Upmeier Regulatory Affairs Manager Moislinger Allee 53-55 Luebeck, Schleswig-Holstein 23542 Germany
Re: K222822
Trade/Device Name: VentStar Helix dual heated N Exten. Kit (MP02280), VentStar Helix heated (N) Plus (MP02608) Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing system heater Regulatory Class: Class II Product Code: BZE Dated: May 17, 2023 Received: May 17, 2023
Dear Jan Upmeier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ethan L. Nyberg -S
for James J. Lee. Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22822
#### Device Name
VentStar Helix Dual heated N Exten. Kit / VentStar Helix heated (N) Plus
#### Indications for Use (Describe)
VentStar Helix dual heated N Exten. Kit (MP02280): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.
VentStar Helix heated (N) Plus (MP02608): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
# 510(k) Premarket Notification Summary
| Submitter: | Drägerwerk AG & Co. KGaA<br>Moislinger Allee 53-55<br>23542 Lübeck, Germany<br>Establishment's registration number: 9611500 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jan Upmeier<br>Regulatory Affairs Manager<br>E-Mail: jan.upmeier@draeger.com<br>Telephone: +49 451 882 8474 |
| Applicant's US Contact Person: | Darlene Thibodeau<br>Head of Quality Assurance and Compliance<br>E-Mail: darlene.thibodeau@draeger.com<br>Telephone: 978 773-1060 |
| Date prepared: | June 14th, 2023 |
| Device Name: | Trade Name: VentStar Helix Dual<br>Heated N Exten. Kit / VentStar Helix Heated (N) Plus<br>Classification Name: Breathing system heater<br>Regulation Number: 21 CFR 868.5270<br>Product Code: BZE<br>Class: II |
| Predicate Device: | Primary predicate for VentStar Helix Dual Heated N<br>Exten. Kit: RT265 Dual Heated Infant Breathing Circuit,<br>Fisher & Paykel HealthCare, cleared via K103767;<br>Secondary predicate for VentStar Helix Heated (N) Plus<br>MP00308 VentStar heated (N), Drägerwerk AG & Co.<br>KGaA, cleared via K103618 |
{4}------------------------------------------------
510(k) Summary
K222822
# Device Description
The subject devices (VentStar Helix Dual Heated N Exten. Kit and VentStar Helix Heated (N) Plus), two inspiratory (and expiratory) heated disposable breathing circuits with humidifier chamber are designed for neonatal patients with a tidal volume of up to 100 mL, for conduction of humidified breathing gas from the humidifier to the patient. VentStar Helix Dual Heated N Exten. Kit is a dual heated breathing circuit and VentStar Helix Heated (N) Plus is a single heated breathing circuit to be connected to a humidifier MR850 by Fisher and Paykel, both tested for system compatibility and released for use with Dräger Babylog VN-Series ventilators.
Additionally, this submission comprises a neonatal humidifier chamber and a double connector for connection to the ventilators as accessories to the above mentioned medical devices.
# Indications for Use
VentStar Helix dual heated N Exten. Kit (MP02280): Inspiratory and expiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.
VentStar Helix heated (N) Plus (MP02608): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.
# Environments of use
The device is intended for stationary use in hospitals and medical rooms or for intrahospital patient transport.
Do not use the device in the following environments of use:
- -Magnetic resonance imaging
- -During electrocautery
### Comparison to Predicate
Drägerwerk AG & Co KGaA concludes that VentStar Helix Dual Heated N Exten. Kit (MP02280) in Table 1 is substantial equivalent to the Fisher & Paykel's product REF RT265 (K103767) and VentStar Helix heated (N) Plus (MP02608) in Table 2 is substantial equivalent to the Drägerwerk AG & Co KGaA's VentStar heated (N) REF MP00308 (K102618).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue.
510(k) Summary
K222822
| Table 1: Comparison Table VentStar Helix Dual Heated N Exten. Kit (MP02280) | | | |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Substantial Equivalence<br>Discussion | Subject of Submission | Primary Predicate Device | Comments |
| Product Name | MP02280 | REF RT265 | |
| 510(k) Number | K222822 | K103767 | |
| Indications for use | VentStar Helix dual heated N Exten. Kit<br>(MP02280): Inspiratory and expiratory heated<br>disposable breathing circuit with humidifier<br>chamber for connection to a humidifier MR850 by<br>Fisher & Paykel, for neonatal patients with a tidal<br>volume of up to 100 mL, for conducting humidified<br>breathing gas from humidifier to patient. | The dual-heated breathing circuits are<br>intended as conduits of breathing gas for<br>ventilation of infant patients, and to maintain<br>the temperature of humidified inspired gas.<br>The RT265 is used for flow rates greater than<br>4 L/min, and the RT266 is for flow rates<br>between 0.3 and 4 L/min. | Similar |
| Description | Dual-heated breathing circuits with humidifier<br>chamber | Dual-heated breathing circuits with humidifier<br>chamber | Same |
| Heater | for connection to a Fisher & Paykel humidifier<br>MR850 | for connection to a Fisher & Paykel humidifier<br>MR850 | Same |
| Patient Population | Neonatal patients with a tidal volume of up to 100<br>mL and a weight up to 8 kg. | Infants, tidal volume <185 mL | Similar |
| Application | Disposable | disposable | Same |
| Hose diameter<br>ID = Inner Diameter<br>OD = Outer Diameter | 11 mm ID | 10 mm ID | Different |
| Color | Inspiratory limb: blue spiral on colorless hose<br>Expiratory limb: white spiral on colorless hose | Inspiratory limb: blue-transparent<br>Expiratory limb: white | Similar |
| Heating Wire | Inspiratory limb: in the wall of the circuit<br>Expiratory limb: in the wall of the circuit | Inspiratory limb: Spiral wire inside the circuit<br>Expiratory limb: Spiral wire inside the circuit | Different |
| Connectors | ISO 5356-1 Conical Connectors | ISO 5356-1 Conical Connectors | Same |
| Patient Connector | Y-piece with ISO cone 15 mm female according<br>to ISO 5356-1 | Y-piece with ISO cone 15 mm female<br>according to ISO 5356-1 | Same |
| Device Connector | 11.5 mm female connector (ID) according to ISO<br>5356-1 | 15 mm male (OD)/11.5 mm (ID) female<br>connector according to ISO 5356-1 | Same |
| Traditional 510(k) | 510(k) Summary | | K22282 |
| Substantial Equivalence<br>Discussion | Subject of Submission | Primary Predicate Device | Comments |
| Product Name | MP02280 | REF RT265 | - |
| Length<br>Inspiratory Line | 1.70 m ±10% (66,9 inch ±10%), including 40 cm<br>(15.74 in) hose extension | 1.60 m (59 inch), including 40 cm (15.74 in)<br>hose extension | Different |
| Expiratory Limb | 1.63 m ± 80 mm (64.17 ±3.15 inch) | 1.60 m (63 inch) | Different |
| Humidifier Connection | 0.50 m ± 40 mm (19.7 ±1.58 inch) | 0.70 m (27.6 inch) | Different |
| Resistance<br>Inspiratory: | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 1.0 mbar<br>at 30 L/min < 3.0 mbar | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.4 mbar<br>at 15 L/min < 1.3 mbar<br>at 30 L/min < 9.6 mbar | Different |
| Expiratory: | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 0.7 mbar<br>at 30 L/min < 2.4 mbar | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 1.4 mbar<br>at 30 L/min < 5.3 mbar | Different |
| Compliance | at 60 mbar (breathing circuit: inspiration and<br>expiration) < 0.8 mL/hPa (mL/mbar) | 1.27 mL/mbar (1.3 cmH2O) | Different |
| Leakage | at 20 mbar < 15 mL/min<br>at 40 mbar < 30 mL/min<br>at 60 mbar < 30 mL/min | @ 60 mbar ≤ 75 mL/min | Different |
| Volume of breathing<br>circuit/ humidifier chamber | Inspiratory volume of the breathing circuit with<br>incubator extension 360 ml ±10%<br>Volume (air) of humidifier chamber with water 210<br>ml ±10%<br>Volume (air) of humidifier chamber without water<br>300 ml ±10% | Inspiratory volume of the breathing circuit<br>with incubator extension 450 ml<br>Volume (air) of humidifier chamber with water<br>190 ml<br>Volume (air) of humidifier chamber without<br>water 350 ml | Different |
| Warm up Time | 30 min | 30 min | Same |
| Substantial Equivalence<br>Discussion | Subject of Submission | Primary Predicate Device | Comment |
| Product Name | MP02280 | REF RT265 | - |
| Humidification Output | Invasive ventilation at 2 to 60 L/min > 33 mg/L | Invasive ventilation up to 60 L/min > 33 mg/L | Same |
| | Non-invasive ventilation at 2 to 60 L/min > 12<br>mg/L | Non-invasive ventilation up to 120 L/min > 10<br>mg/L | Different |
| Breathing gas temperature<br>at Y-Piece | Adjusted and controlled by Fisher & Paykel<br>MR850 humidifier | Adjusted and controlled by Fisher & Paykel<br>MR850 humidifier | Same |
| Electrical Connection Data | 22 V, 2,73 A, 60 W<br>14 Ω inspiratory<br>12 Ω expiratory (MP02650) | 22 V, 2.73 A, 60 W, 50/60 Hz | Same |
| Ventilation modes | VentStar Helix dual heated (N) Plus is suitable for<br>standard ventilation modes, such as<br>Synchronized Intermittend Mandatory Ventilation<br>(SIMV), Mandatory Minute Ventilation (MMV), and<br>Continuous Positive Airway Pressure (CPAP) as<br>well as for High Frequency Oscillation (HFO) | No information is provided in the IFU, suitable<br>for delivery of breathing gas. | Similar |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is in blue, and the two dots above the "a" are also blue. The word is centered in the image and is the only element present.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is in blue color. The two dots above the letter "a" are also in blue color. The background is white.
510(k) Summary
K222822
#### Table 2: Comparison Table VentStar Helix heated (N) Plus (MP02608)
| Substantial Equivalence<br>Discussion | Subject of Submission | Reference Predicate Device | Comments |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Name | MP02608 | MP00308 | - |
| 510(k) Number | K222822 | K102618 | - |
| Indications for use | VentStar Helix heated (N) Plus (MP02608):<br>Inspiratory heated disposable breathing circuit<br>with humidifier chamber for connection to a<br>humidifier MR850 by Fisher & Paykel, for<br>neonatal patients with a tidal volume of up to 100<br>mL, for conducting humidified breathing gas from<br>humidifier to patient. | VentStar heated (N) (MP00308): Disposable<br>inspiratory heated breathing circuit for<br>connection to a Fisher & Paykel MR850<br>humidifier, for conveying moistened breathing<br>gas between the humidifier and neonatal<br>patients with a body weight of up to 5 kg (11<br>lbs). | Similar |
| Substantial Equivalence<br>Discussion | Subject of Submission | Reference Predicate Device | Comments |
| Product Name | MP02608 | MP00308 | - |
| Description | Single-heated breathing circuits with humidifier<br>chamber | Single-heated breathing circuits with<br>humidifier chamber | Same |
| Heater | For connection to a Fisher & Paykel humidifier<br>MR850 | For connection to a Fisher & Paykel<br>humidifier MR850 | Same |
| Patient Population | Neonatal patients | Neonatal patients | Same |
| Application | Disposable | Disposable | Same |
| Hose diameter (Inner<br>Diameter) | 11 mm ID | Heated hose: 14 mm ID<br>Non-heated hose: 10 mm ID | Different |
| Color | Inspiratory limb: blue spiral on colorless hose<br>Expiratory limb: white spiral on colorless hose<br>with water trap | Inspiratory limb: transparent with white<br>heating wire inside the limp<br>Expiratory limb: transparent | Different |
| Heating Wire | Inspiratory limb: in the wall of the circuit<br>Expiratory limb: none | Inspiratory limb: white heating wire inside the<br>limp<br>Expiratory limb: none | Different |
| Connectors | ISO 5356-1 Conical Connectors | ISO 5356-1 Conical Connectors | Same |
| Patient Connector | Y-piece with ISO cone 15 mm female according<br>to ISO 5356-1 | Y-piece with ISO cone 15 mm female<br>according to ISO 5356-1 | Same |
| Device Connector | 11.5 mm female (ID) connector according to ISO<br>5356-1 | 15 mm male (OD)/11.5 mm female (ID)<br>connector according to ISO 5356-1 | Same |
| Length | Inspiratory Limb 1.70 m ±10% (66,9 inch ±10%), including 40 cm<br>(15.74 in) hose extension | 1.2 m ±10% (47 inch ±10%) including 30 cm<br>(11.81 inch ±10%) hose extension | Different |
| | Expiratory Limb 1.40 m ± 90 mm (55 ±3.54 inch) | 1.1 m ± 60 mm (43,30 ±2.36 inch) | |
| | Humidifier Connection 0.50 m ± 40 mm (19.7 ±1.58 inch) | 0.3 m ± 10 mm (11.81 ±0.39 inch) | |
| Substantial Equivalence<br>Discussion | Subject of Submission | Reference Predicate Device | Comments |
| Product Name | MP02608 | MP00308 | - |
| Resistance | Inspiratory:<br>at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 1.0 mbar<br>at 30 L/min < 3.0 mbar | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.4 mbar<br>at 15 L/min < 2 mbar<br>at 30 L/min < 7 mbar | Different |
| | Expiratory:<br>at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 1.0 mbar<br>at 30 L/min < 3.3 mbar | at 2.5 L/min < 0.1 mbar<br>at 5 L/min < 0.2 mbar<br>at 15 L/min < 1.5 mbar<br>at 30 L/min < 5 mbar | |
| Compliance | at 60 mbar (breathing circuit: inspiration and<br>expiration) < 1.0 mL/hPa | at 60 mbar <0.9 mL/mbar | Similar |
| Leakage | at 20 mbar < 15 mL/min<br>at 40 mbar < 30 mL/min<br>at 60 mbar < 30 mL/min | at 60 mbar <50 mL/min | Different |
| Volume of breathing<br>circuit/ humidifier chamber | Inspiratory volume of the breathing circuit with<br>incubator extension 360 ml ±10%<br>Volume (air) of humidifier chamber with water 210<br>ml ±10%<br>Volume (air) of humidifier chamber without water<br>300 ml ±10% | Inspiratory volume of the breathing circuit<br>with incubator extension 350 ml<br>Volume (air) of humidifier chamber with water<br>190 ml<br>Volume (air) of humidifier chamber without<br>water 300 ml | Similar |
| Warm up Time | 30 min | 30 min | Same |
| Humidification Output | Invasive ventilation at 2 to 60 L/min > 33 mg/L<br>Non-invasive ventilation at 2 to 60 L/min > 12<br>mg/L | Invasive ventilation at 4 to 30 L/min >33 mg/L<br>Non-invasive ventilation at 4 to 30 L/min > 12<br>mg/L | Different |
| Breathing gas temperature<br>at Y-Piece | Adjusted and controlled by Fisher & Paykel<br>MR850 humidifier | Adjusted and controlled by Fisher & Paykel<br>MR850 humidifier | Same |
| Substantial Equivalence<br>Discussion | Subject of Submission | Reference Predicate Device | Comments |
| Product Name | MP02608 | MP00308 | - |
| Electrical Connection Data | 22 V, 2.73 A, 60 W<br>$14 \Omega$ inspiratory | 22 V, 2,73 A, 60 W,<br>$16 \Omega$ inspiratory | Same |
| Ventilation modes | VentStar Helix dual heated (N) Plus is suitable for standard ventilation modes, such as Synchronized Intermittend Mandatory Ventilation (SIMV), Mandatory Minute Ventilation (MMV), and Continuous Positive Airway Pressure (CPAP) as well as for High Frequency Oscillation (HFO). | Suitable for standard ventilation modes, such as Synchronized Intermittend Mandatory Ventilation (SIMV), Mandatory Minute Ventilation (MMV), and Continuous Positive Airway Pressure (CPAP) as well as for High Frequency Oscillation (HFO). | Same |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue and are positioned close together.
510(k) Summary
K222822
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue and are slightly larger than the rest of the letters.
510(k) Summary
K222822
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is centered in the image and the background is white.
# 510(k) Summary
K222822
The indications for use and fundamental scientific technology are the same for both proposed and predicate devices.
As demonstrated in the nonclinical testing the different technological characteristics do not raise any new questions of safety and effectiveness when compared to the predicate devices.
{11}------------------------------------------------
510(k) Summary
K222822
# Performance Data
Non-clinical testing of the breathing circuits has been performed covering mechanical, thermal safety, environmental conditions, electrical safety and electromagnetic compatibility, functional verification, and performance capacity and accuracy. Verification and validation testing was conducted in conformance to the FDA recognized standards as listed below. Performance data related to each proposed modification has been tested and evaluated. High level summary reports included in this 510(k) demonstrate that the subject devices are substantially equivalent to the predicate device.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted for VentStar Helix Dual Heated N Exten. Kit and VentStar Helix heated (N) Plus. The devices comply with the IEC 60601-1 for basic safety and performance and the IEC 60601-1-2 standard for EMC.
# Further Performance Testing
To demonstrate performance and functionality, VentStar Helix Dual Heated N Exten. Kit and VentStar Helix heated (N) Plus were tested and meet all applicable requirements of the following standards:
- . ISO 5367: Anaesthetic and respiratory equipment -- Breathing sets and connectors
- . ISO 80601-2-12: Medical electrical equipment. Particular requirements for basic safety and essential performance of critical care ventilators
- . ISO 5356-1: Anaesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones and sockets
- . ISO 80601-2-74: Medical electrical equipment - Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment (Gap Analysis)
{12}------------------------------------------------
#### Traditional 510(k) 510(k) Summary
### K222822
# Biocompatibility Testing
All materials used in the fabrication of the heated breathing systems and also the completed products were evaluated through biological qualification safety tests as outlined in ISO 10993-1 "Biological Evaluation of Medical Devices" and the international standard series of ISO 18562 "Biocompatibility evaluation of breathing gas pathways in healthcare". These materials were also tested in accordance with industry recognized test methods and were found to be acceptable for the intended use. The following tests were performed:
- . Test for emissions of particulate matter (PM) according to ISO 18562-2
- Test for emission of VOC with additional humidity according to ISO 18562-3
- . Test for leachables in condensate according to ISO 18562-4
- . Extracting according to ISO 10993-18 (Exhaustive extraction in Water, Isopropanol; simulated use extraction with Isopropanol)
- . Material characterization according to ISO 10993-18 (2009-08) (organic and inorganic extractables/ leachables)
- . Cytotoxicity according to ISO 10993-5
- Sensitization according to ISO 10993-10 .
- . Irritation according to ISO 10993-10
{13}------------------------------------------------
510(k) Summary
K222822
# List of Consensus Standards
| Standards Number | Standards Title | FDA Recognition<br>No. + date |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| ISO<br>10993-1<br>Fifth Edition<br>2018 | Biological evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process | 2-258 01/14/2019 |
| ISO<br>18562-1<br>First Edition<br>2017 | Biocompatibility evaluation of breathing gas pathways in healthcare<br>applications - Part 1: Evaluation and testing within a risk<br>management process | 1-134<br>06/07/2018 |
| IEC<br>60601-1-2<br>Edition 4.0<br>2014 | Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests | 19-8<br>09/17/2018 |
| ISO<br>18190<br>First Edition<br>2016 | Anaesthetic and respiratory equipment - General requirements for<br>airways and related equipment | 1-120<br>01/14/2019 |
| ANSI AAMI<br>ES60601-1<br>2005/(R)2012 | Medical electrical equipment - Part 1: General requirements for basic<br>safety and essential performance (IEC 60601-1:2005, MOD) and<br>A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) | 19-4<br>07/09/2014 |
| ISO<br>80601-2-12<br>First Edition<br>2011 | Medical electrical equipment - Part 2-12: Particular requirements for<br>the safety of lung ventilators - Critical care ventilators [Including:<br>Technical Corrigendum 1 (2011)] | 1-98<br>01/30/2014 |
| ISO<br>80601-2-74<br>First Edition<br>2017 | Medical electrical equipment - Part 2-74: Particular requirements for<br>basic safety and essential performance of respiratory humidifying<br>equipment | 1-138<br>01/14/2019 |
| IEC<br>60601-1-6<br>Edition 3.1<br>2013 | Medical electrical equipment - Part 1-6: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Usability | 5-89<br>06/27/2016 |
| IEC<br>62366-1<br>Edition 1.0<br>2015 | Medical devices - Part 1: Application of usability engineering to<br>medical devices [Including CORRIGENDUM 1 (2016)] | 5-114<br>12/23/2016 |
| ISO<br>5356-1<br>Third Edition<br>2004 | Anaesthetic and respiratory equipment - Conical connectors: Part 1:<br>Cones and sockets | 1-62<br>03/16/2012 |
| ISO<br>5367<br>Fifth Edition<br>2014 | Anaesthetic and respiratory equipment -- Breathing sets and<br>connectors | 1-103<br>08/14/2015 |
| ISO<br>14971<br>Third Edition<br>2019 | Medical devices - Application of risk management to medical devices | 5-125<br>12/23/2019 |
{14}------------------------------------------------
Traditional 510(k) K222822 510(k) Summary
# Conclusion
Based on the indications for use/intended use, technological characteristics, performance/nonclinical testing, and comparison to the predicate devices, the subject devices are substantially equivalent to the legally marketed predicate devices and raise no new safety and effectiveness questions compared to the predicate devices.
- END -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.