Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient.
Device Story
FlowStar Touch Digital Mixer Flowmeter is a microprocessor-controlled device for precise dosing of medical oxygen and nitrous oxide for patient sedation in dental offices or clinics. Operated by a doctor or dentist via an integrated touchpad, the device mixes gases to a user-defined concentration (up to 70% N2O). It features a fail-safe system to prevent pure N2O delivery if oxygen supply fails, mechanical encodings to prevent gas mix-up, and an O2 flush function. Output connects to a Bag T system, which integrates a reservoir bag and scavenger system for patient delivery. The device monitors gas supply and flow, triggering visual/acoustic alarms for insufficient O2, N2O, or flow obstructions. It provides data storage via USB. By controlling the gas mixture, the clinician manages the patient's sedation level, benefiting the patient through controlled analgesia during procedures.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, including compliance with IEC 60601-1, IEC 60601-1-2, ISO 11195, ISO 5356-1, and biocompatibility testing per ISO 18562-2 and ISO 18562-3. Performance testing verified alarm functionality, pressure differentials, ambient air valve operation, and high-altitude performance.
Technological Characteristics
Microprocessor-controlled valve system; DISS gas inlet connectors; integrated touchpad interface; USB Type A port for data storage. Power: 110-230V 50-60Hz. Materials: Breathing gas pathways tested per ISO 18562-2/3. Standards: IEC 60601-1, IEC 60601-1-2, ISO 11195, CGA V-5, CGA C-9, ISO 5356-1. Software: Moderate level of concern.
Indications for Use
Indicated for conscious, spontaneously breathing adult and pediatric patients requiring nitrous oxide-oxygen sedation. Contraindicated in patients with hindered nasal breathing, impaired communication, severe psychological/behavioral disorders, severe general diseases (MS, ileus), post-eye surgery with intraocular gas, ASA >= 3, Vitamin B12/folic acid deficiency, pregnancy (1st/2nd trimester), hypersensitivity, head injuries with impaired consciousness, maxillofacial injuries, pneumothorax, air embolism, middle ear occlusion/infection, decompression sickness, or abdominal distension/intestinal obstruction.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Below that is the word "ADMINISTRATION" in a smaller font.
July 28, 2023
Baldus Sedation GmbH & Co. KG % Daniel Kamm Principal Engineer Kamm & Assosciates 8870 Ravello Ct Naples, Florida 34114
Re: K222794
Trade/Device Name: FlowStar Touch Digital Mixer Flowmeter Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR Dated: July 26, 2023 Received: July 26, 2023
Dear Daniel Kamm:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
James J. Lee -S
James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222794
Device Name FlowStar Touch Digital Mixer Flowmeter
Indications for Use (Describe)
Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size: 20px;">☐</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary: 510(k) Number K222794
Image /page/3/Picture/1 description: The image shows the logo for Baldus Sedation. The logo features a blue circle on the left side of the word "Baldus", which is written in a bold, sans-serif font. Below the word "Baldus" is the word "Sedation", which is written in a smaller, lighter font. A thin blue line separates the two words.
Baldus Sedation GmbH & Co.KG In der Langfuhr 32 56170 Bendorf Germany www.baldus-sedation.de Phone: 0049 261 96 38 92666 Date Prepared: July 28, 2023 Prepared by: Lisa Baldus Head of Quality Management and Regulatory Affairs
- 1) Identification of the Device: Trade/Device Name: FlowStar Touch Digital Mixer Flowmeter Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: II Product Code: BZR
- 2) Equivalent legally marketed device: K052335 Trade/Device Name: Accutron Digital Ultra Manufacturer: Accutron, Inc. Regulation Name: Breathing Gas Mixer Regulatory Class: II Product Code: BZR
- 3) Indications for Use: Indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen to a conscious, spontaneously breathing patient
#### 4) Description of the Device:
The device will be available in two models: (Details below) FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 50 %, and FLOWSTAR TOUCH DIGITAL MIXER FLOWMETER 70 %.
50% and 70% represent the maximum nitrous oxide concentration. The FlowStar Touch O2- N₂O Mixer for Analgesia is a device which can precisely dose a mixture of medical oxygen and medical nitrous oxide, depending on the adjusted mixture, for the sedation of patients in hospitals and at the dentist. The operator, a doctor or a dentist is able to adjust the total flow and the concentration of nitrous oxide of this flow. Depending on these settings the level of sedation of the patient can be controlled by the operator. The Baldus® Touch O2- N2O Mixer for Analgesia device has been created with the following safety features. The total flow emitted can consist of a maximum of 70 % nitrous oxide, the mix up of gas types is prevented by mechanical encodings and the flow of pure nitrous oxide in the absence of a missing oxygen supply is prevented by a fail-safe device. An integrated Touch pad is used to select the desired N₂O /O2 gas mixture. The total flow can consist of between 0 (100% O2) and 70 (30% O2)
{4}------------------------------------------------
percent N₂O. The minimum output of 30 % O2 prevents from a hypoxic mixture for the patient. The FlowStar TouchO2- N2O Mixer for Analgesia device has an integrated O2 Flush function which delivers at least 30 l/min of pure oxygen to flush the N20 out of the patient circuit after the treatment. The FlowStar TouchO2- N₂O Mixer for Analgesia device has an alarm system which generates an alarm in the case of an insufficient O2 supply. The device will shut off the sedation when the O2 supply reaches a critical level. The gas mixture output of the FlowStar TouchO2- N2O Mixer for Analgesia device is connected to the Bag T. The Bag T is a device with two inputs and two outputs. One input is the gas mixture from the mixing device. The other input is a valve which opens when the adjusted total flow is too low for the patient. The gas mixture is then enriched with additional ambient air. The first output of the Bag T is for the connected breathing bag which is filled when the patient exhales and is emptied when the patient inhales. The second output leads the gas mixture to the scavenger system and finally to the patient. The main accessories for the proposed device are a Breathing Bag and a Double Mask Scavenger System.
The device contraindications (from the labeling) are:
- Hindered or restricted nasal breathing (rhinitis, sinusitis)
- impaired ability to communicate
- Severe psychological or behavioral and personality disorders ●
- Severe general diseases (MS, ileus) ●
- Status post eye operation with an intraocular gas bubble
- ASA >= 3
- Vitamin B12 and folic acid defiency or disorder
- First and second trimester of pregnancy
- Neither female employees in the dental practice nor patients in their first or second trimester of pregnancy.
- Hypersensitivity to the medication
- Head injuries with impaired consciousness
- Maxillofacial injuries
- Artificial, traumatic or spontaneous pneumothorax
- Air embolism
- Middle ear occlusion, ear infection
- Decompression sickness
- Abdominal distension / intestinal obstruction.
#### There are TWO product variants:
#### PRODUCT VARIANT 1:
FlowStar Touch Digital Mixer Flowmeter 70 %
Minimum O2 concentration: 30 %
Maximum O2 concentration: 100 %
Minimum N₂O concentration: 0 %
Maximum N₂O concentration: 70 %
#### PRODUCT VARIANT 2:
#### FlowStar Touch Digital Mixer Flowmeter 50 %
Minimum 02 concentration: 50 % Maximum O2 concentration: 100 %
{5}------------------------------------------------
The intended patient population is adults and pediatrics.
The intended operational environment is dental offices or clinics.
The patient contacting component is a face mask (not part of the mixer device described in this submission). The patient contact is with skin, and the duration is for the length of the patient wearing the mask. The face mask is FDA product code BSJ, 510(k) exempt. Other components are: Mask, Scavenging, product code KHA, 510(k) exempt, and Device Bag, Reservoir, product code BTC, 510(k) exempt. Color coded connectors are employed to prevent switching the oxygen and N₂O connections.
Constructional description: The device includes a microprocessor-controlled valve system that precisely meters Oxygen and Nitrous Oxide gases for analgesic purposes. The inputs of the touchpad influence the gas flow inside of the device. The firmware is considered to be of a moderate level of concern. The user interface is essentially identical to the predicate described in the detailed comparison table below.
| Characteristic | K052335, Accutron Digital Ultra, Accutron,<br>Inc. | FlowStar Touch Digital Mixer Flowmeter<br>(Two models: N2O max 70% or<br>N2O max 50%) |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | To be used in nitrous oxide-oxygen sedation<br>systems for delivering to a patient a mixture of<br>nitrous oxide and oxygen gases with a<br>maximum nitrous oxide concentration of 70%. | Indicated for administering an adjustable mixture<br>of Nitrous Oxide analgesic gas and Oxygen to a<br>conscious, spontaneously breathing patient.<br>(Functionally identical) |
| Range | Between 0 (100% O2) and 70 (30% O2)<br>percent N2O | Between 0 (100% O2) and 70 (30% O2) percent<br>N2O SAME |
| Power Source | 100-240 V AC, 50/60 Hz only. | 110-230 V 50-60 Hz (Essentially the same) |
| Controls | Touch Panel | SAME |
| Photo of Front | Image: Accutron Digital Ultra | Image: FlowStar Touch Digital Mixer Flowmeter |
| Population | Adults and Pediatrics | SAME |
| Gas Delivery | Oxygen Flush: Minimum 20 LPM<br>Oxygen Flow: 1.0 LPM-9.9 LPM<br>Oxygen %: 30%-100%<br>Oxygen Resuscitator Flow: Minimum 100 LPM<br>Nitrous Oxide Flow: 0-6.9 LPM<br>Nitrous Oxide %: 0-70%<br>Mixed Gas Flow: 1.0-9.9 LPM | Oxygen Flush 40-55 l/min<br>Oxygen Flow 1.0 - 18 l/min<br>Oxygen %: 30 % - 100 %<br>Oxygen Resuscitator Flow 100 - 250l/min<br>Nitrous Oxide Flow 0 - 12,6 l/min<br>Nitrous Oxide %: 0 -70 %<br>Mixed Gas Flow 3 – 18 l/min<br>SIMILAR |
| Characteristic | K052335, Accutron Digital Ultra, Accutron,<br>Inc. | FlowStar Touch Digital Mixer Flowmeter<br>(Two models: N₂O max 70% or<br>N₂O max 50%) |
| Safety<br>Standards | IEC 60601-1, "Medical Electrical Equipment -<br>Part 1: General Requirements for Safety"<br>IEC 60601-1-2, "Medical electrical equipment -<br>Part 1-2: General requirements for safety -<br>Collateral standard: Electromagnetic<br>compatibility - Requirements and tests" | SAME |
| Alarms | O₂ Fail Safe System - Offers assurance that N₂O<br>ceases if the O₂ supply is interrupted or reduced.<br>Alarm will sound, FAILURE OXYGEN LEDs will light,<br>and N₂O flow will cease.<br>N₂O Supply Warning – Triggers an alarm if the<br>N₂O supply cannot keep up with user's needs.<br>Alarm will sound and FAILURE NITROUS LEDS will<br>light. Flow switches to 100% oxygen.<br>Mixed Gas Flow to Patient – Monitors pressure in<br>mixed gas line and triggers an alarm if flow to<br>patient is obstructed. Alarm will sound and<br>FAILURE NITROUS OXYGEN LEDS will light. | Check O₂: If no oxygen is being supplied, the unit<br>emits visual and acoustic information signals<br>Check N₂O: If no nitrous oxide is being supplied,<br>the unit emits visual and acoustic information signals.<br>This also happens if the pressure is too low or there is<br>a measurement section failure<br>O₂ minimum flow 3 l/min: Visual and acoustic signals:<br>The O₂ minimum flow is 3 l/min. The mixture<br>administered can never contain less than 1 litre of O₂.<br>Max. flow reached: Once the maximum total flow (18<br>l/min) is reached, this is signaled visually in the top<br>right and a signal tone sounds.<br>Flush flow low: This message appears if the flow rate<br>is less than the minimum value of 18 l/min when the<br>O₂ Flush button is pressed. The alarms are equivalent<br>to those of the predicate. |
| Dimensions/<br>weight | 7 1/2" x 7" x 8 3/4" 6.3 lbs | 5.3″ x 11.8″ x 4 ¾" (H x W x D) 5.5 lbs.<br>SIMILAR SIZE AND WEIGHT |
| Gas Inlet<br>Connectors | DISS connector (Diameter Index Safety<br>System) | DISS connector (Diameter Index Safety System)<br>SAME |
| External<br>interface<br>connector | Type DB-9 Serial Output for printer | USB Type A, accepts USB memory sticks for<br>storage of patient data entered via the front<br>panel. (MORE CONVENIENT THAN PRINTING) |
| Use<br>environment | Hospital, Dental Facility, Healthcare Facility | SAME |
## 5) Substantial Equivalence Chart
{6}------------------------------------------------
- 6) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new device is as safe and effective as the predicate device.
- 7) Summary of non-clinical testing: Laboratory testing confirmed compliance with the applicable portions of the following standards:
IEC 60601-1:2005 + Cor. :2006 + Cor. :2007 + A1:2012), Medical electrical equipment - Part 1: General requirements for basic safety and essential performance FDA Recognition Number: 19-4
IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic
{7}------------------------------------------------
disturbances - Requirements and tests FDA Recognition Number: 19-36
ISO 11195:2018 Gas mixers for medical use – Stand-alone gas mixers, no FDA Recognition Number
CGA V-5:2008 (Reaffirmed 2013), Diameter-Index Safety System (Noninterchangeable Low Pressure Connections for Medical Gas Applications), FDA Recognition Number: 1-81
CGA C-9: 2013 Standard Color Marking of Compressed Gas Containers for Medical Use STANDARD by Compressed Gas Association, FDA Recognition Number: 1-101
ISO 5356-1 Third edition 2004-05-15 Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and sockets FDA Recognition Number: 1-62
Biocompatibility testing: VOCs and particulate matter testing was performed according to: ISO 18562-2 First edition 2017-03 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter, FDA recognition number 1-135, AND: ISO 18562-3 First edition 2017-03, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds, FDA recognition number 1-136.
Cybersecurity measures were taken in compliance with the FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff. The problem is relatively simple to address because the unit has only one method of entry, the USB port.
Additional successful bench testing performed:
All alarms specified in the comparison table above were tested and verified as operational as intended. Silicon sealing testing. Performance ambient air valve Pressure differential Information signals + Alarms Durability of marking Normal operating conditions: Works as expected in the defined pressure ranges. Reverse gas flow Ambient temperature range Accessory compatibility On Demand Flow Performance of the device at 2,260 meter above the sea level (High altitude test) N2O Room Contamination NIOSH O2 Running Empty Test N2O Empty Test
- 8) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
- 9) Conclusion: The comparison tables and the differences between the subject device and the predicate device shows that the subject device is substantially equivalent to the predicate/reference devices in terms of safety, efficacy, indications, and all the other technological
{8}------------------------------------------------
characteristics and do not introduce any potential risks. The subject device and the predicate device has the same intended use and the technological differences do not raise different questions of safety and effectiveness. So, the proposed device can be rated as substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.