K222790 · Lumenis Be, Ltd. · OAP · Jun 14, 2023 · Physical Medicine
Device Facts
Record ID
K222790
Device Name
F65 Laser System
Applicant
Lumenis Be, Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Jun 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222790 · Jun 14, 2023
F65 Laser System
Lumenis Be, Ltd.
Retrospective patient records/files
A retrospective study of 98 patient records was conducted to evaluate the performance and safety of the F65 Laser System for hair loss treatment. The study assessed adverse events and used blinded reviewers to evaluate improvement in scalp hair appearance based on the retrospective data.
Retrospective single arm study; Retrospective single arm study
98 subjects (ages 21-66) seeking treatment for hair loss; 44 with Androgenic Alopecia, 54 with no documented diagnosis; Fitzpatrick skin types II-IV.; Sample Size: 98
Not applicable for this study
Safety (adverse events) and performance (improvement in scalp hair appearance assessed by blinded reviewers).
Indications for Use
The subject F65 Laser System has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows: The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for: o Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and ephelides (freckles) o Cutaneous lesions, including warts, scars and striae o Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations o Removal of unwanted hair and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles o Mild to moderate inflammatory Acne (Acne vulgaris) The Nd: YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd: YAG) is indicated for: o The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg o The removal of unwanted hair and to effect table long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles o The non-ablative treatment of facial wrinkles The F65 module and handpiece, with wavelength of 1565 nm, are indicated for: o Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue o Improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss. The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for: o Removal of dark tattoos o Treatment of pigmented lesions
Device Story
F65 Laser System is a multi-application aesthetic/dermatological device; console includes power supply, laser module, control unit, and PC. System utilizes interchangeable handpieces (IPL, Nd:YAG, 1565 nm Er:Glass fiber-laser) to deliver light energy to target tissue. 1565 nm module uses scanner with two motors/mirrors to direct beam in fractional patterns (up to 500 micro-beams/cm²); tip includes thermo-electric cooler (TEC). Operated by professionals in clinical settings. Output affects tissue via coagulation, hemostasis, or selective melanin targeting for hair reduction/hair loss improvement. Healthcare providers use device to perform skin resurfacing, vascular lesion clearance, and hair growth stimulation. Benefits include improved scalp hair appearance and dermatological condition management.
Clinical Evidence
Two clinical studies evaluated hair loss treatment. Prospective study (n=32): Androgenic alopecia patients showed mean hair density increases of 70.8 hairs/cm² (males) and 45.2 hairs/cm² (females) at 3 months. Retrospective study (n=98): 97.7% of androgenic alopecia cases and 96.9% of total cohort showed improvement in scalp hair appearance as assessed by blinded reviewers (95% CI: 87.8%-99.7% and 91.38%-98.5% respectively). No adverse events reported. Preclinical animal study confirmed hair follicle regeneration.
Technological Characteristics
Er:Glass fiber-laser (1565 nm), Nd:YAG (1064 nm), and IPL (400-1200 nm). 1565 nm module features scanner with two motors/mirrors, up to 500 micro-beams/cm², and TEC cooling. System includes console, power supply, control unit, and PC. Professional use only. No new bench testing required; relies on predicate testing for electrical/EMC/biocompatibility.
Indications for Use
Indicated for adult males and females, Fitzpatrick skin types I-IV, seeking treatment for hair loss; also indicated for aesthetic/dermatological procedures including hair reduction (types I-V), vascular/pigmented lesion treatment, acne, skin resurfacing, and tattoo removal.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 14, 2023
Lumenis Be, Ltd. Shlomit Segman Senior Director RA 6 Hakidma Street PO BOX 240 Yokneam, Yokneam 2069204 Israel
Re: K222790
Trade/Device Name: F65 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: GEX, ONF, ONG Dated: Mav. 2023 Received: May 5, 2023
Dear Shlomit Segman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222790
Device Name F65 Laser System
Indications for Use (Describe)
The subject F65 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:
- o Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and ephelides (freckles)
- o Cutaneous lesions, including warts, scars and striae
- o Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- o Removal of unwanted hair and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- o Mild to moderate inflammatory Acne (Acne vulgaris)
The Nd: YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd: YAG) is indicated for:
- o The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
- o The removal of unwanted hair and to effect table long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- o The non-ablative treatment of facial wrinkles
The F65 module and handpiece, with wavelength of 1565 nm, are indicated for:
- o Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue
- o Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.
The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for:
o Removal of dark tattoos
o Treatment of pigmented lesions
#### *Note
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.
{3}------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 5. 510(K) SUMMARY K222790-S001
#### F65 Laser System and Accessories
| Applicant Name: | Lumenis Be Ltd.<br>9 Hakidma Street PO Box 426<br>Yokneam Industrial Park,<br>Yokneam 2069236, Israel<br>Tel: +972-77-9599000 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Shlomit Segman, Lumenis Be Ltd.<br>9 Hakidma Street PO Box 426<br>Yokneam Industrial Park,<br>Yokneam 2069236, Israel<br>Tel: +972-77-9599230<br>Email:Shlomit.Segman@lumenis.com |
| Date Prepared: | June, 2023 |
| Trade Name: | F65 Laser System and Accessories |
| Common Name: | Powered Laser Surgical Instrument With Microbeam\Fractional Output |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology. |
| Product Code: | GEX, ONF, ONG |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Panel: | General & Plastic Surgery |
| Predicate Devices: | K193500 (Lumenis Stellar M22)<br>Reference: K173846 DermaScalp Laser Cap |
#### Intended Use/ Indications for Use:
The subject F65 Laser System has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
{5}------------------------------------------------
The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:
- o Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and ephelides (freckles)
- o Cutaneous lesions, including warts, scars and striae
- o Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, truncal and leq telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leq veins and venous malformations
- o Removal of unwanted hair and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- o Mild to moderate inflammatory Acne (Acne vulgaris)
The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for:
- o The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
- o The removal of unwanted hair and to effect table long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- o The non-ablative treatment of facial wrinkles
The F65 module and handpiece, with wavelength of 1565 nm, are indicated for:
- o Use in dermatological procedures requiring fractional skin resurfacing and coaqulation of soft tissue
- o Improving the appearance of scalp hair in adult males and females with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.
The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for:
- o Removal of dark tattoos
- o Treatment of pigmented lesions
*Note
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.
{6}------------------------------------------------
### Device Description
The Lumenis Be F65 Laser System is intended to be used for the new indication for Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV seeking treatment for hair loss, as well as, for the identical set of functions as the predicate device. The description here is limited to the system module required to support the new indication which is the subject of this 510(k):
- The laser console includes a power supply unit, a Laser module (+ aiming beam), a control unit and a PC module. The laser beam is transferred to the handpiece via fiber optics.
- . The handpiece includes a scanner module – built of two mirrors that are controlled by two motors directing the Laser beam to desired target tissue.
- . The handpiece uses a tip (user may choose from two types) to align the Laser beam focus with the target tissue, enabling a treatment shape size and a tip thermo-electric cooler (TEC).
The F65 is intended for professional use only.
### Substantial Equivalence Discussion
The key difference between the Lumenis Be F65 Laser System and its predicate is the new indication for improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV seeking treatment for hair loss, which has led to this 510(k). While the F65 Laser System delivers the same light wavelength as the predicate, the intensity and pattern of the delivered light to the scalp are specific to the proposed new indication.
Promoting hair growth through exposure to light is not a new concept, as demonstrated by the clearance of the reference device and other devices under the same classification as the reference device.
Safety questions associated with electrical, electromagnetic compatibility, and biocompatibility, are addressed through testing conducted on the predicate device. Animal testing of the subject device's operating parameters addresses both safety and effectiveness questions for the new proposed indication. Likewise, the safety and effectiveness of the device for the proposed indication were also evaluated through two clinical studies of patients seeking treatment for hair loss.
Table 1 provides a comparison, relevant to the F65 module with the 1565 nm laser, which is related to the new proposed indication, of the technical characteristics of the subject device to its predicate and reference devices.
{7}------------------------------------------------
| Parameter | Subject Device | Predicate Device | Reference Device |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | F65 Laser module System | Stellar M22 1565 nm module<br>(K193500) | DermaScalp Laser Cap (K172846) |
| Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 890.5500 |
| Procode | GEX, ONF, ONG | GEX, ONF, ONG | OAP |
| Rx/OTC | Rx | Rx | OTC |
| Indications for Use | The F65 module and<br>handpiece, with wavelength of<br>1565 nm, are indicated for:<br>• Use in dermatological<br>procedures requiring<br>fractional skin resurfacing<br>and coagulation of soft tissue<br>• Improving the appearance of<br>scalp hair in adult males and<br>females with Fitzpatrick skin<br>type I to IV who are seeking<br>treatment for hair loss. | The 1565 nm Stellar M22<br>Module & Handpiece, with<br>wavelength of 1565 nm is<br>indicated for use in<br>dermatological procedures<br>requiring fractional skin<br>resurfacing and coagulation of<br>soft tissue. | The family of DermaScalp Laser<br>Cap devices are indicated to treat<br>Androgenetic Alopecia and promote<br>hair growth in Males who have<br>Norwood Hamilton Classifications of<br>Ila to V patterns of hair loss and to<br>treat Androgenetic Alopecia and<br>promote hair growth in Females<br>who have Ludwig (Savin) Scale I-1<br>to I-4, II-1, II-2, or frontal; both with<br>Fitzpatrick Skin Types I to IV. |
| Device<br>Configuration | System console and handpiece | System console and handpiece | Power supply and laser diode caps |
| Wavelength | 1565 nm | 1565 nm | 650 nm |
{8}------------------------------------------------
| Type of Laser | Er:Glass Fiber-laser with<br>scanner up to 500 micro-<br>beams per cm² | Er:Glass Fiber-laser with<br>scanner up to 500 micro-beams<br>per cm² | Diode lasers at count of - 50, 80,<br>120, 148, 180, 202, 224, 272 |
|-----------------|------------------------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Energy | Up to 70 mJ per microbeam | Up to 70 mJ per microbeam | Total Laser Energy:<br><5 mW<br>50 diode lasers: <250mW,<br>80: <400mW,<br>120: <600mW,<br>148: <740mW,<br>180: <900mW,<br>202: <1,010mW,<br>224: <1,120mW,<br>272: <1,360mW<br>Fluence: ~ 6 J/cm² |
| Clinic/Home Use | Clinic Use by professional<br>operator | Clinic Use by professional<br>operator | |
{9}------------------------------------------------
### Summary of Clinical Testing and Animal Testing
A preclinical study was conducted to evaluate the Lumenis F65 subject module for hair follicle regeneration capabilities using both Low Level Laser Therapy (LLLT) and untreated controls. The results showed complete tissue repair and formation of hair follicles.
Two clinical studies were conducted to demonstrate that the subject device is safe and effective in patients seeking treatment for hair loss, as indicated. Subjects received multiple treatments over several months, per the protocol described in the Operator's Manual.
In a prospective, single arm study, 32 subjects with hair loss were enrolled, 25 of which completed the entire course of the study. Patients were diagnosed with Androgenic Alopecia (18) or other diagnosis (7) and were evaluated separately. In the Androgenic Alopecia group, the mean age was 32.1 ± 6.4 years. Most of the subjects were female (13/18) and all subjects included were of Fitzpatrick skin type II. For male subjects with androgenic alopecia (n=5) there was an increase of 70.8 ± 35.5 hairs per cm² between baseline and the 3-month follow-up visit. For female subjects with androgenic alopecia (n=18) there was an increase of 45.2 ± 30.2 hairs per cm² between baseline and 3-month follow-up visit. Pain and discomfort during treatment were assessed with a visual analog scale (VAS) and ranged between 1.97 to 2.33 on a scale from 0 to 10. Post treatment subjects had a trace to mild erythema. No adverse events were reported during the study treatment or follow-up period.
In a second retrospective single arm study, a total of 98 subjects were evaluated for performance, with ages ranging from 21 to 66 years and a mean age of 37.2 ± 9.9 years. Of them, in 44 (44/98, 44.9%) of the cases the documented diagnosis was Androgenic Alopecia, and in 54 cases (54/98, 55.1%) there was no documented diagnosis. Females slightly outnumbered males (51 vs. 47), and the majority of subjects had Fitzpatrick skin type III (60/98, 61.2%), with some having skin types II (18/98, 18.4%) or IV (1/98, 1%). All patient records were reviewed for any reports of adverse events during treatment or follow-up period – no adverse events were reported or documented in patient files. For the 44 subjects with a confirmed diagnosis of androgenic alopecia, the assessment of improvement in scalp hair appearance yielded a correct identification in 97.7% (95% Cl: 87.8% -99.7%) of the cases by a group of 3 blinded reviewers. Demonstrating improvement in scalp hair appearance in more than 85% of the patients treated which supports the primary efficacy endpoint of demonstrating improvement in more than 70% of subjects. For the 98 subjects included in the performance analysis, the assessment of improvement in scalp hair appearance yielded a correct identification in 96.9% (95% CI: 91.38 -98.5%) of the cases by a group of 3 blinded reviewers.
### Summary of Testing
No new bench testing was required because the testing of the Lumenis predicate devices also applies to the subject device.
{10}------------------------------------------------
### Conclusion
The proposed new indication for the subject device does not raise different questions of safety or effectiveness and is adequately supported by the validation testing. The conclusions drawn from the animal and clinical testing, and previously conducted testing of the predicate device, demonstrate that the subject device is as safe, as effective, and performs as well as the predicate.
Thus, the subject device is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.