← Product Code [GCJ](/productcode/GCJ) · K222783

# MD Mattress Suture Placement Device (K222783)

_Lsi Solutions, Inc. · GCJ · Jan 12, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222783

## Device Facts

- **Applicant:** Lsi Solutions, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 12, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The MD™ MATTRESS Device is intended for use in the approximation of soft tissue and prosthetic materials.

## Device Story

Nonpowered, hand-operated suture placement device; facilitates horizontal mattress stitches. Input: manual lever actuation by physician. Mechanism: lever-driven system advances dual curved needles across tissue jaw to engage needle caps with pre-loaded suture; retraction pulls suture through tissue. Features: angulation knob (up to 15°), rotational knob (6 positions), jaw width indicator, and needle axle indicator. Used in surgical environments (open or minimally invasive/cannula-based). Output: placed suture stitches. Benefits: enables precise, repeatable suture placement for soft tissue/prosthetic approximation.

## Clinical Evidence

No clinical data. Bench testing only. Verification and validation testing confirmed design inputs met. Biocompatibility testing per ISO 10993-1:2018 (cytotoxicity, sensitization, irritation, systemic toxicity, material-mediated pyrogenicity). Packaging/shelf-life testing per ISO 11607-1:2019 demonstrated 2-year shelf life.

## Technological Characteristics

Nonpowered, hand-operated; stainless steel and polymer construction. Conforms to ISO 10993-1:2018. Features: dual curved needles, angulation/rotational knobs. Sterilization: ethylene oxide (minimum lethality 10^-6). Single-patient-use.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- RD180® - THE RUNNING DEVICE® ([K100593](/device/K100593.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

January 12, 2024

LSI Solutions, Inc. Christopher Miller Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd Victor, New York 14564

Re: K222783

Trade/Device Name: MD Mattress Suture Placement Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Codes: GCJ, GAS Dated: June 21, 2023 Received: June 22, 2023

Dear Christopher Miller:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Tek N. v signed by Tek N Lamichhane -S Date: 2024.01.12 17:17:45 -05'00 Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222783

Device Name MD™ MATTRESS Suture Placement Device

Indications for Use (Describe)

The MD™ MATTRESS Suture Placement Device is indicated for use in the approximation of soft tissue and prosthetic materials.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

| Submitted By:              | LSI SOLUTIONS, Inc.<br>7796 Victor-Mendon Road<br>Victor, NY 14564                                                                                               |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:            | Christopher B. Miller<br>Executive Director of Regulatory Affairs and Quality<br>Phone: (585) 869-6665<br>Fax: (585) 742-8086<br>Email: cmiller@lsisolutions.com |
| Date Prepared:             | 10 JAN 2024                                                                                                                                                      |
| Trade Name:                | MD™ MATTRESS Suture Placement Device                                                                                                                             |
| Common Name:               | Needle Guide                                                                                                                                                     |
| Classification Name:       | Endoscope and Accessories (per 21 CFR 876.1500)                                                                                                                  |
| Classification Regulation: | 21 CFR 876.1500                                                                                                                                                  |
| Product Code:              | GCJ (Primary Product Code)<br>GAS (Secondary Product Code)                                                                                                       |

{4}------------------------------------------------

#### Device Classification: Class II

Predicate Device:

RD180® - THE RUNNING DEVICE® (Cleared under Premarket Notification K100593)

#### Device Description

The MD™ MATTRESS Device is a nonpowered, hand-operated suture placement device that facilitates placement of a stitch, for example horizontal mattress stitches. Each sterile package contains one single-patient-use MD™ MATTRESS Device. The lavender device tip has an integrated tissue jaw incorporating a jaw width indicator that identifies the jaw when it is embedded in tissue. Two curved needles emerge from their protective compartments to advance across the tissue jaw, rotating about the circular needle axle indicator to engage needle caps with attached suture that are loaded into specialized needle cap compartments. An angulation indicator on the tip and the device shaft depicts the direction of the device tip angle, which is adjusted using the white angulation knob. The embossed arrow and word "IN" indicate the direction the device tip moves when the angulation knob is rotated clockwise; the device tip can be angled up to approximately 15° in either direction, as indicated by the 3 embossed radial lines in the angulation indicator The lavender rotational knob can be rotated to orient the device shaft and tip into one of six distinct positions. The rotational knob has an integrated eyelet, for stabilizing a suture tube with indwelling suture, and an indicator fin that is aligned with the device tip tissue jaw. Fully squeezing the lavender lever toward the white handle causes both curved needles in the device tip to advance across the tissue jaw toward the needle cap compartments. Fully releasing the lever causes both needles and their engaged needle caps with

{5}------------------------------------------------

attached suture to retract back across the tissue gap, pulling both suture ends through the targeted tissue within the jaw. Fully squeezing the lever again causes the needles and engaged needle caps with attached suture to advance back into their corresponding needle cap compartments, and an internal mechanism alternates the device into its rearming mode. Fully releasing the lever again causes the needles to retract into their protective compartments, leaving the needle caps and attached suture ends in the needle cap compartments. Subsequently squeezing and releasing the lever will repeat the pickup and rearm cycle.

While the MD™ MATTRESS Device is used for the placement of MD™ QUICK LOAD® SURGICAL SUTURES, this premarket submission does not include a request for clearance of suture materials since MD™ QUICK LOAD® SURGICAL SUTURE is listed on FDA's Establishment Registration and Device Listing database (K203081).

One (1) MD™ MATTRESS Device is provided per package, and the MD™ MATTRESS Device is intended to facilitate the placement of multiple sutures through tissue and prosthetic materials.

#### Intended Use

The MD™ MATTRESS Device is intended for use in the approximation of soft tissue and prosthetic materials.

#### Indications for Use

The MD™ MATTRESS Suture Placement Device is indicated for use in the approximation of soft tissue and prosthetic materials.

{6}------------------------------------------------

#### Technological characteristics (comparison to Predicate Device)

Both the predicate RD180® Device and the subject MD™ MATTRESS Device have the same intended use: "Intended for use in the approximation of soft tissue and prosthetic materials." The predicate RD180® Device is "indicated for use in the approximation of soft tissue and prosthetic materials," and LSI recommends its use with 2-0 polyester suture, among other suture types. The subject MD™ MATTRESS Device has similar indications. The MD™ MATTRESS Device is "indicated for use in the approximation of soft tissue and prosthetic materials." The subject and predicate devices are both designed for sewing suture on similar technical principles. Both devices are single-patient-use, nonpowered, hand-operated devices that incorporate a leverdriven mechanism that pierces needles through tissue, picks up suture, and pulls the suture back through the tissue to facilitate placing a stitch.

Both the subject and predicate devices operate on the same principle to accomplish the tasks the devices are intended to perform. These tasks include: loading suture needle caps into the device tip; passing the device tip through the surgical access site (optionally through a cannula if appropriate); selecting the desired device tip position over a target site; firing needles through tissue and pulling the suture back through the tissue; rearming the device to allow for placement of additional tissue bites (as deemed appropriate by the surgeon); and removing the suture needle caps from the needles. Both the subject and predicate devices are provided sterile, after sterilization with ethylene oxide such that a minimum lethality of 10% is achieved. The predicate RD180® Device is a stainless steel and polymer suture delivery device. The MD™ MATTRESS Device is made from similar, often identical, materials. Both devices conform to the requirements of ISO 10993. The predicate device is packaged in a rigid thermoformed PETG

{7}------------------------------------------------

blister tray with a Tyvek® cover. The subject device is packaged in the similar configuration with similar materials: a blue-tint PETG tray and retainer inserted into a Tyvek®/Nylon pouch.

The MD™ MATTRESS Device and the predicate RD180® Device are substantially equivalent. The subject and predicate devices are comparable, and the differences do not introduce any new risks, and have no negative impact on the safety and efficacy of the MD™ MATTRESS Device.

A summary of the comparison between the subject MD™ MATTRESS Device and predicate RD180® Device is provided in the table below.

| Table 6: Substantial Equivalence Comparison |                                                                                       |                                                                                       |
|---------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Subject Area                                | Predicate Device:<br>RD180® - THE RUNNING<br>DEVICE®                                  | Subject Device:<br>MD™ MATTRESS Device                                                |
| 510(k) Number:                              | K100593                                                                               | K222783                                                                               |
| Device Class:                               | II                                                                                    | II                                                                                    |
| Classification<br>Regulation:               | 21 CFR 876.1500                                                                       | 21 CFR 876.1500                                                                       |
| Product Code:                               | GCJ                                                                                   | GCJ                                                                                   |
| Intended Use:                               | Intended for use in the<br>approximation of soft tissue and<br>prosthetic materials.  | Intended for use in the<br>approximation of soft tissue and<br>prosthetic materials.  |
| Indications for Use:                        | Indicated for use in the<br>approximation of soft tissue and<br>prosthetic materials. | Indicated for use in the<br>approximation of soft tissue and<br>prosthetic materials. |

{8}------------------------------------------------

| Table 6: Substantial Equivalence Comparison |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Area                                | Predicate Device:<br>RD180® - THE RUNNING<br>DEVICE®                                                                                                                                                                                                                                                                                            | Subject Device:<br>MD™ MATTRESS Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Single Use/<br>Reusable:                    | Single Use                                                                                                                                                                                                                                                                                                                                      | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Design:                                     | Device tip (including a tissue<br>jaw, single straight needle, and a<br>needle cap compartment)                                                                                                                                                                                                                                                 | Device tip (including a tissue<br>jaw, jaw width indicator, dual<br>curved needles, needle axle<br>indicator, needle cap<br>compartments, and an<br>angulation indicator)                                                                                                                                                                                                                                                                                                                                                                                       |
|                                             | Device shaft                                                                                                                                                                                                                                                                                                                                    | Device shaft (including an<br>angulation knob, rotational knob<br>with an integrated eyelet and<br>indicator fin)                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                             | Pink lever                                                                                                                                                                                                                                                                                                                                      | Lavender lever                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                             | White handle                                                                                                                                                                                                                                                                                                                                    | White handle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Table 6: Substantial Equivalence Comparison |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Subject Area                                | Predicate Device:<br>RD180® - THE RUNNING<br>DEVICE®                                                                                                                                                                                                                                                                                            | Subject Device:<br>MD™ MATTRESS Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Technological<br>Characteristics:           | Nonpowered, hand-operated; incorporating a lever-driven<br>mechanism utilizing needle-and-<br>needle-cap technology to place<br>suture through tissue or<br>prosthetic materials using a<br>single straight needle.<br><br>The predicate device does not<br>contain any articulation or<br>rotational knobs or functions.                       | Nonpowered, hand-operated; incorporating a lever-driven<br>mechanism utilizing needle-and-<br>needle-cap technology to place<br>suture through tissue or<br>prosthetic materials using dual<br>curved needles.<br><br><span style="vertical-align:middle">COMPARABLE</span><br><br>The subject device contains an<br>articulation knob, which can be<br>used to angulate the device tip<br>inward or outward, and a<br>rotational knob, which enables<br>the device shaft and<br>correspondingly the device tip<br>to rotate through six distinct<br>positions. |
| Target Population:                          | Any patient requiring the<br>approximation of soft tissue or<br>prosthetic materials.                                                                                                                                                                                                                                                           | Any patient requiring the<br>approximation of soft tissue or<br>prosthetic materials.<br><br><span style="vertical-align:middle">SAME</span>                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Table 6: Substantial Equivalence Comparison |                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Subject Area                                | Predicate Device:<br>RD180® - THE RUNNING<br>DEVICE®                                                                                                                                                                                                                                                                                            | Subject Device:<br>MD™ MATTRESS Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Environment of<br>Use:                      | Suitable for use in any<br>environment where surgical<br>procedures occur under the<br>supervision of a physician.<br><br>Compatible with open surgeries,<br>smaller access/minimally<br>invasive surgeries, and surgery<br>through a cannula or other<br>access ports with seals to<br>maintain positive CO2 pressure<br>at the surgical site. | Suitable for use in any<br>environment where surgical<br>procedures occur under the<br>supervision of a physician.<br><br>Compatible with open surgeries,<br>smaller access/minimally<br>invasive surgeries, and surgery<br>through a cannula or other<br>access ports with seals to<br>maintain positive CO2 pressure<br>at the surgical site.                                                                                                                                                                                                                 |
| Sterility                                   | To be sterilized with ethylene<br>oxide such that a minimum<br>lethality of 10-6 is achieved.                                                                                                                                                                                                                                                   | To be sterilized with ethylene<br>oxide such that a minimum<br>lethality of 10-6 is achieved.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Materials and<br>Biocompatibility           | Comprising stainless steel and<br>polymers.<br>Conforms to the requirements of<br>ISO 10993-1:2009.                                                                                                                                                                                                                                             | Comprising stainless steel and<br>polymers.<br>Conforms to the requirements of<br>ISO 10993-1:2018.                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Packaging                                   | Packaged in a rigid<br>thermoformed PETG tray with a<br>Tyvek® cover.                                                                                                                                                                                                                                                                           | Packaged in a PETG tray and<br>retainer lid which are sealed in a<br>Tyvek®/Nylon pouch.                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{9}------------------------------------------------

{10}------------------------------------------------

# Performance Testing Summary

Bench top performance testing was conducted to verify that the MD™ MATTRESS Device will perform as intended and to ensure the device will perform equivalently to the predicate device.

{11}------------------------------------------------

The following non-clinical tests were conducted, and all results met the performance and riskbased acceptance criteria:

- . Functional Design Verification and Validation Testing to ensure the Design Input Requirements were met.
- Biocompatibility per ISO 10993-1:2018 and in accordance with FDA Guidance for . Industry Use of International Standard ISO 10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process, September 4, 2020. The MD™ MATTRESS Device can be classified as an externally communicating medical device that has limited contact (≤24 hours) with tissue. The battery of testing included:
	- Chemical characterization O
	- Cytotoxicity o
	- Sensitization O
	- Irritation o
	- Material-mediated pyrogenicity o
	- 0 Acute systemic toxicity
- . Packaging/Shelf-life Testing per ISO 11607-1:2019 demonstrating a 2 year shelf life.

## Clinical Testing

Not applicable. Neither the predicate nor the subject product require clinical testing. The substantial performance testing discussed above confirms the reliability and excellent performance of the MD™ MATTRESS Device. The performance testing included multiple experiments that do not require clinical testing in patients.

{12}------------------------------------------------

## Substantial Equivalence

The subject device and predicate device have the same intended use and technological characteristics. Non-clinical performance data has demonstrated the subject device is substantially equivalent to the predicate device.

### Conclusion

The subject MD™ MATTRESS Device is substantially equivalent to the predicate RD180® Device, submitted under Premarket Notification K100593 and cleared in 2010. Extensive testing demonstrates that the proposed product consistently provides excellent suture placement and can be dependably manufactured. The MD™ MATTRESS Device will be safe and effective for the approximation of soft tissue and prosthetic materials.

---

**Source:** [https://fda-staging.innolitics.com/device/K222783](https://fda-staging.innolitics.com/device/K222783)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
