K222770 · Top-Rank Health Care Co., Ltd. · GYB · Dec 13, 2022 · Neurology
Device Facts
Record ID
K222770
Device Name
Conductive Gel
Applicant
Top-Rank Health Care Co., Ltd.
Product Code
GYB · Neurology
Decision Date
Dec 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for use with electric stimulation therapy devices, such as TENS and EMS. Conductive Gel is used with external electrodes to reduce the impedance of the contact between the electrode and the skin.
Device Story
Conductive gel applied to intact skin surface; used as interface between external electrodes and skin for TENS/EMS therapy. Gel reduces electrical impedance; ensures even current distribution. Applied by user or clinician; washed off after use. Benefits patient by improving electrical contact efficiency during stimulation therapy.
Clinical Evidence
No clinical data. Bench testing only. Testing included in vitro cytotoxicity (ISO 10993-5), skin sensitization (ISO 10993-10), skin irritation (ISO 10993-10), usability (IEC 62366-1), and shelf-life validation (ASTM F1980-16). All tests passed.
Technological Characteristics
Electroconductive gel; composition: 98.25% purified water, 0.5% Carbopol, 1% glycerin, 0.2% sodium hydroxide, 0.05% triclosan. pH 6.5-7.5; impedance ≤500Ω at 1 MHz. Non-sterile; 2-year shelf life. Complies with ISO 10993-5/10 for biocompatibility.
Indications for Use
Indicated for use with TENS and EMS therapy devices to reduce electrode-to-skin contact impedance. For use on intact skin.
Regulatory Classification
Identification
Electroconductive media are the conductive creams or gels used with external electrodes to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
December 13, 2022
Top-Rank Health Care Co.,Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technoloy Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, Guangdong 510000 China
Re: K222770
Trade/Device Name: Conductive Gel Regulation Number: 21 CFR 882.1275 Regulation Name: Electroconductive media Regulatory Class: Class II Product Code: GYB Dated: September 14, 2022 Received: September 14, 2022
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222770
Device Name Conductive Gel
Indications for Use (Describe)
Intended for use with electric stimulation therapy devices, such as TENS and EMS. Conductive Gel is used with external electrodes to reduce the impedance of the contact between the electrode and the skin.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Date of the summary prepared: December 8, 2022
#### 2. Submitter's Information
Sponsor Name: Top-Rank Health Care Co. Ltd.
Address: Mashan Village, Dongguan Street, Shangyu District, Shaoxing City, Zhejiang
Province, China. 312300
Establishment Registration Number: Applying
Contact name: Gangfeng Lou
Title: Chairman of the board
Tel: +86-0575-82912999
Fax: +86-0575-82912999
E-mail: info@etop-rank.com
#### Application Correspondent:
Contact Person: Ms. Cassie Lee Company: Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
## 3. Subject Device Information
Type of 510(k): Traditional Company Name: Top-Rank Health Care Co. Ltd. Trade/Device Name: Conductive Gel
{4}------------------------------------------------
Model Name: TG01-50, TG01-65, TG01-90, TG01-100, TG01-125, TG01-160, TG01-200, TG01-250 510k Review Panel: Neurology Classification Name: Media, Electroconductive Product Code: GYB Regulation Number: 21 CFR 882.1275 Regulatory Class: II
#### Intended Use / Indications for Use:
Intended for use with electric stimulation therapy devices, such as TENS and EMS. Conductive Gel is used with external electrode to reduce the impedance of the contact between the electrode and the skin.
#### 4. Predicate Device Information
#### Predicate Device:
Sponsor: Guangzhou Xinbo Electronic Co., Ltd. Common Name: Media, Electroconductive Trade/Device Name: DR-HO'S Electro Therapy Conductive Gel Classification Name: Media, Electroconductive 510(K) Number: K200402 510k Review Panel: Neurology Product Code: GYB Regulation Number: 21 CFR 882.1275 Regulatory Class: II
#### Intended Use / Indications for Use:
DR-HO'S Electro Therapy Conductive Gel is intended for use with TENS (transcutaneous electrical nerve stimulation) and EMS (electric muscle stimulation) therapy. The Conductive Gel is used with external to reduce the impedance of the contact between the electrode surface and the skin.
{5}------------------------------------------------
## 5. Device Description
Conductive Gel can be used with any of dozens of stimulating devices that are legally sold, to reduce the impedance between the skin and the stimulating device. It consists of Purified water: 98.25%, Carbopol: 0.5%, Glycerin: 1%, Sodium hydroxide: 0.2% and Triclosan: 0.05%. And the Purified water used as the solvent, the Carbopol as a gel forming material, the Glycerin as a Moisturizing, the Sodium hydroxide as a Buffering and the Triclosan as a preservative. The Conductive Gel is used on intact skin surfaces. The entire surface of Conductive Gel is very conductive, smeared evenly on the stimulation device, so that the current is evenly distributed. The gel is to be generously applied to the area where an electrode will be used. The gel can be washed off the skin after use.
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Device Name<br>and Model | Conductive Gel<br>Model: TG01-50, TG01-65, TG01-90, TG01-100,<br>TG01-125, TG01-160,<br>TG01-200, TG01-250 | DR-HO'S Electro Therapy<br>Conductive Gel<br>Model: DHGEL | -- |
| 510(k) Number | K222770 | K200402 | -- |
| Product Code | GYB | GYB | SE |
| Intended Use /<br>Indications for<br>Use | Intended for use with<br>electric stimulation therapy<br>devices, such as TENS<br>and EMS. Conductive Gel<br>is used with external<br>electrode to reduce the<br>impedance of the contact | DR-HO'S Electro Therapy<br>Conductive Gel is intended<br>for use with TENS<br>(transcutaneous electrical<br>nerve stimulation) and EMS<br>(electric muscle stimulation)<br>therapy. The Conductive Gel | Similar<br>Note 1 |
| | | | |
| | between the electrode and<br>the skin. | is used with external to<br>reduce the impedance of the<br>contact between the<br>electrode surface and the<br>skin. | |
| Regulation<br>Number | 882.1275 | 882.1275 | SE |
| Body contact | Intact Skin | Intact Skin | SE |
| Sterilization | Non-Sterile | Non-Sterile | SE |
| Shelf life | 2 years | 2 years | SE |
| Impedance (at 1<br>MHz) | $\leq500\Omega$ | 500Ω | Similar<br>Note 2 |
| Material | Purified water (solvent);<br>Carbopol (gel forming);<br>Glycerin (Moisturizing);<br>Sodium hydroxide<br>(Buffering); Triclosan<br>(preservative) | Purified water (solvent);<br>Carbopol (gel forming);<br>Glycerin (Moisturizing);<br>Sodium hydroxide<br>(Buffering); Triclosan<br>(preservative) | SE |
| Percent<br>concentration of<br>each ingredient | - Purified water:<br>98.25%;<br>- Carbopol: 0.5%;<br>- Glycerin: 1%;<br>- Sodium hydroxide:<br>0.2%;<br>- Triclosan: 0.05% | - Purified water: 98.25%<br>- Carbopol: 0.5%<br>- Glycerin: 1%<br>- Sodium hydroxide: 0.2%<br>- Triclosan: 0.05% | SE |
| pH | 6.5 ~ 7.5 | 7.0-7.5 | Similar<br>Note 3 |
| Preservative | Triclosan | Triclosan | SE |
| Biocompatibility | Complied with ISO 10993-<br>5, ISO 10993-10 | Complied with ISO 10993-5,<br>ISO 10993-10 | SE |
| Cytotoxicity | Yes | Yes | SE |
| Irritation | Yes | Yes | SE |
| Sensitization | Yes | Yes | SE |
| Operating<br>Environment | Temperature: 5~40°C<br>Humidity: ≤80%RH<br>Atmospheric Pressure:<br>86~106kPa | Temperature: 5~40°C<br>Humidity: ≤80%RH<br>Atmospheric Pressure:<br>86~106kPa | SE |
| Storage<br>Environment | Temperature: 5~40°C<br>Humidity: ≤95% RH<br>Atmospheric Pressure:<br>50~106 kPa | Temperature: 5~40°C<br>Humidity: ≤95% RH<br>Atmospheric Pressure:<br>50~106 kPa | SE |
#### 6. Comparison to predicate device and conclusion
{6}------------------------------------------------
{7}------------------------------------------------
## Comparison in Detail(s):
## Note 1:
Although the "Intended Use / Indications for Use" of subject device is slightly different from predicate device, it's just a difference in how they are described. So, the slight difference will not raise any safety or effectiveness issue.
## Note 2:
Although the "Impedance" of the subject device is lower than the predicate device, it does not affect the effectiveness of the product in use. So, the slight difference in Impedance will not raise any safety or effectiveness issues.
Note 3:
{8}------------------------------------------------
Although the "pH" of the subject device is a little different from the predicate device, both of them are close to the pH value of the human skin surface, the pH is closed to 7 (neutral). So, the slight difference in pH will not raise any safety or effectiveness issues.
## 7. Test Summary
## 7.1 Summary of Non-Clinical Testing
Conductive Gel (Model: TG01-50, TG01-65, TG01-90, TG01-100, TG01-125, TG01-160, TG01-200, TG01-250) has been evaluated the safety and performance by lab bench testing as following:
| Test Item | Purpose of the test | Reference Standard | Acceptance criteria | Test results |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| In vitro<br>Cytotoxicity<br>Test | Under the research conditions, determine whether the target device extract is cytotoxic. | ISO 10993-5:2009<br>Biological evaluation of medical devices-<br>Part 5: Tests for in vitro cytotoxicity | Under the conditions of this study, the Test Sample were non-cytotoxicity to L929 cells. | Pass |
| Skin<br>Sensitization<br>Test | Under the research conditions, to evaluate the possibility of skin sensitization after topical applications of the test sample on the skin of guinea pig. | ISO 10993-10:2010<br>Biological evaluation of medical devices-<br>Part 10: Tests for irritation and skin sensitization | Under the conditions of this study, the frequency of positive challenge results in sample extract and Negative control animals are 0%, the Positive control is 100%. | Pass |
| Skin Irritation<br>Test | Under the<br>research<br>conditions, to<br>evaluate the<br>possibility of skin<br>irritation after<br>single topical<br>applications of the<br>test sample on the<br>skin of rabbits. | ISO 10993-10:2010<br>Biological evaluation<br>of medical devices-<br>Part 10: Tests for<br>irritation and skin<br>sensitization | Under the<br>conditions of this<br>study, the test<br>sample induce<br>negligible irritation<br>in a rabbit skin<br>single exposure<br>test. | Pass |
| Usability<br>Study | To study how<br>usability will be<br>performed in<br>subject device in<br>order to comply<br>with IEC 62366-1 | IEC 62366-1 Edition<br>1.0 2015-02, Medical<br>devices - Part 1:<br>Application of usability<br>engineering to medical<br>devices [Including<br>CORRIGENDUM 1<br>(2016)] | The subject device<br>can meet the<br>usability goal of<br>IEC 62366-1<br>standards. | Pass |
| Shelf Life<br>Test | To study whether<br>the performance<br>of subject device<br>can meet the 2-<br>year shelf life<br>requirements | ASTM F1980-16:<br>Standard Guide for<br>Accelerated Aging of<br>Sterile Barrier<br>Systems for Medical<br>Devices;<br>Guidance document<br>for the "Shelf Life of<br>Medical Devices"<br>issued in April 1991 | All items (Exterior,<br>Package, pH,<br>Impedance and<br>conductivity)<br>tested on both<br>before and after<br>aging samples<br>meet the<br>performance<br>required. | Pass |
{9}------------------------------------------------
{10}------------------------------------------------
# 7.2 Summary of Non-Clinical Testing
No clinical study is included in this submission.
## 8. Final Conclusion
The subject device "Conductive Gel, (Model: TG01-50, TG01-65, TG01-90, TG01-100, TG01-125, TG01-160, TG01-200, TG01-250)" is as safe and effective and substantially equivalent to the legally marketed predicated K200402.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.