K222760 · Biomet, Inc. · MBB · Dec 28, 2022 · Orthopedic
Device Facts
Record ID
K222760
Device Name
StageOne™ Select Hip Cement Spacer Molds
Applicant
Biomet, Inc.
Product Code
MBB · Orthopedic
Decision Date
Dec 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The StageOne™ Select Hip Cement Spacer Molds are intended to create a temporary hemi-hip replacement prosthesis as part of a two-stage revision procedure.
Device Story
Sterile, single-use silicone molds with stainless steel reinforcement stems, adapters, and inserts. Used by surgeons in clinical settings to create temporary hemi-hip prostheses. Process: Refobacin Bone Cement R injected into molds; assembled with neck length adapter; inserted into femoral medullary cavity following debridement. Acts as temporary spacer during two-stage revision for septic hip. Remains in situ ≤180 days until permanent implant surgery. Benefits: maintains joint space and provides local antibiotic delivery during infection treatment. Requires patient use of mobility aids due to cement material mechanical limitations.
Clinical Evidence
No clinical data provided. Substantial equivalence established via non-clinical bench testing, including biocompatibility, packaging, shelf-life, sterilization validation, MRI analysis, fatigue performance, antibiotic elution, and pyrogenicity testing.
Technological Characteristics
Silicone molds with stainless steel reinforcement stems, head inserts, and neck length adapters. Single-use. Sterile. Double sterile barrier packaging. Compatible with Refobacin Bone Cement R. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing two-stage revision for septic process. Used to mold temporary hemi-hip prosthesis from Refobacin Bone Cement R, inserted into femoral medullary cavity/acetabular cavity after implant removal and debridement. Used with systemic antimicrobial therapy. Not for use >180 days; must be explanted. Indicated only for patients using mobility assist devices (crutches, walkers).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
StageOne™ Select Disposable Cement Spacer Molds for Making Temporary Hemi-Hip Prosthesis with Reinforcement Stem (K161166)
Reference Devices
StageOne™ Select Cement Spacer Molds for Temporary Hip Replacement (K080979)
Submission Summary (Full Text)
{0}------------------------------------------------
December 28, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biomet, Inc. Meredith Reed Regulatory Affairs Senior Specialist 56 East Bell Drive, P.O. Box 587 Warsaw, Indiana 46581
Re: K222760
Trade/Device Name: StageOne™ Select Hip Cement Spacer Molds Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: MBB, KWY, KWL Dated: December 1, 2022 Received: December 5, 2022
Dear Meredith Reed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laura C. Rose -S
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222760
Device Name
StageOne™ Select Hip Cement Spacer Molds
#### Indications for Use (Describe)
StageOne™ Select Hip Cement Spacer Molds with stainless steel reinforcement stems, adapters and inserts are indicated for use to mold a temporary hemi-hip replacement for skeletally mature patients undergoing a two-stage revision procedure due to a septic process. The temporary prosthesis is molded using Refobacin Bone Cement R, assembled and inserted into the femoral medullary cavity following removal of the existing femoral and acetabular replacement implants and debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
The hemi-hip prosthesis made from the StageOne™ Select Cement Spacer Molds is not intended for use more than 180 days, at which time it must be explanted and permanent devices implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.).
Due to the inherent mechanical limitations of the hemi-hip prosthesis material (Refobacin Bone Cement R), the temporary hemi-hip prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
| Department of Health and Human Services | |
|-----------------------------------------|--|
| Food and Drug Administration | |
| Office of Chief Information Officer | |
| Paperwork Reduction Act (PRA) Staff | |
| PRAStaff@fda.hhs.gov | |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the StageOne™ Select Hip Cement Spacer Molds 510(k) Premarket Notification. The submission was prepared in accordance with the FDA guidance document, "Format for Traditional and Abbreviated 510(k)s", issued September 13, 2019.
| Sponsor: | Biomet, Inc.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, Indiana 46581, USA<br>Establishment Registration Number: 1825034 |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Meredith Reed<br>Regulatory Affairs Senior Specialist<br>Telephone: (574) 209-6028 |
| Date: | 27 December 2022 |
| Subject Device: | Trade Name: StageOne™ Select Hip Cement Spacer Molds<br>Common Name: Temporary Bone Cement Hemi-Hip Prosthesis<br>Classification Name:<br>• MBB – Polymethylmethacrylate (PMMA) bone cement (21 CFR 888.3027)<br>• KWY – Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented or Uncemented (21 CFR 888.3390)<br>• KWL – Prosthesis, Hip, Hemi-, Femoral, Metal (21 CFR 888.3360) |
| Predicate Device: | K161166 StageOne™ Select Disposable<br>Cement Spacer Molds for Making<br>Temporary Hemi-Hip Prosthesis with<br>Reinforcement Stem Biomet, Inc. |
| Reference Device: | K080979 StageOne™ Select Cement Spacer<br>Molds for Temporary Hip<br>Replacement Biomet, Inc. |
| Purpose and Device Description: | The purpose of this submission is to obtain clearance for the proposed<br>modifications to the manufacturing process, sterile packaging<br>configuration, product labeling including adding MR Unsafe labeling,<br>engineering drawings and indications for use regarding bone cement<br>marketed for use with the StageOne™ Select Hip Cement Spacer Molds. |
{4}------------------------------------------------
| | The StageOne™ Select Hip Cement Spacer Molds are sterile, single use<br>medical devices made of silicone with a stainless steel reinforcement<br>stem, head insert and neck length adapter. The device is used to create a<br>temporary hip implant component made from antibiotic bone cement,<br>Refobacin® Bone Cement R by injecting with a dispenser/gun into the<br>mold. After removal of the initial femoral and acetabular implants, the<br>prepared cement spacers are assembled using the neck length adapter and<br>placed into the femoral joint space using Refobacin® Bone Cement R as<br>the first stage of a two-stage revision surgical procedure. The temporary<br>hemi-hip prosthesis remains in place (180 days or less) until the second<br>stage of the two-stage revision procedure is performed to implant a<br>conventional hip joint prosthesis. |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The StageOne™ Select Hip Cement Spacer Molds are intended to create<br>a temporary hemi-hip replacement prosthesis as part of a two-stage<br>revision procedure. |
| Indications for Use: | StageOne™ Select Hip Cement Spacer Molds with stainless steel<br>reinforcement stems, adapters and inserts are indicated for use to mold a<br>temporary hemi-hip replacement for skeletally mature patients<br>undergoing a two-stage revision procedure due to a septic process. The<br>temporary prosthesis is molded using Refobacin Bone Cement R,<br>assembled and inserted into the femoral medullary canal and acetabular<br>cavity following removal of the existing femoral and acetabular<br>replacement implants and debridement. The device is intended for use in<br>conjunction with systemic antimicrobial antibiotic therapy (standard<br>treatment approach to an infection). |
| | The hemi-hip prosthesis made from the StageOne™ Select Hip Cement<br>Spacer Molds is not intended for use more than 180 days, at which time it<br>must be explanted and permanent devices implanted or another<br>appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.). |
| | Due to the inherent mechanical limitations of the hemi-hip prosthesis<br>material (Refobacin Bone Cement R), the temporary hemi-hip prosthesis<br>is only indicated for patients who will consistently use traditional<br>mobility assist devices (e.g. crutches, walkers) throughout the implant<br>period. |
| Summary of Technological<br>Characteristics: | The rationale for substantial equivalence is based on consideration of the<br>following characteristics:<br>• Intended Use: Identical to the predicate device.<br>• Indications for Use: The subject device has similar indications for<br>use as the identified predicate device. Subject device is now only<br>indicated for use with Refobacin® Bone Cement R cleared under<br>K171540. There are no changes to the intended clinical use of the<br>device. |
{5}------------------------------------------------
- . Materials: The subject device is made of the same raw materials as the identified predicate device.
- Design Features: The subject device has the same design features as ● the identified predicate device.
- Variants/Sizes: The subject device has the same variants and sizes as the identified predicate device.
- Sterilization: Identical to the predicate device. ●
- Packaging: The subject device has similar packaging to the predicate ● device. Both the subject and predicate devices are packaged using a double sterile barrier configuration.
- Manufacturing Process: The subject device is manufactured using a ● manufacturing process similar to the predicate device.
#### Summary of Performance Data (Nonclinical and/or Clinical):
- Non-Clinical Testing: Non-clinical performance testing was ● conducted to support the proposed modifications for the subject StageOne Select Hip Cement Spacer Molds and establish substantial equivalence between the subject device and the identified predicate.
- Biocompatibility Evaluation o
- Packaging testing O
- Shelf-life testing O
- O Sterilization Validation
- Magnetic Resonance Imaging (MRI) Analysis O
- Fatigue performance testing O
- Antibiotic elution testing O
- BET and pyrogenicity testing O
- Clinical Testing: Clinical data and conclusions were deemed not ● necessary to establish substantial equivalence between the subject StageOne Select Hip Cement Spacer Molds and the identified predicate for the proposed device modifications.
The subject device has the same intended use, similar clinical indications for use as the predicate device. There are no changes to the design features, materials, operating principle, shelf-life or sterilization method. Except for the modifications described in this submission the subject device is identical to the predicate device, and the performance data and analyses demonstrate that:
- O Any differences do not raise new questions of safety and effectiveness; and
- The proposed device is at least as safe and effective as the legally o marketed predicate device.
## Substantial Equivalence Conclusion:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.