K222750 · Neocis, Inc. · QRY · Dec 8, 2022 · Dental
Device Facts
Record ID
K222750
Device Name
Yomi Robotic System
Applicant
Neocis, Inc.
Product Code
QRY · Dental
Decision Date
Dec 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Yomi Robotic System is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System.
Device Story
Yomi Robotic System is a dental stereotaxic instrument providing robotic navigational guidance for dental implantation and alveoplasty. Input data consists of patient CBCT scans processed via YomiPlan software to create 3-D anatomical models for virtual surgical planning. During surgery, the system uses a robotic arm to hold and guide third-party powered bone-cutting instruments. The system tracks patient position via physical linkages (Patient Tracker, Kinematic Mount, Tracker End Effector) attached to splints secured to the patient's mouth. The system provides physical, visual, and audible haptic feedback to the surgeon, restricting instrument movement to the pre-defined surgical volume. Used in dental clinics by clinicians; the system enhances surgical precision by ensuring instrument placement aligns with the pre-operative plan, potentially improving implant outcomes.
Clinical Evidence
Bench testing only. Performance validation included total system accuracy, kinematic mount repeatability, and drill jig accuracy. High-level disinfection validation was performed per AAMI TIR12 and FDA reprocessing guidance. Usability validation testing was conducted with dental clinicians to evaluate critical tasks related to the modified disinfection instructions.
Technological Characteristics
Dental stereotaxic instrument; 6-degree-of-freedom robotic arm; haptic feedback. Materials: TEE main body (Aluminum 7075), TEE thumb screws (Stainless Steel 316). Patient tracking via physical linkages (Patient Tracker, Kinematic Mount, Tracker End Effector) to splints. Reprocessing: High-level disinfection. Connectivity: YomiPlan software for pre-operative planning on third-party PCs.
Indications for Use
Indicated for partially edentulous and fully edentulous adult patients who qualify for dental implants, requiring assistance in pre-operative planning and intra-operative robotic navigational guidance for dental implantation or guided bone reduction (alveoplasty).
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Neocis Inc. Joshua Davis Regulatory Affairs Manager 2800 Biscayne Blvd Suite 600 Miami, Florida 33137
Re: K222750
Trade/Device Name: Yomi Robotic System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: QRY, PLV Dated: September 29, 2022 Received: September 30, 2022
Dear Joshua Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K222750
Device Name Yomi Robotic System
#### Indications for Use (Describe)
Yomi Robotic System is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a gray horizontal line that extends to the right, where it intersects with a vertical line. At the intersection of the two lines, there is a small circle.
510(k) Summary K22750
#### I. Submitter
Neocis Inc. 2800 Biscayne Blvd. Suite 600 Miami, FL 33137 Tel: 1-855-9NEOCIS
Contact Person: Joshua Davis, Regulatory Affairs Manager Date Prepared: December 7, 2022
## II. Device
| Trade Name: | Yomi Robotic System |
|----------------------|-----------------------------------------------------------|
| Common Name: | Dental Stereotaxic Instrument |
| Classification Name: | Bone cutting instrument and accessories (21 CFR 872.4120) |
| Classification: | Class II |
| Product Code: | QRY, PLV |
## III. Predicate Devices
Primary Predicate: Yomi Robotic System (K222049) Reference Device: Neocis Guidance System (K191605) Reference Device: Neocis Guidance System (K161399) Reference Device: Resert XL HD High-Level Disinfectant (K091022)
#### IV. Indications for Use
Yomi Robotic System is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre- operative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System.
#### V. Device Description
Yomi Robotic System is a dental stereotaxic instrument and a powered surgical device for bone cutting. Yomi Robotic System is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a thin gray line that extends horizontally. To the right of the word, there is a gray crosshair symbol with a small circle in the center.
surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
The Yomi Robotic System allows the user to plan the surgery virtually in YomiPlan, cleared for use alone on third-party PCs for preplanning. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used for the system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The Yomi robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument.
The patient tracking portion of Yomi is comprised of linkages from the patient to Yomi, which include the Clamped Chairside Patient Splint (C-CPS) or YomiLink Bone (YLB), the Tracker End Effector (TEE) and the Patient Tracker (PT). The Patient Splint is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials prior to the presurgical CBCT scan. The EPS is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required).
The subject of this submission is to modify the design and reprocessing method for the Tracker End Effector (TEE) of the Yomi Robotic System. All other aspects of the Yomi Robotic System remain unchanged from prior clearances.
## VI. Comparison of Technological Characteristics
The following Table 1 provides a summary of the subject Yomi Robotic System features compared to the predicate device, Yomi Robotic System (K222049).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a golden color. Above the word is a gray horizontal line that extends to the left and right of the word. A gray circle is above the "i" in "neocis", and a vertical gray line extends upwards from the circle.
| Technological<br>Characteristics | Subject Device: Yomi Robotic<br>System | Primary Predicate: Yomi<br>Robotic System (K222049) | Comparison |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use (IFU) | Yomi Robotic System is a<br>computerized robotic navigational<br>system intended to provide<br>assistance in both the planning<br>(pre-operative) and the surgical<br>(intra-operative) phases of dental<br>implantation surgery. The system<br>provides software to preoperatively<br>plan dental implantation<br>procedures and provides robotic<br>navigational guidance of the<br>surgical instruments. The system<br>can also be used for planning and<br>performing guided bone reduction<br>(also known as alveoplasty) of the<br>mandible and/or maxilla. Yomi is<br>intended for use in partially<br>edentulous and fully edentulous<br>adult patients who qualify for<br>dental implants.<br>When YomiPlan software is used<br>for preplanning on third party PCs,<br>it is intended to perform the<br>planning (pre-operative) phase of<br>dental implantation surgery.<br>YomiPlan provides pre-operative<br>planning for dental implantation<br>procedures using the Yomi Robotic<br>System. The output of YomiPlan is<br>to be used with the Yomi Robotic<br>System. | Yomi Robotic System is a<br>computerized robotic<br>navigational system intended<br>to provide assistance in both<br>the planning (pre-operative)<br>and the surgical (intra-<br>operative) phases of dental<br>implantation surgery. The<br>system provides software to<br>preoperatively plan dental<br>implantation procedures and<br>provides robotic navigational<br>guidance of the surgical<br>instruments. The system can<br>also be used for planning and<br>performing guided bone<br>reduction (also known as<br>alveoplasty) of the mandible<br>and/or maxilla. Yomi is<br>intended for use in partially<br>edentulous and fully<br>edentulous adult patients who<br>qualify for dental implants.<br>When YomiPlan software is<br>used for preplanning on third<br>party PCs, it is intended to<br>perform the planning (pre-<br>operative) phase of dental<br>implantation surgery. Yomi<br>Plan provides pre-operative<br>planning for dental<br>implantation procedures using<br>the Yomi Robotic System.<br>The output of Yomi Plan is to<br>be used with the Yomi<br>Robotic System. | Equivalent |
| Principles of<br>Operation | The Tracker End Effector<br>physically attaches to, and provides<br>linkage between, the Patient<br>Tracker of the Yomi Robotic<br>System and the Patient Splint. The<br>Tracker End Effector is connected<br>to the Patient Tracker via a screw at<br>the Patient Tracker Flange. The | The Tracker End Effector<br>physically attaches to, and<br>provides linkage between, the<br>Patient Tracker of the Yomi<br>Robotic System and the Patient<br>Splint. The Tracker End<br>Effector is connected to the<br>Patient Tracker via a screw at | Equivalent |
| Technological<br>Characteristics | Subject Device: Yomi Robotic<br>System | Primary Predicate: Yomi<br>Robotic System (K222049) | Comparison |
| | Tracker End Effector is then also<br>connected via screw to the Patient<br>Splint. | the Patient Tracker<br>Flange. The Tracker End<br>Effector is then also connected<br>via screw to the Patient Splint. | |
| Robotic Guide<br>Arm | Guided robotic arm | Guided robotic arm | Equivalent |
| Movement<br>Direction | Guided Robotic Arm holds a surgical<br>instrument and provides haptic<br>feedback on position with respect to<br>the plan restricting movement outside<br>of volume predefined during<br>planning.<br>6 degrees of freedom | Guided Robotic Arm holds a<br>surgical instrument and<br>provides haptic feedback on<br>position with respect to the<br>plan restricting movement<br>outside of volume predefined<br>during planning.<br>6 degrees of freedom | Equivalent |
| Patient affixed<br>tracking parts | Splints with arrays | Splints with arrays | Equivalent |
| Patient Tracking<br>Mechanism | Physical linkage to patient via Patient<br>Tracker (PT), Kinematic Mount<br>(KM), and Tracker End Effector<br>(TEE) connected to splints | Physical linkage to patient via<br>Patient Tracker (PT),<br>Kinematic Mount (KM), and<br>Tracker End Effector (TEE)<br>connected to splints | Equivalent |
| Fiducials for CT<br>scan | Fiducial Array (FA) attached to splint | Fiducial Array (FA) attached<br>to splint | Equivalent |
| Patient Contact | No contact | No contact | Equivalent |
| Reprocessing<br>Classification | Non-critical | Non-critical | Equivalent |
| Reprocessing<br>Method for<br>Tracker End<br>Effector | High level disinfection | Sterilization | Equivalent<br>Disinfection<br>validation testing<br>has demonstrated<br>disinfection of the<br>subject device<br>provides adequate<br>microbicidal<br>reprocessing |
| Mating<br>Component<br>Design | V-coupled design | Fiducial pins and spheres | Equivalent<br>Verification testing<br>has demonstrated the<br>modified design has<br>no impact on |
| Technological<br>Characteristics | Subject Device: Yomi Robotic<br>System | Primary Predicate: Yomi<br>Robotic System (K222049) | Comparison |
| | | | substantial<br>equivalence<br>Equivalent |
| Materials | TEE main body: Aluminum 7075<br>TEE thumb screws: Stainless Steel<br>316 | TEE main body: Aluminum<br>6061<br>TEE thumb screws: Stainless<br>Steel 316 | Verification<br>testing has<br>demonstrated the<br>modified material<br>has no<br>impact on substantial<br>equivalence |
| Performance<br>Testing | Total System Accuracy<br>Kinematic Mount<br>Repeatability<br>Drill Jig Accuracy<br>Disinfection Validation<br>Reprocessing Instruction<br>Validation | Total System Accuracy<br>Kinematic Mount<br>Repeatability<br>Drill Jig Accuracy | Equivalent |
## Table 1: Comparison of technological characteristics to the predicate
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Neocis. The word "neocis" is written in a sans-serif font in gold color. A thin gray line is above the word, and a thin gray line intersects the word at the "i", forming a crosshair-like symbol. A small gray circle is at the intersection of the lines.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Neocis. The logo consists of the word "neocis" in a sans-serif font, with the letters in a light orange color. Above the word is a thin gray horizontal line that extends to the left of the "n" and to the right of the "s". A thin gray vertical line intersects the horizontal line above the "i", forming a cross shape with a small circle at the intersection.
#### VII. Performance Testing
The following testing has been fully executed to ensure that the subject device functions as intended:
- Total System Accuracy Verification
- Kinematic Repeatability Verification
- Drill Jig Accuracy Verification ●
- High-Level Disinfection Validation was performed in accordance with recommended evaluations as listed in AAMI TIR12 and Guidance for Industry and FDA Staff -Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling
Usability validation testing for the modified reprocessing instructions has been performed. Usability testing included dental clinician users who were evaluated on critical tasks for the disinfection steps per the disinfection instructions for use and provided responses to open-ended questions.
#### VIII. Conclusion
This subject of this submission is to modify the design and reprocessing method for the Tracker End Effector (TEE) of the Yomi Robotic System. There are no changes to the intended use compared to the predicate device. There are no fundamental changes to the technology. The performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.