Disposable Nitrile Examination Glove (Tested for use with Chemotherapy Drug)
K222733 · Inner Mongolia Boming Medical Supplies Co., Ltd. · LZA · Dec 11, 2022 · General Hospital
Device Facts
Record ID
K222733
Device Name
Disposable Nitrile Examination Glove (Tested for use with Chemotherapy Drug)
Applicant
Inner Mongolia Boming Medical Supplies Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Dec 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, powder-free, ambidextrous nitrile examination gloves; blue color; beaded cuff. Intended for medical use to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Available in sizes XS, S, M, L, XL. Non-sterile. Used in clinical settings by healthcare providers. Provides barrier protection during patient examinations and handling of chemotherapy agents.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated per ASTM D6319-19 (physical dimensions, tensile strength, elongation), ASTM D5151-19 (watertightness/holes), ASTM D6124-06 (powder content), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).
Indicated for use as a disposable patient examination glove worn on the examiner's hands to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for Use with Chemotherapy Drugs) (K201390)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
December 11, 2022
Inner Mongolia Boming Medical Supplies Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No. 161 East Lujiazui Rd.,Pudong Shanghai, 200120 China
Re: K222733
Trade/Device Name: Disposable Nitrile Examination Glove (Tested for use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ Dated: November 14, 2022 Received: November 14, 2022
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Qian., M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K222733
#### Device Name
Disposable Nitrile Examination Glove(Tested for Use with Chemotherapy Drugs)
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
|----------------------------|------------------------|----------------------------------------|
| Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 23.7 Minutes |
| Cisplatin | 1.0 mg/ml(1,000 ppm) | 240 Minutes |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml(20,000 ppm) | 240 Minutes |
| Dacarbazine (DTIC) | 10.0 mg/ml(10,000 ppm) | 240 Minutes |
| Doxorubicin HCl | 2.0 mg/ml(2,000 ppm) | 240 Minutes |
| Etoposide | 20.0 mg/ml(20,000 ppm) | 240 Minutes |
| Fluorouracil | 50.0 mg/ml(50,000 ppm) | 240 Minutes |
| Methotrexate | 25 mg/ml(25,000 ppm) | 240 Minutes |
| Mitomycin C | 0.5 mg/ml(500 ppm) | 240 Minutes |
| Paclitaxel | 6.0 mg/ml(6,000 ppm) | 240 Minutes |
| Thio Tepa | 10.0 mg/ml(10,000 ppm) | 45.7 Minutes |
| Vincristine Sulfate | 1.0 mg/ml(1,000 ppm) | 240 Minutes |
Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.7 Minutes Thio-Tepa 10.0 mg/ml 45.7 Minutes
Caution: Testing showed an average breakthrough time of 45.7 minutes with Thio-Tepa
WARNING: Do not use with Carmustine Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K222733
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
#### 1.0 Submitter's Information
| Name: | Inner Mongolia Boming Medical Supplies Co., Ltd. | |
|----------|-------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | Room 326, New Industrial Park Management Committee Office,<br>Tumet Right Banner, Baotou City, Inner Mongolia Autonomous<br>Region, China | |
| Contact: | Guo Hua | |
| Tel: | +86-13485097856 | |
#### Designated Submission Correspondent
| Contact: | Mr. Boyle Wang |
|----------|------------------------------------------------------------------------------------------------------------------------------|
| Address: | Shanghai Truthful Information Technology Co., Ltd.<br>Room 1801, No.161 East Lu Jiazui Rd., Pudong,Shanghai,<br>200120 China |
| Tel: | +86-21-50313932 |
| Email: | info@truthful.com.cn |
Date submitted: Dec.7,2022
#### 2.0 Device Information
| Trade name: | Disposable Nitrile Examination Glove (Tested for Use with<br>Chemotherapy Drugs) |
|----------------------|----------------------------------------------------------------------------------|
| Common name: | Patient Examination Gloves |
| Classification name: | Non-powdered patient examination glove |
| Model(s): | XS, S, M, L, XL |
| Production code: | LZA,LZC, OPJ |
| Regulation number: | 21CFR880.6250 |
| Classification: | Class I |
| Panel: | General Hospital |
#### 3.0 Predicate Device Information
| Manufacturer: | Medline Industries, Inc. |
|---------------|-----------------------------------------------------------------------------------------------|
| Device: | Medline Powder-Free Light Blue Nitrile Exam Glove (Tested for<br>Use with Chemotherapy Drugs) |
{4}------------------------------------------------
510(k) number: K201390
#### 4.0 Indication for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs, per ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug | Concentration | Breakthrough Detection<br>Time in Minutes |
|-------------------------------|------------------------|-------------------------------------------|
| Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 23.7 |
| Cisplatin | 1.0 mg/ml(1,000 ppm) | > 240 |
| Cyclophosphamide<br>(Cytoxan) | 20.0 mg/ml(20,000 ppm) | > 240 |
| Dacarbazine (DTIC) | 10.0 mg/ml(10,000 ppm) | > 240 |
| Doxorubicin HCI | 2.0 mg/ml(2,000 ppm) | > 240 |
| Etoposide | 20.0 mg/ml(20,000 ppm) | > 240 |
| Fluorouracil | 50.0 mg/ml(50,000 ppm) | > 240 |
| Methotrexate | 25 mg/ml(25,000 ppm) | > 240 |
| Mitomycin C | 0.5 mg/ml(500 ppm) | > 240 |
| Paclitaxel | 6.0 mg/ml(6,000 ppm) | > 240 |
| Thio Tepa | 10.0 mg/ml(10,000 ppm) | 45.7 |
| Vincristine Sulfate | 1.0 mg/ml(1,000 ppm) | > 240 |
Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.7 Minutes;
Thio Tepa 10.0 mg/ml 45.7 Minutes.
Caution: Testing showed an average breakthrough time of 45.7 minutes with Thio-Tepa
WARNING: Do not use with Carmustine.
#### 5.0 Device Description
The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered non-sterile and in five sizes: extra-small, small, medium, large, and extra-large.
#### 6.0 Technological Characteristic Comparison Table
{5}------------------------------------------------
| Item | Subject Device<br>(K222733) | Predicated Device<br>(K201390) | Remark |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | LZA,LZC, OPJ | LZA,LZC | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use /<br>Indication for Use | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner.<br>These gloves were tested<br>for use with<br>chemotherapy drugs as<br>per ASTM D6978-<br>05(2019) Standard<br>Practice for Assessment<br>of Resistance of Medical<br>Gloves to Permeation<br>by Chemotherapy Drugs. | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner.<br>These gloves were tested<br>for use with chemotherapy<br>drugs as per ASTM<br>D6978-05(2019) Standard<br>Practice for Assessment of<br>Resistance of Medical<br>Gloves to Permeation by<br>Chemotherapy Drugs. | Same |
| Powdered or<br>Powered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Labeling<br>Information | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, non-Sterile,<br>a statement of standard<br>ASTM D6978-05<br>compliance and a<br>summary of the testing<br>results | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, non-Sterile, a<br>statement of standard<br>ASTM D6978-05<br>compliance and a<br>summary of the testing<br>results. | Same |
## Table2 Device Dimensions Comparison
| Predicate<br>Device(K201390) | Designation | Size | | | | | Tolerance |
|------------------------------|-------------|------|-----|-----|-----|-----|-----------|
| | Length, mm | XS | S | M | L | XL | |
| | Width, mm | NA | 240 | 240 | 240 | 240 | min |
| | | NA | 85 | 95 | 105 | 115 | ±10 |
| Thickness, mm: | | | | | | | |
{6}------------------------------------------------
| | Finger | 0.16 | min | | | | |
|-----------------------------|----------------|------|-----|-----|-----|-----|-----------|
| | Palm | 0.14 | min | | | | |
| Subject Device<br>(K222733) | Designation | Size | | | | | |
| | | XS | S | M | L | XL | Tolerance |
| | Length, mm | 220 | 220 | 230 | 230 | 230 | min |
| | Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | min | | | | |
| | Palm | 0.05 | min | | | | |
| Remark | Different | | | | | | |
Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.
| Item | Subject device<br>(K222733) | Predicated device<br>(K201390) | Remark | | |
|------------------------|-----------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------|------------|-----------|
| Colorant | Blue | Light Blue | Different | | |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14Mpa, min | 17Mpa, min | Different |
| | | Ultimate<br>Elongation | 500% min | 500% min | Same |
| | After<br>Aging | Tensile<br>Strength | 14Mpa, min | 14Mpa, min | Same |
| | | Ultimate<br>Elongation | 400%min | 400%min | Same |
| Comply with ASTM D6319 | | Comply with ASTM D6319 | Comply with ASTM D6319 | Same | |
| Freedom from Holes | | Be free from holes when tested in accordance with<br>ASTMD5151 AQL=2.5 | Be free from holes when tested in accordance with<br>ASTMD5151<br>AQL=2.5 | Same | |
| Powder Content | | Meet the requirements of<br>ASTM D6124<2.0mg | Meet the<br>requirements of<br>ASTM D6124 | Same | |
| | | Carmustine (BCNU) 3.3<br>mg/ml: 23.7 Minutes | Carmustine (BCNU)<br>3.3 mg/ml:<br>25.3 Minutes | Different | |
| | | Cisplatin 1.0 mg/ml: > 240<br>Minutes | Cisplatin 1.0 mg/ml:<br>≥240<br>Minutes | Same | |
#### Table3 Performance Comparison
{7}------------------------------------------------
| | Cyclophosphamide<br>(Cytoxan) 20.0 mg/ml: > 240 Minutes | Cyclophosphamide<br>(Cytoxan)<br>20.0 mg/ml: ≥240 Minutes | Same |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------------|-----------|
| Chemotherapy<br>Drugs<br>Tested with Minimum<br>Breakthrough<br>Detection Time as<br>Tested per ASTM D<br>6978 | Dacarbazine (DTIC) 10.0 mg/ml: > 240 Minutes | Dacarbazine (DTIC)<br>10.0 mg/ml:<br>≥240 Minutes | Same |
| | Doxorubicin HCl 2.0 mg/ml: > 240 Minutes | Doxorubicin<br>Hydrochloride 2.0<br>mg/ml: ≥240 Minutes | Same |
| | Etoposide 20.0 mg/ml: > 240 Minutes | Etoposide (Toposar)<br>20.0<br>mg/ml: ≥240 Minutes | Same |
| | Fluorouracil 50.0 mg/ml: > 240 Minutes | Fluorouracil 50.0<br>mg/ml: ≥240<br>Minutes | Same |
| | Methotrexate 25 mg/ml: > 240 Minutes | Methotrexate 25<br>mg/ml: ≥240<br>Minutes | Same |
| | Mitomycin C 0.5 mg/ml: > 240 Minutes | Mitomycin C 0.5<br>mg/ml: ≥ 240<br>Minutes | Same |
| | Paclitaxel 6.0 mg/ml: >240 Minutes | Paclitaxel (Taxol)<br>6.0 mg/ml:<br>≥240 Minutes | Same |
| | Thio Tepa 10.0 mg/ml: 45.7 Minutes | Thio-Tepa 10.0<br>mg/ml: 43.7<br>Minutes | Different |
| | Vincristine Sulfate 1.0<br>mg/ml: > 240 Minutes | Vincristine Sulfate<br>(Oncovin) 1.0<br>mg/ml: ≥240 Minutes | Same |
#### Table4 Safety Comparison
| Item | Subject device (K222733) | Predicated device (K201390) | Remark |
|---------------------------------------------------------------------------------------------|----------------------------------------------------|-----------------------------|--------|
| Material | Nitrile | Nitrile | Same |
| Irritation (ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests For | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | Same |
{8}------------------------------------------------
| | Irritation And Skin<br>Sensitization) | | | |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------|
| Biocompatibility | Sensitization (ISO<br>10993-10:2010<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin<br>Sensitization) | Under<br>conditions of<br>the study, not<br>a sensitizer. | | |
| | Cytotoxicity (ISO<br>10993-5:2009<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 5: Tests For In<br>Vitro Cytotoxicity) | Under<br>conditions of<br>the<br>study,<br>device extract<br>is<br>not<br>cytotoxic | Under<br>conditions of<br>the<br>study,<br>device<br>extract<br>is<br>cytotoxic | Different |
| | Acute Systemic<br>Toxicity(ISO<br>10993-<br>11:2017,Biological<br>evaluation of<br>medical devices -<br>Part 11: Tests for<br>systemic toxicity.) | N/A | Under<br>conditions of<br>the<br>study,<br>device<br>extract<br>is<br>non-toxic | Different |
#### 8.0 Summary of Non-clinical Testing
The biocompatibility evaluation for Disposable Nitrile Medical Examination Glove (Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards:
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
#### Performance Testing (Bench)
Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
{9}------------------------------------------------
Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.
In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:
- ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual -Powder on Medical Gloves
- -ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
- -ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
- -ASTM D 6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
| Test Method | Purpose | Acceptance Criteria | Results |
|-------------|--------------------------------|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D6319 | Physical<br>Dimensions<br>Test | Length(mm): | Length: >230 |
| | | XS/S:≥220; | Width: |
| | | M/L/XL: ≥230; | XS: 75-80; |
| | | Width(mm): | S: 85-87 |
| | | XS: 70±10; | M: 95-98 |
| | | S: 80±10; | L: 105-107 |
| | | M: 95±10; | XL: 115-117 |
| | | L: 110±10; | |
| | | XL: 120±10 | Pass |
| | | Thickness (mm):<br>Finger: ≥0.05<br>Palm: ≥0.05 | XS:<br>Finger: 0.10~0.11<br>Palm: 0.06~0.10<br>S:<br>Finger: 0.09~0.12<br>Palm: 0.07~0.09<br>M:<br>Finger: 0.08~0.12<br>Palm: 0.07~0.09 |
#### Table 5: Summary of Non-clinical Testing Table
{10}------------------------------------------------
| | | | | | L:<br>Finger: 0.09~0.12<br>Palm: 0.06~0.09<br>XL:<br>Finger: 0.11~0.12<br>Palm: 0.06~0.09 |
|---------------|-----------------------------------------------------------------------------------|------------------------------------------------|------------------------|--------|---------------------------------------------------------------------------------------------|
| ASTM D5151 | Watertightness<br>Test<br>for<br>Detection<br>of<br>Holes | Meet the requirements of<br>ASTM D5151 AQL 2.5 | | | Pass<br>XS:0/125 leaks<br>S:1/125 leaks<br>M:0/125 leaks<br>L:2/125 leaks<br>XL:2/125 leaks |
| ASTM D6124 | Powder Content | Meet the requirements of<br>ASTM D6124 < 2.0mg | | | Pass<br>XS:0.08 mg;<br>S:0.12mg;<br>M:0.09mg;<br>L:0.10mg;<br>XL:0.11. |
| ASTM D412 | Physical<br>properties | Before<br>Aging | Tensile<br>Strength | ≥14MPa | Pass<br>XS:18~21<br>S:18~25<br>M:18~26<br>L:16~24<br>XL:16~25 |
| | | | Ultimate<br>Elongation | ≥500% | Pass<br>XS: 520-536<br>S:518~517<br>M:515~575<br>L:529~561<br>XL:523~578 |
| | | After<br>Aging | Tensile<br>Strength | ≥14MPa | Pass<br>XS:15 ~21<br>S:16~21<br>M:15~24 |
| | | | | | L:16~23<br>XL:16~24 |
| | | | | | Pass |
| | | | Ultimate<br>Elongation | ≥400% | XS:500~525<br>S:501~522<br>M:501~550<br>L:501~551<br>XL:503~569 |
| | | | | | Pass |
| ASTM<br>D6978 | Chemotherapy<br>Drugs<br>Tested with<br>Minimum<br>Breakthrough<br>Detection Time | | | | Carmustine (BCNU) 3.3 mg/ml: 23.7 Minutes |
| | | | | | Cisplatin 1.0 mg/ml: > 240 Minutes |
| | | | | | Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes |
| | | | | | Dacarbazine (DTIC) 10.0 mg/ml: > 240 Minutes |
| | | | | | Doxorubicin HCI 2.0 mg/ml: > 240 Minutes |
| | | | | | Etoposide 20.0 mg/ml: > 240 Minutes |
| | | | | | Fluorouracil 50.0 mg/ml: >240 Minutes |
| | | | | | Methotrexate 25 mg/ml: >240 Minutes |
| | | | | | Mitomycin C 0.5 mg/ml: > 240 Minutes |
| | | | | | Paclitaxel 6.0 mg/ml: > 240 Minutes |
| | | | | | Thio Tepa 10.0 mg/ml: 45.7 Minutes |
| | | | | | Vincristine Sulfate 1.0 mg/ml: > 240 Minutes |
| ISO 10993-5 | Cytotoxicity | Non-cytotoxic | | | Under conditions of<br>the study, did not<br>show potential<br>toxicity to L-929 cells. |
| | | | | | Pass |
| ISO 10993-10 | Irritation | Non-irritating | | | Under the conditions<br>of the study, not an<br>irritant. |
| | | | | | Pass |
| ISO 10993-10 | Sensitization | Non-sensitizing | | | Under conditions of<br>the study, not a<br>sensitizer. |
{11}------------------------------------------------
{12}------------------------------------------------
#### 9.0 _Summary of Clinical Testing
Clinical testing is not needed for this device.
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Examination Glove (Tested for Use with Chemotherapy Drugs), is as safe, as effective, and performs as well as or better than the legally marketed predicate device under K201390.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.