K222698 · Curexo, Inc. · OLO · Oct 7, 2022 · Neurology
Device Facts
Record ID
K222698
Device Name
CUVIS-spine
Applicant
Curexo, Inc.
Product Code
OLO · Neurology
Decision Date
Oct 7, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.
Device Story
Mobile robotic system for pedicle screw placement; comprises robotic arm, main console, and optional staff console. Inputs: intraoperative 2D/3D images (O-arm/C-arm). Operation: surgeon performs surgical planning on console; system uses optical tracking (infrared camera) to register patient anatomy and track instruments in real-time. Robotic arm provides forced-controlled movement (hand guide function) to position guide bush for instrument trajectory. Output: visual guidance and physical positioning of surgical tools. Used in OR by surgeons. Benefits: precise anatomical localization and instrument orientation during spinal procedures.
Clinical Evidence
Bench testing included pose accuracy and repeatability validation. Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10. Electrical safety and EMC tested per ES 60601-1 and IEC 60601-1-2. A cadaveric study was performed to validate robotic-assisted pedicle screw placement.
Technological Characteristics
Mobile robotic system; optical tracking (infrared camera); forced-controlled movement. Materials: patient marker biocompatibility per ISO 10993. Connectivity: console-based. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 14971 (risk management), IEC 60601-1-6 (usability).
Indications for Use
Indicated for use as an aid for precisely locating anatomical structures and spatial positioning/orientation of guide bush for navigating surgical instruments in open or percutaneous surgical procedures, provided markers and rigid patient anatomy are identifiable on O-arm or C-arm.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2022
Curexo, Inc. % Do Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 68-gil 5, Gangnam-gu Seoul, 06210 Korea, South
Re: K222698
Trade/Device Name: CUVIS-spine Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 5, 2022 Received: September 7, 2022
#### Dear Do Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K222698
Device Name
CUVIS-spine (Model name: CS100)
#### Indications for Use (Describe)
CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# CUVIS-spine 510(k) Summary K222698
# 5. 510(k) Summary
# 5.1. General Information
| Applicant/Submitter: | CUREXO, INC. |
|----------------------|----------------------------------------------------------------------------------------------------------------------|
| Address: | 577, Gangnam-Daero, Seocho-Gu<br>Seoul, 06530, Republic of Korea<br>Tel: +82-31-788-7935 |
| Contact Person: | Do Hyun Kim |
| Address: | Unit 904, Eonju-ro 86gil 5, Gangnam-gu<br>Seoul 06210, Korea.<br>Tel: +82-2-538-9140<br>Email: ceo@btsolutions.co.kr |
| Preparation Date: | September 5, 2022 |
# 5.2. Device Name and Code
| Device Trade Name: | CUVIS-spine |
|----------------------|----------------------------|
| Model Name: | CS100 |
| Common Name: | Pedicle Screw Guide System |
| Classification Name: | Stereotaxic instrument |
| Product Code: | OLO |
| Regulation Number: | 21 CFR 882.4560 |
| Classification: | II |
| Review Panel: | Orthopedic |
# 5.3. Predicate Device
CUVIS-spine is substantially equivalent to the following legally marketed predicate devices
Table 1 Primary Predicate device
| Applicant | Device Name | 510(k) Number |
|-------------|-------------|---------------|
| CUREXO, INC | CUVIS-spine | K201569 |
# 5.4. Device Description
The CUVIS-spine is a mobile system mainly comprising the robotic arm, the main console and
{4}------------------------------------------------
#### CUVIS-spine K22698
the staff console as an option. The robotic arm is positioned on the floor near the side of the surgical table. The location of the main console or the staff console is appropriately determined considering the user preference and the environments.
CUVIS-spine is a pedicle screw guide system which consists of Robotic Arm, Main Console, Staff Console, Guide bush, Source Calibrator, Registration Tool, Registration Tool Adapter, Robotic Arm Drape, Tool Drape, Detector Drape, Marker Ball, Patient Marker, Marker Driver, Detector Calibrator, Dilator, Serration-tip Dilator, Drill Bit, Tapper, Stylet Tapper, Screwdriver, Instrument Container.
### 5.5. Indications / Intended Use
CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous surgical procedures provided that the required markers and rigid patient anatomy can be identified on O-arm or Carm.
#### 5.6. Technical Characteristics in Comparison to Predicate Devices
CUVIS-spine is substantially equivalent to the following legally marketed predicate devices
| | Proposed Device | Predicate Device |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | CUVIS-spine | CUVIS-spine |
| | | (K201569) |
| Device description and indications for use | | |
| Manufacturer | CUREXO, INC | CUREXO, INC |
| Classification | Class II | Class II |
| Product Code | OLO | OLO |
| Regulation No. | 882.4560 | 882.4560 |
| General Device description | The CUVIS-spine is a mobile<br>system mainly comprising the<br>robotic arm, the main console and<br>the staff console as an option. The<br>robotic arm is positioned on the<br>floor near the side of the surgical | The CUVIS-spine is a mobile<br>system mainly comprising the<br>robotic arm, the main console and<br>the staff console as an option. The<br>robotic arm is positioned on the<br>floor near the side of the surgical |
| | Proposed Device | Predicate Device |
| Device | CUVIS-spine | CUVIS-spine<br>(K201569) |
| | table. The location of the main<br>console or the staff console is<br>appropriately determined<br>considering the user preference<br>and the environments. | table. The location of the main<br>console or the staff console is<br>appropriately determined<br>considering the user preference<br>and the environments |
| | CUVIS-spine is intended for use<br>as an aid for precisely locating<br>anatomical structures and for the<br>spatial positioning and orientation | CUVIS-spine is intended for use<br>as an aid for precisely locating<br>anatomical structures and for the<br>spatial positioning and orientation |
| | of guide bush to be used by<br>surgeons for navigating and/or | of guide bush to be used by<br>surgeons for navigating and/or |
| Indications<br>for use | guiding compatible surgical<br>instruments in open or | guiding compatible surgical<br>instruments in open or |
| | percutaneous surgical procedures<br>if the required markers and rigid | percutaneous surgical procedures<br>if the required markers and rigid |
| | patient anatomy can be identified<br>on O-arm or C-arm. | patient anatomy can be identified<br>on O-arm or C-arm. |
| Technical Characteristics | | |
| | - Intraoperative images | - Intraoperative images |
| | - Patient registration | - Patient registration |
| Principle of<br>operation | - Surgical planning | - Surgical planning |
| | - Real-time tracking of navigated<br>instruments | - Real-time tracking of<br>navigated instruments |
| | - Guidance of instruments | - Guidance of instruments |
| Image | O-arm, C-arm | O-arm, C-arm |
| Input Images | - 3D intra-operative exam | - 3D intra-operative exam |
| | - 2D intra-operative exam | - 2D intra-operative exam |
| Tracker | Optical Tracking System | Optical Tracking System |
| Guide | Dilator and Tapper | Dilator and Tapper |
| Target<br>Tracking | YES | YES |
| Software<br>Information | - SRC (Control Software)<br>- SPN (Planner Software) | - SRC (Control Software)<br>- SPN (Planner Software) |
| Save/load<br>Planning | YES | YES |
| Merge images | YES | YES |
| | Proposed Device | Predicate Device |
| Device | CUVIS-spine | CUVIS-spine<br>(K201569) |
| functionality | | |
| Trajectory planning<br>parameters | - Entry point | - Entry point |
| | - Target point | - Target point |
| | - Instrument length/diameter | - Instrument length/diameter |
| Localization means | Optical system (infrared camera) | Optical system (infrared camera) |
| Image-guided | YES | YES |
| Controller | Forced-controlled movement | Forced-controlled movement |
| | allowing robot arm positioning | allowing robot arm positioning |
| | (called hand guide function) | (called hand guide function) |
| Patient registration<br>method | - Intra-op CT : Registration tool<br>- Fluoroscopy : Source calibrator | - Intra-op CT : Registration tool<br>- Fluoroscopy : Source calibrator |
| Real time display of<br>instrument position | YES | YES |
| Accessories | - Registration instruments<br>(Guide bush, Source calibrator,<br>Registration tool, Registration<br>tool adapter, Detector<br>calibrator) | - Registration instruments<br>(Guide bush, Source calibrator,<br>Registration tool, Registration<br>tool adapter, Detector<br>calibrator) |
| | - Patient reference instruments<br>(Patient marker, Clamp, Pin,<br>Marker ball, Adapter, Pin Vise,<br>Slide Hammer, Probe) | - Patient reference instruments<br>(Patient marker, Clamp,<br>Marker ball) |
| | - Surgical instruments<br>(Marker Driver, Dilator,<br>Serration-tip dilator, Bur, Stylet<br>tapper, Tapper, Screwdriver,<br>Robotic arm drape, Tool drape,<br>Detector drape, Instrument<br>Container) | - Surgical instruments<br>(Marker Driver, Dilator,<br>Serration-tip dilator, Bur,<br>Stylet tapper, Tapper,<br>Screwdriver, Robotic arm<br>drape, Tool drape, Detector<br>drape, Instrument Container) |
| Electrical Safety and<br>Electromagentic<br>compatibility | - IEC60601-1<br>- IEC60601-1-2 | - IEC60601-1<br>- IEC60601-1-2 |
| Accuracy | Yes | Yes |
| | Proposed Device | Predicate Device |
| Device | CUVIS-spine | CUVIS-spine<br>(K201569) |
| verification on<br>anatomical<br>landmarks | | |
| Bio-compatibility | The biocompatibility evaluation<br>for the patient marker has been<br>conducted in accordance with EN<br>ISO 10993 standards. | The biocompatibility evaluation<br>for the patient marker has been<br>conducted in accordance with EN<br>ISO 10993 standards. |
| Parts of<br>Contact with<br>Patient | Patient marker | Patient marker, Clamp* |
| Nature of<br>Body Contact | Implant device / bone | Implant device / bone |
| Patient Fixation | Reference is fixed to patient's<br>bony structure for tracking system | Reference is fixed to patient's<br>bony structure for tracking system |
{5}------------------------------------------------
## CUVIS-spine K222698
{6}------------------------------------------------
## CUVIS-spine K222696
{7}------------------------------------------------
*For Clamp corresponds to the Body of the Patient Marker of the initial approved device. The structure of Patient Marker, the initial approved device, of the Maker Frame and Body were disassembled, and the name was changed for each structure. This change was documented on August 14, 2021.
# 5.7. Performance Data
Non-clinical tests: Pose accuracy and Repeatability of the CUVIS-spine were tested and validated. A cadaveric study of the proposed device had been performed for the robotic-assisted pedicle screw placement.
Biocompatibility were tested using following consensus standards:
- Tests for in vitro cytotoxicity were tested and evaluated according to the FDA-recognized consensus standard, ISO 10993-5.
- -Tests for irritation, skin sensitization and intracutaneous reactivity were tested and evaluated according to the FDA-recognized consensus standard, ISO 10993-10.
Electromagnetic compatibility and electrical safety, etc, were tested using following consensus standards:
- Basic safety and essential performance of the CUVIS-spine is tested and evaluated -
{8}------------------------------------------------
according to the FDA-recognized consensus standard, ES 60601-1.
- Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2.
- Risk management was recorded by referring to ISO 14971. -
- Usability was documented by referring to IEC 60601-1-6. -
#### 5.8. Substantial Equivalence
CUVIS-spine is not based on a new technology. The predicate device is already available in the US market. There is a slight difference between CUVIS-spine and the predicate device. However, it does not affect the technical, clinical and biological aspects. It can be claimed that CUVIS-spine is substantially equivalent to the predicate device.
### 5.9. Conclusions
On the basis of the information provided in this Summary, CUREXO, INC. believes that CUVIS-spine is substantially equivalent to legally commercialized predicate device for the purposes of this 510 (k) submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.