K222663 · Dongguan TT Medical, Inc. · FGE · Mar 9, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K222663
Device Name
MultiStage Balloon Dilatation Catheter
Applicant
Dongguan TT Medical, Inc.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Mar 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
Device Story
MultiStage Balloon Dilatation Catheter is an endoscopic device used to dilate strictures in the alimentary tract and the Sphincter of Oddi. The device consists of an inflatable Pebax balloon mounted on a catheter shaft with multiple lumens for inflation and guidewire passage. It features 'three-in-one' technology for successive, gradual dilation. Two radiopaque tantalum marker bands on the balloon shoulders allow for fluoroscopic positioning. Available in two configurations: MSW (wireguided) for simultaneous introduction with a guidewire, and MSO (non-wireguided) for use over a pre-positioned guidewire. Operated by physicians in a clinical setting, the device is passed through an endoscope working channel (minimum 2.8mm for alimentary, 3.7mm for biliary). The physician monitors the balloon position fluoroscopically and inflates it to dilate the stricture, potentially improving patient outcomes by restoring patency to the tract.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5, -10), shelf life (ASTM F1980), sterile barrier integrity (ISO 11607-1), sterilization validation (ISO 11135), and performance testing (visual inspection, tension, guidewire/channel compatibility, balloon fatigue, rated burst pressure, kink resistance, leakage, and radiopacity).
Indicated for adult and adolescent patients for endoscopic dilation of alimentary tract strictures; also indicated for adults for endoscopic dilation of the Sphincter of Oddi, with or without prior sphincterotomy.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 9, 2023
Dongguan TT Medical, Inc. % Mingzi Hussey Principal Regulatory Consultant Zi-medical. Inc. 93 Springs Rd Bedford, MA 01730
Re: K222663
> Trade/Device Name: MultiStage Balloon Dilatation Catheter Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: FGE Dated: February 10, 2023 Received: February 10, 2023
Dear Mingzi Hussey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any
{1}------------------------------------------------
Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image contains the text "Je An -S" in a large, sans-serif font. The letters are black and bolded, and they are arranged horizontally. The background is white, with a faint watermark-like design in a light blue color.
Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222663
Device Name MultiStage Balloon Dilatation Catheter
Indications for Use (Describe)
The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Checked Box" src="checkbox_checked.png"/> </div> <div style="display:inline-block;">Repackaging (Part 201.22(a) Subject Shipment)</div> </div></td></tr><tr><td><div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Unchecked Box" src="checkbox_unchecked.png"/> </div> <div style="display:inline-block;">Own-Use Controlled (201.22(a) Subject Shipment)</div> </div></td></tr></table> | <div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Checked Box" src="checkbox_checked.png"/> </div> <div style="display:inline-block;">Repackaging (Part 201.22(a) Subject Shipment)</div> </div> | <div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Unchecked Box" src="checkbox_unchecked.png"/> </div> <div style="display:inline-block;">Own-Use Controlled (201.22(a) Subject Shipment)</div> </div> |
| <div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Checked Box" src="checkbox_checked.png"/> </div> <div style="display:inline-block;">Repackaging (Part 201.22(a) Subject Shipment)</div> </div> | | |
| <div style="display:inline-block; vertical-align:middle;"> <div style="display:inline-block;"> <img alt="Unchecked Box" src="checkbox_unchecked.png"/> </div> <div style="display:inline-block;">Own-Use Controlled (201.22(a) Subject Shipment)</div> </div> | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 510(k) Summary
## [As required by 21 CFR 807.92]
#### 1. Submission Sponsor
Dongguan TT Medical, Inc. Address: BLD#1, 1 Taoyuan Road, Songshan Lake, Dongguan, Guangdong, 523808 China Phone: +86 15553341910 Contact: Yuying Bi, Regulatory Affairs Specialist Email: ybi @ttmedicalinc.com
## 2. Submission Correspondent
Mingzi Hussey Zi-medical, Inc. Address: 93 Springs Rd, Bedford, MA 01730 US Phone: 206-981-0675 Email: mingzi@zi-medical.com
## 3. Date Prepared
Aug 10th, 2022
## 4. Device Identification
Trade Name: MultiStage Balloon Dilatation Catheter Regulation name: Biliary Catheter And Accessories Product Code: FGE Regulation Number: 21 CFR 876.5010 Regulation Class: Class II Review Panel: Gastroenterology/Urology
## 5. Predicate Devices
The proposed device is substantially equivalent to the following predicate device:
| Applicant | Device name | 510(k) Number | Product code |
|---------------------------|-------------------------------------------------|---------------|--------------|
| Boston Scientific<br>Corp | CRETM Wireguided Balloon<br>Dilatation Catheter | K112994 | FGE |
#### 6. Device Description
The MultiStage Balloon Dilatation Catheters are indicated for use to endoscopically dilate strictures of the alimentary tract and the Sphincter of Oddi with or without prior sphincterotomy. It consists of an inflatable balloon on a catheter shaft with multiple lumens for inflation and passage of a guidewire.
{4}------------------------------------------------
The MultiStage balloon is designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon is made of Pebax material. The MultiStage Balloon Dilatation is designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through an minimum working channel of a 2.8mm alimentary tract, or a 3.7mm biliary tract. The hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture. Accordingly, product models, are divided into those with a wireguided (MSW) and non-wireguided (MSO) component.
The MSW model allows the guidewire and catheter to be introduced into the endoscope together. The MSO model as a traditional operation method, the guidewire is first inserted into the endoscope for positioning and then the catheter is inserted through the guidewire for good passability. The physician can choose the model as needed.
#### 7. Indication for Use Statement
The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
## 8. Comparison of Technological Characteristics
The proposed device is compared with the predicate devices in terms of intended use, technological characteristics and principles of operation. The proposed device shares the similar indications for use, similar materials, same device operation, and same overall technical capabilities as the predicate device. The MSW model has same components with the predicate device while the MSO model does not contain a guidewire for the users to install a guidewire by their choice. The proposed device also has the same standards and requirements as the predicate device, and has been verified for its safety and effectiveness to establish substantially equivalence.
| | Nominal Balloon Diameter | | | | | |
|-------------------|--------------------------|-----------|-----------|-----------|-----------|-----------|
| Nominal<br>Length | 6mm | 8mm | 10mm | 12mm | 15mm | 18mm |
| 30mm | MSO063008 | MSO083008 | MSO103008 | MSO123008 | MSO153008 | MSO183008 |
| 30mm | MSO063017 | MSO083017 | MSO103017 | MSO123017 | MSO153017 | MSO183017 |
| 30mm | MSO063023 | MSO083023 | MSO103023 | MSO123023 | MSO153023 | MSO183023 |
| 55mm | MSO065508 | MSO085508 | MSO105508 | MSO125508 | MSO155508 | MSO185508 |
| 55mm | MSO065517 | MSO085517 | MSO105517 | MSO125517 | MSO155517 | MSO185517 |
| 55mm | MSO065523 | MSO085523 | MSO105523 | MSO125523 | MSO155523 | MSO185523 |
| 30mm | MSW063008 | MSW083008 | MSW103008 | MSW123008 | MSW153008 | MSW183008 |
| 30mm | MSW063017 | MSW083017 | MSW103017 | MSW123017 | MSW153017 | MSW183017 |
MultiStage Balloon Dilatation Catheter has 72 models:
{5}------------------------------------------------
| 30mm | MSW063023 | MSW083023 | MSW103023 | MSW123023 | MSW153023 | MSW183023 |
|------|-----------|-----------|-----------|-----------|-----------|-----------|
| 55mm | MSW065508 | MSW085508 | MSW105508 | MSW125508 | MSW155508 | MSW185508 |
| 55mm | MSW065517 | MSW085517 | MSW105517 | MSW125517 | MSW155517 | MSW185517 |
| 55mm | MSW065523 | MSW085523 | MSW105523 | MSW125523 | MSW155523 | MSW185523 |
## 9. Description of Non-clinical Testing
The non-clinical tests of the subject device and predicate device are in compliance with the following standards and guidance.
The biocompatibility evaluation for MultiStage Balloon Dilatation Catheter was conducted in accordance with ISO 10993-1.
- · Cytotoxicity (ISO 10993-5)
- · Sensitization (ISO 10993-10)
- · Intracutaneous reactivity test (ISO 10993-10)
Result: All tests were passed.
Shelf life for MultiStage Balloon Dilatation Catheter test is conducted based on ASTM F1980: · Shelf life test report
- Shell life test report
Result: All tests were passed.
Sterile barrier systems for MultiStage Balloon Dilatation Catheter were evaluated in accordance with ISO 11607-1:2019.
Sterilization Process has been validated accordance with ISO 11135:2014.
Result: The MultiStage Balloon Dilatation Catheter is sterile with a validation of
of the sterilization process that demonstrates achievement of the required SAL of 10% and is in accordance with the half cycle overkill approach.
Technological characteristics for MultiStage Balloon Dilatation Catheter have been tested for its functions as intended including verification of performance characteristics per performances characteristics relevant to functions as intended:
- · Visual Inspection
- · Bacteriostasis performance of pouch
- · Dye penetration
- Pouch
- · Tension test
- · Guidewire compatibility test
- · Channel compatibility
{6}------------------------------------------------
- Product Dimensional Inspection
- . Compliance (deployed balloon dimensional analysis)
- Balloon fatigue; No leakage and damage when inflation .
- · Balloon rated burst pressure (RBP)
- · Peak tensile
- · Kink resistance
- Leakage Test
- · Hydratability
- Hub
- · Corrosion resistance
- · X-ray detectability
- · Chemical properties
Result: All tests were passed.
The results of Non-Clinical Performance testing demonstrate that the MultiStage Balloon Dilatation Catheter is considered as safe and effective as the predicate.
### 10. Performance Data-Clinical
No clinical study is included in this submission.
### 11. Conclusion
The differences in technological characteristics between the proposed device and the predicate device do not raise any different questions of safety or effectiveness. Performance testing and compliance with voluntary standards demonstrate the subject device is substantially equivalent to the relevant aspects of the predicate device in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
Therefore, the proposed devices are determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.