K222644 · Nihon Kohden Orangemed, Inc. · CBK · May 31, 2023 · Anesthesiology
Device Facts
Record ID
K222644
Device Name
Nihon Kohden NKV-440 Ventilator System
Applicant
Nihon Kohden Orangemed, Inc.
Product Code
CBK · Anesthesiology
Decision Date
May 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.
Device Story
Servo-controlled ventilator providing continuous invasive/non-invasive respiratory support; neonate to adult patients. System comprises Breath Delivery Unit (BDU) and Graphic User Interface (GUI). BDU uses internal blower and proportional valve to mix ambient air and oxygen (21-100% concentration); microprocessor controls gas delivery based on user-defined parameters (PEEP, inspiratory pressure). GUI displays waveforms, numeric values, and alarm settings. Used in hospitals/hospital-type facilities by trained clinicians for ventilation therapy; supports intra-hospital transport. Output informs clinical decision-making regarding respiratory support levels; benefits patients by providing mandatory/spontaneous ventilation modes (A/CMV, SIMV, SPONT) and monitoring.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including performance of ventilation modes, power/battery testing, environmental testing, waveform comparison, biocompatibility, cleaning/disinfection, EMC/electrical safety, and human factors/usability.
Technological Characteristics
Servo-controlled ventilator; AC/internal Li-ion battery power. Materials: breathing gas pathways compliant with ISO 18562-1. Sensing/actuation: internal blower, proportional valves, microprocessor control. Connectivity: standalone/networked. Sterilization: cleaning/disinfection protocols. Software: IEC 62304 compliant. Standards: ISO 80601-2-12, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring invasive or non-invasive respiratory support in hospitals, hospital-type facilities, or during in-hospital transport. Contraindicated for use in the presence of flammable anesthetics and MRI applications.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
May 31, 2023
Nihon Kohden OrangeMed, Inc. Sheryl Higgins Vice President of Regulatory Affairs & Quality Assurance 1800 E. Wilshire Avenue Santa Ana, California 92705
Re: K222644
Trade/Device Name: Nihon Kohden NKV-440 Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: May 4, 2023 Received: May 4, 2023
Dear Sheryl Higgins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan E. Nyberg -S
for James Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222644
Device Name Nihon Kohden NKV-440 Ventilator System
### Indications for Use (Describe)
The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Nihon Kohden OrangeMed, Inc. The logo consists of the company's name in two lines of text. The first line, "NIHON KOHDEN", is in blue, while the second line, "Nihon Kohden OrangeMed, Inc.", is in orange. To the left of the company name is a blue graphic element.
# 510(k) Summary
#### 1. Date
May 04, 2023
#### 2. Submitter / Manufacturing Location
Nihon Kohden OrangeMed, Inc. 1800 E. Wilshire Avenue Santa Ana, CA 92705 USA
#### 3. Company Contact
| Primary: | Sheryl Higgins<br>Vice President of Regulatory Affairs and Quality Assurance |
|----------|------------------------------------------------------------------------------|
| Email: | shiggins@orange-med.com |
| Phone: | (949) 502-6448 x7012 |
| | Secondary: Jacqueline Villanueva |
|--------|----------------------------------|
| | Director of Quality Assurance |
| Email: | jvillanueva@orange-med.com |
| Phone: | (949) 502-6448 x7045 |
#### Common Name of Device 4.
Critical Care Ventilator
#### 5. Trade Name
Nihon Kohden NKV-440 Ventilator System
#### Classification Name 6.
| Product Code: | CBK - Ventilator, Continuous, Facility Use |
|--------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 868.5895 Continuous Ventilator |
| Device Class: | II |
| Review Panel: | Anesthesiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Nihon Kohden OrangeMed, Inc. The logo consists of a blue graphic element on the left, followed by the text "NIHON KOHDEN" in blue. Below this, the text "Nihon Kohden OrangeMed, Inc." is displayed in orange. The logo is simple and professional, likely representing a medical or technology company.
#### 7. Predicate Device
| Product: | Nihon Kohden NKV-550 Ventilator System |
|---------------|----------------------------------------|
| 510k #: | K192307 |
| Manufacturer: | Nihon Kohden OrangeMed, Inc. |
#### 8. Reference Device
| Product: | Hamilton-C3 |
|---------------|---------------------|
| 510k #: | K201306 |
| Manufacturer: | Hamilton Medical AG |
#### 9. Description of Device
The NKV-440 is a servo-controlled ventilator that is designed to meet the gas delivery and performance requirements for neonate through adult patients. The NKV-440 design is comprised of two major components, a Breath Delivery Unit (BDU) and a Graphic User Interface (GUI). The GUI allows clinicians to set ventilator control parameters such as PEEP and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, and to view waveforms. The BDU assembly contains a blower and the electronics required to perform breath delivery. Ambient air is taken into the blower and mixed with oxygen which is flow rate controlled by a proportional valve. The mixed gas is provided to the patient. The microprocessor controls the blower and the proportional valve to deliver the pressure and oxygen concentration which are set by the user. It also provides various alarms and other design features to maximize patient safety.
### 10. Product Intended Function
The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation using medical oxygen and an internal source of air to deliver oxygen concentrations of 21 to 100%. Ventilatory support is intended to be delivered invasively or non-invasively to patients who require Assisted/Control Mandatory Ventilation (A/CMV), Synchronized Intermittent Mandatory Ventilation (SIMV), or Spontaneous Ventilation (SPONT).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nihon Kohden OrangeMed, Inc. The logo consists of a blue square with a white design on the left, followed by the text "NIHON KOHDEN" in blue. Below that, the text "Nihon Kohden OrangeMed, Inc." is written in orange.
### 11. Indication for Use (Intended Medical Indication)
The Nihon Kohden NKV-440 Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-440 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-440 is intended for use in hospitals and hospitaltype facilities, as well as, for in-hospital transportation.
# 12. Intended Patient Population
The intended patient population includes neonate through adult patients who require invasive or non-invasive respiratory support.
# 13. Intended Part of the Body or Type of Tissue Applied to or Interacted with
The device does not come into direct contact with the patient. The device is considered to be externally communicating with the patient airway because it delivers air/oxygen to the patient respiratory system.
### 14. Intended Environment of Use (Use Environment)
The device is intended for use in hospitals and hospital-type facilities, which provide respiratory care for patients requiring respiratory support.
The device may be used for intra-hospital transport within a hospital or hospital-type facility. The device is not intended for transport between hospitals or hospital-type facilities, therefore is not a transport ventilator as defined by ISO 80601-2-12 Section 201.1.1.
The device is not to be used in the presence of flammable anesthetics and MRI applications.
### 15. Summary of Technical Characteristics
The technological characteristics of the Nihon Kohden NKV-440 Ventilator System are substantially equivalent to the predicate device as compared and summarized in the table below.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue abstract symbol on the left, resembling a stylized wave or mountain range. To the right of the symbol is the company name, "NIHON KOHDEN," also in blue. The font is sans-serif and appears to be a bold typeface.
Nihon Kohden OrangeMed, Inc.
| Table 12-1 | Nihon Kohden<br>OrangeMed, Inc. | Nihon Kohden<br>OrangeMed, Inc. | Comparison |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | | | |
| Establishment | 3014631252 | 3014631252 | |
| MODEL | NKV-440 | NKV-550 | |
| 510(k) | This Submission | K192307 | |
| Indications for<br>Use | The Nihon Kohden NKV-<br>440 Ventilator System is<br>intended to provide<br>continuous ventilation for<br>adult, pediatric and neonatal<br>patients who require invasive<br>or noninvasive respiratory<br>support. The NKV-440<br>offers mandatory and<br>spontaneous ventilation<br>modes as well as respiratory<br>monitoring. The NKV-440 is<br>intended for use in hospitals<br>and hospital-type facilities,<br>as well as, for in-hospital<br>transportation. | The Nihon Kohden NKV-<br>550 Series Ventilator System<br>is intended to provide<br>continuous ventilation for<br>adult, pediatric and neonatal<br>patients who require invasive<br>or noninvasive respiratory<br>support. The NKV-550<br>offers mandatory and<br>spontaneous ventilation<br>modes as well as respiratory<br>monitoring. The NKV-550 is<br>intended for use in hospitals<br>and hospital-type facilities,<br>as well as, for in-hospital<br>transportation. | Same |
| Clinical<br>Conditions | Hospitals, hospital-type<br>facilities and in-hospital<br>transportation for patients<br>who need ventilation therapy | Hospitals, hospital-type<br>facilities and in-hospital<br>transportation for patients<br>who need ventilation therapy | Same |
| Anatomical Site | Patient airways | Patient airways | Same |
| Target<br>Population | Adult, pediatric and neonatal<br>patients | Adult, pediatric and neonatal<br>patients | Same |
| Users | Trained clinicians, not lay<br>users | Trained clinicians, not lay<br>users | Same |
| Clinical<br>Performance | Met ISO 80601-2-12<br>requirements on essential<br>performance of critical care<br>ventilator | Met ISO 80601-2-12<br>requirements on essential<br>performance of critical care<br>ventilator | Same |
| Waveform<br>Comparison | Comparable waveform test<br>result | Comparable waveform test<br>result | Similar |
| Table 12-1 | Nihon Kohden<br>OrangeMed, Inc. | Nihon Kohden<br>OrangeMed, Inc. | |
| Company | OrangeMed, Inc. | OrangeMed, Inc. | |
| Establishment | 3014631252 | 3014631252 | Comparison |
| MODEL | NKV-440 | NKV-550 | |
| 510(k) | This Submission | K192307 | |
| Design | Consists of a graphic user<br>interface to set and monitor<br>ventilation, breath delivery<br>unit, breathing circuit, and<br>cart;<br>Controls air and oxygen<br>deliveries by a blower and<br>proportional valve through<br>microprocessors | Consists of a graphic user<br>interface to set and monitor<br>ventilation, breath delivery<br>unit, breathing circuit, and<br>cart;<br>Controls air and oxygen<br>deliveries by proportional<br>valves through<br>microprocessors | Similar<br>In the NKV-440, a built-<br>in blower draws the room<br>air in and controls the air<br>delivery. In the NKV-<br>550, an external air<br>compressor is the air gas<br>source, and a built-in<br>proportional valve<br>controls the air delivery. |
| Conditions of<br>Use | Hospitals, hospital-type<br>facilities and in-hospital<br>transportation | Hospitals, hospital-type<br>facilities and in-hospital<br>transportation | Same |
| Chemicals<br>Deployed to<br>Patient | Air and oxygen | Air and oxygen | Same |
| Delivery method<br>to Patient | Positive pressure | Positive pressure | Same |
| Energy Used for<br>Device | AC Power and internal DC<br>Power (Li-ion battery) | AC Power and internal DC<br>Power (Li-ion battery) | Same<br>Exactly same battery<br>packs are used in both<br>NKV-440 and NKV-550. |
| Control<br>principle | time-cycled, volume-<br>constant, pressure-controlled | time-cycled, volume-<br>constant, pressure-controlled | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo of NIHON KOHDEN. The logo consists of a blue square with a white abstract design inside, resembling a stylized wave or a rising sun. To the right of the square is the text "NIHON KOHDEN" in a bold, sans-serif font, also in blue.
### Nihon Kohden OrangeMed, Inc.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue square with a white curved line running through it, followed by the text "NIHON KOHDEN" in blue. The text is in a sans-serif font and is aligned to the right of the square.
Nihon Kohden OrangeMed, Inc.
| Table 12-1 | Nihon Kohden | Nihon Kohden | Comparison |
|------------------------------------|---------------------------|---------------------------|------------|
| Company | OrangeMed, Inc. | OrangeMed, Inc. | |
| Establishment | 3014631252 | 3014631252 | |
| MODEL | NKV-440 | NKV-550 | |
| 510(k) | This Submission | K192307 | |
| Specifications:<br>Breathing Modes | Invasive Ventilation: | Invasive Ventilation: | |
| | A/CMV-PC | A/CMV-PC | Same |
| | A/CMV-VC | A/CMV-VC | |
| | A/CMV-PRVC | A/CMV-PRVC | |
| | SIMV-PC-PS | SIMV-PC-PS | |
| | SIMV-VC-PS | SIMV-VC-PS | |
| | SIMV-PRVC-PS | SIMV-PRVC-PS | |
| | SPONT-CPAP | SPONT-CPAP | |
| | SPONT-PS | SPONT-PS | |
| | SPONT-VS | SPONT-VS | |
| | APRV | APRV | |
| | Non-Invasive Ventilation: | Non-Invasive Ventilation: | |
| | A/CMV-PC | A/CMV-PC | |
| | SIMV-PC-PS | SIMV-PC-PS | |
| | SPONT-CPAP | SPONT-CPAP | |
| | SPONT-PS | SPONT-PS | |
| | APRV | APRV | |
| | CPAP | CPAP | |
| | O2 Therapy | O2 Therapy | |
### 16. Summary of Non-Clinical Performance Data
Performance of the Nihon Kohden NKV-440 Ventilator was demonstrated by the following:
- Agency Testing to Applicable Standards
- Performance of Ventilation Modes . and Control Settings
- Device/Software Functionality
- . Power performance with AC power, External Battery, and Backup Battery
- . Essential Performance and Worst Case VBS
- Environmental
- Waveform Comparison with Predicate
- . Product Endurance/Reliability
- . Biocompatibility
- . Cleaning & Disinfection
- EMC and Electrical Safety ●
- Human Factors/Usability
- Risk Management
- Compatibility with 3rd Party Devices ●
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image features the logo for Nihon Kohden. The logo consists of a blue abstract graphic to the left of the company name, "NIHON KOHDEN," which is also in blue. Below the main logo, in orange text, is the text "Nihon Kohden OrangeMed, Inc."
Where applicable, the Nihon Kohden NKV-440 Ventilator has been tested in compliance with the following standards:
| Standard | Title |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AIM 7351731 | Medical Electrical Equipment & System Electromagnetic Immunity Test for<br>RFID Readers |
| Rev. 2 2017-02 | |
| ANSI AAMI ES 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety |
| 2005+AC1;A2 (R2012) | and essential performance (IEC 60601-1:2005, MOD) |
| IEC 60601-1 | Medical Electrical Equipment - Part 1: General Requirements For Basic Safety<br>And Essential Performance |
| Ed. 3.1 2012-08 | |
| IEC 60601-1-2 | Medical Electrical Equipment - Part 1-2: General Requirements for Safety -<br>Collateral Standard: Electromagnetic Compatibility - Requirements and Tests |
| Ed. 4.1 2020-09 | |
| IEC 60601-1-6 | Medical electrical equipment - Part 1-6: General requirements for safety -<br>Collateral Standard: Usability |
| Ed. 3.1 2013-10 | |
| IEC 60601-1-8 | Medical Elec. Equip. - Part 1-8: General Req. for Basic Safety & Essential<br>Perf. - Collateral Standard: General Req., Tests & Guidance for Alarm<br>Systems in Medical Elec. Equip. & Medical Elec. Systems |
| Ed 2.1 2012-11 | |
| IEC 62133 | Secondary cells and batteries containing alkaline or other non-acid electrolytes<br>- Safety requirements for portable sealed Secondary cells, and for batteries<br>made from them, for use in portable applications |
| Ed. 2.0 2012-12 | |
| IEC 62304 | Medical device software - Software life cycle processes |
| Ed. 1.1 2015-06 | |
| IEC 62366-1 | Medical devices - Application of usability engineering to medical devices |
| Ed. 1.0 2015-02 | |
| ISO 10993-1 | Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing<br>within a Risk Management Process |
| Ed. 5 2018-08 | |
| ISO 14971 | Medical devices - Application of risk management to medical devices |
| Ed. 3 2019-12 | |
| ISO 18562-1 | Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare<br>Applications |
| Ed. 1 2017-03 | |
| ISO 80601-2-12 | Medical Electrical Equipment - Part 2-12: Particular Requirements for Basic<br>Safety and Essential Performance of Critical Care Ventilators |
| Ed. 2 2020-02 | |
| ISO 80601-2-55 | Medical Electrical Equipment – Part 2-55: Particular Requirements for The<br>Basic Safety and Essential Performance of Respiratory Gas Monitor |
| Ed. 2 2018-02 | |
| ISO 80601-2-61 | Medical electrical equipment - Part 2-61: Particular requirements for basic<br>safety and essential performance of pulse oximeter equipment |
| Ed 2 2017-12 | |
{10}------------------------------------------------
# 17. Summary of Animal Performance Data
Not Applicable - Animal performance data was not required to demonstrate substantial equivalence.
### 18. Summary of Clinical Performance Data
Not Applicable – Clinical performance data was not required to demonstrate substantial equivalence.
### 19. Conclusion
The evaluation and testing performed demonstrates that the Nihon Kohden NKV-440 Ventilator System is substantially equivalent the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.