← Product Code [KCT](/productcode/KCT) · K222640

# Sweden & Martina Surgical Trays (K222640)

_Sweden & Martina S.P.A. · KCT · May 26, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K222640

## Device Facts

- **Applicant:** Sweden & Martina S.P.A.
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** May 26, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

SWEDEN & MARTINA SURGICAL TRAYS are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. Soiled-used tools should never be placed back into the tray or silicone grommets. The Sweden & Martina trays are available in two (2) formats, nominal dimensions Size M (189x140 mm h 61.5 mm) and Size L (270x150 mm h 62 mm), and the worst-case loading configuration have been validated for a maximum load of 740 grams for all device configurations. The tray is to be enclosed in an FDA cleared steam sterilization validation was completed by: - Method Autoclave (Pre-vacuum Dynamic-Air-Removal Cycles) at a temperature of 134 °C (273 °F) with an exposure of four (4) minutes and a minimum drying time of twenty (20) minutes.

## Device Story

Sweden & Martina Surgical Trays are containers designed to hold dental surgical drills and tools; used to organize, steam sterilize, and transport instruments. Trays feature silicone grommets for instrument retention. Available in two sizes (M and L). Operated by dental staff in clinical environments. Trays are loaded with instruments, enclosed in FDA-cleared sterilization wrap, and processed via pre-vacuum steam autoclave. Facilitates instrument management and sterilization compliance; ensures organized storage and safe transport between procedures.

## Clinical Evidence

Bench testing only. Validation performed for steam sterilization using pre-vacuum dynamic-air-removal cycles at 134 °C for 4 minutes with 20 minutes drying time. Maximum load capacity validated at 740 grams.

## Technological Characteristics

Sterilization tray for dental instruments; includes silicone grommets. Available in two sizes: Size M (189x140x61.5 mm) and Size L (270x150x62 mm). Maximum load capacity: 740 grams. Sterilization method: Pre-vacuum steam autoclave (134 °C).

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 26, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Sweden & Martina S.p.A. Alessia Pezzato Quality Assurance Manager Via Veneto, 10 Due Carrare, Padova 35020 Italv

Re: K222640

Trade/Device Name: Sweden & Martina Surgical Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: October 27, 2022 Received: April 21, 2023

Dear Alessia Pezzato:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows a digital signature. The name "Eileen Cadel -S" is prominently displayed on the left side of the image. To the right, the text "Digitally signed by Eileen Cadel -S Date: 2023.05.26 14:38:09-04'00'" is present, indicating the date and time of the digital signature.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K222640

Device Name SWEDEN & MARTINA SURGICAL TRAYS

## Indications for Use (Describe)

SWEDEN & MARTINA SURGICAL TRAYS are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. Soiled-used tools should never be placed back into the tray or silicone grommets.

The Sweden & Martina trays are available in two (2) formats, nominal dimensions Size M (189x140 mm h 61.5 mm) and Size L (270x150 mm h 62 mm), and the worst-case loading configuration have been validated for a maximum load of 740 grams for all device configurations.

The tray is to be enclosed in an FDA cleared steam sterilization validation was completed by: - Method

Autoclave (Pre-vacuum Dynamic-Air-Removal Cycles)

at a temperature of 134 °C (273 °F) with an exposure of four (4) minutes and a minimum drying time of twenty (20) minutes.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K222640](https://fda-staging.innolitics.com/device/K222640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com